K141852 · Amo Manufacturing USA, LLC · GEX · Jan 6, 2015 · General, Plastic Surgery
Device Facts
Record ID
K141852
Device Name
IFS ADVANCED FEMTOSECOND LASER SYSTEM
Applicant
Amo Manufacturing USA, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea. In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments. In patients undergoing ophthalmic surgery or other treatment requiring arcuate cuts/incisions in the cornea, penetrating and/or intrastromal. In lamellar IEK and corneal harvesting. In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea. In the creation of a lamellar cut/resection of the cornea for lamellar IEK and for the creation of a penetrating cut/incision for penetrating IEK. In patients undergoing ophthalmic surgery or other treatment requiring the creation of corneal channels for placement/insertion of a corneal inlay device.
Device Story
Ophthalmic surgical laser; uses focused femtosecond laser pulses to create corneal incisions/tissue separation. Operates via precise micro-photodisruption of tissue at laser focus; thousands of pulses per second scanned to produce continuous cuts. System includes disposable applanation lens contacting cornea; eye fixated under low vacuum. Used by physicians in clinical settings. Output is controlled by computer-guided delivery system; provides surgeon with graphical user interface to select incision patterns. Benefits patient by enabling precise, automated corneal tissue resection for various refractive and therapeutic procedures.
Clinical Evidence
Bench testing only. Evaluated channel widths, depths, and cut quality using glass microscope slides, Agarose gel, and porcine eyes. Acceptance criteria for depth were identical to previously established performance for existing resection patterns.
Indicated for patients undergoing ophthalmic surgery requiring corneal lamellar resection, corneal flap creation (LASIK), corneal ring segment tunnel creation, arcuate incisions, lamellar/penetrating IEK, corneal harvesting, or corneal channel creation for inlay device insertion.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized human figure with three faces in profile, representing the department's focus on people and health. The emblem is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The text is in uppercase letters and is positioned around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 8, 2015
AMO Manufacturing USA, LLC Mr. Rodney Huang Regulatory Affairs PDP 510 Cottonwood Drive Milpitas, CA 95035
Re: K141852
Trade/Device Name: iFS Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Powered Laser Surgical Instrument Regulatory Class: Class II Product Code: GEX, HNO Dated: November 24, 2014 Received: November 26, 2014
Dear Mr. Huang:
This letter corrects our substantially equivalent letter of January 6, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) Unknown
Device Name iFS Laser System Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Indications for Use (Describe)
· In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea
· In patients undergoing surgery or other trequiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments
· In patients undergoing ophthalmic surgery or other treatment requiring arcuate cuts/incisions in the cornea, penetrating and/or intrastromal
· In lamellar IEK and corneal harvesting
· In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
· In the creation of a lamellar cutresection of the cornea for lamellar IEK and for the creation of a penetrating cutincision for penetrating IEK
· In patients undergoing ophthalmic surgery or other treation of corneal channel for placement/ insertion of a corneal inlay device
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
iFS Laser System 510(k) Premarket Notification AMO Manufacturing USA, LLC AMO Confidential
{4}------------------------------------------------
# Section 5 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### Submitter's name, address, telephone number, contact person, and date summary prepared:
| a. Applicant: | AMO Manufacturing USA, LLC<br>510 Cottonwood Drive<br>Milpitas, CA 95035<br>Tel: 408-273-4159 |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| b. Contact Person: | Rodney Huang<br>Regulatory Affairs PDP<br>510 Cottonwood Drive<br>Milpitas, CA 95035<br>Tel: 408-273-4159<br>Fax: 408-273-5966<br>Email: rodney.huang@amo.abbott.com |
| c. Date of Summary Preparation: | July 7, 2014 |
#### Name of device, including trade name and classification name:
Trade/Proprietary Name: iFS Laser System Common/Usual Name: Laser Classification Name: Powered Laser Surgical Instrument Classification Code(s): GEX (21 CFR 878.4810) HNO (21 CFR 886.4370)
#### Identification of the predicate device or legally marketed device or devices to which substantial equivalence is being claimed:
| 510(k) # | Trade Name | Manufacturer |
|----------|---------------------------------------------|--------------------------------------------|
| K131207 | FEMVTO LDV Z6<br>Femtosecond Surgical Laser | SIE AG,<br>Surgical Instrument Engineering |
| | | |
| K113151 | iFS Laser System | AMO Manufacturing USA, LLC |
{5}------------------------------------------------
A description of the device that is the subject of the 510(k), including explanation of how the device functions, basic scientific concepts, significant physical and performance characteristics (design, material, physical properties):
The iFS Laser System is an ophthalmic surgical laser designed for use as an ophthalmic surgical laser indicated for use as follows:
- . In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea
- . In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments
- . In patients undergoing ophthalmic surgery or other treatment requiring arcuate cuts/incisions in the cornea, penetrating and/or intrastromal
- In lamellar IEK and corneal harvesting
- In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
- . In the creation of a lamellar cut/resection of the cornea for lamellar IEK and for the creation of a penetrating cut/incision for penetrating IEK
- In patients undergoing ophthalmic surgery or other treatment requiring the creation of corneal channels for placement/insertion of a corneal inlay device
The iFS Laser System uses focused femtosecond laser pulses to create incisions and separate tissues in the cornea. Corneal dissection with the iFS Laser is achieved through precise individual micro-photodisruptions of tissue, which are controlled by repeatedly repositioning the laser focus. The light pulse is focused into a sufficiently small spot in order to achieve photodisruption of the tissue inside the focus. A tiny volume of tissue, a few microns in diameter, is thereby photodisrupted at the laser focus. The surgical effect is produced by scanning thousands of individual pulses per second to produce continuous incisions or tissue separation. These laser pulses are delivered through a disposable applanation lens that contacts the cornea while fixating the eye under low vacuum.
### Statement of how the technological characteristics of the device compare to those of the predicate or legally marketed device:
The technological characteristics of the iFS Laser System are substantially equivalent to those cleared indications under K131207 and K113151 for corneal resections and incisions. The subject and predicate iFS Laser System (K113151) and the Femto LDV Z6 Femtosecond Surgical Laser (K131207) share the same design principle and mode of operation in that they all deliver femtosecond pulses through a computer-controlled delivery system to produce a pattern of photodisruption to create cuts / separation in ophthalmic tissue. The means of fixation of the patient contact portion of the devices are all substantially equivalent in that a suction vacuum affixes a suction ring to the corneal surface prior to use.
iFS Laser System 510(k) Premarket Notification
{6}------------------------------------------------
No changes to the iFS Laser System operating principles, hardware, system specifications, or disposable applanation lens were required for the additional indication for use in patients requiring the creation of corneal channels for placement/insertion of a corneal inlay device. Software updates were required to include the addition of an inlay channel pattern and the associated graphical user interface functionality
## Brief summary of nonclinical test and results:
Verification and validation of the software updates were documented per FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005, and demonstrated that the functional and safety critical requirements are fulfilled and no outstanding safety critical issues remain open.
| Standard | Description |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2005<br>3rd edition | Medical Electrical Equipment - Part 1: General<br>Requirements for Basic Safety and Essential Performance |
| IEC 60601-1-2:2007<br>3rd Edition | Medical Electrical Equipment - Part 1-2: General<br>requirements for basic safety and essential performance –<br>Collateral standard: Electromagnetic compatibility –<br>Requirements and tests |
| IEC 60601-2-22:2007<br>3rd Edition | Medical Electrical Equipment - Part 2-22: Particular<br>requirements for basic safety and essential performance<br>of surgical, cosmetic, therapeutic and diagnostic laser<br>equipment |
The iFS Laser System underwent medical electrical equipment testing and was in compliance with applicable safety standards as listed in the following table.
Laboratory verification bench studies evaluated the channel widths, depths, and cut quality using glass microscope slides, Agarose gel, and porcine eyes, respectively. The acceptance criteria for depth were identical to that previously used to establish the acceptable performance of the currently available resection patterns. These studies demonstrated the acceptable performance of the iFS Laser System in creating corneal channel resection patterns associated with the proposed indication for inlay channels.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.