Aqua Naina Sterile Saline Solution

K193441 · Chemtex USA, Inc. · LPN · May 26, 2020 · Ophthalmic

Device Facts

Record IDK193441
Device NameAqua Naina Sterile Saline Solution
ApplicantChemtex USA, Inc.
Product CodeLPN · Ophthalmic
Decision DateMay 26, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5928
Device ClassClass 2

Indications for Use

Aqua Naina Sterile Saline Solution is indicated only for rinsing soft contact lenses after cleaning and disinfection before use.

Device Story

Aqua Naina Sterile Saline Solution is a sterile, preservative-free, buffered, isotonic aqueous solution containing sodium chloride, boric acid, and sodium borate in purified water USP. It is intended for use by contact lens wearers to rinse soft contact lenses after cleaning and disinfection, prior to insertion. The device is supplied in a 4 fl oz (118ml) high-density polyethylene (HDPE) container with a tamper-evident seal. It functions as a chemical rinse to remove residual cleaning/disinfection agents from the lens surface. The solution is pH-balanced (6.30–7.80) and isotonic (280–320 mOsmol/kg) to ensure compatibility with the ocular environment and lens material. It is an over-the-counter product used by the patient at home.

Clinical Evidence

Bench testing only. Testing included sterility (USP <71>), bacteriostasis (USP <71>), ocular irritation, and cytotoxicity (Direct contact and MEM Elution tests). Testing was conducted in accordance with ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010. Results demonstrated the solution is non-toxic and comparable to currently marketed soft contact lens solutions.

Technological Characteristics

Preservative-free, buffered, isotonic aqueous solution (NaCl, boric acid, sodium borate in purified water USP). pH 6.30–7.80; Osmolality 280–320 mOsmol/kg. Packaging: 118ml HDPE container. Biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10.

Indications for Use

Indicated for rinsing soft contact lenses after cleaning and disinfection before use. For use by soft (hydrophilic) contact lens wearers.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text. May 26, 2020 Chemtex USA, Inc. % Thomas Padula VP Regulatory Compliance Schiff & Company, Inc. 583 Mountain Avenue North Caldwell, NJ 07006 Re: K193441 Trade/Device Name: Aqua Naina Sterile Saline Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN Dated: April 22, 2020 Received: April 22, 2020 Dear Thomas Padula: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # J Angelo Green J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K193441 Device Name Aqua Naina Sterile Saline Solution Indications for Use (Describe) Aqua Naina Sterile Saline Solution is indicated only for rinsing soft contact lenses after cleaning and disinfection before use. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(K) PREMARKET NOTIFICATION FOR AQUA NAINA STERILE SALINE SOLUTION CHEMTEX USA INC. #### 510(k) Summary (as required by 807.92) #### (1) SUBMITTER: CHEMTEX USA, INC. 27-29 DWIGHT PLACE FAIRFIELD, NJ 07004-3303 Contact person: HARIBABU TALASILA, PRESIDENT Telephone: 973-447-4373 Email: htalasila@chemtexusa.com Contact Person: Mr. Hemant Shastri Telephone No: 905 -212-9949 Fax No: 905 -212-7776 Email: hshastri@chemtexinternational.com Date prepared: December 9, 2019 (2) DEVICE NAME: Trade Name: Aqua Naina Sterile Saline Solution Common Name: Contact Lens Saline Solution Classification Name: Soft contact lens care products Device Classification: Class II Regulation Number: 21 CFR 886.5928 Product Code: LPN - (3) PREDICATE DEVICE: Substantial equivalence is based on following legally marketed devices. Eye-Cept Sterile Saline Solution K110221 (Primary Predicate) Purilens Saline Solution - K002319 Unisol Saline Solution - P790011 #### (4) DESCRIPTION OF THE DEVICE: Aqua Naina Sterile Saline Solution is a sterile, preservative-free, buffered, isotonic, clear colorless aqueous solution containing isotonic solution of sodium chloride, boric acid and sodium borate in purified water USP. Aqua Naina Sterile Saline Solution has pH of 6.30 to 7.80 and Osmolality of 280 to 320 mOsmol/kg The Sterile Saline Solution is packed in 4 fl oz (118ml) high density polyethylene (HDPE) container with a tamper evident seal. - (5) INDICATION FOR USE: Aqua Naina Sterile Saline Solution is indicated only for rinsing soft contact lenses after cleaning and disinfection before use. {4}------------------------------------------------ ## 510(K) PREMARKET NOTIFICATION FOR AQUA NAINA STERILE SALINE SOLUTION CHEMTEX USA INC. - (6) COMPARISON WITH PREDICATE DEVICES: Following table is a comparison of Aqua Naina Sterile Saline Solution and predicate devices. Aqua Naina Sterile Saline Solution is substantially equivalent in terms of it actions and indications for use, to Purilens Saline solution -K002319, Unisol Saline Solution P790011, Eye-Cept Saline Solution K110221 (Primary Predicate), cleared for marketing under 510(K). Aqua Naina Sterile Saline Solution meets the guideline set forth in FDA's May 1,1997 Guidance for Industry, Premarket Notification 510(K) Guidance Document for Contact Lens Care Products. | Equivalency | Aqua Naina<br>Sterile Saline<br>Solution | Purilens Saline<br>Solution K002319 | Unisol<br>P790011 | Eye-Cept Saline<br>Solution K110221<br>(Primary<br>Predicate) | |--------------------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------|----------------------------------------------------------------------------|---------------------------------------------------------------| | Manufacturer | Niagara<br>Pharmaceuticals,<br>Inc. | Purilens Inc | Alcon | Optics Laboratory | | Product Code | LPN | LPN | LPN | LPN | | Regulation<br>Number | 21 CFR 886.5928 | 21 CFR 886.5928 | 21 CFR 886.5928 | 21 CFR 886.5928 | | Intended use | Only for rinsing | For cleaning | For heat | For rinsing soft | | (Indications for | soft contact | disinfection and | disinfecting | (hydrophilic) | | use) | lenses after<br>cleaning and for<br>wetting soft<br>contact lenses<br>after disinfection<br>before use. | storing of contact<br>lenses | rinsing, storage and<br>wetting of soft<br>(hydrophilic)<br>contact lenses | contact lens after<br>cleaning and<br>wetting | | Preservative free | Yes | Yes | Yes | Yes | | Sterile | Yes | Yes | Yes | Yes | | Aqueous Solution<br>of NaCl and<br>Borates | Yes | Yes | Yes | Yes | | pH balanced | Yes | Yes | Yes | Yes | | Volume | 4oz (118 ml) | 4oz (118 ml) | 4oz (118 ml) | 0.34oz (10ml) | #### (7) PERFORMANCE STSTANDARDS APPLIED: A Series of Studies were completed to demonstrate the substantial equivalence of Aqua Naina Sterile Saline Solution to the predicate devices. All testing was conducted in accordance with and in conformance to applicable device regulations and guidance. Results of all testing demonstrate the solution is non-toxic, is comparable to other currently marketed soft contact lens solution and is substantially equivalent to legally marketed predicates. These studies followed the "Testing Matrix for {5}------------------------------------------------ ## 510(K) PREMARKET NOTIFICATION FOR AQUA NAINA STERILE SALINE SOLUTION CHEMTEX USA INC. Saline Solution" of the Guidance for Industry-Premarket Notification (510K) Guidance Document for Contact Lens Care Products, issued May 1, 1997" and included: - 1. Sterility USP<71> - 2. Bacteriostasis USP<71> - Ocular irritation 3. - Cytotoxicity -Direct contact test 4. - ನ. Cytotoxicity- MEM Elution test | Standard | Title of Standard | Applied Section | |----------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------| | ISO 10993-1:2009 | Biological Evaluation of Medical Devices - Part 1:<br>Evaluation and Testing within a Risk Management Process | Section 15.<br>Biocompatibility | | ISO 10993-5:2009 | Biological Evaluation of Medical Devices - Part 5: Tests for<br>In Vitro Cytotoxicity | | | ISO<br>10993-10:2010 | Biological Evaluation of Medical Devices - Part 10: Tests for<br>Irritation and Skin Sensitization | | - (8) CONCLUSION: Aqua Naina Sterile Saline Solution has the same intended use and technology characteristics as the predicate devices. Aqua Naina Sterile Saline Solution is as safe, as effective, and performs as well as the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...