LensGiene Sterile Saline Solution

K161622 · Optics Laboratory, Inc. · LPN · Aug 11, 2016 · Ophthalmic

Device Facts

Record IDK161622
Device NameLensGiene Sterile Saline Solution
ApplicantOptics Laboratory, Inc.
Product CodeLPN · Ophthalmic
Decision DateAug 11, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5928
Device ClassClass 2

Indications for Use

The LensGiene™ Sterile Saline Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The LensGiene™ Sterile Saline Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the cav to rinse contact lenses.

Device Story

LensGiene™ Sterile Saline Solution is a preservative-free, aqueous solution of sodium chloride, boric acid, and sodium borate. Provided in single-use 10 ml plastic containers. Used by contact lens wearers to rinse debris and cleaning solution from soft, rigid gas permeable, or hard lenses prior to insertion. Also functions as an insertion solution for scleral lenses and for rinsing lens cases. The solution is applied topically to the lens; it provides a natural environment for the cornea compared to preserved multipurpose solutions. The device is intended for over-the-counter use.

Clinical Evidence

No clinical data. Clinical studies were deemed unnecessary as the solution was previously cleared for rinsing soft contact lenses under K110221.

Technological Characteristics

Preservative-free aqueous solution; ingredients: sodium chloride, boric acid, sodium borate. pH: 7.2-7.8; tonicity: 290-320 mOsm/kg. Form factor: 10 ml single-use plastic resin container with twist-off cap. Sterility: Sterile.

Indications for Use

Indicated for rinsing soft (hydrophilic), rigid gas permeable, and hard contact lenses following disinfection; also for use as an insertion solution for scleral lenses and for rinsing contact lens cases.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping heads or faces. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 11, 2016 Optics Laboratory, Inc. % Mr. Paul Kramsky President Rockin' Regulatory, Inc. 21831 Tumbleweed Circle Lake Forest, CA 92630 Re: K161622 Trade/Device Name: LensGiene™ Sterile Saline Solution Regulation Number: 21 CFR 886.5918 Regulation Name: Rigid Gas Permeable Contact Lens Care Products Regulatory Class: Class II Product Code: MRC, LPN Dated: June 10, 2016 Received: June 13, 2016 Dear Mr. Kramsky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {2}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Kesia Alexander for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161622 Device Name LensGiene(TM) Sterile Saline Solution #### Indications for Use (Describe) The LensGiene™ Sterile Saline Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The LensGiene™ Sterile Saline Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the cav to rinse contact lenses. | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------|--|--| | | | | | | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### 510(k) Summary for the LensGiene™ Sterile Saline Solution ### "Submission Sponsor: Optics Laboratory, Inc. 9480 Telstar Ave Ste 3 El Monte, CA 91731-2988 Telephone: 626-350-1926 Fax: 626-350-1906 FDA Establishment Registration #: 3001918419 # Submission Correspondent: Rockin' Regulatory, Inc. 21831 Tumbleweed Circle Lake Forest, CA 92630 Telephone: 949-636-1464 Contact: Paul Kramsky, President Email: pkramsky@cox.net # Date Prepared: June 10, 2016 ### Device Name: | Trade/Proprietary Name: | LensGiene™ Sterile Saline Solution | |-------------------------|------------------------------------------------------------------| | Classification Name: | Contact Lens Care Products | | Product Code: | MRC and LPN | | Regulation Number: | 21 CFR 886.5918 (Rigid Gas Permeable Contact Lens Care Products) | | | 21 CFR 886.5928 (Soft (hydrophilic) Contact Lens Care Products) | | Device Classification: | II | | Review Panel: | Ophthalmic | ### Predicate Devices: The LensGiene™ Sterile Saline Solution is substantially equivalent to the Menicon Co., Ltd. Saline Rinse Solution (K151768) and the Optics Laboratory, Inc. EyeCept™ Sterile Saline Solution (K110221). # Device Description: The LensGiene™ Sterile Saline Solution is a sterile, preservative-free, aqueous solution of sodium chloride, boric acid, and sodium borate provided in single does, carry-on size 10 ml bottles. The sterile saline solution has a pH 7.2-7.8 and a tonicity of 290-320 mOsm/kg. It rinses loose debris and cleaning solution off soft (hydrophilic), rigid gas permeable (RGP) and hard contact lenses prior to insertion. The rinsing solution removes {5}------------------------------------------------ ### Optics Laboratory, Inc. LensGiene™ Sterile Saline Solution Traditional 510(k) Premarket Notification debris and bacteria following proper disinfection as recommended by the eye care practitioner. The sterile saline solution can be used to rinse contact lens cases, rinse lenses as needed throughout the day and to fill the concave posterior surface of scleral lenses prior to insertion to provide a more natural environment than currently approved multipurpose solutions which contain preservatives and osmolarity agents that are not designed to be held against the cornea for extended periods of time. The LensGiene™ Sterile Saline Solution is a clear solution and should not be used if cloudy or discolored. The individual 10 ml bottle should be discarded after opening and use and should not be used after its labeled expiry date. ### Indications for Use: The LensGiene™ Sterile Saline Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The LensGiene™ Sterile Saline Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses. # Performance Data: Non-Clinical Data Non-clinical tests were unnecessary for this application because this testing was previously submitted for this solution under cleared 510(k) K110221. # Clinical Data Clinical studies were unnecessary for this application because this solution was previously cleared for rinsing soft (hydrophilic) contact lenses under K110221. # Conclusion Based upon the composition of the saline solution and previous test data submitted in K110221, the LensGiene™ Sterile Saline Solution is as safe, as effective and performs as well as the predicate devices. A comparison of the LensGiene™ Sterile Saline Solution and the predicate devices is presented in Table 1. # Substantial Equivalence: The claim of substantial equivalence to the previously cleared Menicon Co., Ltd. Saline Rinse Solution (K151768) and the Optics Laboratory, Inc. EyeCept™ Sterile Saline Solution (K110221) is supported by the following Comparison of Characteristics in Table 1. The LensGiene™ Sterile Saline Solution and the predicates are similar in composition and intended use. Therefore, the LensGiene™ Sterile Saline Solution is substantially equivalent to the predicate devices. {6}------------------------------------------------ | Table 1 Comparison of Characteristics | | | | |---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------| | | LensGiene™ Sterile Saline Solution | Menicon Saline Rinse Solution | EyeCept™ Sterile Saline Solution | | Device Name | Contact Lens Care Product | SAME | SAME | | Trade Name | LensGiene™ Sterile Saline Solution | Menicon Saline Rinse Solution | EyeCept™ Sterile Saline Solution | | Document Number | | K151768 | K110221 | | Classification | Ophthalmic | SAME | SAME | | Product Code | MRC and LPN | SAME | LPN | | Regulation Number | 21 CFR 886.5918<br>21 CFR 886.5928 | SAME | 21 CFR 886.5928 | | Class | II | II | II | | Intended Use | For use following proper lens<br>disinfection as recommended<br>by the eye care practitioner.<br>The LensGiene™ Sterile<br>Saline Solution is for rinsing<br>soft (hydrophilic), rigid gas<br>permeable and hard contact<br>lenses prior to lens insertion.<br>This solution may also be used<br>as an insertion solution for<br>large diameter (scleral) contact<br>lenses, as a rinse for contact<br>lens cases, and may be used as<br>needed throughout the day to<br>rinse contact lenses. | SAME | For rinsing soft<br>contact lenses after<br>cleaning and for<br>wetting soft contact<br>lenses after<br>disinfection before<br>use. | | Volume | 10 ml | 5 ml | 10 ml | | Preservative Free | Yes | SAME | SAME | | Container Usage | Single use | SAME | SAME | | Sterility | Sterile | SAME | SAME | | Materials | Plastic resin container with twist<br>off cap | SAME | SAME | | Biocompatibility | Evaluated in accordance with<br>FDA Guidance for Contact Lens<br>Care Products | SAME | SAME |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...