VISTAKON (NARAFILCON B) CONTACT LENS VISIBILITY TINTED WITH UV BLOCKER

K100349 · Johnson & Johnson Vision Care, Inc. · LPL · May 21, 2010 · Ophthalmic

Device Facts

Record IDK100349
Device NameVISTAKON (NARAFILCON B) CONTACT LENS VISIBILITY TINTED WITH UV BLOCKER
ApplicantJohnson & Johnson Vision Care, Inc.
Product CodeLPL · Ophthalmic
Decision DateMay 21, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Intended Use

The VISTAKON® (narafilcon B) Soft Contact Lens (spherical) is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism. The VISTAKON® (narafilcon B) Multifocal Soft Contact Lens is indicated for daily wear single use only for the optical correction of distance and near vision in presbyopic, phakic or aphakic persons with non-diseased eyes who may have 4.00D of ADD power or less and 0.75D of astigmatism or less. The VISTAKON® (narafilcon B) Toric Soft Contact Lens is indicated for daily wear single use only for the optical correction of visual acuity in phakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00D or less of astigmatism. The VISTAKON® (narafilcon B) Multifocal-Toric Soft Contact Lens is indicated for daily wear single use only for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eves who may have 4.00D of ADD power or less and 10.00D of astigmatism or less. VISTAKON® (narafilcon B) Contact Lenses contain a UV Blocker to help protect against transmission of harmful UV radiation to the cornea and into the eye. The Eye Care Professional should prescribe the lenses for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required.

Device Story

VISTAKON® (narafilcon B) contact lens; silicone hydrogel material with internal wetting agent; spherical, multifocal, toric, and toric-multifocal designs. Features blue visibility tint and benzotriazole UV-absorbing monomer. Supplied sterile in buffered saline with methyl ether cellulose. Prescribed by eye care professionals for daily wear, single-use; discarded after removal. Provides refractive correction for phakic/aphakic patients; UV blocker reduces transmission of UV radiation to cornea/eye.

Clinical Evidence

One-month clinical study, 48 subjects, 1:1 ratio (test vs. 1-DAY ACUVUE® control). Endpoints: slit lamp findings, symptoms, visual acuity, wear time, adverse reactions, keratometry, and discontinuation rates. Results showed similar performance in vision and health compared to concurrent controls.

Technological Characteristics

Silicone hydrogel (narafilcon B); 52% polymer, 48% water. Contains internal wetting agent, Reactive Blue Dye #4, and benzotriazole UV blocker. Transmittance: <1% UVB (280-315nm), <10% UVA (316-380nm). Molded construction; moist heat sterilization. Hemispherical/hemitoric shell.

Indications for Use

Indicated for daily wear, single-use optical correction of myopia, hyperopia, and presbyopia in phakic or aphakic persons with non-diseased eyes. Supports astigmatism correction up to 10.00D and ADD power up to 4.00D depending on lens design.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K100349 MAY 2 ] 2010 # 510(k) Summary . | Submitter<br>Information | Company: | Johnson & Johnson Vision Care, Inc.<br>7500 Centurion Parkway<br>Suite 100<br>Jacksonville, FL 32256 | | |---------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--| | | Contact Person: | Catherine Dillon | | | | Email: | cdillon9@its.jnj.com | | | | Telephone: | 904-443-3180 | | | | FAX: | 904-443-1424 | | | | Date Prepared: | February 8, 2010 | | | Identification of<br>the Device | Common Name: | Soft Contact Lens | | | | Device Name: | VISTAKON® (narafilcon B) Contact Lens | | | | Classification Name: | Soft Hydrophilic Contact Lens, Daily Wear | | | | Device Classification: | Class II, 21 CFR 886.5925 (b) (1). | | | Predicate<br>Device(s) | Material<br>VISTAKON® (narafilcon A) Contact Lens – K073485<br>(FDA Group I; low water, nonionic polymer) | | | | | Indication, Wear Schedule<br>VISTAKON® (etafilcon A) Contact Lenses – K962804<br>(Daily wear, single use) | | | . {1}------------------------------------------------ | Description of Device | The VISTAKON® (narafilcon B) Contact Lens Visibility Tinted with UV Blocker is available as a spherical lens, multifocal lens, toric lens, and toric-multifocal lens. The lenses are made of a silicone hydrogel material containing an internal wetting agent. The VISTAKON® (narafilcon B) Contact Lens is tinted blue using Reactive Blue Dye #4 to make the lens more visible for handling. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 1% in the UVB range of 280-315nm and less than 10% in the UVA range of 316-380nm. The VISTAKON® (narafilcon B) Contact Lens is a hemispherical or hemitoric shell. The lens is supplied in a sterile state, packaged in a buffered saline solution with methyl ether cellulose. The composition of the lens is 52% narafilcon B and 48% water by weight when hydrated and stored in the buffered saline solution. | |-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| Continued on next page . {2}------------------------------------------------ ・ Ind | Indications for<br>Use | Lens Design | Indication | |------------------------|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Spherical | The VISTAKON® (narafilcon B) Contact Lens is<br>indicated for daily wear single use only for the optical<br>correction of refractive ametropia (myopia and<br>hyperopia) in phakic or aphakic persons with non-<br>diseased eyes who may have 1.00D or less of<br>astigmatism. | | | Multifocal | The VISTAKON® (narafilcon B) Contact Lens is<br>indicated for daily wear single use only for the optical<br>correction of distance and near vision in presbyopic,<br>phakic or aphakic persons with non-diseased eyes who<br>may have 4.00D of ADD power or less and 0.75D of<br>astigmatism or less. | | | Toric | The VISTAKON® (narafilcon B) Contact Lens is<br>indicated for daily wear single use only for the optical<br>correction of visual acuity in phakic or aphakic persons<br>with non-diseased eyes that are hyperopic or myopic<br>and may have 4.00D of ADD power or less and 10.00D<br>or less of astigmatism. | | | Multifocal<br>Toric | The VISTAKON® (narafilcon B) Contact Lens is<br>indicated for daily wear single use only for the optical<br>correction of distance and near vision in presbyopic<br>phakic or aphakic persons with non-diseased eyes who<br>may have 4.00D of ADD power or less and 10.00D of<br>astigmatism or less. | - VISTAKON® (narafilcon B) Contact Lenses contain a UV . Blocker to help protect against transmission of harmful UV . radiation to the cornea and into the eye. - . The Eye Care Professional should prescribe the lenses for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required. {3}------------------------------------------------ #### Technological Characteristics The technological characteristics of the VISTAKON® (narafilcon B) Contact Lenses are compared to the characteristics of the predicate device, VISTAKON® (narafilcon A) Contact Lens, in the following tables. | | Predicate Device | Subject Device | |----------------------|---------------------------------------------|---------------------------------------------| | Product Name | VISTAKON®<br>(narafilcon A)<br>Contact Lens | VISTAKON®<br>(narafilcon B)<br>Contact Lens | | Material USAN Name | narafilcon A | narafilcon B | | 510(k) Number | K073485 | TBD - This submission | | FDA Category (Group) | Group I | Group I | | Manufacturing Method | Molded | Molded | | Sterilization | Moist Heat | Moist Heat | | Packaging | Blister | Blister | | Visibility Tint | Blue | Blue | | Parameter Comparison | | | | | |-------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------|----------|---------| | | Predicate Device<br>VISTAKON®<br>(narafilcon A) Contact<br>Lens-K073485 | Subject Device<br>VISTAKON®<br>(narafilcon B)<br>Contact Lens | | | | | Measured | Labeled | Measured | Labeled | | Water Content, % | 47 | 46 | 48 | 48 | | Refractive Index @ 20°C | 1.40 | 1.41 | 1.41 | 1.41 | | Dk, edge corrected | 96 | 100 | 52 | 55 | | Base Curve, mm | 8.53 | 8.5 | 8.49 | 8.5 | | Diameter, mm | 14.26 | 14.2 | 14.19 | 14.2 | | Power, D | -0.92 | -1.00 | -0.98 | -1.00 | Polarographic Method, Dk units: 10 -11(cm+/sec)(ml O2/ml x mmHg) {4}------------------------------------------------ | Non-clinical<br>Testing | A series of in-vitro and in-vivo preclinical toxicology and<br>biocompatibility tests were performed to assess the safety and<br>effectiveness of the contact lens. All tests were conducted in<br>accordance with the GLP regulation (21 CFR Part 56) or according to<br>valid scientific protocols. | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The results of the non-clinical testing/evaluation demonstrate that: | | | the lens material and/or extracts are non-toxic, non-irritating and<br>non-sensitizing under the experimental conditions; and the lens physical and material properties are consistent with<br>currently marketed lenses. | | Clinical Testing | A one-month clinical study was completed to evaluate the safety and<br>efficacy of the VISTAKON® (narafilcon B) Contact Lens for single.<br>use daily wear only. | | | The study evaluated 48 subjects with a 1:1 ratio between the test lens<br>and the control lens (1·DAY ACUVUE® Brand Contact Lenses). The<br>primary endpoints were slit lamp findings, symptoms, problems and<br>complaints, visual acuity and average wear time. Additional | parameters measured included adverse reactions, keratometry changes, reasons for discontinuation, and the number and reasons for unscheduled lens replacements. The clinical evaluation demonstrated similar overall performance in the clinically relevant areas of vision and health as compared to concurrent controls when used under daily wear single use conditions. {5}------------------------------------------------ #### Conclusions Drawn from Studies | Validity of<br>Scientific Data | A contract laboratory under Good Laboratory<br>Practice Regulations conducted toxicology studies.<br>Microbiology, chemistry, shelf-life stability, and<br>leachability studies were conducted by VISTAKON®<br>laboratories and followed scientific protocols. The<br>data were determined to be scientifically valid under<br>21 CFR 860.7. | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Substantial<br>Equivalence | Information presented in this Premarket Notification<br>establishes that the VISTAKON® (narafilcon B)<br>Contact Lens is as safe and effective as the predicate<br>device when used in accordance with the labeled<br>directions for use and for the requested indication. | | Risk and<br>Benefits | The risks of the subject device are the same as those<br>normally attributed to the wearing of soft<br>(hydrophilic) contact lenses on a daily wear single<br>use basis. The benefits to the patient are the same as<br>those for other soft (hydrophilic) contact lenses. | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem featuring a stylized eagle or bird-like figure with its wings spread, positioned above a wavy line. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-0609 Silver Spring, MD 20993-0002 Johnson & Johnson Vision Care, Inc. C/O Catherine C. Dillon Regulatory Affairs 7500 Centurion Parkway, Suite 100 Jacksonville, FL 32256 Re: K100349 Trade/Device Name: VISTAKON® (narafilcon B) Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: May 4, 2010 Received: May 5, 2010 Dear Ms. Dillon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. MAY 2 1 2010 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {7}------------------------------------------------ Page 2 – Ms. Catherine C. Dillon forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 70 http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kesia Alexander Image /page/7/Picture/8 description: The image shows a handwritten word that appears to be "for". The script is cursive and the letters are connected. The "f" has a prominent loop extending below the baseline, and the "o" and "r" are joined together at the top. Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ### Indications for Use Statement 510(k) Number (if known): Device Name: VISTAKON® (narafilcon B) Contact Lens Visibility Tinted, with UV Blocker Indications for Use: The VISTAKON® (narafilcon B) Soft Contact Lens (spherical) is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism. 100349 The VISTAKON® (narafilcon B) Multifocal Soft Contact Lens is indicated for daily wear single use only for the optical correction of distance and near vision in presbyopic, phakic or aphakic persons with non-diseased eyes who may have 4.00D of ADD power or less and 0.75D of astigmatism or less. The VISTAKON® (narafilcon B) Toric Soft Contact Lens is indicated for daily wear single use only for the optical correction of visual acuity in phakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00D or less of astigmatism. The VISTAKON® (narafilcon B) Multifocal-Toric Soft Contact Lens is indicated for daily wear single use only for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eves who may have 4.00D of ADD power or less and 10.00D of astigmatism or less. VISTAKON® (narafilcon B) Contact Lenses contain a UV Blocker to help protect against transmission of harmful UV radiation to the cornea and into the eye. The Eye Care Professional should prescribe the lenses for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH. Office of Device Evaluation (ODE) Kaur Warburton vision Sign-Off Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K100349
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...