Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses
K211603 · Pegavision Corporation · LPL · Feb 2, 2022 · Ophthalmic
Device Facts
| Record ID | K211603 |
| Device Name | Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses |
| Applicant | Pegavision Corporation |
| Product Code | LPL · Ophthalmic |
| Decision Date | Feb 2, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 886.5925 |
| Device Class | Class 2 |
Intended Use
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Sphere designs are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Toric designs are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and astigmatic corrections from -0.25 to -3.50 diopters. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Multifocal designs are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and with non-diseased eyes who may require a reading addition form +0.25 to +3.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Multifocal-Toric designs are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters who may need up to +3.00 diopters of ADD power and have -3.50 diopters of astigmatism or less. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses are intended for frequent/planned replacement wear with cleaning, disinfecting and scheduled replacement as prescribed by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
Device Story
Ionic hydrogel contact lenses; Etafilcon A material (42% polymer, 58% water); cast-molded; visibility-tinted with Reactive Blue 19; contains benzotriazole UV absorber. Available in sphere, asphere, toric, multifocal, and multifocal-toric designs. Supplied in sterile blister packs with borate buffered saline, Tween 80, Sodium Hyaluronate, and Polyethylene Glycol. Prescribed by eye care professionals for daily wear; daily disposable or frequent/planned replacement (with chemical disinfection). Corrects refractive errors; improves visual acuity; provides UV protection.
Clinical Evidence
No clinical data. Safety and effectiveness established via non-clinical bench testing including physicochemical properties, biocompatibility (ISO 10993-5, -10, -11), extractables, shelf-life, and sterility testing. Predicate devices provide historical clinical evidence for the Etafilcon A material.
Technological Characteristics
Material: Etafilcon A (2-HEMA/MAA copolymer cross-linked with EGDMA/TMPTMA). Group IV high water ionic hydrogel. Water content: 58%. Refractive index: 1.402. Oxygen permeability: 19.73 x 10^-11 (cm²/sec)(mlO2/ml-mmHg). UV blocking: <5% UVB (280-315nm), <50% UVA (315-380nm). Production: Cast-molded. Sterilization: Sterile blister pack.
Indications for Use
Indicated for daily wear correction of ametropia (myopia/hyperopia), astigmatism, and presbyopia in aphakic/non-aphakic persons with non-diseased eyes. Power ranges: +6.00 to -12.25 diopters; astigmatism up to -3.50 diopters; reading addition +0.25 to +3.00 diopters.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
- Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses and Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses (K200296)
- ACUVUE® (etafilcon A) Soft Contact Lens for Presbyopia, Clear and Tinted (Visibility and/or Cosmetically) with UV (ultraviolet) Blocker for Daily Wear (K141670)
Related Devices
- K211448 — Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses · Pegavision Corporation · Dec 6, 2021
- K140025 — AQUAMAX (ETAFILCON A) BI-WEEKLY DISPOSABLE SOFT (HYDROPHILIC) CONTACT LENSES · Pegavision Corporation · Oct 15, 2014
- K132146 — SAVIEW-AQUA 58 UV (ETAFILCON A) VISIBILITY TINTED SOFT (HYDROPHILIC) CONTACT LENS, SAVIEW-AQUA 58 UV TORIC (ETAFILCON A) · St. shine Optical Co., Ltd. · Aug 30, 2013
- K030167 — MIERU (ETAFILCON A) DAILY WEAR CONTACT LENS CLEAR AND VISIBILITY TINT, WITH UV BLOCKER · Clearlab Pte , Ltd. · Apr 2, 2003
- K200296 — Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses, Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses · Pegavision Corporation · Oct 15, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
February 2, 2022
Pegavision Corporation Estela Lin Regulatory Affair Senior Engineer 2F-1, No. 5 Shing Yeh St Taoyuan, Guishan Dist. 33341 Taiwan
Re: K211603
Trade/Device Name: Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL. MVN Dated: December 23, 2021 Received: December 27, 2021
Dear Estela Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation titled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K211603
#### Device Name
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses
#### Indications for Use (Describe)
#### Sphere and Asphere
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Sphere designs are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
#### Toric
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Toric designs are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and astigmatic corrections from -0.25 to -3.50 diopters.
#### Multifocal
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Multifocal designs are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and with non-diseased eyes who may require a reading addition form +0.25 to +3.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
#### Multifocal-Toric
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Multifocal-Toric designs are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters who may need up to +3.00 diopters of ADD power and have -3.50 diopters of astigmatism or less.
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses are intended for frequent/planned replacement wear with cleaning, disinfecting and scheduled replacement as prescribed by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
The following 510(K) Summary is being submitted as required by 21CFR 807.92(a).
| Submitter Information | |
|-----------------------|---------------------------------------------------------------------------------------------|
| Company: | PEGAVISION CORPORATION<br>2F-1 No.5, Shing Yeh St., Guishan Dist., Taoyuan City 333, Taiwan |
| Contact Person: | Mr. TS Yang, President |
| Phone: | 886-3-329-8808 |
| Fax: | 886-3-329-8897 |
| E-Mail: | TSYang@pegavision.com |
| Date Prepared: | January 28, 2022 |
# Submitter Information
# Identification of Device
| Trade Name: | Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses<br>Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Soft (hydrophilic) Contact Lenses (daily wear) |
| Classification Name: | Lenses, Soft Contact, Daily Wear 21CFR. 886.5925, Product Code LPL<br>Lens. Soft Contact (Disposable). 21CFR. 886.5925, Product Code MVN |
| FDA Classification: | Class II |
| Predicate Device Name: | K200296, Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses, Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses<br>K141670, ACUVUE® (etafilcon A) Soft Contact Lens for Presbyopia, Clear and Tinted (Visibility and/or Cosmetically) with UV (ultraviolet) Blocker for Daily Wear |
# Description of Device
The Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
{5}------------------------------------------------
and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses are clear, visibility-tinted, UV absorber-containing and are available in sphere, asphere, toric, multifocal and multifocal-toric designs. It is an ionic hydrogel lens derived from Etafilcon A material which is a co-polymer of 2-Hydroxyethyl Methacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) and 1,1,1-Trimethylolpropane Trimethacrylate (TMPTMA) and made by photo-polymerization. The lens consists of 42% Etafilcon A and 58% water by weight when immersed in buffered saline solution. The lens polymer further contains a benzotriazole, a UV absorbing monomer, thus is able to block the UV radiation to the ocular. The average transmittance characteristics of theses lenses are less than 5% in the UVB range of 280-315 nm and less than 50% in the UVA range of 315-380mm. The lens is visibly tined with "Reactive Blue19" color additive, 21 CFR part 73.3121. Lenses are supplied sterile in the sealed blister package containing sterile isotonic borate buffered saline solution with Tween 80, Sodium Hyaluronate and Polyethylene Glycol.
# Indications for use
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses
### Sphere and Asphere
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Sphere and Asphere designs are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
### Toric
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Toric designs are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and astigmatic corrections from -0.25 to -3.50 diopters.
### Multifocal
{6}------------------------------------------------
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Multifocal designs are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and with non-diseased eyes who may require a reading addition form +0.25 to +3.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
### Multifocal-Toric
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses with Multifocal-Toric designs are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters who may need up to +3.00 diopters of ADD power and have -3.50 diopters of astigmatism or less.
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfecting and scheduled replacement as prescribed by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
# Summary of Clinical Study
Etafilcon A lenses have been used widely. Their safety and effectiveness have been well documented. Their safety and effectiveness can be further exemplified by two lenses cleared by FDA.
- Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses and Pegavision ● (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses, K200296 Submitted by Pegavision Corporation, Taiwan.
- ACUVUE® (etafilcon A) Soft (hydrophilic) Contact Lens for Presbyopia, Clear and Tinted (Visibility and/or Cosmetically) with UV (ultraviolet) Blocker for Daily Wear, K141670 Submitted by Johnson & Johnson Vision Care, Inc.
{7}------------------------------------------------
Clinical studies for Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses of the present device are not required for the premarket notification as the USAN name and process are the same as the above mentioned predicate devices.
# Summary of Non-clinical Study
### a) Non-Clinical Testing performed
The following tests were conducted as recommended by the Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, revised May 1994. Includes:
- -Physicochemical Properties Test
- Biocompatibility Test -
| Contact Lenses (Etafilcon A) | • Cytotoxicity Test (according to ISO 10993-5) |
|--------------------------------------|------------------------------------------------------------|
| | • Ocular Irritation Test (according to ISO 10993-10) |
| | • Acute Systemic Toxicity Test (according to ISO 10993-11) |
| Packaging Solution | • Cytotoxicity Test (according to ISO 10993-5) |
| | • Ocular Irritation Test (according to ISO 10993-10) |
| Plastic Container (Foil& PP Blister) | • Cytotoxicity Test (according to ISO 10993-5) |
| | • Ocular Irritation Test (according to ISO 10993-10) |
| | • Acute Systemic toxicity test (according to ISO 10993-11) |
- Extractable Test -
- Shelf Life Test and Sterility Test -
The non-clinical performance tests had been performed to demonstrate the safety and effectiveness of Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses, which have the same lens materials, packaging solution, and manufacturing process with predicate lenses Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses (K200296). In addition, the Multifocal and Multifocal-Toric lens designs are the same with predicate lenses ACUVUE® (etafilcon A) Soft Contact Lens for Presbyopia, Clear and Tinted (Visibility and/or Cosmetically) with UV Blocker for Daily Wear (K141670).The evidence of substantial equivalence to the predicate lenses is described below.
{8}------------------------------------------------
### b) Technological characteristics studies
The technological characteristics of Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses are illustrated in the following Table.
| | Proposed Device | K200296 Predicate | K141670 Predicate |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------|
| Production Method | Cast-Molded | Cast-Molded | Cast-Molded |
| Lens Designs | Sphere and Asphere, Toric,<br>Multifocal, Multifocal-Toric | Sphere and Asphere, Toric,<br>Multifocal | Multifocal,<br>Multifocal-Toric |
| USAN Name | Etafilcon A | Etafilcon A | Etafilcon A |
| Material<br>Classification | Group IV<br>high water ionic | Group IV<br>high water ionic | Group IV<br>high water ionic |
| Water Content (%) | 58% | 58% | 58% |
| Refractive Index | 1.402 | 1.402 | 1.402 |
| Oxygen<br>Permeability (edge<br>corrected) @ 35°C | $19.73 x10^{-11}$<br>(cm²/sec)(mlO2/ml-mmHg) | $19.73 x 10^{-11}$<br>(cm²/sec)(mlO2/ml-mmHg) | $21.4 x 10^{-11}$<br>(cm²/sec)(mlO2/ml-mmHg) |
| Cosmetically tinted | N/A | To enhance or alter the<br>appearance of the eye | To enhance or alter the<br>appearance of the eye |
| Percent<br>Transmittance<br>% T at 380~780nm<br>% T at 315-380nm<br>% T at 280-315nm | > 95%<br>< 50%<br>< 5% | > 95%<br>< 50%<br>< 5% | > 95%<br>< 30%<br>< 5% |
| Packaging Solution | Saline A: Borate buffered saline<br>Saline B: Borate buffered saline<br>(with Tween 80, Sodium<br>Hyaluronate, and Polyethylene<br>Glycol) | Borate buffered saline<br>(with Tween 80, Sodium<br>Hyaluronate, Polyethylene<br>Glycol) | Borate buffered saline |
### Substantial Equivalence Statement
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses submitted in this 510(k) file is substantially equivalent in intended use, main materials, design, safety and performance claims to the predicate devices (K200296 & K141670).
Successful results from the chemical/physical, stability, biocompatibility tests, extractable test and care solution compatibility confirm the lenses are within established finished product specification, remain stable, and are non-toxic and biocompatible with the ocular environment.
{9}------------------------------------------------
# Conclusion
A series of pre-clinical tests were performed to demonstrate the safety and effectiveness of the Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses. Information submitted in the 510(k) establishes that the proposed devices have the same materials and manufacture process with the predicate devices - Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses and Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses (K200296), and the same lens designs of Multifocal and Multifocal-Toric with the ACUVUE® (etafilcon A) Soft (hydrophilic) Contact Lens for Presbyopia, Clear and Tinted (Visibility and/or Cosmetically) with UV (ultraviolet) Blocker for Daily Wear (K141670). Therefore, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses do not raise questions of safety and effectiveness, and substantial equivalence to the predicate devices.