NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses

K150385 · Visioneering Technologies, Inc. · LPL · Apr 30, 2015 · Ophthalmic

Device Facts

Record IDK150385
Device NameNaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
ApplicantVisioneering Technologies, Inc.
Product CodeLPL · Ophthalmic
Decision DateApr 30, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Indications for Use

NaturalVue (etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear.

Device Story

NaturalVue contact lenses are daily disposable, visibility-tinted, ionic hydrogel lenses. Manufactured via cast molding using etafilcon A (copolymer of 2-HEMA and MAA) cross-linked with EGDMA and TMPTMA. Lenses incorporate a benzotriazole UV-absorbing monomer to block UV radiation. Supplied sterile in blister packs with isotonic borate buffered saline. Intended for daily wear by patients to correct refractive errors (myopia, hyperopia, astigmatism, presbyopia). Prescribed by eye care professionals. Lenses provide vision correction and UV protection. Single-use design eliminates need for cleaning/disinfection.

Clinical Evidence

Bench testing only. No new clinical studies were conducted. Safety and effectiveness were established by demonstrating equivalence to the predicate device (K140025) and leveraging historical clinical use data for etafilcon A lenses.

Technological Characteristics

Material: etafilcon A (42% copolymer, 58% water). Manufacturing: Cast-molded. UV protection: Benzotriazole monomer. Oxygen permeability: 19.3 x 10^-11 (cm²/sec)(ml O2/ml-mmHg). Refractive index: 1.4023. Sterilization: Steam sterilization (SAL 10^-9). Packaging: Sealed blister with isotonic borate buffered saline.

Indications for Use

Indicated for daily wear for correction of ametropia (myopia, hyperopia, astigmatism) and presbyopia in aphakic and non-aphakic persons with non-diseased eyes. Powers range from -20.00 to +20.00 diopters; astigmatic correction up to -10.00 diopters; reading addition up to +3.00D.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 30, 2015 Visioneering Technologies, Inc. c/o Ms. Penny Northcutt, RAC, FRAPS, CQA Regulatory Consultant and Official Correspondent 174 Watercolor Way Suite 103-403 Santa Rosa Beach, Florida 32459 Re: K150385 Trade/Device Name: Naturalvue (etafilcon A) Daily Disposable Soft (hydrophilic) Contact Lenses NaturalVue (etafilcon A) SPHERE and ASPHERE NaturalVue (etafilcon A) Toric NaturalVue (etafilcon A) Multifocal NaturalVue (etafilcon A) Multifocal Toric Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: March 9, 2015 Received: March 10, 2015 Dear Ms. Northcutt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {1}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kesia Y. Alexander -A for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150385 #### Device Name NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses # Indications for Use (Describe) #### Spherical and Aspheric NaturalVue (etafilcon A) Sphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Toric NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. #### Multifocal Spherical NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Multifocal Toric NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510k Summary This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | DATE | February 13, 2015 | |-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | APPLICANT | Visioneering Technologies, Inc.<br>Stephen Snowdy, CEO<br>4555 Mansell Road<br>Suite 300<br>Alpharetta, GA 30022<br>770-521-4358 (office phone)<br>404-863-8892 (cell phone)<br>ssnowdy@vtivision.com (email) | | OFFICIAL<br>CORRESPONDENT | Penny Northcutt, RAC, FRAPS, CQA<br>174 Watercolor Way, Suite 103-403<br>Santa Rosa Beach, FL 32459<br>678-428-6978 (phone)<br>866-630-4082 (fax)<br>pennynorthcutt@theregsolutions.com (email) | | TRADE NAME | NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic)<br>Contact Lenses | | COMMON NAME | Soft (hydrophilic) Contact Lenses (daily wear)<br>Soft (hydrophilic) Contact Lenses (disposable) | | DEVICE<br>CLASSIFICATION<br>AND PRODUCT<br>CODE | Class II<br>Lenses, Soft Contact, Daily Wear, per 21 CFR 886.5925,<br>Product Code: LPL | | | Class II<br>Lens, Contact, (Disposable), per 21 CFR 886.5925, Product<br>Code: MVN | | PREDICATE<br>DEVICE | K140025 Pegavision Corporation, NaturalVue (etafilcon A)<br>Daily Disposable Soft (Hydrophilic) Contact Lenses | # SUBSTANTIAL EQUIVALENCE: It is Visioneering Technologies, Inc.'s conviction that the data submitted in this 510(k) provides evidence of substantial equivalence for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses with the same established safety profile and effectiveness of the predicate device K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses. {5}------------------------------------------------ # DESCRIPTION OF THE DEVICE: The NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are visibility-tinted, containing a UV blocker and are available in a spherical, aspheric, toric, multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380mm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution. # INTENDED USE/INDICATIONS FOR USE: # NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses ## Spherical and Aspheric NaturalVue (etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. ## Toric NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. ## Multifocal NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eves in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. ## Multifocal Toric NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear. {6}------------------------------------------------ # TECNOLOGICAL CHARACTERISTICS: # Historical Clinical Use etafilcon A lenses have been used widely. Their safety and effectiveness have been well established and documented. Their safety and effectiveness can be further exemplified by the following lenses, as well as many others as discussed in the 510(k), cleared by FDA: - ACUVUE (etafilcon A) Contact lens, clear and visibility tint with UV blocker, K962804 ● submitted by Vistakon USA - ACUVUE (etafilcon A) contact lens, clear and visibility tint with UV blocker, K991134 ● submitted by Vistakon USA - Aquamax (etafilcon A) Disposable Soft (Hydrophilic) Contact Lenses, K120028 . submitted by Pegavision Corporation, Taiwan - NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, . K140025 submitted by Pegavision Corporation, Taiwan NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses from the previous FDA cleared K140025 did not require clinical studies as the USAN name and manufacturing processes are the same as the above-mentioned predicate (and reference) devices. # Non-Clinical Study All tests were conducted in accordance with the May 1994 FDA Guidance Document for Daily Wear Contact Lenses. The non-clinical performance tests have been performed to demonstrate the safety and effectiveness of NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and establish substantial equivalence to the predicate NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses (K140025). The evidence of substantial equivalence to the predicate lenses is described below. # a) Material Properties Testing The material properties of NaturalVue (etafilcon A) Soft (Hydrophilic) Contact Lenses were tested as illustrated in the following table. | | Proposed Device | K140025 Predicate | |------------------------------------------------|------------------------------------------|------------------------------------------| | Production Method | Cast-Molded | Cast-Molded | | USAN Name | etafilcon A | etafilcon A | | Material Classification | Group 4 high water ionic | Group 4 high water ionic | | Water Content (%) | 58% | 58% | | Refractive Index | 1.4023 | 1.4023 | | Oxygen Permeability<br>(edge corrected) @ 35°C | 19.3 x 10-11<br>(cm²/sec)(ml O2/ml-mmHg) | 19.3 x 10-11<br>(cm²/sec)(ml O2/ml-mmHg) | | Percent Transmittance | | | | % T at 593nm | > 95% | > 95% | | % T at 380-315nm | < 30% | < 30% | | % T at 315-280nm | < 5% | < 5% | # b) Biocompatibility The standard cytotoxicity, maximization, and ocular irritation tests were carried out for the predicate K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and therefore apply to NaturalVue (etafilcon A) Daily Disposable Soft {7}------------------------------------------------ (Hydrophilic) Contact Lenses. Negative responses were recorded for all tests. The validity of blister packaging for these lenses was demonstrated by passing the standard extraction tests. # c) Microbiologv The steam sterilization process has been validated to deliver a minimum SAL of 10-9, thereby complying with the requirement of FDA Group 4. There is shelf-life stability supporting that these lenses remain sterile through the expiration date claimed for the product. # d) Bacteriostatic Validation The steam sterilizer was tested for effectiveness by measuring and demonstrating the uniformity of temperature at different locations inside the sterilizer over test period. Tested microorganisms were killed under tested conditions as compared to control. Lenses remained sterilized and there was no microbial growth for a period of 4 years tested under accelerated conditions. Seal of lens packages remained tight for a period of 4 years as demonstrated by the constant peeling strength tested under accelerated conditions. ## e) Leachability Studies were conducted to determine the leachable materials from the finished lenses. The results show that, at the levels of the detection reported, there are no leachable monomers and additive residues. # CONCLUSION: Based on the performance and manufacturing verification studies, it can be concluded that the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are equivalent to the predicate device K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses with respect to intended use, principles of operation, and technological characteristics.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...