DISCON (ETAFILCON A) SOFT (HYDROPHILIC) CONTACT LENS FOR DAILY WEAR
K051129 · Innova Vision, Inc. · LPL · Aug 12, 2005 · Ophthalmic
Device Facts
Record ID
K051129
Device Name
DISCON (ETAFILCON A) SOFT (HYDROPHILIC) CONTACT LENS FOR DAILY WEAR
Applicant
Innova Vision, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Aug 12, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear are indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopia or hyperopia and may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity. The eye care practitioner may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement. The contact lens may be disinfected using a chemical (not heat) disinfection system only.
Device Story
Discon (Etafilcon A) soft contact lens; spin-cast hydrogel copolymer (HEMA and MAA); cross-linked with TMPTMA and EGDMA via UV photo-polymerization; tinted blue with Reactive Blue Dye #19 for handling. Supplied sterile in blister packs with phosphate buffered saline; steam-sterilized. Prescribed by eye care practitioners for daily wear; requires patient-performed cleaning, rinsing, and chemical disinfection. Corrects visual acuity for myopia, hyperopia, and astigmatism. Benefits include vision correction and handling visibility.
Clinical Evidence
Clinical study of 65 human eyes over 6 months. Results: nearly 100% of participants achieved corrected vision; high satisfaction with lens wear and care. No severe complications, significant side effects, or complaints reported. Procedures remained stable.
Technological Characteristics
Material: Etafilcon A (HEMA/MAA copolymer). Water content: 56-60%. Oxygen permeability: 24x10^-11 (cm³/sec)(ml O2/ml-mmHg) @ 35°C. Refractive index: 1.407. Light transmittance: >93%. Diameter: 13.8-14.2 mm. Power: +6.00D to -12.00D. Manufacturing: Spin-casting, UV photo-polymerization. Sterilization: Steam autoclave.
Indications for Use
Indicated for non-aphakic patients with non-diseased eyes requiring correction of myopia, hyperopia, or refractive astigmatism up to 2.00 diopters. For daily wear with frequent replacement schedule.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
SuperView (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear (K033136)
ACUVUE (etafilcon A) Contact Lens Visibility tint (K962804)
Related Devices
K053191 — DISCON DAILY (ETAFILCON A) ONE-DAY SOFT (HYDROPHILIC) CONTACT LENS · Innova Vision, Inc. · May 12, 2006
K042824 — FREQUENCY 38 AND SILVER 07 (POLYMACON) SOFT (HYDROPHILIC) CONTACT LENS · CooperVision, Inc. · Nov 16, 2004
K994125 — BAUSCH & LOMB 2-WEEK (HILAFILCON B) VISIBILITY TINTED CONTACT LENSES · Bausch & Lomb, Inc. · Mar 3, 2000
K082883 — EZVIEW (POLYMACON) SOFT (HYDROPHILLIC) CONTACT LENS FOR DAILY WEAR · I-See Vision Technology, Inc. · Jun 16, 2009
K083288 — DISCON PLUS (ETAFILCON A) CONTACT LENS VISIBILITY TINT WITH UV BLOCKER · Innova Vision, Inc. · Sep 3, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
AUG 1 2 2005 51129
## 510 (K) Summary
## SUMMARY OF SAFETY AND EFFECTIVENESS FOR
## Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear Submitter Information:
| Company: | INNOVA VISION INC. |
|-----------------|---------------------------------------------|
| | No. 231-1, Wen-Te Road, Chiung-lin Village, |
| | Hsin-Chu County, Taiwan. |
| Contact Person: | Mr. Michael w. Hsu |
| | Executive Vice President |
| Phone: | 886-3-592-7299 |
| FAX: | 886-3-592-6213 |
| E-Mail: | mikehsu@innovavision.com.tw |
| Date Prepared | 15 Apr. / 2005 |
## ldentification of Device:
| Classification Name: | Soft hydrophilic contact lens, per 21 CFR. 886.5925 |
|-----------------------|------------------------------------------------------------------------|
| Trade Name: | Discon (Etafilcon A) Soft (hydrophilic) Contact Lens<br>for Daily Wear |
| Common or usual Name: | Soft (hydrophilic) Contact lens (daily wear) |
| FDA Classification: | Class II |
| Registration Number | 3003746024 |
## Predict Device:
SuperView (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear (K033136)-made from Innova Vision Inc., Taiwan. ACUVUE (etafilcon A) Contact Lens Visibility tint
(K962804) -made from Johnson & Johnson Co. USA.
#### Indications for Use:
Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopia or hyperopia and may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity. The eye care practitioner may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement. The contact lens may be disinfected using a chemical (not heat) disinfection system only. Description of Device:
{1}------------------------------------------------
Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear are available as non-spherical lenses manufactured by spin-casting method. The model illuminated with high water content (58 %). The hydrogel lens' material is a random copolymer composed of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which was cross-linked with 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) and Ethylene Glycol Dimethacrylate (EGDMA) via UV photo- polymerization. The Discon Contact lens with visible tint is tinted blue using Reactive Blue Dye #19 to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing sterile isotonic phosphate buffered saline. The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam-sterilized in a validated autoclave.
### Summary of Clinical Study:
The Discon Lenses were tested in 65 human eves separately within 6 months. Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses. In general, these products are good and safe for customers.
Discon lenses have been wide-used around the world, including Taiwan, China, Europe, etc. Among the users being daily worn the Discon lenses, all the procedures were in generally stable condition without severe complication. There are no significant side effects and complaints to be observed.
### Nonclinical Studies:
A series of nonclinical performance tests were performed to demonstrate the safety and effectiveness of the Discon Soft Contact lens, and establish substantial equivalence to predicate lenses-SuperView Lens (K033136): ACUVUE Lens (K962804). All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(K) Guidance Document for Class IV Contact lenses, and in conformance to applicable device regulations. The evidence of substantial equivalent to the predicate lens described as follow:
### a) Technological characteristics studies
Discon contact lens designs in the following parameter ranges:
Diameter range; 13.8 to 14.2 mm
Power range: +6.00D to -12.00D
Center thickness: varies with power (0.08 to 0.12 mm for -3.00D)
Lenses have the following properties:
Refractive index: 1.407 (hydrated)
{2}------------------------------------------------
Light transmittance: >93%
Water content: 56 to 60 %
Oxygen permeability (edged corrected) : 24x10'11 [(cm³/sec)(ml O2/ml-mmHg)] @ 35°C
There characterizations of Discon Contact lenses are equivalent and comparable to those of predicate lenses.
- b) Biocompatibilitv
In accordance with the May 1994 Guidance Document for daily wear contact lenses, toxicity studies have been conducted on the model: Discon Contact lenses. The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor irritated. The Cytotoxicity testing demonstrates the lens is not cytotoxic under the conditions of the study.
### c) Microbiology
Steam sterilization process has been validated to deliver a minimum SAL of 10°, thereby complying with the requirement of FDA Group IV. There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product.
- d) Leachability
Studies were conducted to determine the leachable materials from the finished lens. The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues.
## Substantial equivalence Statement:
Testing performed on the Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices- SuperView (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear (K033136) & ACUVUE (etafilcon A) Contact Lens visibility tint-K962804), when used in accordance with the instructions for use. It is due to the facts that The risks and benefits of the subject device are the same as soft contact lenses for to the daily wear.
In conclusion, it is Innova's conviction that data submitted in this 510(k) to validate the claim of substantial equivalency, substantiates our ability to manufacture a soft contact lens, the Discon Soft (hydrophilic) Contact Lens for Daily Wear, with the same established safety profile and effectiveness as the predicate device-SuperView (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear (K033136) & ACUVUE (etafilcon A) Contact Lens visibility tint (K962804).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings or feathers.
AUG 1 2 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Innova Vision, Inc. c/o Ms. Jennifer Reich Harverst Consulting Corp. 3892 South America West Trail Flagstaff, AZ 86001
Re: K051129
Trade/Device Name: Discon (etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: July 11, 2005 Received: July 13, 2005
Dear Ms. Reich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
David M. Whipple
Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
#### K051129 510(k) Number:
Device Name: Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear
Indications for Use:
Discon (Etafilcon A) Soft (hydrophilic) Contact Lens for Daily Wear are indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopia or hyperopia and may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The eye care practitioner may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement. TThe contact lens may be disinfected using a chemical (not heat) disinfection system only.
Prescription Use X OR
(Part 21 CFR 801 Subpart D)
Over-The-Counter (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number. Ko 51129
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.