BAUSCH & LOMB 2-WEEK (HILAFILCON B) VISIBILITY TINTED CONTACT LENSES
K994125 · Bausch & Lomb, Inc. · LPL · Mar 3, 2000 · Ophthalmic
Device Facts
Record ID
K994125
Device Name
BAUSCH & LOMB 2-WEEK (HILAFILCON B) VISIBILITY TINTED CONTACT LENSES
Applicant
Bausch & Lomb, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Mar 3, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, that exhibit refractive astigmatism up to 10.00 diopters. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D. Replacement schedules may vary from patient to patient, and should be decided by eye care practitioners in consultation with their patients. The lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lens may be disinfected using a chemical disinfection system.
Device Story
Hemispherical flexible contact lens; covers cornea and portion of sclera. Material: copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidinone (hilafilcon B); 59% water content; tinted with Reactive Blue Dye 246. Used for daily wear vision correction of refractive ametropia. Prescribed by eye care practitioners; patient cleans, rinses, and disinfects lens daily using chemical disinfection system. Lens discarded after practitioner-determined wearing period. Benefits include vision correction for myopic, hyperopic, and astigmatic patients.
Clinical Evidence
One-month randomized clinical study comparing subject lens to Softlens one day (control) in myopic, phakic patients. Primary endpoints: proportion of grade 2+ slit lamp findings (safety) and visual acuity 20/40 or better (efficacy). Results showed no significant differences (p > 0.05) between test and control, meeting safety and efficacy endpoints.
Technological Characteristics
Hydrophilic contact lens; material: hilafilcon B (copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidinone); 59% water content; refractive index 1.4036; oxygen permeability (Dk) 22 x 10^-11. Dimensions: 13.5-15.0mm diameter, 0.05-0.75mm center thickness, 7.8-9.5mm base curve. Cast molded. Compatible with chemical disinfection systems.
Indications for Use
Indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism up to 10.00D) in aphakic and non-aphakic persons with non-diseased eyes. Prescribed in spherical powers +20.00D to -20.00D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Softlens one day disposable (hilafilcon A) Visibility Tinted Contact Lens (K974780)
Phoenix (hilafilcon A) Visibility Tinted Contact Lens (K983894)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "MAR - 3 2000" on the left and "K994125" on the right. The date indicates March 3, 2000. The alphanumeric string "K994125" appears to be a code or identifier.
510(k) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS
FOR
## BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens
#### Submitter Information: 1.
Bausch & Lomb Incorporated Global Vision Care Division 1400 North Goodman Street Rochester, NY 14603-0450
Contact Person: Glenn A. Davies, O.D. Director, Regulatory Affairs Telephone No.: (716) 338-8215
#### Device Name: 2.
Classification Name: Soft (hydrophilic) contact lens
Proprietary Name: BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens
#### 3. Predicate Devices:
Soft ens™ one day disposable (hilafilcon A) Visibility Tinted Contact Lens (K974780)
Phoenix (hilafilcon A) Visibility Tinted Contact Lens (K983894)
{2}------------------------------------------------
#### DESCRIPTION OF DEVICE 4.
The BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens is a hemispherical flexible shell which covers the cornea and may cover a portion of the adjacent sclera. It consists of a copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidinone, and is 59% water by weight when immersed in a sterile saline solution. This lens is tinted blue with Reactive Blue Dye 246 ((1,4-Bis[4-(2-methacryloxyethyl) phenylamino] anthraquinone). The color additive conforms with 21 CFR Part 73.3106. The lens may also be supplied clear (no tint).
The physical / optical properties of the lens are:
| Specific Gravity: | 1.119 |
|---------------------------|--------------------------------------------------------------------------------|
| Refractive Index: | 1.4036 |
| Light Transmittance: | C.I.E. Y value - at least 97% |
| Water Content: | 59% |
| Oxygen Permeability (Dk): | 22 x 10-11 [cm3O2(STP) x cm]/(sec x cm2 x mmHg)@35°C<br>(Polarographic Method) |
The BAUSCH & LOMB 2-Week (hilafilcon B) Visibility Tinted Contact Lens is a hemispherical shell of the following dimensions:
| • Diameter: | 13.5mm to 15.0mm |
|-----------------------|--------------------|
| • Center Thickness: | 0.05mm to 0.75mm |
| • Base Curve: | 7.8mm to 9.5mm |
| • Powers (Spherical): | +20.00D to -20.00D |
| • Toric (Cylinder): | 0 to 10 diopters |
| • Toric Axis: | 0° to 180° |
Each BAUSCH & LOMB 2-Week (hilafilcon B) Visibility Tinted Contact Lens is supplied in a plastic blister container with a saline solution. The container is marked with the manufacturing lot number of the lens, the base curve, sphere power, diameter and expiration date.
{3}------------------------------------------------
# 5. INDICATIONS FOR USE
The BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, that exhibit refractive astigmatism up to 10.00 diopters. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D.
Replacement schedules may vary from patient to patient, and should be decided by eye care practitioners in consultation with their patients. The lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lens may be disinfected using a chemical disinfection system.
#### DESCRIPTION OF SAFETY AND SUBSTANTIAL EQUIVALENCE ર-
A series of non-clinical laboratory testing and clinical testing was performed to demonstrate the safety and effectiveness of the BAUSCH & LOMB 2-Week (hilafficon B) Contact Lens. A summary of results from the preclinical and clinical tests is provided below.
### Non-Clinical Laboratory Testing:
A series of in vitro and in vivo preclinical toxicology and biocompatibility testing was performed to assess the safety and effectiveness of the contact lens device. Testing was performed in accordance with FDA guideline titled Premarket Notification (510(k)) Guidance Document for Daily Wear Contact Lenses, May 1994. All non-chinical laboratory studies were conducted in compliance with the GLP regulation.
The results of the preclinical testing on the BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens demonstrate that:
The hilafilcon B contact lens material is similar to the hilafilcon A contact lens material. The differences between the two lens materials are primarily due to the difference in water content. The extracts of the lens material do not show any significant quantities of monomer components and toxicity testing results of lens extracts and lens material demonstrated no toxicity or irritation. The lenses are compatible with chemical disinfection systems (including hydrogen peroxide.
Stability testing has demonstrated a three year expiration date for the sterile lenses.
{4}------------------------------------------------
### Clinical Testing:
A one month, randomized clinical study was completed to evaluate the safety and efficacy of the Bausch & Lomb 2-Week Contact Lens (Test) compared to the Bausch & Lomb SofLens one day™ Contact Lens (Control) when worn by myopic, phakic patients on a daily wear basis with no scheduled replacements.
The primary endpoints were:
- Statistical equivalence in the proportion of total grade 2 or greater positive slit . Safety: lamp findings, between the Test and Control lenses, was considered clinically acceptable.
- Statistical equivalence in the proportion of lens visual acuities at the level of Efficacy: . 20/40 or better, between Test and Control lenses, was considered clinically acceptable.
There were no significant differences in the proportion of total grade 2 or greater slit lamp findings, for the category of Any Finding, between the Test and Control lenses (all p-values > 0.05). Therefore, the safety end-point defined in the protocol was achieved.
There were no significant differences in the proportion of lens visual acuities at the level of 20/40 or better between the Test and Control lenses (all p-values > 0.05). Therefore, the efficacy end-point defined in the protocol was achieved.
The Sponsor concludes, based on the data presented, that the protocol-specified endpoints were achieved, and that the Bausch & Lomb 2-Week Contact Lens is equivalent in safety and efficacy to the SoftLens one day™ Contact Lens, when worn on a daily wear basis. The Bausch & Lomb 2-Week Contact Lens is a safe and effective means of daily wear vision correction.
### Substantial Equivalence
The BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens is similar to the BAUSCH & LOMB® SofLens™ one day disposable (hilafilcon A) Visibility Tinted Contact Lens, in that both fall into the same FDA material classification grouping (Group II) and both are manufactured by the same manufacturing process (cast molding). The BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens is different from the BAUSCH & LOMB® SofLens™ one day disposable (hilafilcon A) Visibility Tinted Contact Lens in that it has a different USAN name suffix (hilafilcon B vs. hilafilcon A). The differences between these two lenses will not have any negative effect on the safety and effectiveness of the device.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around the perimeter. In the center of the circle is a stylized image of an eagle with three wavy lines extending from its head, representing the department's mission to protect and promote the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 3 2000
Bausch & Lomb Incorporated Glenn A. Davies, O.D. Director, Regulatory Affairs 1400 N. Goodman Street Rochester, NY 14603-0450
Re: K994125 Trade Name: Bausch & Lomb 2-Week (hilafilcon B) Visibility Tinted Contact Lens Regulatory Class: II Product Code: 86 LPL Dated: December 6, 1999 Received: December 7, 1999
Dear Dr. Davies:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{6}------------------------------------------------
Page 2 - Glenn A. Davies, O.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Roer L. Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Bausch & Lomb Incorporated 1400 North Goodman Street Rochester, NY 14692-0450
### Indications for Use Statement
510(k) Number (if known): K994125
BAUSCH & LOMB® 2-Week (hilafilcon B)Visibility Tinted Contact Lens Device Name:
Indications for Use:
The BAUSCH & LOMB® 2-Week (hilafilcon B) Visibility Tinted Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, that exhibit refractive astigmatism up to 10:00 diopters. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D.
Replacement schedules may vary from patient to patient, and should be decided by eye care practitioners in consultation with their patients. The lens is to be cleaned, rinsed and disinfected each time it is removed from the patient's eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lens may be disinfected using a chemical disinfection system.
### CRLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X | OR | Over-The-Counter-Use | |
|------------------|---|----|----------------------|--|
|------------------|---|----|----------------------|--|
(Division Sign-Off)
Division of Ophthalmic Devices
| 510(k) Number | K994125 |
|---------------|---------|
|---------------|---------|
PH.D.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.