BOSTON XO® (hexafocon A) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery). Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. Boston XO Scleral Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject: 1. cannot be adequately corrected with spectacle lenses 2. requires a rigid gas permeable contact lens surface to improve vision 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities. Common causes of corneal distortion include but are not limited to corneal infections as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The Boston XO Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error. The lenses may be disinfected using a chemical disinfection (not heat) system only.
Device Story
Fluoro silicone acrylate copolymer rigid gas permeable contact lenses; available in spherical, aspherical, toric, and scleral designs. Function as refracting medium to focus light on retina; scleral designs provide physical protection, aqueous hydrated environment, and saline bath for ocular surface disorders. Used in clinical settings by eye care professionals. Output is improved visual acuity or therapeutic protection for irregular/distorted corneas and ocular surface diseases. Benefits include vision correction and management of complex corneal conditions.
Clinical Evidence
Bench testing only. No new clinical data required as the application was for labeling revisions to existing cleared materials and designs.
Technological Characteristics
Fluoro silicone acrylate copolymer; lathe-cut; rigid gas permeable; water content <1%; oxygen permeability (Dk) 100 (XO) and 141 (XO2); refractive index 1.415-1.424; chemical disinfection only.
Indications for Use
Indicated for daily wear for correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia) in aphakic/non-aphakic persons. Also indicated for management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-penetrating keratoplasty, post-refractive surgery) and therapeutic use for ocular surface disease (Graft-versus-Host, Sjögren's, dry eye, Filamentary Keratitis, limbal stem cell deficiency, skin disorders, neurotrophic keratitis, corneal exposure). XO2 additionally indicated for orthokeratology (myopia reduction up to 5.00D).
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
K240618 — Acuity 200 (fluoroxyfocon A) Rigid Gas Permeable Contact Lens; Acuity 100 (hexafocon A) Rigid Gas Permeable Contact Lens; Acuity 200 with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens; Acuity 100 with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lens · Acuity Polymers, Inc. · Apr 3, 2024
K172314 — Ampleye Scleral RGP Lens (roflufocon D, roflufocon E, hexafocon A, paflufocon D) · Art Optical Contact Lens, Inc. · Sep 20, 2017
K241398 — SDJMASTERVISION (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses · Tianjin Mastervision Technology Co., Ltd. · Oct 22, 2024
K183175 — BostonSight Scleral · Bostonsight · Jan 4, 2019
K180988 — Acuity 18 (enflufocon A) Rigid Gas Permeable Contact Lens, Acuity 58 (enflufocon B) Rigid Gas Permeable Contact Lens, Acuity 85 (Oprifocon A) Rigid Gas Permeable Contact Lens, Acuity 100 (hexafocon A) Rigid Gas Permeable Contact Lens · Acuity Polymers, Inc. · Jun 8, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 17, 2017
BAUSCH + LOMB Incorporated % Ellen M. Beucler Vice President Foresight Regulatory Strategies, Inc. 187 Ballardvale Street, Suite A250 Wilmington, MA 01887
Re: K171404
Trade/Device Name: BOSTON XO® (hexafocon A), BOSTON XO2® (hexafocon B) Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HOD Dated: May 10, 2017 Received: May 12, 2017
Dear Ellen M. Beucler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K171404
#### Device Name
BOSTON XO® (hexafocon A) Rigid Gas Permeable Contact Lens
#### Indications for Use (Describe)
BOSTON XO® (hexafocon A) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery).
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs.
Boston XO Scleral Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:
- 1. cannot be adequately corrected with spectacle lenses
- 2. requires a rigid gas permeable contact lens surface to improve vision
- 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).
The Boston XO Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error.
The lenses may be disinfected using a chemical disinfection (not heat) system only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K171404
#### Device Name
BOSTON XO2® (hexafocon B) Rigid Gas Permeable Contact Lens
### Indications for Use (Describe)
BOSTON XO2® (hexafocon B) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery).
BOSTON XO2 Contact Lenses are also indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs.
Boston XO2 Scleral Contact Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:
- 1. cannot be adequately corrected with spectacle lenses
- 2. requires a rigid gas permeable contact lens surface to improve vision
- 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).
The Boston XO2 Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error.
The lenses may be disinfected using a chemical disinfection (not heat) system only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
## 510(k) Summary
## 1. Applicant Information
Bausch + Lomb Inc. 1400 North Goodman Street Rochester, NY 14609 USA
| Contact: | Glenn Davies, O.D.<br>Senior Director Regulatory Affairs |
|----------------|----------------------------------------------------------|
| Phone: | (585) 338-8215 |
| Fax: | (585) 338-0702 |
| email: | glenndavies@bausch.com |
| Date Prepared: | July 11, 2017 |
### 2. Device Information
| Classification name: | Rigid Gas Permeable Contact Lens |
|------------------------|--------------------------------------------------------------------------------------------|
| Device classification: | Class II |
| Regulation number: | 21 CFR 886.5916 (Rigid Gas Permeable Contact Lenses) |
| Product code: | HQD |
| Proprietary name: | Boston XO® (hexafocon A) and Boston XO2®<br>(hexafocon B) Rigid Gas Permeable Contact Lens |
## 3. Predicate Devices
Bausch + Lomb Inc. claims substantial equivalence to Boston Scleral Contact Lens, PMA No. P860022/S40 approved March 1, 1994 and BostonSight PD Prosthetic Device cleared in K161461 on July 25, 2016.
## 4. Description of Device
The Boston XO and Boston XO2 Rigid Gas Permeable (RGP) Contact Lenses are fluoro silicone acrylate copolymer rigid gas permeable contact lenses available in a variety of tints and may contain an ultraviolet light absorber. The hemispherical shells are available in several lens designs including spherical, aspherical, toric, scleral and irregular cornea designs.
{7}------------------------------------------------
## 5. Indications for Use
Boston XO (hexafocon A) and Boston XO2 (hexafocon B) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and non aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery).
The Boston XO2 (hexafocon B) Contact Lenses are also indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs.
The Boston XO and XO2 Scleral Lens designs for daily wear are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:
- 1. cannot be adequately corrected with spectacle lenses
- 2. requires a rigid gas permeable contact lens surface to improve vision
- 3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).
{8}------------------------------------------------
## 5. Indications for Use (continued)
The Boston XO and XO2 Scleral Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment.
When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Boston Scleral Lenses may concurrently provide correction of refractive error.
The lenses may be disinfected using a chemical disinfection (not heat) system only.
### 6. Performance Data
There are no changes to the Boston XO (hexafocon A) and Boston XO2 (hexafocon B) Rigid Gas Permeable contact lens material or lens designs as part of this application. These materials have been previously cleared for scleral lens designs. This application is for labeling revisions. No performance data is required.
## 7. Substantial Equivalence
In support of the expanded indications for Boston XO and XO2 to include scleral lens indications the following statements apply.
- Boston XO and XO2 Rigid Gas Permeable Scleral Contact Lenses are fluoro silicone acrylate contact lenses that use similar actions and technological characteristics as the predicate devices (see Table 1 and Table 2).
- . The Boston XO and XO2 Rigid Gas Permeable Scleral Lenses present the same risks and benefits as the predicate devices and are as safe and effective as the predicate devices when used according to the labeled directions for use and requested indications.
Therefore, the Boston XO and XO2 Rigid Gas Permeable Scleral Lenses are substantially equivalent to the BostonSight PD Prosthetic Device and the Boston Scleral Contact Lens.
{9}------------------------------------------------
| omnannen<br>3<br>villum<br>C<br>) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| |
| <br>ાદ તા<br>ﺎ<br>) |
| l<br>8 |
| I |
| 000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000<br>œ<br>uly uncil<br>1 1 6 0 0 0 0 1 1<br>H<br>1 |
| D VEUT DUWUV<br>1<br>Thetmittel<br>P<br>All I<br>C<br>2 |
| 1<br>1<br>4<br>œ |
| | Boston XO & XO2 Scleral Lens<br>(hexafocon A and hexafocon B) | BostonSight PD<br>Prosthetic Device<br>(oprifocon A) | Boston Scleral Lens<br>(oprifocon A) | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| | K171404 | K161461 | P860022/S040 | |
| | Subject Device | Predicate Device | Predicate Device | |
| Indication for Use | BOSTON XO (hexafocon A) and XO2® (hexafocon B) Contact Lenses are indicated for daily wear for the<br>correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and non<br>aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes<br>that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus,<br>pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK surgery).<br>BOSTON XO2 Contact Lenses are also indicated for daily wear in an orthokeratology fitting program for the<br>temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. Note: To maintain the<br>orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.<br>Furthermore, eyes suffering fromcertain ocular surface disorders may benefit fromthe physical protection,<br>aqueous hydrated environment and the saline bath provided by scleral lens designs.<br>Boston Xo and XO2 Scleral Contact Lens designs for daily wear are indicated for therapeutic use for the<br>management of irregular and distorted corneal surfaces where the subject:<br>1. cannot be adequately corrected with spectacle lenses<br>2. requires a rigid gas permeable contact lens surface to improve vision<br>3. is unable to wear a comeal rigid gas permeable lens due to corneal distortion or surface irregularities<br>Common causes of corneal distortion include but are not limited to comeal infections, trauma, tractions as a<br>result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or comeal<br>transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid<br>marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy,<br>granular comeal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy). The Boston XO and XO2 Scleral<br>Lens designs for daily wear are also indicated for therapeutic use in eyes with ocular surface disease (e.g.<br>ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal<br>stemcell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal bums), disorders of the<br>skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial<br>Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit fromthe presence of an<br>expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use<br>for a distorted comea or ocular surface disease, the Boston Scleral Lenses may concurrently provide<br>correction of refractive erroг.<br>lenses may be disinfected using a chemical disinfection (not heat) system only. | | | |
| Actions/Operational<br>Principles | When placed on the eye the Rigid Gas Permeable Contact Lens acts as a<br>refracting mediumto focus light rays on the retina to improve visual acuity. | Indicated for therapeutic use in eyes with<br>ocular surface disease from dry eye (e.g. ocular<br>Graft-versus-Host disease, Sjögren's<br>syndrome, dry eye syndrome), limbal stem cell<br>deficiency (e.g.Stevens-Johnson syndrome,<br>chemical and thermal burns, radiation),<br>disorders of the skin (e.g. atopy, ectodermal<br>dysplasia), neurotrophic keratitis (e.g. Herpes<br>simplex, Herpes zoster, Familial Dysautonomia),<br>and corneal exposure (e.g. anatomic, paralytic)<br>that might benefit from the presence of an<br>expanded tear reservoir and protection against<br>an adverse environment. When prescribed for<br>therapeutic use for distorted cornea or ocular<br>surface disease, the BostonSight PD Prosthetic<br>Device may incidentally provide correction of<br>refractive erroг.<br>When placed on the eye the Rigid Gas<br>Permeable Contact Lens acts as a refracting<br>medium to focus light rays on the retina to<br>improve visual acuity. | Indicated for eyes having significantly reduced<br>vision due to the presence of a distorted<br>comeal surface that:<br>1) precludes satisfactory spectacle lens<br>correction 2) demonstrates significantly<br>improved rigid contact lens corrected vision 3)<br>is incapable of wearing traditional corneal<br>lenses because of the inability to achieve<br>adequate lens centration/stability and/or<br>tolerance to physical contact with a lens.<br>Furthermore, eyes suffering from certain ocular<br>surface disorders may benefit from the physical<br>protection and the saline bath provided by a<br>scleral lens.<br>When placed on the eye the Rigid Gas<br>Permeable Contact Lens acts as a refracting<br>medium to focus light rays on the retina to<br>improve visual acuity. | |
| Product Code | HQD | HQD | HQD | |
| Common name | Contact Lens, Rigid Gas Permeable | Contact Lens, Rigid Gas Permeable | Contact Lens, Rigid Gas Permeable | |
| Device Class | II | II | II* | |
| CFR Reference | 21 CFR 886.5916 | 21 CFR 886.5916 | 21 CFR 886.5916 | |
| Production Method | Lathe Cut | Lathe cut | Lathe cut | |
| FDA Group # | Group # 3 Fluom Silicone Acrylate | Group # 3 Fluoro Silicone Acrylate | Group # 3 Fluoro Silicone Acrylate | |
| | Boston XO & XO2<br>Contact Lens<br>K171404 |…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.