PLANCON MICROLAMELLAR KERATOME (SINGLE PIECE KERATOME HEAD WITH ACCESSORIES)

K970377 · Plancon Instruments · HNO · Mar 31, 1997 · Ophthalmic

Device Facts

Record IDK970377
Device NamePLANCON MICROLAMELLAR KERATOME (SINGLE PIECE KERATOME HEAD WITH ACCESSORIES)
ApplicantPlancon Instruments
Product CodeHNO · Ophthalmic
Decision DateMar 31, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4370
Device ClassClass 1
AttributesTherapeutic

Indications for Use

To perform microlamellar keratoplasty for the correction of myopia or hyperopia

Device Story

The Plancon Instruments Microlamellar Keratome is a surgical instrument used by ophthalmologists to perform microlamellar keratoplasty. The system consists of a gas-powered turbine motor, a fixed single-piece keratome head, a keratome blade, pneumatic fixation rings, and applanation lenses. The device is powered by a vacuum pump unit that regulates gas pressure. During surgery, the pneumatic fixation ring secures the eye, and the keratome head, equipped with a blade, creates a lamellar corneal incision. The fixed-head design limits the cut depth to a single predetermined value, enhancing safety compared to adjustable-head models. The device is used in a clinical or surgical setting to facilitate refractive surgery, potentially improving patient vision by correcting myopia or hyperopia.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

System includes gas-powered turbine motor (0-20,000 RPM), stainless steel/titanium/aluminum construction, pneumatic fixation rings, and applanation lenses. Operates via gas pressure regulation. Fixed single-piece keratome head design. No gear system.

Indications for Use

Indicated for patients requiring microlamellar keratoplasty for the correction of myopia or hyperopia.

Regulatory Classification

Identification

A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant.

Predicate Devices

Submission Summary (Full Text)

{0} MAR 31 1997 K970377 # SAFETY AND EFFECTIVENESS SUMMARY K970377 p. 1/2 Plancon Instruments, 15 Rue Georges Besse, 92160 Antony, France Contact person: Mr. Alain DUPRAT, General Manager Date: December 13, 1996 Tel.: (33-1) 46 74 46 74 | | Predicated device ① 510 K Document number K 960395 | Predicated device ② 510 document number : Not known | Predicated device ③ 510 K document number : K 935342 | Submitted device | | --- | --- | --- | --- | --- | | Designation / Feature | PLANCON INSTRUMENTS Microlamellar Keratome | EYE TECHNOLOGY | SCMD | PLANCON INSTRUMENTS Microlamellar Keratome (single piece head) | | Intended use | To perform microlamellar keratoplasty for the correction of myopia or hyperopia | To perform micro lamellar keratoplasty for the correction of myopia or hyperopia | To perform micro lamellar keratoplasty for the correction of myopia or hyperopia | To perform microlamellar keratoplasty for the correction of myopia or hyperopia | | System composition | Power unit Turbine motor Adjustable keratome head Keratome blade Pneumatic fixation rings Applanation lenses Plate checking gauge | Power unit Turbine motor Adjustable keratome head Keratome blade Pneumatic fixation rings Applanation lenses Plate checking gauge | Power unit Turbine motor Adjustable keratome head Keratome blade Pneumatic fixation rings Applanation lenses Plate checking gauge | Power unit Turbine motor Fixed keratome head Keratome blade Pneumatic fixation rings Applanation lenses | | Power unit | Vacuum pump Battery Vacuum release valve Knob to regulate gas pressure Gas & vacuum pressure gauges Vacuum & turbine quick connectors Battery charger unit | Vacuum pump Battery Vacuum release valve Knob to regulate gas pressure Gas & vacuum pressure gauges Vacuum & turbine quick connectors Battery charger unit | Vacuum pump Battery Vacuum release valve Knob to regulate gas pressure Gas & vacuum pressure gauges Vacuum & turbine quick connectors Battery charger unit | Vacuum pump Battery Vacuum release valve Knob to regulate gas pressure Gas & vacuum pressure gauges Vacuum & turbine quick connectors Battery charger unit | {1} K970377 SAFETY AND EFFECTIVENESS SUMMARY K970377 p. 2/2 | | Predicated device ① 510 K Document number K 960395 | Predicated device ② 510 document number : Not known | Predicated device ③ 510 K document number : K 935342 | Submitted device | | --- | --- | --- | --- | --- | | Designation / Feature | PLANCON INSTRUMENTS Microlamellar Keratome | EYE TECHNOLOGY | SCMD | PLANCON INSTRUMENTS Microlamellar Keratome (single piece head) | | Turbine motor | Gas powered Adjustable speed 0 to 20 000 RPM (recommended 14 000 RPM) Material : stainless steel, titanium and aluminium. No gear system | Gas powered Adjustable speed Material : stainless steel No gear system | Gas powered Adjustable speed from 10 to 24 000 RPM Pre-set speed 14 000 RPM Material : stainless steel and brass No gear system | Gas powered Adjustable speed 0 to 20 000 RPM (recommended 14 000 RPM) Material : stainless steel, titanium and aluminium. No gear system | | Keratome | Material : stainless steel Adjustable Safety : maximum cut depth, and checking gauge | Material : stainless steel Adjustable Permanent adjustable head with adjustment knob Safety : maximum cut depth and checking gauge | Material : stainless steel Adjustable Safety : maximum cut depth and checking gauge | Material : stainless steel Fixed single piece head Safety : only one cut depth allowed by head principle | | Pneumatic fixation rings | Material : stainless steel Double dovetails | Material : stainless steel Double dovetails | Material : stainless steel Double dovetails | Material : stainless steel Double dovetails | | Applanation lenses | Material : clear plastic + stainless handle Engraved inner reticle diameter | Material : clear plastic + stainless handle Engraved inner reticle diameter | Material : clear plastic + stainless handle Engraved inner reticle diameter | Material : clear plastic + stainless handle Engraved inner reticle diameter |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...