LXS · Laser, Neodymium:Yag, Ophthalmic For Posterior Capsulotomy And Cutting Pupilla

Ophthalmic · 21 CFR 886.4392 · Class 2

Overview

Product CodeLXS
Device NameLaser, Neodymium:Yag, Ophthalmic For Posterior Capsulotomy And Cutting Pupilla
Regulation21 CFR 886.4392
Device ClassClass 2
Review PanelOphthalmic
3rd-Party ReviewableYes

Identification

The Nd:YAG laser for posterior capsulotomy and peripheral iridotomy consists of a mode-locked or Q-switched solid state Nd:YAG laser intended for disruption of the posterior capsule or the iris via optical breakdown. The Nd:YAG laser generates short pulse, low energy, high power, coherent optical radiation. When the laser output is combined with focusing optics, the high irradiance at the target causes tissue disruption via optical breakdown. A visible aiming system is utilized to target the invisible Nd:YAG laser radiation on or in close proximity to the target tissue.

Classification Rationale

Class II (special controls). Design Parameters: Device must emit a laser beam with the following parameters: wavelength = 1064 nanometers; spot size = 50 to 100 micros; pulse width = 3 to 30 nanoseconds; output energy per pulse = 0.5 to 15 millijoules (mJ); repetition rate = 1 to 10 pulses; and total energy = 20 to 120 mJ.

Special Controls

*Classification.* Class II (special controls). Design Parameters: Device must emit a laser beam with the following parameters: wavelength = 1064 nanometers; spot size = 50 to 100 micros; pulse width = 3 to 30 nanoseconds; output energy per pulse = 0.5 to 15 millijoules (mJ); repetition rate = 1 to 10 pulses; and total energy = 20 to 120 mJ.

Recent Cleared Devices (7 of 7)

RecordDevice NameApplicantDecision DateDecision
K023045MICRORUPTER VMeridian AGSep 26, 2002SESE
K901605PHOENIX LASER SYSTEMS OPHTHALMIC SURGICAL WORKSTA.Phoenix Laser Systems, Inc.Dec 13, 1990SESE
K895864MICRORUPTOR III SYST, Q-SWITCHED ND:YAG LASER COMPLasag AGDec 29, 1989SESE
K893987NIDEK MODEL YC-1200Nidek, Inc.Aug 18, 1989SESE
K892204NIDEK INC. MODELS YC-2000 & 3000 OPHTHALMIC LASERSNidek, Inc.Jun 22, 1989SESE
K892058ISL QC 2000Intelligent Surgical Lasers, Inc.Jun 15, 1989SESE
P840047CILCO ND:YAG LASERCoopervision-CilcoAug 2, 1985APRL

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...