21 CFR 886.4392 — Laser, Neodymium:Yag, Ophthalmic For Posterior Capsulotomy And Cutting Pupilla
Ophthalmic (OP) · Part 886 Subpart E—Surgical Devices · § 886.4392
Identification
The Nd:YAG laser for posterior capsulotomy and peripheral iridotomy consists of a mode-locked or Q-switched solid state Nd:YAG laser intended for disruption of the posterior capsule or the iris via optical breakdown. The Nd:YAG laser generates short pulse, low energy, high power, coherent optical radiation. When the laser output is combined with focusing optics, the high irradiance at the target causes tissue disruption via optical breakdown. A visible aiming system is utilized to target the invisible Nd:YAG laser radiation on or in close proximity to the target tissue.
Classification Rationale
Class II (special controls). Design Parameters: Device must emit a laser beam with the following parameters: wavelength = 1064 nanometers; spot size = 50 to 100 micros; pulse width = 3 to 30 nanoseconds; output energy per pulse = 0.5 to 15 millijoules (mJ); repetition rate = 1 to 10 pulses; and total energy = 20 to 120 mJ.
Product Codes
| Product Code | Device Name | Class | Devices | Attributes |
|---|---|---|---|---|
| LXS | Laser, Neodymium:Yag, Ophthalmic For Posterior Capsulotomy And Cutting Pupilla | 2 | 7 | 3rd Party |
Special Controls
LXS — Laser, Neodymium:Yag, Ophthalmic For Posterior Capsulotomy And Cutting Pupilla
*Classification.* Class II (special controls). Design Parameters: Device must emit a laser beam with the following parameters: wavelength = 1064 nanometers; spot size = 50 to 100 micros; pulse width = 3 to 30 nanoseconds; output energy per pulse = 0.5 to 15 millijoules (mJ); repetition rate = 1 to 10 pulses; and total energy = 20 to 120 mJ.
eCFR