K232273 · Retmap, Inc. · HLZ · Dec 7, 2023 · Ophthalmic
Device Facts
Record ID
K232273
Device Name
RM Electrode (RMH 23-01)
Applicant
Retmap, Inc.
Product Code
HLZ · Ophthalmic
Decision Date
Dec 7, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1220
Device Class
Class 2
Indications for Use
The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal dysfunctions. It is indicated for use in patients aged 12 years and above undergoing diagnostic full-field electroretinogram (ERG) recording procedures. The RM Electrode® is a single use, disposable device.
Device Story
RM Electrode is a passive, single-use, disposable contact lens-type corneal electrode; used for recording bioelectric signals during full-field electroretinography (ERG) procedures. Device consists of contact lens substrate, conductive electrode ring, lead wire with 1.5mm touchproof connector, and heat shrink tubing. Operated by ophthalmologists, optometrists, medical technicians, or vision researchers in hospitals, clinics, physician offices, or research labs. Device is filled with lubricating ophthalmic solution, placed on cornea, and connected to external ERG recording system. Passive sensor transmits retinal electrical signals to recorder; healthcare provider interprets recorded ERG waveforms to diagnose retinal dysfunctions. Benefits include improved positional stability, prevention of blinking during testing, and clear pupil area compared to predicate.
Clinical Evidence
Clinical study in 15 healthy subjects compared subject and predicate devices using full-field flash and 30 Hz flicker ERG protocols. Results showed equivalent amplitudes (a-wave, b-wave, flicker) and signal-to-noise ratios. Human ocular irritation study (n=10) with 60-minute wear time showed equivalent or lower redness/abrasion compared to predicate. No adverse events reported.
Technological Characteristics
Passive corneal electrode; contact lens substrate with conductive electrode ring. Materials are biocompatible. Connects via 1.5mm touchproof connector to standard ERG recording systems. Single-use, disposable. No internal power source or software. Designed for full-field flash ERG protocols.
Indications for Use
Indicated for patients aged 12 years and older undergoing diagnostic full-field electroretinogram (ERG) recording procedures to support diagnosis of retinal dysfunctions.
Regulatory Classification
Identification
A corneal electrode is an AC-powered device, usually part of a special contact lens, intended to be applied directly to the cornea to provide data showing the changes in electrical potential in the retina after electroretinography (stimulation by light).
K961805 — TOMEY DTL ELECTRODE · Tomey Corporation USA · May 9, 1997
K153019 — Disposable Adhesive Surface Electrodes, Disposable/Reusable Cup Electrodes · Bio Protech, Inc. · Mar 21, 2016
K173944 — Endocular Viewing Lenses and Silicone Ring · Phakos · Oct 25, 2018
K192564 — Disposable Gold Cup EEG Electrodes · Dymedix Diagnostics, Inc. · Jan 21, 2020
K180232 — DAEHAN Disposable CUP Electrodes · Daehan Medical Systems Co., Ltd. · Apr 3, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
December 7, 2023
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are displayed side-by-side.
RetMap, Inc. Shresta Patangay Chief of Staff 832 W Superior St, Suite 202 Chicago, Illinois 60642
Re: K232273
Trade/Device Name: RM Electrode (RMH 23-01) Regulation Number: 21 CFR 886.1220 Regulation Name: Corneal electrode Regulatory Class: Class II Product Code: HLZ Dated: November 3, 2023 Received: November 3, 2023
Dear Shresta Patangay:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Elvin Y.
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K232273
Device Name
RM Electrode (RMH 23-01)
Indications for Use (Describe)
The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal dysfunctions. It is indicated for use in patients aged 12 years and above undergoing diagnostic full-field electroretinogram (ERG) recording procedures.
The RM Electrode® is a single use, disposable device.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
# K232273
| Submitter's Name | RetMap, Inc.<br>832 W Superior St., Suite 202<br>Chicago, IL 60642<br>Telephone: (312) 224-8938 |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------|
| Contact Person | Shresta Patangay, PhD<br>Chief of Staff<br>RetMap, Inc.<br>832 W Superior St., Suite 202<br>Chicago, IL 60642 |
| Date Prepared | July 28, 2023 |
| Device Trade Name | RM Electrode |
| Device Common Name | Corneal or ERG Electrode |
| Classification Name | Corneal Electrode |
| Product Code | HLZ |
| Device Classification | Class II |
| Panel | Ophthalmology |
| Regulatory Classification | 21 CFR 886.1220 |
| Type of 510(k) Submission | Traditional |
| Legally Marketed Predicate Device | ERG Jet Electrode (K813399) |
## Premarket Notification (510(k)) Number
{4}------------------------------------------------
#### Description
The RM Electrode® is a contact lens-type corneal electrode for use in electroretinography (ERG) procedures. Like other corneal electrodes classified under 21 CFR 886.1220 and cleared under product code HLZ, the RM Electrode is used for the measurement and recording of ERG signals, to support the diagnosis of retinal dysfunctions. It has four components, including (1) the contact lens (substrate), which supports the conductive element (electrode ring) and allows light from the stimulus source to enter the eye; (2) the electrode ring, which is the conductive element of the device; (3) the lead wire, which is connected to the electrode ring extension and terminates in an industry standard 1.5mm female "touchproof" connector that connects to a commercially available ERG recording system (not part of the subject device); and (4) heat shrink tubing which covers the electrode ring extension and junction with the lead wire. The device materials were demonstrated to be biocompatible.
The RM Electrode® is not a powered device. As with the predicate device (ERG Jet Electrode), it is a passive sensor for recording bioelectric signals from the surface of the eye and transmitting them to an ERG recording system. From an electrical perspective, the device is a single conductor that connects the eye surface to a data recorder.
#### Indications for Use
The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of electroretinography (ERG) signals, to support the diagnosis of retinal dysfunctions. It is indicated for use in patients aged 12 years and above undergoing diagnostic full-field electroretinogram (ERG) recording procedures.
The RM Electrode® is a single use, disposable device.
#### Predicate Device
The RM Electrode® is substantially equivalent to the ERG Jet Electrode cleared under K813399. Both devices are electrodes intended to be applied on the cornea for the measurement and recording of ERG signals, to support the diagnosis of retinal dysfunctions. Both the subject and predicate devices have the same intended use and similar technological characteristics, but the RM Electrode includes a softer eye contact material and design features that help prevent the wearer from blinking during the ERG test, improve positional stability, and keep the area of the pupil clear of obstructions.
{5}------------------------------------------------
Please refer to the Substantial Equivalence Summary Comparison Table below for a summary of the similarities and differences between the subject and predicate devices.
#### Description of Operation
As part of an ERG procedure, the device is filled with a lubricating ophthalmic solution and placed on the eye. The lead wire is gently draped over the ipsilateral ear, secured with tape if needed, and the connector is plugged into a compatible ERG recording system. The device can be used for any full-field ERG protocol to record bioelectric signals from the surface of the eye and transmit them to the ERG data recording system. After the ERG test procedure, the device is removed from the patient's eye.
#### Technological Characteristics Summary
The RM Electrode is substantially equivalent to the ERG Jet Electrode (K813399) with respect to intended use, design, principle of operation, technological characteristics and performance. Both devices include similar components, including a contact lens (eye contact) portion, electrode, and lead wire with connector. The RM Electrode includes a softer eye contact material and design features that help prevent the wearer from blinking during the ERG test, improve positional stability, and keep the area of the pupil clear of obstructions. These differences do not raise new questions of safety or effectiveness.
| Features | RM Electrode<br>(Subject Device) | ERG Jet Electrode<br>Predicate Device<br>K813399 | Comparison |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Corneal (contact lens) electrode to transmit the retinal electrical signal in diagnostic ERG procedures | Corneal (contact lens) electrode to transmit the retinal electrical signal in diagnostic ERG procedures | Identical |
| Product Code | HLZ | HLZ | Identical |
| Indications for Use | The RM Electrode® is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of | The ERG-Jet is a contact lens electrode intended to be applied on the cornea and to be used for the measurement and recording of | Equivalent. While the predicate device is intended for use in both full-field flash and multi-focal ERG procedures, the subject device is |
| | electroretinography<br>(ERG) signals, to<br>support the diagnosis of<br>retinal dysfunctions. It<br>is indicated for use in<br>patients aged 12 years<br>and above undergoing<br>diagnostic full-field<br>electroretinogram<br>(ERG) recording<br>procedures.<br><br>The RM Electrode® is a<br>single use, disposable<br>device. | electroretinography<br>(ERG) signals, to<br>support the diagnosis of<br>retinal dysfunctions.<br><br>The ERG-Jet is not<br>intended for multiple<br>use. If re-used, an<br>infection of the cornea<br>can occur, and the<br>measuring results are no<br>longer guaranteed. | currently intended<br>only for use in full-<br>field flash ERG<br>procedures. Full-<br>field flash protocols<br>represent the<br>majority<br>(approximately<br>80%) of the ERG<br>procedures<br>conducted in the<br>United States,<br>therefore this<br>difference is not<br>critical to the<br>intended use of the<br>device and does not<br>affect the safety or<br>effectiveness of the<br>device when used as<br>labeled. The device<br>labeling clearly<br>identifies the<br>intended use for full-<br>field flash<br>procedures only. |
| Target Population | Patients aged 12 years<br>and above undergoing<br>ERG recording<br>procedures | Patient population not<br>specified | Both devices are used<br>in patients undergoing<br>ERG procedures |
| Types of ERG<br>Procedures | Full-field flash ERG<br>protocols | Either full-field or<br>multifocal ERG protocols | The RM Electrode is<br>not intended for ERG<br>protocols that use<br>patterned visual<br>stimuli, such as the<br>multi-focal ERG<br>(mfERG). Full-field<br>flash protocols<br>represent<br>approximately 80% of<br>ERG procedures that |
| | | | are conducted in the |
| | | | US thus this |
| | | | difference is not |
| | | | critical to the intended |
| | | | use of the device and |
| | | | does not affect the |
| | | | safety or effectiveness |
| | | | of the device when |
| | | | used as labeled. |
| Wear Duration | Typically 20 minutes.<br>< 1 hour | Typically 20 minutes.<br>< 1 hour | Identical |
| Intended Users | Ophthalmologists,<br>optometrists, trained<br>medical technicians and<br>professionals and vision<br>science researchers | Ophthalmologists and<br>optometrists practicing in<br>ophthalmic hospitals,<br>clinics and medical<br>settings | Equivalent intended<br>users |
| Environment of<br>Use | Hospitals, clinics,<br>physician offices,<br>research labs | Ophthalmic hospitals,<br>clinics and medical<br>settings | Equivalent medical<br>setting use<br>environments |
| Components | Contact lens, electrode<br>ring, lead wire with<br>1.5mm touchproof<br>connector that connects to<br>ERG data recorder | Contact lens, electrode,<br>lead wire with 1.5mm<br>touchproof connector that<br>connects to ERG data<br>recorder | Equivalent<br>components |
| Materials | Biocompatible materials | Biocompatible materials | Equivalent materials |
| Compatible ERG<br>Recorders | ERG recording system<br>accepting 1.5mm<br>touchproof connector<br>that complies with all<br>applicable device safety<br>standards, such as IEC<br>60601-1 | ERG recording system<br>accepting 1.5mm<br>touchproof connector<br>that complies with all<br>applicable device safety<br>standards, such as IEC<br>60601-1 | Identical |
#### Substantial Equivalence Summary Comparison Table
{6}------------------------------------------------
{7}------------------------------------------------
#### Performance Testing
As described below, a comprehensive battery of biocompatibility, bench, and clinical studies was submitted to confirm product conformance with device requirements. Performance data demonstrated biocompatibility, absence of ocular irritation in both animals and humans, and substantially equivalent performance to the predicate device in terms of ERG signal quality, lead
{8}------------------------------------------------
wire integrity and electrical impedance. The biocompatibility, bench, shelf life, and clinical studies summarized below concluded that the device is as safe, as effective, and performs as well or better than the legally marketed predicate device.
#### Biocompatibility Testing
Biocompatibility testing was conducted in accordance with ISO 10993. Testing was successfully completed for cytotoxicity, sensitization and ocular irritation. In addition, human ocular irritation was completed following 60 minutes of wear time (approximately 3 times the typical 20-minute wear duration) in 10 subjects with both the subject and predicate devices. Bulbar redness, limbal redness, tarsal redness, and corneal abrasion associated with wear of the subject device were found to be equivalent to, or less than, that associated with wear of the predicate device.
#### Bench Testing
The subject and predicate device were tested under tensile load to determine the failure load for the lead wire-electrode ring junction. The subject device failure load was significantly higher (better) than that for the predicate device. In addition, all samples of the subject device met the prospectively defined acceptance criteria for the minimum failure load.
The subject and predicate device were also evaluated for electrical impedance measured at three frequencies within the passband of the ERG signal. At all frequencies, the electrical impedance of the subject device was equivalent to, or less than (better) that of the predicate device.
#### Shelf Life Testing
The device and its packaging successfully completed evaluations after accelerated aging to support the labeled shelf life. Integrity of the sterile packaging was confirmed before and after accelerated aging through seal strength and bubble leak testing. Device properties were confirmed before and after accelerated aging through evaluation of material hardness of the eye contact portion of the device using a calibrated durometer and through evaluation of the electrical impedance of the device measured at three frequencies within the passband of the ERG signal. All samples met the acceptance criteria demonstrating no significant changes over the labeled shelf life.
#### Clinical Testing
Clinical performance testing was compared using the subject and predicate devices in 15 subjects at two sites who sequentially wore both devices in the same eye in a randomized order. Subjects
{9}------------------------------------------------
were healthy with no history of retinal disease or corneal surgery. Full-field flash (ISCEV LA 3.0 stimulus) and 30 Hz flicker ERG protocols were followed. Amplitudes (a-wave, b-wave and flicker) and signal to noise ratios were substantially equivalent for both devices. There were no adverse events or complications related to the subject device.
In addition, as noted in the Biocompatibility section above, human ocular irritation was completed following 60 minutes of wear time (approximately 3 times the typical wear duration) in 10 subjects with both the subject and predicate devices. Subjects were normally sighted aged 18 years and above. Bulbar redness, limbal redness, tarsal redness, and corneal abrasion associated with wear of the subject device were found to be equivalent to, or less than, that associated with wear of the predicate device. There were no adverse events or complications related to the study device.
#### Conclusions
The nonclinical and clinical studies demonstrated the subject device is (1) biocompatible (lack of cytotoxicity, ocular irritation in animals or humans, or sensitization); (2) maintains device (material hardness of the eye contact portion and electrical impedance) and package integrity (seal strength and bubble leak) after a 1-year shelf life; (3) has substantially equivalent or better performance with respect to lead wire integrity and electrical impedance compared to the predicate device; and (4) can be used successfully in full-field ERG protocols with comparable results (amplitudes and signal to noise ratios) as the predicate device. These results demonstrate that the RM Electrode is as safe, as effective, and performs as well or better than the legally marketed device identified above.
#### Summary
In summary, the RM Electrode is substantially equivalent to the predicate ERG Jet Electrode with respect to intended use, technological characteristics and performance. The biocompatibility, bench, shelf life, and clinical studies concluded that the device is as safe, as effective, and performs as well or better than the legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.