TOMEY DTL ELECTRODE

K961805 · Tomey Corporation USA · HLZ · May 9, 1997 · Ophthalmic

Device Facts

Record IDK961805
Device NameTOMEY DTL ELECTRODE
ApplicantTomey Corporation USA
Product CodeHLZ · Ophthalmic
Decision DateMay 9, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1220
Device ClassClass 2

Intended Use

The Tomey DTL Electrode is intended for use in the measurement and recording of ERG signals from the ocular surfaces of the eye. These measurements are only to be performed by licensed practitioners and trained technicians within a medical environment.

Device Story

Device consists of soft silver-impregnated nylon fiber threads placed under the lower eyelid to record ERG signals. Fibers (3-10, 5-20 microns diameter) are stabilized with adhesive attachments. Thin wire connects to an adaptor cable for integration with ERG recording systems. Used in medical environments by licensed practitioners or trained technicians. Provides electrical signal input to external ERG recording systems to assist in clinical assessment of ocular function.

Clinical Evidence

Bench testing only. Includes elution testing per Japanese Ministry of Health and Welfare requirements and rabbit eye irritation studies confirming no irritation. Biocompatibility supported by literature review of similar electrode types.

Technological Characteristics

Silver-impregnated nylon fiber threads (3-10 fibers, 5-20 microns diameter). Adhesive attachments. Adaptor cable for connection to ERG recording systems. Non-powered, passive electrode. Supplied in factory-sealed package; intended for user-performed disinfection.

Indications for Use

Indicated for patients requiring electroretinography (ERG) signal measurement and recording from the ocular surface. Used by licensed practitioners and trained technicians in medical environments.

Regulatory Classification

Identification

A corneal electrode is an AC-powered device, usually part of a special contact lens, intended to be applied directly to the cornea to provide data showing the changes in electrical potential in the retina after electroretinography (stimulation by light).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} MAY - 9 1997 K961805 # SECTION VI ## 510(k) SUMMARY ### Tomey DTL Electrode Common/Classification Name: Electrode Corneal, 21 CFR 886.1220 Applicant: Tomey Corporation 2-22-10 Noritakeshinnmachi Nishi_ku, Nagoya 451 Japan Attn: Mr. Susumu Nozowa (Manager /Regulatory) (052) 581-5595 (052) 581-5596 (FAX) Correspondent: Tomey Corporation (USA) 325 Vassar Street, Cambridge, MA 02139 Contact: Dave MacLellan (Engineer) (617) 864-6488 (617) 864-6897 (FAX) Prepared: February 25, 1997 ## A. LEGALLY MARKETED PREDICATE DEVICES The Tomey DTL Electrode is substantially equivalent in design and function to the DTL Electrode as manufactured by Sauquoit Industries, Inc., Scranton, PA (as cleared in K844409), the Gold Foil ERG Electrodes as manufactured by Cadwell Laboratories, Inc., Kennewick, WA (as cleared in K820254), and the ERG-JET Electrode as manufactured by Life Tech, Houston, TX (as cleared in K823767). The Tomey DTL Electrode is also similar to the ERG electrode that was included with the Tomey PE-400/PS-400 Portable ERG system under K932571. Tomey Corporation DTL Electrode, Section VI 1/3 {1} Tomey Corporation DTL Electrode, Section VI 2/3 ## B. DEVICE DESCRIPTION Tomey DTL Electrode consists of soft silver impregnated nylon fiber threads that are normally placed under the lower eye lid. The electrodes are constructed with 3 to 10 individual fibers between 5 and 20 microns in diameter. The fibers are stabilized at both ends with adhesive attachments. A thin wire is connected to an adaptor cable that permits the electrodes to be used with various ERG Recording systems including the Tomey PE-400/PS-400. ## C. INTENDED USE The Tomey DTL Electrode is intended for use in the measurement and recording of ERG signals from the ocular surfaces of the eye. These measurements are only to be performed by licensed practitioners and trained technicians within a medical environment. There are no differences with respect to the predicate devices for indications for use, or target population. ## D. SUBSTANTIAL EQUIVALENCE SUMMARY The Tomey Corporation device has the same intended use and target population as the predicate devices, and has equivalent effectiveness for its intended use. ## E. NON-CLINICAL TESTING This section summarizes the performance testing that Tomey Corporation carried out on the Tomey DTL Electrode. This testing addressed the following issues: 1. Elution testing The Tomey DTL Electrode passed the Elution test performed according to the requirements of the Ministry of Health and Welfare in Japan. 2. Rabbit Eye irritation testing An eye irritation test was performed by Juridical Foundation, Japan Food Research Laboratory which indicates that the Tomey DTL Electrode does not cause irritation to the eye when used according to prescribed instructions. {2} 3. Sterilization The Tomey DTL Electrode is supplied to the user in a factory sealed package and is intended to be disinfected by the end user employing procedures recommended in the product labeling. 4. Biocompatibility The Tomey DTL Electrode is for short term ERG recording purposes. Extensive use by many investigators, including Dawson, Prager and Lachapelle, has not revealed any biocompatibility problems with this type of an electrode. Eye Irritation studies performed by Tomey, also indicate that short term biocompatibility is not a problem for the Tomey DTL Electrode. 5. Other Safety and Compatibility The Tomey DTL Electrode, by means of adaptor cables, is compatible with commercially available ERG recording systems that employ protected electrode leads. The Tomey DTL Electrode is unable to be mistakenly connected to an a/c or d/c power source. F. CONCLUSIONS Tomey Corporation has demonstrated that its evaluation of the Tomey DTL Electrode and its review of the literature shows equivalent safety and effectiveness with respect to performance, biocompatibility issues, comfort and corneal surface damage, although no specific claim for these attributes is being made. Tomey Corporation DTL Electrode, Section VI 000053
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...