ARGOS

K191051 · Santec Corporation · MXK · May 16, 2019 · Ophthalmic

Device Facts

Record IDK191051
Device NameARGOS
ApplicantSantec Corporation
Product CodeMXK · Ophthalmic
Decision DateMay 16, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1850
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White-to-white (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool. It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.

Device Story

ARGOS is a non-invasive, non-contact optical biometer using swept-source optical coherence tomography (SS-OCT) to measure ocular parameters; device captures axial length, corneal thickness, anterior chamber depth, lens thickness, K-values, astigmatism, white-to-white diameter, and pupil size. It also provides reference image capture for preoperative/postoperative documentation. Operated by ophthalmologists or eye-care professionals in clinical settings; device performs automated calculations for IOL power and type selection. Output includes ocular measurements and reference images, which can be transferred to external image-guided surgical systems to assist in preoperative planning. Device benefits patients by providing precise biometric data necessary for accurate IOL selection during cataract or refractive surgery.

Clinical Evidence

Bench testing only. Performance evaluated against standards including ISO 15004-1, ANSI Z80-36, IEC 60825, and ISO 10343. Hardware validation confirmed accuracy for axial/lateral distance measurements, signal-to-noise ratio, and depth attenuation. Usability engineering performed per IEC 62366. Software verification and validation conducted per FDA guidance. No clinical data presented.

Technological Characteristics

SS-OCT biometer; 1060nm swept laser (3kHz rate); 16-LED video keratometry (850nm); white LED for reference imaging; 630nm fixation LED. Class 1 laser product. Complies with ANSI/AAMI ES 60601-1, IEC 60529 (IP20), ANSI Z80-36, IEC 62471, and ISO 10343. Connectivity to external surgical planning systems (Verion, ORA).

Indications for Use

Indicated for ophthalmologists, physicians, and eye-care professionals to perform non-invasive ocular measurements (axial length, corneal thickness, anterior chamber depth, lens thickness, K-values, astigmatism, white-to-white, pupil size) and IOL power calculations for patients undergoing intraocular lens placement.

Regulatory Classification

Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 16, 2019 Santec Corporation % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, NJ 07059 Re: K191051 Trade/Device Name: Argos Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slitlamp biomicroscope Regulatory Class: Class II Product Code: MXK Dated: April 17, 2019 Received: April 19, 2019 Dear Dave Yungvirt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K191051 Device Name ARGOS (ARGOS ver1.5) #### Indications for Use (Describe) ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Clamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool. It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician. #### Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ ## 510(k) SUMMARY The information contained in this premarket notification 510(k) summary is submitted as required by 21 CFR 807.92: ### 1. Type of Submission 510k Traditional submission for New Devices #### 2. Applicant Company: SANTEC CORPORATION 5823 Ohkusa-Nenjozaka, Komaki, Aichi 485-0802, JAPAN Phone: 81 (0568) 79 3535 www.santec.com #### 3. Applicant (Contact Person) Name: Changho Chong # 4. Date Summary Prepared: May 15, 2019 #### 5. Device Trade / Proprietary Name (Model name): ARGOS (ARGOS ver1.5) ## 6. Common Name: Optical Biometer ## 7. Regulation Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Bio-Microscope Regulatory Class: Class II Product Code: MXK {4}------------------------------------------------ ### 8. Indications for Use: ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, K - values (Radii of flattest and steepest meridians). Astigmatism. White-to-white (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool. It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician. The indications for use are similar between the proposed and predicate ARGOS devices. The proposed ARGOS device has additional reference image functionality as minor difference. #### 9. Device Descriptive The ARGOS is substantially equivalent to the predicate device identified previously: - the ARGOS (Santec Corporation) that was cleared by the FDA on October 2nd, 2015(K150754) as the primary predicate as it has the most similar intended use and characteristics - the IOLMaster 700 (Carl Zeiss Meditec) that was cleared by the FDA on June 29, ● 2015 (K143275, K170171) as an additional (secondary) predicate to support a new feature of reference image capture on the version of ARGOS in this submission. The predicate devices are Class 2 devices to premarket notification, as defined per regulation number 21 CFR 886.1850. In addition, the predicate devices have product codes of MXK(ARGOS), and HJO(IOLMaster700). The version of ARGOS in this submission is a modified version of the Argos cleared under K150754 which is substantially equivalent with regard to intended use, operating principle, function, materials, and energy source. The differences from the predicate ARGOS (K150754) that are subject of this 510(k) submission are: - An additional feature of reference image capture function ● - Labeling change including change in the intended use adding the feature of reference . image capture. - . This image can be transferred to image guided devices in order to support the execution of preoperative plan. {5}------------------------------------------------ The changes described in this submission do not affect how the hardware is used to acquire measurements as a biometer, nor do these changes affect the principle of operation of the device. | Compliant standard | Description | Classification | |----------------------|------------------------------------------------------------------------|---------------------| | ANSI/AAMI ES 60601-1 | Electrical safety | Class 1 Type B | | IEC 60601-1-2 | Electrical safety | Refer to Appendix 4 | | IEC 60529 | Housing protection | IP20 | | ANSI.Z80-36 | Ophthalmics - Light Hazard<br>Protection For Ophthalmic<br>Instruments | Group 2 | | IEC 62471 | LED light hazard | Exempt group | | IEC 60825-1 | Laser product safety | Laser Class 1 | | ISO 10343 | Ophthalmometer | | #### 10. Guidance on the Recognition and Use of Consensus Standards: ARGOS complies with the requirements of listed FDA Recognized Consensus Standards {6}------------------------------------------------ # 11. Substantial equivalence to the predicate device ## 11.1 Comparisons of technological characteristics with Predicate ARGOS devices | | | Table of Comparison between the ARGOS and the Predicate Device | | | | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--| | | New Device (Subjective device) | Primary Predicate Device | Secondary Predicate device | Comparison | | | Manufacturer | santec corporation | santec corporation | Zeiss Meditech | - | | | Model | ARGOS ver1.5 | ARGOS | IOLMaster 700 | - | | | 510(k) number | | K150754 | K143275, K170171 | - | | | Class | II | II | II | Same | | | Classification<br>panel | Ophthalmic device panel | Ophthalmic device panel | Ophthalmic device panel | Same | | | Product code | MXK | MXK | HJO | Same | | | Classification<br>number | AC-Powered Slitlamp Bio-<br>Microscope | Device, Analysis, Anterior<br>Segment | AC-Powered Slitlamp Bio-<br>Microscope | Same | | | Type | Optical biometer | Optical biometer | Biometer | Same | | | Intended Use | ARGOS is a non-invasive, non-contact<br>biometer based on swept-source<br>optical coherence tomography (SS-<br>OCT). The device is intended to<br>acquire ocular measurements as well<br>as perform calculations to determine<br>the appropriate intraocular lens (IOL)<br>power and type for implantation during<br>intraocular lens placement. ARGOS<br>measures the following 9 parameters:<br>Axial Length, Corneal Thickness,<br>Anterior Chamber Depth, Lens<br>Thickness, K-values (Radii of flattest<br>and steepest meridians), Astigmatism,<br>White-to-white (corneal diameter) and<br>Pupil Size. The Reference Image<br>functionality is intended for use as a<br>preoperative and postoperative image<br>capture tool.<br>It is intended for use by<br>ophthalmologists, physicians, and<br>other eye-care professionals and may<br>only be used under the supervision of<br>a physician. | ARGOS is a non-invasive, non-<br>contact biometer based on swept-<br>source optical coherence<br>tomography (SS-OCT). The device<br>is intended to acquire ocular<br>measurements as well as perform<br>calculations to determine the<br>appropriate intraocular lens (IOL)<br>power and type for implantation<br>during intraocular lens placement.<br>ARGOS measures the following 9<br>parameters: Axial Length, Corneal<br>Thickness, Anterior Chamber<br>Depth, Lens Thickness, K-values<br>(Radii of flattest and steepest<br>meridians), Astigmatism, White-to-<br>white (corneal diameter) and Pupil<br>Size.<br>lt is intended for use by<br>ophthalmologists, physicians, and<br>other eye-care professionals and may<br>only be used under the<br>supervision of a physician. | The indications for use are identical<br>between the subject and predicate<br>IOLMaster 700 devices.<br>The IOLMaster 700 is intended for<br>biometric measurements and<br>visualization of ocular structures. The<br>measurements and visualization assist<br>in the determination of the appropriate<br>power and type of intraocular lens.<br>The IOLMaster 700 measures:<br>• Lens thickness<br>• Corneal curvature and thickness<br>• Axial length<br>• Anterior chamber depth<br>• Pupil diameter<br>• White-to-white distance (WTW)<br>For visualization, the IOLMaster 700<br>employs optical coherence<br>tomography (OCT) to obtain two-<br>dimensional images of ocular<br>structures of the anterior and posterior<br>segments of the eye.<br>The Reference Image functionality is<br>intended for use as a preoperative and<br>postoperative image capture tool. | Similar to Primary,<br>additional feature<br>identical to<br>secondary<br>predicate | | | Measured<br>parameters | Axial Length, Corneal Thickness,<br>Anterior Chamber Depth, Lens<br>Thickness, K-values (Radii of<br>flattest and steepest meridians),<br>Astigmatism (Toric Angle), White-<br>to-white (corneal diameter), and<br>Pupil Size. | Axial Length, Corneal<br>Thickness, Anterior Chamber<br>Depth, Lens Thickness, K-<br>values (Radii of flattest and<br>steepest meridians),<br>Astigmatism (Toric Angle),<br>White-to-white (corneal<br>diameter), and Pupil Size. | Lens thickness, Corneal<br>curvature and thickness, Axial<br>length, Anterior chamber<br>depth, Pupil diameter, White-to-<br>white distance (WTW) | Same parameters | | | In-vivo<br>repeatability | Axial length: range 14-38mm, SD<br>0.01mm | Axial length: range 14-38mm,<br>SD 0.01mm | Axial length: range 14-38mm,<br>SD 0.005mm | Same between<br>subjective and<br>primary, Similar<br>between subjective<br>and secondary<br>predicate | | | | Corneal thickness: range 200-<br>120um, SD 10um | Corneal thickness: range 200-<br>120um, SD 10um | Corneal thickness: range 200-<br>120um, SD 2.5um | | | | | Anterior Chamber Depth: range<br>0.7-8.0mm SD 0.01mm | Anterior Chamber Depth:<br>range 0.7-8.0mm SD 0.01mm | Anterior Chamber Depth: range<br>0.7-8mm SD 0.007mm | | | | | Lens Thickness: range 0.5-<br>10.0mm SD 0.02mm | Lens Thickness: range 0.5-<br>10.0mm SD 0.02mm | Lens Thickness: range 1-10.0mm<br>SD 0.006mm | | | | | Keratometry: range 5.5-10.0mm<br>SD 0.02mm | Keratometry: range 5.5-<br>10.0mm SD 0.02mm | Keratometry: range 5-11.0mm SD<br>0.09D(~0.02mm) | | | | | Astigmatism: range 0-180deg, SD<br>5deg (Cylinder>1D) | Astigmatism: range 0-180deg,<br>SD 5deg (Cylinder>1D) | Astigmatism: range 0-180deg, SD<br>3.8deg (Cylinder>0.75D) | | | | | Pupil size: range 2-13mm SD<br>0.09mm | Pupil size: range 2-13mm SD<br>0.09mm | Pupil size: (no specification<br>available) | | | | | White-to-White: range 7-15mm<br>SD 0.06mm | White-to-White: range 7-15mm<br>SD 0.06mm | White-to-White: range 8-16mm<br>SD 0.044mm | | | | Method for<br>biometry | Interferometry (SS-OCT) | Interferometry (SS-OCT) | Interferometry (SS-OCT) | Same | | | Method for<br>keratometry | Video keratometry (16 LED) | Video keratometry (16 LED) | Video keratometry (IR LED) | Same | | | Optical<br>radiation | | | | | | | Type of light<br>source | Wavelength swept laser (3kHz<br>swept rate)<br>Wavelength 1060nm<br>Power <0.72mW (<0.1mW<br>average at 1mm aperture during<br>B-scan) | Wavelength swept laser (3kHz<br>swept rate)<br>Wavelength 1060nm<br>Power <0.72mW (<0.1mW<br>average at 1mm aperture<br>during B-scan) | Wavelength swept laser (2kHz<br>swept rate)<br>Wavelength 1055nm<br>unavailable | Same | | | Laser class | Class 1 (IEC 60825)<br>ANSI-Z80.36-2016 Group2 | Class 1 (IEC 60825)<br>ISO15004-2 Group2 | Class 1 (IEC 60825)<br>ISO15004-2 Group2 | Same<br>(ANSI.Z80-36 is<br>equivalent to<br>ISO15004-2) | | | Keratometry | L…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...