Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye, as well as for management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.
Device Story
Harmony is a software-only ophthalmic image management and processing system (PACS). It imports, stores, and displays clinical images/videos from various ophthalmic diagnostic instruments (OCT, retinal cameras, slit lamps, scanning laser ophthalmoscopes, ultrasound). The system manages patient demographics and clinical reports. Modifications include PixelSmart (moving average algorithm for OCT B-scan noise reduction/contrast enhancement), internationalization, analytical thickness grids (sectorial average thickness from OCT data), hanging protocols (customizable display arrangements), and automatic image smoothing during zoom. Used in clinical settings by eye care professionals to aid clinical workflows. Output is displayed on a user interface for review and analysis, supporting clinical decision-making by providing organized, enhanced, and measured ophthalmic data.
Clinical Evidence
Bench testing only. Software verification and validation were performed to confirm the modified Harmony functions equivalently to the predicate. Testing adhered to IEC 62304, NEMA PS 3.1-3.20 (DICOM), ISO 10918-1, and ISO 14971 standards.
Technological Characteristics
Software-only PACS. Connectivity via direct instrument connection or computerized networks. Supports DICOM formats. Implements PixelSmart (moving average algorithm), analytical thickness grids (sectorial averages), and image display/smoothing features. Complies with IEC 62304, NEMA PS 3.1-3.20, ISO 10918-1, and ISO 14971.
Indications for Use
Indicated for use in importing, processing, measurement, analysis, and storage of clinical eye images/videos and management of patient/diagnostic data and reports from ophthalmic instruments. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
November 20, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Topcon Healthcare Solutions % Lena Sattler President Orasi Consulting LLC 226 1st Street Bonita Springs, Florida 34134
Re: K232555
Trade/Device Name: Harmony Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: NFJ Dated: October 23, 2023 Received: October 23, 2023
Dear Lena Sattler:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Elvin
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices
{3}------------------------------------------------
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232555
Device Name Harmony
Indications for Use (Describe)
Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye as well as in management of patient data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> | <span style="text-decoration: overline;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## K232555: Special 510(k) SUMMARY
## Topcon Healthcare Solutions, Inc. Harmony
#### 510(k) Owner
Topcon Healthcare Solutions, Inc. 111 Bauer Drive Oakland, NJ 07436 Contact Person: Tobias Schwarz, Senior Director Regulatory Affairs +1-(201)-599-5100 Phone: E-mail: tschwarz@topcon.com
### Submission Correspondent
Lena Sattler President, Orasi Consulting, LLC. 226 1st Street Bonita Springs, FL 34134 +1-(440) 554-3706 Phone: E-mail: lena@orasiconsulting.com
Date Prepared: November 16, 2023
Trade Name of Device Harmony
Predicate Device Topcon Harmony (K182376)
#### Classification Information
| Regulation Number: | 21 CFR 892.2050 |
|----------------------|------------------------------------------------|
| Classification name: | Medical image management and processing system |
| Device Class: | Class II |
| Common name: | System, Image Management, Ophthalmic |
| Product Codes, Name: | NFJ |
# Intended Use / Indications for Use
Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye as well as in management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.
{6}------------------------------------------------
# Device Description
Harmony is a modification to the existing Harmony cleared in K182376. The differences between the new version and the currently cleared version are modifications to the graphical user interface consisting of PixelSmart Technology, Internationalization support, Analytical thickness grids, Hanging protocols, and Automatic image smoothing while zooming in.
Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye, as well as for management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.
Harmony is used together with a number of computerized digital imaging devices, including:
- Optical Coherence Tomography devices .
- Mydriatic retinal cameras .
- Non-mydriatic retinal cameras .
- Biomicroscopes (slit lamps)
In addition, Harmony collects and manages patient demographics, image data, and clinical reports from a range of medical devices, including:
- Scanning Laser Ophthalmoscope images and videos .
- Non Radiometric Ultrasound devices ●
- Video image sources ●
- TWAIN compliant imaging sources ●
- Compliant data sources placed in network accessible folders and directories
- . Images of known format from digital cameras and scanners
- . Printer files of known format form computerized diagnostic devices
- Electronic information complying to accepted DICOM formats
- Other devices connected in proprietary formats ●
There are 5 notable device modifications subject of this submission: PixelSmart Technology, International support, Analytical thickness grids, Hanging protocols, and Automatic image smoothing while zooming in, along with some minor modifications.
PixelSmart is an optional post-processing image enhancement algorithm performing a moving average across OCT B-scans, reducing speckle noise and improving contrast by applying smoothing.
{7}------------------------------------------------
International support adds the possibility to use the Harmony user interface and online user manual in Spanish, in addition to the standard English software.
Analytical thickness grids offer the same functionality as the existing, cleared thickness grids in Topcon's IMAGEnet 6, now also in Harmony. The grids show sectorial average thickness values as derived from OCT segmentation data.
Hanging protocols allows a customizable image display arrangement in the Harmony user interface, resembling the arrangement of physical images on a light box.
Automatic image smoothing while zooming in is an optional display feature that will cause OCT B-scan images on higher zoom levels to look less pixelated.
# Performance Data
Software validation and verification demonstrate that Harmony performs as intended and meets its specifications, using methods equivalent to the predicate device.
## Substantial Equivalence
The modified Harmony is a device modification to the cleared Harmony (K182376) predicate device. Technological detail characteristics of the device are unchanged except for the modification as stated above. The modified Harmony has the same Indications for Use and maintains the same fundamental scientific technology as the predicate device.
The modified Harmony measures the same basic ophthalmic features as the cleared Harmony in K182376. The changes applied to the Harmony since the clearance in K182376 do not change the intended patient populations, the type of imported, displayed and analyzed images, or that the Harmony may be used as an aid to clinical workflows.
The basic functionality of the device is unchanged, which is a picture archiving and communication system (PACS). The device is still software only, which is unchanged by the modifications subject of this submission.
The modifications subject of this submission, PixelSmart Technology, Internationalization support, Analytical thickness grids, Hanging protocols, and Automatic image smoothing while zooming in, do not change the indications for use of the device, and do not change the manner in which images are imported, analyzed, and stored, or the operation of the device by the intended user.
{8}------------------------------------------------
Risk assessment was conducted on the modified Harmony as per existing design control, and the impact of the design modifications were assessed on the predicate 510(k) cleared device. Newly identified risks or modified existing risks are mitigated, and no unacceptable risk was identified remaining for the device after risk mitigation.
Topcon performed software validation and verification activities to confirm that the modified Harmony functions equivalently to the predicate Harmony.
Non-clinical performance testing was conducted on the modified Harmony to verify and validate that the device is safe and effective for its intended use and indications for use. The modified Harmony was evaluated according to the requirements of FDA recognized consensus standards:
- IEC 62304 Edition 1.1 2015-06: Medical Device Software Software . Life Cycle Processes
- NEMA PS 3.1 3.20 (2021), Digital Imaging And Communications In . Medicine (DICOM) Set
- ISO IEC 10918-1 First edition 1994-02-15: Information technology -. Digital compression and coding of continuous-tone still images: Requirements and guidelines [Including: Technical Corrigendum 1 (2005)]
- ISO 14971 Third Edition 2019-12: Medical devices Application of risk . management to medical devices
and was found to meet the requirements of the applicable parts, demonstrating that the safety and efficacy of the modified device is comparable to the predicate.
The modified Harmony is manufactured and tested in the exact manner as the predicate 510(k) cleared device.
The modifications to the device do not raise issues of safety and effectiveness. A comparison of technological characteristics and non-clinical performance testing demonstrate that the modified Harmony device is substantially equivalent to the unmodified predicate device.
# Conclusion
Comparison of technological characteristics and evaluation of non-clinical performance testing show that the modifications to the Harmony do not introduce any new potential safety risk and the device is as safe and effective as the predicate devices, therefore supporting a determination of substantial
{9}------------------------------------------------
## #TOPCON Healthcare
equivalence. Topcon's Harmony has the same intended use and indications for use as the previously cleared predicate device. In addition, Harmony has the same technological characteristics as its predicate. The modifications to the cleared version of Harmony are small and do not impact the substantial equivalence, thus the modified device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.