K182376 · Topcon Medical Systems, Inc. · NFJ · Sep 28, 2018 · Radiology
Device Facts
Record ID
K182376
Device Name
Harmony
Applicant
Topcon Medical Systems, Inc.
Product Code
NFJ · Radiology
Decision Date
Sep 28, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye as well as in management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.
Device Story
Harmony is a software-only ophthalmic image management platform. It imports clinical images, videos, and diagnostic data from ophthalmic devices (OCT, retinal cameras, slit lamps) via direct connection or network. It provides tools for viewing (zoom, pan, enhancement), measurement (line, area, Cup-to-Disc ratio, MPS), and data management. Used by clinicians in clinical settings to organize and analyze patient diagnostic information. The system facilitates clinical decision-making by centralizing data and reports from various diagnostic instruments. It supports web-based access and DICOM communication with PACS. The platform is a modification of the Synergy ODM, featuring a responsive GUI and a transition from Microsoft Silverlight to HTML5, allowing execution on Windows and iPad platforms.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by software verification and validation testing confirming the device performs as intended and meets specifications.
Technological Characteristics
Software-only image management system. Web-based architecture; front-end migrated from Microsoft Silverlight to HTML5. Supports DICOM communication, web-based data access, and responsive GUI. Compatible with Windows XP or above and iPad. Functionality includes image import, viewing (zoom, pan, enhancement), measurement (line, area, Cup-to-Disc, MPS), and archive/backup.
Indications for Use
Indicated for use in importing, processing, measurement, analysis, and storage of clinical eye images/videos and management of patient/diagnostic data and reports from ophthalmic instruments. Intended for clinical environments using ophthalmic diagnostic devices.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 28, 2018
Topcon Medical Systems. Inc. % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864
Re: K182376
Trade/Device Name: Harmony Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: NFJ Dated: August 29, 2018 Received: August 31, 2018
Dear Maureen O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
# Bradley S. Cunningham -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182376
Device Name Harmony
Indications for Use (Describe)
Harmony is a comprehensive software platform intended for use in importing, processing, measurement, and storage of clinical images and videos of the eye as well as in management of patient data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY
#### Topcon Medical Systems, Inc. Harmony
#### 510(k) Owner
Topcon Medical Systems, Inc. 111 Bauer Drive Oakland, NJ 07436 Phone: (201) 599-5208 Contact Person: James Lorkowski
#### Submission Correspondent
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 (978) 207-1245 Phone: Email: Maureen@oconnellregulatory.com
Date Prepared:
# Trade Name of Device
Harmony
#### Common or Usual Name
System, image management, ophthalmic
#### Classification Name
21 C.F.R. 892.2050 Picture and Archiving communications system
#### Predicate Device
Topcon Corporation Synergy ODM (K151952)
#### Intended Use / Indications for Use
Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye as well as in management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments
{5}------------------------------------------------
through either a direct connection with the instruments or through computerized networks.
## Device Description
Harmony is a modification to Synergy ODM cleared in K151952. The differences between the new version and the currently cleared version are modifications to the GUI using a responsive design and a change to the front end platform from Microsoft Silverlight to HTML5.
Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye, as well as for management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.
Harmony is used together with a number of computerized digital imaging devices, including:
- . Optical Coherence Tomography devices
- Mydriatic retinal cameras .
- Non-mydriatic retinal cameras .
- Biomicroscopes (slit lamps) ●
## Performance Data
No performance data was required or provided. Software validation and verification demonstrate that Harmony performs as intended and meets its' specifications.
## Substantial Equivalence
Harmony is substantially equivalent to Topcon's Synergy ODM cleared in K151952. Harmony has the same intended use and indications for use, similar principles of operation, and similar technological characteristics as the previously cleared predicate device. The intended use of the Harmony and intended use of the Synergy ODM cleared in K151952 are identical and both systems also have the identical indications for use statement.
Harmony has the same technological characteristics as the Synergy ODM (K151952). Both devices are software only image management systems which have the same acquisition, importing, viewing, measurement and analysis; network and security; print, archive and backup functionality.
Regarding acquisition. Harmony and the Synergy ODM (K151952) do not offer capture components. Both Harmony and Synergy ODM import digital
{6}------------------------------------------------
images, patient data, diagnostic data and clinical information from other software capture systems and directly from ophthalmic devices.
Regarding importing, Harmony and the Synergy ODM (K151952) allow importing and management of files and images from a range of ophthalmic diagnostic devices including DICOM files, image files of known format, video images, and printer files.
Regarding viewing, Harmony and the Synergy ODM (K151952) allow standard viewing operations such as zoom in/out, panning, etc. and standard image enhancements such as contrast and brightness adjustment, and drawing tools
Regarding measurement and analysis. Harmony and the Synergy ODM (K151952) provide line and area measurement capabilities and provide Cup to Disc ratio and MPS (Macular Photocoagulation Study) measurements.
Regarding network and security, both Harmony and the Synergy ODM (K151952) provide web-based access to all data files. Harmony and the Synergy ODM (K151952) provide DICOM communication with other PACS.
Regarding printing, Harmony and the Synergy ODM (K151952) provide similar customizable print templates. Regarding archiving and backup, Harmony and the Synergy ODM (K151952) include archive and backup functionality. The operating system for review stations running each of the systems is Windows XP or above. Harmony can also be executed on an iPad while Synergy ODM (K151952) can be used on a review station with Mac OS X operating system. This minor difference does not impact substantial equivalence.
Modifications were made as the front end platform was changed from Microsoft Silverlight to HTML5 and the design of the GUI was changed to a responsive design. Software documentation is provided in Section VIII in support of these changes. It is not believed that any of these changes impact the substantial equivalence of the Harmony to the prior Synergy ODM.
In conclusion. Harmony shares similar technological characteristics with the predicate device, both in terms of the manner in which images are imported, analyzed, and stored, as well as the operation of the device by the intended user.
## Conclusion
Topcon's Harmony has the same intended use and indications for use as the previously cleared predicate device. In addition, Harmony has the same technological characteristics as its predicate. The modifications to the
{7}------------------------------------------------
cleared version of Synergy ODM are minor and do not impact the substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.