OPTOS ADVANCE

K113696 · Optos Plc. · NFJ · May 25, 2012 · Radiology

Device Facts

Record IDK113696
Device NameOPTOS ADVANCE
ApplicantOptos Plc.
Product CodeNFJ · Radiology
Decision DateMay 25, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The Optos Advance Software is a web-based software system application intended for use in storing, managing, and displaying patient data, diagnostic data, videos and images from computerized diagnostic instruments or video documentation systems.

Device Story

Optos Advance is a web-based PACS application for ophthalmic image management. It interfaces with diagnostic instruments (Scanning Laser Ophthalmoscope, Fundus Camera, OCT) via secure network connections. Data is transmitted to a central Optos Advance Server, where a watchdog service processes and stores DICOM-compliant images and patient data. The server obfuscates patient information. Clinicians access the system via web browsers without requiring client-side software. The application allows viewing of studies, bilateral image comparison, manual magnification (zoom), and manual annotation (arrows/text). It does not perform automated image analysis or highlighting. The system supports archiving and secure remote sharing of data over HTTPS. By centralizing diagnostic data, it facilitates clinical review and decision-making for ophthalmic conditions.

Clinical Evidence

Bench testing only. Software and system verification and validation were performed according to the manufacturer's Software Development Lifecycle Process.

Technological Characteristics

Web-based PACS application; DICOM compatible; HTTPS protocol for secure remote data transfer; server-side data obfuscation; watchdog service for DICOM processing; browser-based client interface (no local installation); supports Scanning Laser Ophthalmoscope, Fundus Camera, and OCT inputs.

Indications for Use

Indicated for use by clinicians to store, manage, and display patient data, diagnostic data, videos, and images from computerized diagnostic instruments or video documentation systems.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 2 5 2012 ### 510(k) SUMMARY #### Optos' Optos Advance Software ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared | Sponsor: | Optos plc | |----------|--------------------------| | | Queensferry House | | | Carnegie Business Campus | | | Dunfermline | | | Fife KY11 8GR | | | United Kingdom | Phone: 01144 1383 843300 Fax: 01144 1383 843333 Contact Person: Graham McLeod MSc. Date Prepared: 16 December 2011 Name of Device: Optos Advance Picture Archiving and Communications System Common Name: Picture Archiving and Communications System Classification: (per 21 C.F.R. § 892.2050) NFJ Product Code: System, Image Management, Ophthalmic Device Device: Carl Zeiss Surgical GmbH FORUM Software (K090439) Predicate Device: #### Indications for use The Optos Advance Software is a web-based software system application intended for use in storing, managing, and displaying patient data, diagnostic data, videos and images from computerized diagnostic instruments or video documentation systems. #### Technological Characteristics The Optos Advance Software is an image review application that allows the user to view DICOM compatible images. The Optos Advance Software organizes the digitized images into studies and sessions. The digitized studies and sessions may be viewed by using an overall or bilateral display (i.e., display provides comparisons between old and new images). The software also allows {1}------------------------------------------------ physicians or technicians to create on-line notes with the ability to zoom in or out (i.e., enlarge or decrease magnification of images) and annotate such images with arrows or text boxes to highlight areas the user determines by training and experience to be of interest. The Optos Advance Software does not automatically highlight, annotate, or otherwise alter the images. In addition, the digitized studies and sessions and on-line notes may be viewed, archived on a central file server, or electronically shared in a secure manner with other health care professionals. The Optos Advance Software allows the user to interface a Scanning Laser Ophthalmoscope, Fundus Camera, OCT or other DICOM compliant diagnostic camera with the Optos Advanced Software via a secure network connection to the Optos Advance Server. The server has a watchdog service which processes the DICOM information sent via the network. The information is then processed and placed in local storage on the Optos Advance Server. The data is obfuscated at the server side to protect any patient information transferred. The Optos Advance Software does not require any customized software installed on the client PC. The images, annotations and DICOM data are all stored on the Optos Advance Server, which handles the archiving, reporting, and retrieval of the data. This server can be set up remotely over an HTTPS connection which allows secure remote transfer of the stored data for archive purposes. #### Performance Testing A full software and system verification and validation was performed as per the Software Development Lifecycle Process in Optos PLC. #### Substantial Equivalence The Optos Advance Software has the same intended use and substantially similar indications for use as the cleared Carl Zeiss Surgical GmbH FORUM Software (K090439). Minor differences in technological characteristics allow the user to access the software from any web enabled browser without the need for any customizable software installed on their personal computer and to archive their data to a remote server for retrieval over an HTTPS protocol driven server. Both Optos Advance and FORUM Software are intended for the review, storage and retrieval of images and image data supplied by ophthalmic medical devices. Thus, the minor differences between the Optos Advance Software and its predicate device raise no new questions of safety and effectiveness and the Optos Advance is substantially equivalent. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Optos Plc c/o Mr. Howard M. Holstein, Partner Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington DC 20004 # MAY 2 5 2012 Re: K113696 Trade/Device Name: Optos Advance Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: NFJ Dated: May 8, 2012 Received: May 8, 2012 # Dear Mr. Holstein: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dr has Intatutes and regulations administered by other Federal agencies. You must or uny I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of K F ar 807), acomig (21 CFR 803); good manufacturing practice requirements as set de rico-related adverse overies) (2 - Salation (21 CFR Part 820); and if applicable, the electronic {3}------------------------------------------------ Page 2 - Mr. Howard M. Holstein product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Kesia Alexander Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K 113696 Device Name: Optos Advance Software Indications for Use: The Optos Advance Software is a web-based software system application intended for use in storing, managing, and displaying patient data, diagnostic data, videos and images from computerized diagnostic instruments or video documentation systems. Prescription Use _ > (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sem (Division Sign-Off) (Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Page 1 of 1 510(k) Number_ 15
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...