OptosAdvance 4.0 is a standalone, browser-based software application intended for use by healthcare professionals to import, store, manage, display, analyze and measure data from ophthalmic diagnostic instruments, including: patient data, clinical images and information, reports, videos, and measurement of DICOM-compliant images.
Device Story
OptosAdvance 4.0 is a browser-based PACS software for ophthalmic diagnostic data. It imports DICOM-compliant images, reports, and videos from ophthalmic instruments (e.g., scanning laser ophthalmoscopes, fundus cameras, OCT units) via a network path to a server. Clinicians use a web browser to access, view, compare, annotate, and perform linear/area measurements on images. The system supports secure data transfer, local/remote archiving, and disaster recovery. By providing multi-dimensional visualization and measurement tools, it aids clinicians in analyzing ocular anatomy and pathology, facilitating clinical decision-making and patient history review.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed software performance, including measurement accuracy using DICOM objects with known dimensions. Comparison of image display between custom software and web client confirmed equivalent resolution and clarity. No clinical studies were required.
Technological Characteristics
Standalone, browser-based software application. Operates on standard computer systems via web browser. Supports DICOM (PS 3.1-3.20) standards. Connectivity via network storage and HTTPS for remote server access. Implements software life cycle processes per IEC 62304 and risk management per ISO 14971.
Indications for Use
Indicated for use by healthcare professionals to manage, display, and measure ophthalmic diagnostic data and DICOM-compliant images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three heads in profile, representing health, services, and people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 8, 2017
Optos Plc % Mr. Randy Prebula Partner Hogan Lovells U.S. LLP 555 13th Street NW Washington, DC 20004
Re: K162039
Trade/Device Name: OptosAdvance 4.0 (OA4) Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: NFJ Dated: December 29, 2016 Received: December 29, 2016
Dear Mr. Prebula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K162039
Device Name
OptosAdvance 4.0
Indications for Use (Describe)
OptosAdvance 4.0 is a standalone, browser-based software application intended for use by healthcare professionals to import, store, manage, display, and measure data from ophthalmic diagnostic instruments, including: patient data, diagnostic data, clinical images and information, reports, videos, and measurement of DICOMcompliant images.
Type of Use (Select one or both, as applicable)
区Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
### Optos Plc's OptosAdvance 4.0 Software
In accordance with 21 CFR §807.92 the following summary of information is provided:
| SUBMITTER: | Optos Plc<br>Queensferry House<br>Carnegie Campus<br>Enterprise Way, Dunfermline<br>Fife, KY11 8GR,<br>United Kingdom<br>FDA Registration Number: 9671767 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone:<br>Facsimile: | 011 44 1383 843300<br>011 44 1383 843333 |
| Primary Contact Person: | Paul Burns<br>Vice President, Regulatory Affairs & Quality Assurance<br>Optos Plc<br>Phone: +44 1383 843470<br>Email: pburns@optos.com |
| Secondary Contact Person: | Randy Prebula<br>U.S Agent<br>Hogan Lovells U.S. LLP<br>555 13th Street NW<br>Washington, DC 20004<br>U.S.A.<br>Phone: +1 202-637-6548<br>Facsimile: +1 202-637-5910<br>Email: randy.prebula@hoganlovells.com |
| Date Prepared: | July 22, 2016 |
| DEVICE: | |
| Trade Name: | OptosAdvance 4.0 (OA4) Software |
| Common or Usual Name: | Picture Archiving and Communication System (PACS) |
| Classification Name: | System, Image Management, Ophthalmic<br>(per 21 C.F.R. § 892.2050) |
| Device Classification: | Class II |
| Product Code: | NFJ |
{4}------------------------------------------------
## LEGALLY MARKETED PREDICATE DEVICES
Predicate Device Name: OptosAdvance Software 510(k) Number: K113696 Manufacturer: Optos Plc
Predicate Device Name: Synergy ODM 510(k) Number: K151952 Topcon Medical Systems, Inc Manufacturer:
### INTENDED USE / INDICATIONS FOR USE
OptosAdvance 4.0 is a standalone, browser-based software application intended for use by healthcare professionals to import, store, manage, display, analyze and measure data from ophthalmic diagnostic instruments, including: patient data, clinical images and information, reports, videos, and measurement of DICOM-compliant images.
### DEVICE DESCRIPTION
The OA4 software application provides multi-dimensional visualization of digital images to aid clinicians in their analysis of anatomy and pathology. The OptosAdvance user interface follows typical clinical workflow patterns to process, review, and analyze digital images.
The key features of OptosAdvance 4.0 include the ability to:
- · Acquire, store, retrieve and display DICOM image data;
- · Access patient data securely:
- · Search patient studies and select images for closer examination:
- Interactively manipulate an image to visualize anatomy and pathology;
- · Select multiple images for comparison;
- · Annotate, tag and record selected views;
- · Measure distance (linear) and area of DICOM images;
- Manage, backup and archive data;
- · Import and export data to network storage devices;
- · Securely access and transfer data: and
- Output selected views to printers.
The software relies on images being provided to a specified network path on the OptosAdvance Server by the connected ophthalmic device (Scanning Laser Ophthalmoscope, Fundus Camera, Optical Coherence Tomography unit, etc.) in a DICOM-compliant format. The software will then place the image and associated data on the network storage unit in a format which will allow the image to be available via a securely connected web browser. Locally archived studies will be securely pushed to the remote archive server for storage. The archive in the remote secure server serves as disaster recovery storage as well as access to the patient history.
#### PERFORMANCE TESTING
OptosAdvance 4.0 is designed and tested according to the software development lifecycle process implemented at Optos Plc. Software testing was completed to ensure the new feature operates according to requirements and without impact to existing functionality. Testing included verification, validation, and evaluation of previously acquired medical images.
{5}------------------------------------------------
For additional reassurance, equivalence tests were performed to verify that the OptosAdvance 4.0 software application provides equivalent measurements by loading DICOM objects that contain features with known dimensions, displaying these objects to the user, having the user measure these features, and comparing their measurement with the known dimensions.
#### Risk Management
Each risk pertaining to OptosAdvance 4.0 has been individually assessed to determine if the benefits outweigh the risk. Every risk has been reduced as low as possible and has been evaluated to have a probability of occurrence of harm of no more than "Remote". All risks for the updated software feature were collectively reviewed to determine if the benefits outweigh the risk.
The proposed device has been designed and tested taking into account potential cybersecurity risks, to ensure confidentiality, integrity, availability and accountability.
### Standards Compliance
| Standard No. | Standards organization | Standards Title |
|-----------------------------|------------------------|-----------------------------------------------------------------------------------|
| PS 3.1 - 3.20 (2014c) | NEMA | Digital Imaging and Communications in Medicine (DICOM) Set |
| 14971:2012 | AAMI / ANSI / ISO | Medical Devices – Applications of risk management to medical devices |
| 62304:2006 | AAMI / ANSI / IEC | Medical Device Software - Software life cycle processes |
| 62366-1 Edition 1.0 2015-02 | IEC | Medical Devices – Part 1: Application of usability engineering to medical devices |
The OptosAdvance 4.0 software complies with the following voluntary standards:
#### Clinical Data
OptosAdvance 4.0 is a PACS system that processes DICOM data. Therefore, any clinical comparison with a predicate that operates per the same DICOM standard will produce the same Thus, clinical studies are not required to support the subject device's safety and result. effectiveness; the non-clinical objective test methods used for evaluation demonstrate that the software's performance is equivalent to that of the legally marketed predicates. Nevertheless, the anticipated clinical equivalence was confirmed with a clinical case.
### SUBSTANTIAL EQUIVALENCE
The proposed device is of the same type, common name, classification, and product code as the cleared predicate devices. The devices have the same intended use, and the indications for use of OptosAdvance 4.0 are very similar to the indications for use of the predicate software devices.
OptosAdvance 4.0 is also functionally equivalent to these devices. The proposed and predicate devices are all software-only image management systems with similar acquisition, importing, viewing, measurement and analysis, network and security, print, archive and backup functionalities. The minor technological differences between OptosAdvance 4.0 and its predicate devices raise no new issues of safety or effectiveness, as described below.
{6}------------------------------------------------
The key difference between the predicate OptosAdvance software (K113696) and the new version of OptosAdvance is the introduction of features to measure distance (linear) and area of DICOM images. The predicate Synergy ODM device (K151952) provides such line and area measurement capabilities, so the inclusion of this feature in the subject software does not raise different questions of safety or effectiveness.
The devices also differ slightly in their approach to data management backup and archiving. While the cleared Synergy ODM uses locally installed customized software to allow the user to view images stored on the local or networked computer system, OptosAdvance 4.0 uses the computer system's web browser with no other software installation required to allow the user to view DICOMcompliant images stored on the server. The verification and validation testing for the subject software covered the comparison of images on custom software and the web client, which were found to display at the same resolution and clarity. OptosAdvance 4.0 also allows the user to connect to a remote secure server to allow greater archiving flexibility. The only minor difference to OptosAdvance 4.0 from Synergy ODM in this respect is that the user can set up an HTTPS connection to the remote server to transfer data to and from the archive, providing greater flexibility on storage. A similar mechanism exists in the Synergy ODM software, where data can be stored on an external network location and accessed externally. In order to do this remotely in the OptosAdvance 4.0 software, a secure protocol is introduced to protect data integrity.
### CONCLUSION
OptosAdvance 4.0 has the same intended use and indications for use as the Optos predicate device (K113696), and has very similar principles of operation and technological characteristics as both predicate devices (K113696 and K151952). Any minor differences noted have been explained and do not raise any different questions of safety or effectiveness when the software is used as labelled. The implemented design controls, risk management activities, labelling, and performed tests (software verification & validation) demonstrate the safety and efficacy of the device in comparison to the predicates. Therefore, based on the comparison and test data, the OptosAdvance 4.0 software application is as safe and effective as, and substantially equivalent to, the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.