K150722 · Bioptigen, Inc. · HLI · Dec 2, 2015 · Ophthalmic
Device Facts
Record ID
K150722
Device Name
EnFocus 2300, EnFocus 4400
Applicant
Bioptigen, Inc.
Product Code
HLI · Ophthalmic
Decision Date
Dec 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1570
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Bioptigen EnFocusTM device is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT). The EnFocusTM is indicated for use as an aid in the visualization of physiologic and pathologic conditions of the eye through non-contact optical imaging. It is indicated for use on patient populations from premature and neonatal infants to adult. The system is indicated for use in supine imaging, mounted to a surgical microscope, with cooperative patients or patients under anesthesia.
Device Story
EnFocus is a non-contact, non-invasive ophthalmic imaging system using Spectral Domain Optical Coherence Tomography (SD-OCT) and a near-infrared light source. It consists of a scan head, system computer, UPS, mobile cart, and InVivoVue software. The scan head mounts to a surgical microscope (e.g., Leica M844) to provide real-time OCT imaging during ophthalmic surgery without interrupting the procedure or repositioning the microscope. The device captures depth-resolved images of ocular tissue microstructure. Surgeons view the output on a display to visualize ocular physiology and pathology, including Doppler blood flow. It aids clinical decision-making by providing intraoperative visualization of ocular structures. Benefits include non-invasive imaging, integration into existing surgical workflows, and the ability to perform measurements using on-screen calipers.
Clinical Evidence
Clinical study of 24 eyes from 12 adult subjects compared EnFocus 2300/4400 to the Envisu 2300 predicate. Primary endpoints included visualization of ocular physiology (anterior/posterior segments), Doppler blood flow, and manual caliper measurements. Results showed high agreement with the predicate for identifying physiologic features (e.g., 100% agreement for RPE, inner nuclear layer). Caliper measurements met the equivalence margin of +/- 0.015 mm at the 95% CI. Linearity analysis showed R^2 of 99.8% for measurements between 20-630 um.
Technological Characteristics
Spectral Domain OCT; near-infrared light source; Class 1 laser product (IEC 60825-1). Includes scan head, system computer, UPS, and mobile cart. Electromechanical focus/zoom control. Compatible with surgical microscopes. Software: InVivoVue 2.6. Connectivity: Standalone system. Sterilization: Not applicable (non-contact).
Indications for Use
Indicated for visualization of physiologic and pathologic ocular conditions in patients ranging from premature/neonatal infants to adults. Used in supine position, mounted to a surgical microscope, for cooperative patients or those under anesthesia.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 3, 2015
Bioptigen, Inc. Ms. Tammy B. Carrea Vice President, Quality and Regulatory Affairs 633 Davis Drive, Suite 480 Morrisville, North Carolina 27560
Re: K150722
Trade/Device Name: Enfocus 2300. Enfocus 4400 Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: HLI, OBO Dated: October 23, 2015 Received: October 29, 2015
Dear Ms. Carrea:
This letter corrects our substantially equivalent letter of December 12, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150722
Device Name EnFocus
Indications for Use (Describe)
The Bioptigen EnFocusTM device is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT).
The EnFocusTM is indicated for use as an aid in the visualization of physiologic conditions of the eye through non-contact optical imaging. It is indicated for use on patient promotions from premature and neonant lintants to adult. The system is indicated for use in supine imaging, mounted to a superiod microscope, with coperative patients or patients under anesthesia.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review the salleding data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "bioptigen" with three horizontal bars in shades of blue and green to the left of the name. Below the logo is the text "510(k) SUMMARY (per 21 CFR 807.92)". The text is in bold font.
# K150722
Date Prepared: November 24, 2015
### SUBMITTER INFORMATION l.
| Sponsor: | Bioptigen, Inc. |
|----------------------|------------------------------------------------------|
| Registration Number: | 3006809695 |
| Address: | 633 Davis Drive, Suite 480<br>Morrisville, NC 27560 |
| Phone: | 919-595-1260 |
| Contact Person: | Tammy B. Carrea<br>VP Quality and Regulatory Affairs |
### II. DEVICE IDENTIFICATION
| Trade or proprietary name: | EnFocus™ |
|----------------------------|------------------------------------------------------------------------|
| Common or usual name: | Optical Coherence Tomography |
| Classification Name: | Ophthalmoscope, AC Powered<br>Optical Coherence Tomography |
| Regulation: | 21 CFR 886.1570 |
| Classification: | Class II |
| Product Codes: | HLI (Ophthalmoscope, AC-powered)<br>OBO (Tomography, Optical Coherence |
| Classification Panel: | Ophthalmic |
### PREDICATE DEVICE lll.
Envisu™ SDOIS (Models C2200 and C2300) K120057
#### IV. DEVICE DESCRIPTION
The EnFocus™ is a non-contact, noninvasive ophthalmic imaging device that includes an OCT Engine, a scan head, a System Computer, an Uninterruptible Power Supply (UPS), a mobile Security Cart and the System Software. The EnFocus™ uses Spectral Domain Optical Coherence Tomography (SD-OCT) and a near infrared light source to image ocular tissue microstructures
The EnFocus is coupled to a surgical micrsoscope for OCT imaging during ophthalmic surgical procedures. The EnFocus has been validated and found to be compatible for use with the Leica
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "bioptigen". The logo consists of a stylized "B" made of three horizontal bars in different shades of blue and green, followed by the word "bioptigen" in a bold, sans-serif font. The colors of the "B" are, from top to bottom, blue, teal, and green.
M844 Surgical Microscope and the Insight Instruments Super View™ Wide Angle Viewing System™ for retina visualization.
The software, InVivoVue™ Version 2.6, works with the hardware and the hardware controller to offer intuitive, flexible system control for high-speed volume data acquisition and imaging.
The EnFocus™ system includes two OCT-compatible objective lenses for use with the surgical microscope: a 175mm lens and 200mm lens. The system also offers a choice of accessory masks that may be deployed in the Leica M844 filter port to manage illumination glare artifacts when necessary.
Using the EnFocus™, OCT imaging may be acquired during the surgical procedure, without stopping a procedure or repositioning the surgical microscope. The surgical microscope position is stationary relative to the surgical procedure, and the surgical view is unaltered by the scanning of the OCT beam.
#### INDICATIONS FOR USE V.
The Bioptigen EnFocus™ device is intended to acquire, process, display and save depthresolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT).
The EnFocus™is indicated for use as an aid in the visualization of physiologic and pathologic conditions of the eye through non-contact optical imaging. It is indicated for use on patient populations from premature and neonatal infants to adult. The system is indicated for use in supine imaging, mounted to a surgical microscope, with cooperative patients under anesthesia.
#### VI. SUBSTANTIAL EQUIVALENCE
Like its predicate, the EnFocus™ is a non-contact, non-invasive ophthalmic imaging device that can be used to view ocular tissue physiology and pathology through the interaction of light with the optical scattering properties of structures of the eye using SDOCT and a Class 1 near-infrared laser light source to capture depth-resolved images of ocular tissue microstructure. Both devices utilize SDOCT, include a scan head, a System Computer, an Uninterruptible Power Supply (UPS), a mobile Security Cart and the InVivoVue System Software. The design modifications represented by the subject EnFocus are, as follows:
# Scan Head
The scan head is mounted to the microscope's optical carrier using a mounting bracket that is bolt-secured to the carrier's accessory plate and shares an objective lens with the microscope.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for bioptigen. The logo consists of a stylized graphic element on the left and the word "bioptigen" in bold, black font on the right. The graphic element is composed of three overlapping shapes in different shades of blue and green, creating a sense of depth and movement.
### Focus and Zoom Control
The predicate device utilizes a manual system for focusing the OCT beam. Focus of the predicate device is manually achieved by mechanical displacement of lenses in the system objective lens, similar to manual operation of a SLR camera lens.
The subject device provides an electromechanical system for focusing the OCT beam and for changing the numerical aperture of the OCT beam. Focus and numerical aperture are usercontrolled via an input beam zoom setting that is accessed through the software user interface.
### Retina Viewing Lens
Retina viewing with the predicate device is accomplished using the General Retina Lens, one of multiple lenses supplied by Bioptigen with the Envisu SDOIS system. Retina viewing with the EnFocus is accomplished using a commercially available Fundus (Retina) Viewing Lens.
### Spectrometer
The predicate device is available with one of two spectrometers (2200 or 2300) and one of two internal light sources (HR or VHR). The EnFocus device is available in two models: EnFocus 2300 includes the 2300 spectrometer of the predicate and the VHR superluminescent diode (SLD) light source of the predicate. EnFocus 4400 includes a 4400 spectrometer and utilizes an HR superluminescent light source equivalent to the HR source available with the predicate.
### Engine Control System
The predicate system relied on manual control functionality of reference arm path length, reference arm attenuation level and polarization balance through mechanical linkages in the system. The EnFocus replaces theses mechanical linkages with a motor-driven electromechanical system that can be controlled through a menu on the front panel of the EnFocus engine or through the software user interface.
### Software User Interface and Function
No changes have been made to the predicate's underlying method of image acquisition, signal processing or display technology.
Software modifications include migration to the Microsoft 7 64-bit Operating System and simplification of both user interface and workflow. Examples of these simplifications include Office 2010 and 2013 support for reports, enabling simultaneous exporting of multiple exams, rearranging the user interface layout to follow standard workflows while reducing the steps to begin a new patient exam and associating an exam with patient demographic data.
Two new user interface features have been added including a configuration drop down menu and a prominently displayed "Quick Start" grid. The drop down menu simplifies switching
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "bioptigen". The logo consists of a stylized "b" made up of three stacked, rounded rectangles in blue and green, followed by the word "bioptigen" in black, sans-serif font. The rectangles are arranged with the top one in blue, the middle one in a darker green, and the bottom one in a lighter green.
between system configurations for anterior and posterior imaging without requiring restarting InVivoVue software.
Any technological differences between the EnFocus and Envisu devices have been mitigated via bench testing or clinical testing.
#### VII. NON-CLINICAL PERFORMANCE DATA
# Medical Electrical Safety Testing
Electrical safety and EMI/EMC testing were conducted on the EnFocus and demonstrate the system complies with the IEC 60601-1 and IEC 60601-1-2 standards, 30d edition. Testing included review of the software, risk management file, and usability file.
# Optical Safety and Emissions
IEC 60825-1 Class 1 laser product safety testing and ISO 15004-2 Group 2 light hazard test results together confirm the system to be eye safe when used as labeled.
# Software Verification and Validation Testing
Software testing was validated and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and in accordance with the software lifecycle requirements of ISO 62304.
# Key Performance Attributes
Optical performance testing was conducted in accordance with established key performance attributes. All testing demonstrated compliance with applicable product requirements.
# Surgical Microscope Compatibility
The EnFocus was connected to the microscope and the two EnFocus device models were evaluated for compatibility with the microscope. Compatibility was assessed by evaluating the ability of the microscope and EnFocus to work independently and by ensuring that the EnFocus did not result in any form of interference with the microscope that could impact surgical performance. The microscope and EnFocus were tested together for optical radiation safety to ISO 15004-2, mechanical connectivity, mechanical stability, ergonomics and working distance, and general safety. Microscope properties were evaluated for optical performance including illumination field of view, illumination uniformity, optical distortion, and glare. Installation and removal, including draping, were evaluated. The surgical microscope labeling was evaluated to ensure that the microscope was not contraindicated for pediatric use. Illumination field of view and uniformity were evaluated with and without the accessory glare masks installed. All tests for compatibility passed acceptance criteria.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Bioptigen. The logo consists of a blue and green graphic to the left of the word "bioptigen" in black font. The graphic is made up of three stacked shapes, with the top shape being blue, the middle shape being a darker green, and the bottom shape being a lighter green.
# Fundus Viewing System Compatibility
The optical resolution of the surgical microscope with the fundus viewing system installed was evaluated with and without the EnFocus installed. Both of the two EnFocus objectives were evaluated at the maximum and minimum microscope magnifications demonstrated equivalent fundus imaging optical resolution with and without the EnFocus device installed.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "bioptigen". The logo consists of a stylized "b" symbol made up of three horizontal bars in different shades of blue and green, followed by the word "bioptigen" in a bold, sans-serif font. The color scheme is modern and professional, with the blue and green hues suggesting a connection to science or technology.
# Performance Specifications
| Attribute | Predicate Envisu Device<br>K120057<br>Models 2200/2300 | Proposed EnFocus Device<br>K150722<br>Models 2300/4400 | |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Bioptigen SDOIS is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT). The Envisu SDOIS in indicated for use as an aid in the diagnosis of physiological and pathologic conditions of the eye through non-contact optical imaging. Imaging of the various tissues of the eye is supported through the use of interchangeable lenses. It is indicated for use on patient population from premature and neonatal infants to adult, and is suitable for patients ambulatory or confined. The system is indicated for use in upright or supine imaging, handheld or mounted, and is suited for imaging patients under anesthesia. | The Bioptigen EnFocus™ device is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT). The EnFocus™ is indicated for use as an aid in the visualization of physiologic and pathologic conditions of the eye through non-contact optical imaging. It is indicated for use on patient populations from premature and neonatal infants to adult. The system is indicated for use in supine imaging, mounted to a surgical microscope, with cooperative patients or patients under anesthesia. | |
| Regulation Number | 21 CFR 886.1570 | Same | |
| Class | II | Same | |
| Method of Operation | Spectral Domain Optical Coherence Tomography | Same | |
| Light Source | SLD<br>(Superluminescent diode) | Same | |
| Scanner Type | Galvonometric mirror pair | Same | |
| Optical Power | < 750 μW | Same | |
| Resolution, Lateral | Retina: 20 µm in tissue<br>Anterior: 9, 12, 25 µm in tissue | < 31.0 µm (175 mm Obj., low NA)<br>< 15.1 µm (175 mm Obj., high NA)<br>< 35.4 µm (200 mm Obj., low NA)<br>< 17.3 µm (200 mm Obj., high NA) | |
| Resolution, Axial | | HR: ≤ 6 μm in tissue<br>VHR: <4 μm in tissue | HR: ≤ 9μm<br>VHR: ≤ 4 μm |
| Field of View<br>(air/tissue)<br>(aka depth range) | | Model C2200: 2.3/1.7 mm<br>Model C2300: 3.4/2.5 mm | Model 2300: 3.4/2.5 mm<br>Model 4400: 15.3/11.1 mm |
| Scan<br>Pixels | Axial | Model C2200: 1024<br>Model C2300: 1024 | Model 2300: 1024<br>Model 4400: 2048 pixels |
| | Lateral | User selectable<br>A-scans/B-scan: 5000 maximum<br>Maximum: 150,000 total A-scans | User selectable<br>A-scans/B-scan: 2000 maximum<br>Maximum A-scans/volume<br>Model 2300: ≥1,000,000<br>Model 4400: ≥ 500,000 |
| Scan Rate | | ≥32,000 A-scans/s | Model 2300: ≥32,000 A-scans/s /<br>Model 4400: ≥18,000 A-scans/sec |
| Scan Patterns | | Line, rectangular volume, circle, concentric rings, radial lines | Same |
| Calipers | | Manual placement of on-screen calipers | Same |
| Doppler | | Qualitative blood flow visualization with<br>color Doppler OCT | Same |
| Software | | InVivoVue 1.4 | InVivoVue 2.6 |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Bioptigen. The logo consists of a stylized graphic element on the left and the word "bioptigen" on the right. The graphic element is composed of three curved shapes stacked on top of each other, colored in blue and green. The word "bioptigen" is written in a bold, sans-serif font and is colored in black.
#### CLINICAL DATA VIII.
Clinical data were collected in an IRB approved study to support the indications for use statement for the EnFocus 2300 and EnFocus 4400 systems and to demonstrate substantial equivalence as compared to the predicate Envisu 2300 device with regard to the ability to:
- a) visualize ocular physiology of the anterior and posterior segments of the eye
- b) visualize vascular blood flow in the retina with Doppler OCT
- c) perform measurements of ocular features using manual placement of on-screen calipers.
The study evaluated 24 eyes of twelve adult subjects. Subjects were imaged in an office setting beneath a Leica M844 microscope, without mydriasis. Images were coded and randomized. Following collection of images, three ophthalmic graders independently reviewed the images and documented evaluations
### Ocular Physiology and Presence of Doppler Flow
Three ophthalmic graders evaluated four volumetric images acquired of each subject eve and identified the presence or absence of physiologic features in a binary test of agreement. The ability to visualize and identify the physiologic structures are presented in the following table:
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for bioptigen. The logo consists of a graphic element on the left and the word "bioptigen" on the right. The graphic element is made up of three overlapping shapes in different shades of blue and green. The word "bioptigen" is written in a bold, sans-serif font.
| ID | Feature | N | Percent Agreement with Predicate | N | Percent Agreement with Predicate |
|-----|----------------------------|---------|----------------------------------|---------|----------------------------------|
| | | ENF2300 | | ENF4400 | |
| Q2 | Inner limiting membrane | 71 | 81.7% | 72 | 80.6% |
| Q3 | Parafov. nerve fiber layer | 70 | 100.0% | 72 | 100.0% |
| Q4 | Inner nuclear layer | 71 | 100.0% | 72 | 100.0% |
| Q5 | Outer plexiform layer | 71 | 100.0% | 72 | 100.0% |
| Q6 | Ext. limiting membrane | 71 | 100.0% | 72 | 100.0% |
| Q7 | IS/OS Ellipsoids | 71 | 100.0% | 72 | 100.0% |
| Q8 | End Tips Photoreceptor | 71 | 98.6% | 72 | 94.4% |
| Q9 | Retina pigment epithelium | 71 | 100.0% | 72 | 100.0% |
| Q10 | Choriocapillaris | 71 | 100.0% | 71 | 100.0% |
| Q11 | Chorioscleral interface | 72 | 66.7% | 72 | 62.5% |
| Q14 | Cornea epithelium | 70 | 100.0% | 71 | 100.0% |
| Q15 | Bowman's layer | 70 | 94.3% | 71 | 98.6% |
| Q16 | Cornea endothelium | 70 | 98.6% | 71 | 98.6% |
| Q19 | Scleral corneal junction | 72 | 98.6% | 72 | 97.2% |
| Q20 | Schlemm's canal | 70 | 60.0% | 71 | 57.7% |
| Q21 | Iridocorneal angle | 71 | 81.7% | 72 | 63.9% |
| Q22 | Doppler flow, superior | 72 | 93.1% | 72 | 95.8% |
| Q23 | Doppler flow, central | 72 | 88.9% | 72 | 88.9% |
| Q24 | Doppler flow, inferior | 72 | 95.8% | 72 | 95.8% |
# Caliper Measurements
Graders were asked to use InVivoVue calipers to measure a series of features in the same images. The images were presented as acquired, without magnification, or image processing or image enhancement. The caliper measurements from images acquired with each of the EnFocus 2300 and EnFocus 4400 were analyzed for agreement with images acquired with the predicate Envisu 2300. Agreement was tested against a target equivalence margin of +/- 15 um (0.015 mm) at the 95% confidence level. Results are tabulated in the following table:
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "bioptigen" in a bold, sans-serif font. To the left of the word is a graphic consisting of three stacked, rounded rectangles. The top rectangle is blue, the middle is a darker green, and the bottom is a lighter green. The word "bioptigen" is in black.
| New<br>Device | Measurement | N | Mean Difference<br>(Predicate - New<br>Device (mm)) | SD | Lower<br>95% CI | Upper<br>95% CI | 95% CI Meets<br>Equivalence Margin?<br>(+/- 0.015 mm) |
|---------------|---------------------------------------------------------|----|-----------------------------------------------------|--------|-----------------|-----------------|-------------------------------------------------------|
| ENF2300 | Parafoveal<br>peak, inferior | 24 | 0.0064 | 0.0047 | 0.0045 | 0.0084 | Y |
| | Parafoveal<br>peak,<br>superior | 24 | 0.0036 | 0.0056 | 0.0013 | 0.0060 | Y |
| | Fovea | 24 | 0.0013 | 0.0064 | -0.0014 | 0.0040 | Y |
| | Nerve Fiber<br>Layer, 1mm<br>from<br>parafoveal<br>peak | 24 | 0.0026 | 0.0047 | 0.0006 | 0.0046 | Y |
| | Cornea | 24 | 0.0085 | 0.0063 | 0.0059 | 0.0112 | Y |
| | Cornea<br>Epithelium | 24 | 0.0044 | 0.0022 | 0.0035 | 0.0053 | Y |
| ENF4400 | Parafoveal<br>peak, inferior | 24 | -0.0009 | 0.0045 | -0.0028 | 0.0010 | Y |
| | Parafoveal<br>peak,<br>superior | 24 | -0.0026 | 0.0054 | -0.0049 | -0.0004 | Y |
| | Fovea | 24 | -0.0051 | 0.0085 | -0.0087 | -0.0015 | Y |
| | Nerve Fiber<br>Layer, 1mm<br>from<br>parafoveal<br>peak | 24 | -0.0025 | 0.0049 | -0.0045 | -0.0004 | Y |
| | Cornea | 24 | -0.0028 | 0.0081 | -0.0062 | 0.0006 | Y |
| | Cornea<br>Epithelium | 24 | 0.0020 | 0.0030 | 0.0008 | 0.0033 | Y |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for bioptigen. The logo consists of a stylized "b" made up of two overlapping shapes in blue and green, followed by the word "bioptigen" in black, sans-serif font. The blue shape is on top and the green shape is on the bottom.
The linearity in agreement between measures across the range of measurements (20 um to 630 um) was also evaluated. The best linear fits (Device2 = A*Device1 + B) and correlation coefficient R^2 are presented in the following table.
| Accuracy | A (%) | B (μm) | R2 (%) | |
|-------------------|----------------|--------|--------|-------|
| Pooled<br>Graders | ENF23=f(Pred) | 99.0% | -2.0 | 99.8% |
| | ENF44=f(Pred) | 100.2% | 1.4 | 99.8% |
| | ENF44=f(ENF23) | 101.2% | 3.5 | 99.8% |
#### IX. CONCLUSION
The EnFocus™ device shares the same fundamental technology as the predicate Envisu SDOIS and has a similar intended use.
The conclusion drawn from the test data is that the EnFocus device is as safe and effective as the predicate device and performs similarly to the predicate device. Any differences in technology have been tested and verified and do not raise any new issues of safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.