ENVISU SPECTRAL DOMAIN OPHTHALMIC IMAGING SYSTEM (SDOIS)
K120057 · Bioptigen, Inc. · HLI · May 11, 2012 · Ophthalmic
Device Facts
Record ID
K120057
Device Name
ENVISU SPECTRAL DOMAIN OPHTHALMIC IMAGING SYSTEM (SDOIS)
Applicant
Bioptigen, Inc.
Product Code
HLI · Ophthalmic
Decision Date
May 11, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1570
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K120057 · May 11, 2012
ENVISU SPECTRAL DOMAIN OPHTHALMIC IMAGING SYSTEM (SDOIS)
Bioptigen, Inc.
Peer-reviewed clinical literature
The sponsor utilized a review of published clinical literature to demonstrate the safety and effectiveness of the handheld SDOIS device for pediatric populations, as no independent clinical testing was conducted by the sponsor.
Pediatric population; Handheld imaging; Clinical literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Review of existing peer-reviewed clinical literature; Literature review
Pediatric populations
Not applicable for this study
Safety and effectiveness of handheld SDOIS
Indications for Use
Bioptigen Envisu™ Spectral Domain Ophthalmic Imaging System (Envisu™ SDOIS) is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT). The Envisu™ SDOIS is indicated for use as an aid in the diagnosis of physiologic and pathologic conditions of the eye through non-contact optical imaging. Imaging of the various tissues of the eye is supported through the use of interchangeable lenses. It is indicated for use on patient populations from premature and neonatal infants to adult, and is suitable for patients ambulatory or confined. The system is indicated for use in upright or supine imaging, handheld or mounted, and is suited for imaging patients under anesthesia.
Device Story
Envisu SDOIS is a mobile, non-contact, near-infrared spectral domain optical coherence tomography (SD-OCT) system. It uses a low-power broadband SLED light source (760-930 nm) and a spectrometer with a line-scan camera to capture backscattered radiation from ocular tissues. The system transforms these signals into depth-resolved images of retinal, corneal, scleral, and conjunctival microstructures. Operated by clinicians in clinical settings, the device features a lightweight handheld scanner or a mounted configuration for patient comfort. InVivoVue Clinic software processes and displays images, allowing for registration, averaging, manual caliper measurements, 3D visualization, and report generation. Output assists clinicians in diagnosing ocular conditions by visualizing tissue scattering properties. Benefits include flexible imaging for pediatric, supine, or anesthetized patients without requiring a chin rest.
Clinical Evidence
No independent clinical trials conducted by the manufacturer. Evidence relies on a review of peer-reviewed clinical literature, which reported the handheld SDOIS as a safe, non-invasive, and effective method for evaluating the eye in pediatric populations. Bench testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), optical emission safety (IEC 60825-1, ISO 15004-2), image resolution, and software validation.
Technological Characteristics
Spectral Domain OCT; SLED light source (760-930 nm); Class 1 LED; Galvanometric mirror scanner; Transmission grating/spectrometer detection; 32,000 A-scans/s; Axial resolution <4-6 µm; Lateral resolution 9-25 µm; Imaging depth up to 3.4 mm (air); Windows-based software (InVivoVue Clinic); Mobile cart with UPS; Handheld or mounted form factor.
Indications for Use
Indicated for aid in diagnosis of physiologic and pathologic eye conditions via non-contact optical imaging of ocular tissue microstructure. Applicable to all patient populations from premature and neonatal infants to adults; suitable for ambulatory, confined, or anesthetized patients.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
Spectral Domain Ophthalmic Imaging System (SDOIS) (K063343)
Submission Summary (Full Text)
{0}------------------------------------------------
K120057
New Traditional 510(k)
bioptigen
MAY 1 1 2012
#### 510(K) SUMMARY (PER 21 CFR 807.92) -
Date Prepared: December 23, 2011
#### SUBMITTER INFORMATION
| 510(k) Owner and Device Manufacturer: | Bioptigen, Inc. |
|---------------------------------------|--------------------------------------------------|
| Registration Number: | 3006809695 |
| Address: | 104 T.W. Alexander, Building 2, Durham, NC 27709 |
| Phone: | +1 919 314 5500 |
| Fax: | +1 919 314 5501 |
Contact:
Dawn Reilly-O'Dell, RAC Full Circle Regulatory Consulting, LLC +1 919 410 7497 (Phone) dreilly@bioptigen.com (Email)
#### DEVICE IDENTIFICATION
Trade or proprietary name:
Envisu™ Spectral Domain Ophthalmic Imaging System (SDOIS) Spectral Domain Optical Coherence Tomography (SD-OCT) Common or usual name: Ophthalmoscope, AC-powered; Classification name: Tomography, Optical Coherence ни, ово Product Codes: 886.1570 Regulation Number:
#### INTENDED USE
Bioptigen Envisu™ Spectral Domain Ophthalmic Imaging System (SDOIS) is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SD-OCT).
The Envisu SDOIS is indicated for use as an aid in the diagnosis of physiologic and pathologic conditions of the eye through non-contact optical imaging of the various tissues of the eye is supported through the use of interchangeable lenses. It is indicated for use on patient populations from premature and neonatal infants to adult, and is suitable for patients ambulatory or confined. The system is indicated for use in upright or supine imaging, handheld or mounted, and is suited for imaģing patients under anesthesia.
{1}------------------------------------------------
New Traditional 510(k)
# bioptigen
#### SUBSTANTIAL EQUIVALENCE
The Envisu™ SDOIS is substantially equivalent to the predicate SDOIS in technology, safety and effectiveness. The Envisu™ SDOIS incorporates a number of design changes, but it shares the same fundamental spectral domain optical coherence tomography imaging technology and incoherent superluminiscent light emitting diode technology as the predicate SDOIS. The differences between the 2006 predicate, K063343, and the current subject SDOIS are the following design modifications incorporated to enhance device usability and performance:
- . Upgraded spectrometer line-scan camera allowing 32,000 A-scans/second for faster imaging speeds with no increase in patient irradiance
- . Added VHR light source option for improved axial resolution without increase in patient irradiant exposure
- . Revised scan head with enhanced ergonomics including handheld operation for improved clinical flexibility and patient comfort
- Mobile transport cart with Uninterruptable Power Supply (UPS)
- . Increased computer memory and processing speed
- . Additional objective lens options for cornea and related anterior imaging
- Updated Information for Use labeling to parallel the system changes and a new training manual to address use of the handheld scanner
- Added spectrometer design option for increased imaging depth to 3.4 mm in air (2.5 mm in tissue) for greater depth of imaging field
- . Upgraded IVVC software post-processing and visualization functionality including four new. features, which are the option to register and average multiple frames, the manual Onscreen Calipers, the 3D view and the option to generate Word-based reports
| Predicate Device | 510(k) Holder | 510(k) | Clearance Date |
|-----------------------------------------------------|-----------------|---------|-------------------|
| Spectral Domain Ophthalmic Imaging System (SDOIS) . | Bioptigen, Inc. | K063343 | December 13, 2006 |
{2}------------------------------------------------
New Traditional 510(k)
#### Envisu SDOIS
## bioptigen
Predicate
SDOIS
#### Technological Comparison:
#### Medical Device
Indications for Use
Spectral Domain Imaging The System (SDOIS) is intended to acquire, process, display and save depth-resolved images of ocular microstructure tissue using Spectral Domain Optical Coherence Tomography (SD-OCT). It is primarily intended for the imaging of retinal tissue, but the cornea, sclera and conjunctiva can also be imaged by changing the focal position. Indications for use evaluation include the of ophthalmic tissue in routine clinical examinations and as an aid in the diagnosis of conditions that affect the optical scattering properties of ocular tissue.
#### Subject Envisu™ SDOIS
Bioptigen Envisu™ Spectral Domain Ophthalmic Imaging system is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT). The Envisu™ SDOIS is indicated for use as an aid in the diagnosis of physiologic and pathologic conditions of the eye through non-contact optical imaging. Imaging of the various tissues of the eye is supported through the use of interchangeable lenses. It is indicated for use on patient populations from premature and neonatal infants to adult, and is suitable for patients ambulatory or confined. The system is indicated for use in upright or supine imaging, handheld or mounted, and is suited for imaging patients under anesthesia.
#### Categories
Method of Operation Light Source
Light Source Classification
Optical Power
Resolution, Lateral
Resolution, Axial
Depth Range (in tissue/air)
Scanner Type
#### Specifications
| SD-OCT | Same as predicate | |
|-------------------------------------------------------|----------------------------------------------------------------|--|
| SLED | SLED | |
| High Resolution (HR) | HR: Same as predicate<br>VHR: Very High Resolution | |
| Class 1 LED | HR: Same as predicate | |
| | VHR: Class 1 LED | |
| < 750 µW at cornea | Same as predicate | |
| Retina: 20 µm in tissue<br>Anterior: ~25 um in tissue | Retina: same as predicate<br>Anterior Segment: 9, 12 and 25 um | |
| < 6 um in tissue | HR: Same as predicate<br>VHR: <4 µm in tissue | |
| 1.7 / 2.2 mm | Model C2200: 1.7 /2.3<br>Model C2300: 2.5 / 3.4 mm | |
| Galvanometric mirror pair | Same as predicate | |
{3}------------------------------------------------
| Scan Patterns Available | Line, rectangular volume, circle, concentric rings, radial lines | Same as predicate |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Scan Pixels | Axial (depth): 512 or 1024<br>Lateral : User Selectable<br>Maximum 5,000 A-scans / B-scan<br>Maximum 150,000 total A-scans | Same as predicate |
| Scan Rate | 20,000 A-scans/s | 32,000 A-scans/s |
| Detection | Transmission Grating<br>Spectrometer / Line-Scan Camera | Same as predicate |
| Scanner Ergonomics | Mounted (tabletop) | Same as predicate or handheld |
| Patient Interface | Chin Rest Assembly | Same as predicate or None (Chin<br>Rest Assembly now optional) |
| Footprint | Stationary: 15" x 18"<br>Engine, 12" x 15" scanner | Mobile: 24" x 22" x 37.5" |
| Scanner Dimensions | 12" (h) x 6" (w) x 9"(d) | 7" (h) x 3" (w) x 9" (d)<br>h: incl handle / d: incl lens<br>Weight: ≤ 3.5 lbs |
| Software | InVivoVue™ 1.5 | InVivoVue™ Clinic 1.4 |
| Operating System | Windows XP | Same as predicate |
| Processor | Dual 3.4 GHz Xeon | Dual 2.0 GHz Quad Core |
| Memory | 2 GB | 4 GB |
#### DEVICE DESCRIPTION
New Traditional 510(k)
The Envisu™ SDOIS is a non-contact, non-invasive and mobile ophthalmic imaging device that can be used to view ocular tissue physiology and pathology through the interaction of light with the optical scattering properties of structures of the eye.
Like its predicate, the Envisu™ Spectral Domain Ophthalmic Imaging System (SDOIS) is a nearinfrared scanning imaging medical device designed to capture depth-resolved images of ocular tissue microstructure. It is available in "high (HR)" and "very-high (VHR)" resolution based on the SLED light source chosen. Each option includes an optical engine, a scanning head, an interchangeable lens set, a computer and the InVivoVue™ Clinic software for imaging of ocular tissue microstructures. The smaller system footprint and mobile security cart easily accommodate any clinical environment.
{4}------------------------------------------------
| Envisu™<br>System | Maximum<br>Detector<br>Bandwidth (nm) | Maximum Image<br>Depth (mm in<br>tissue)1 | Pixel Resolution<br>(µm in tissue, 1024<br>pixels full depth) | Detection Type |
|-------------------|---------------------------------------|-------------------------------------------|---------------------------------------------------------------|-----------------------------------|
| C2200 | 160 | 1.7 | 1.6 | Transmission |
| C2300 | 110 | 2.5 | 2.4 | Spectrometer,<br>Line-scan Camera |
The optical engine includes a low power broadband LED light source with bandwidths of 40 - 100 nm operating in the wavelength range between 760 - 930 nm and a spectrometer that detects backscattered radiation. The scanning head is flexibly designed to be used mounted or handheld for the mutual convenience and comfort of patient and ciinician. Handheld use of the light-weight (< 3.5 lbs), compact handheld scanner makes it possible to image pediatric patients, supine patients, or any patient that finds it difficult to sit upright or where it is clinically preferable to image without the constraint of a chin rest. Tabletop, chin-rest mediated imaging is enabled though a mounting accessory.
#### NON-CLINICAL PERFORMANCE AND SAFETY DATA
IEC-60601-1 electrical, IEC 60601-1-2 electromagnetic compatibility, IEC 60825-1 and ISO 15004-2 Group I optical emission and Class 1 LED safety, image resolution, image comparison and software validation testing have been conducted in accordance with established protocols. All testing has demonstrated compliance with applicable product requirements.
#### CLINICAL LITERATURE REVIEW
A review of existing peer-reviewed clinical literature publications has been completed. Published clinical data reported that the handheld SDOIS is a safe, noninvasive and effective method for evaluation of the eye for pediatric populations. Clinical testing has not been conducted independently by Bioptigen.
#### CONCLUSION
New Traditional 510/k)
The Envisu™ SDOIS has the same fundamental technology, method of operation and technological characteristics as the predicate SDOIS. It introduces new characteristics of mobility and ergonomic utility, including handheld imaging operation, which impact indications for use. Results of performance and safety testing verify that these modifications present no new issues regarding the Envisu™'s safety or effectiveness, and that the Envisu™ SDOIS remains a Class 1 LED , group 1, nonhazardous device eye safe under all operating conditions. In conclusion, the Envisu™ SDOIS is substantially equivalent to the predicate SDOIS device when used as labeled.
<sup>1</sup> Values are approximate, based on an average refractive index of 1.38
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 1 1 2012
Bioptigen, Inc. c/o Ms. Dawn Reilly-O'Dell, RAC Principal Consultant Full Circle Regulatory Consulting, LLC 104 T.W. Alexander Drive, P.O. Box 13569 Durham, North Carolina 27709
#### Re: K120057
Trade/Device Name: Envisu Spectral Domain Opthalmic Imaging System (SDOIS) C2200 and C2300 Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Codes: HLI, OBO Dated: May 4, 2012 Received: May 7, 2012
Dear Ms. Reilly-O'Dell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
机_
Sincerely yours,
Kesia Alexander
Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
K120057
Device Name:
Envisu "M SDOIS (Spectral Domain Ophthalmic Imaging System)
Indications for Use:
Bioptigen Envisu™ Spectral Domain Ophthalmic Imaging System (Envisu™ SDOIS) is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SDOCT). The Envisu™ SDOIS is indicated for use as an aid in the diagnosis of physiologic and pathologic conditions of the eye through non-contact optical imaging. Imaging of the various tissues of the eye is supported through the use of interchangeable lenses. It is indicated for use on patient populations from premature and neonatal infants to adult, and is suitable for patients ambulatory or confined. The system is indicated for use in upright or supine imaging, handheld or mounted, and is suited for imaging patients under anesthesia.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
K120057
310(k) Number
Bioptigen, Inc. ~ Confidential
Page 189 of 176
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.