The ANTERION is a non-contact ophthalmic imaging and analysis device for the eye. It is intended for visualization of the anterior segment.
Device Story
ANTERION is a non-contact diagnostic imaging device using swept-source optical coherence tomography (SS-OCT) and an infrared (IR) camera to visualize the anterior segment of the eye. Operated by clinicians in a clinic setting, the device features a motorized chin/forehead rest for patient positioning. The system captures cross-sectional OCT images and en-face IR images. The software (PC-based) includes an acquisition module (AQM) for scan control and a viewing module (VWM) for image display and analysis. The device provides high-resolution visualization of ocular structures (cornea to posterior lens surface) to assist in identifying pathologies and surgical evidence (e.g., keratoplasty, LASIK, IOLs). Output is displayed on a touchscreen GUI for clinician review, aiding in clinical decision-making and management of anterior segment conditions. Benefits include non-invasive, high-quality imaging without pupil dilation or eye contact.
Clinical Evidence
Prospective clinical study (n=87, age ≥22) compared ANTERION to CIRRUS HD-OCT 5000. Cohorts included normal and abnormal anterior segments (e.g., glaucoma, corneal surgeries). Endpoints: image quality grade and agreement in abnormality identification using slit lamp exam as reference. Results: ANTERION demonstrated superior image quality across key anatomical structures (cornea, lens, iris, angle, scleral spur) and equal or superior ability to identify pre-specified abnormalities compared to CIRRUS. No adverse events reported.
Technological Characteristics
Swept-source OCT (1310 nm) and IR camera. Hardware includes motorized chin/forehead rest, laser/LED optics, and touchscreen. Scan patterns: Line, Volume, Arc, Radial. Axial resolution <10 µm, lateral resolution 30 µm. Connectivity: PC-based. Biocompatibility per ISO 10993-5/18. Electrical safety per ANSI AAMI ES60601-1; electromagnetic compatibility per IEC 60601-1-2. Laser safety per IEC 60825-1 and ANSI Z80.36.
Indications for Use
Indicated for non-contact ophthalmic imaging and visualization of the anterior segment of the eye in patients 22 years or older, including those with normal anterior segments and those with anterior segment abnormalities such as glaucoma or post-surgical conditions.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
November 5, 2021
Image /page/0/Picture/1 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's acronym in a blue square, followed by the full name "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.
Heidelberg Engineering GmbH % Lena Sattler Consultant Orasi Consulting, LLC. 226 1st Street Bonita Springs, Florida 34134
Re: K211817
Trade/Device Name: Anterion Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO Dated: September 24, 2021 Received: September 27, 2021
#### Dear Lena Sattler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211817
Device Name
ANTERION
Indications for Use (Describe)
The ANTERION is a non-contact ophthalmic imaging and analysis device for the eye. It is intended for visualization of the anterior segment.
Type of Use (Select one or both, as applicable)
| Descriptive Use (Part 21 CFR 601 Subpart D) | <div style="display:inline-block;vertical-align:top;"> <svg height="16" width="16"> <path d="M2 2h12v12H2z" fill="none" stroke="black" stroke-width="2"></path> <path d="M4 8l4 4l6-6" fill="none" stroke="black" stroke-width="2"></path> </svg> </div> |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 601 Subpart C) | <div style="display:inline-block;vertical-align:top;"> <svg height="16" width="16"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12" x="2" y="2"></rect> </svg> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a red square to the left of the first line and to the right of the second line.
# 510(K) SUMMARY
#### Date Prepared
June 10, 2021
#### SPONSOR/510(K) OWNER/ MANUFACTURER
Heidelberg Engineering GmbH Max-Jarecki-Strasse 8 69115 Heidelberg, Germany Telephone: +49 6221 / 64 63 0 Facsimile: +49 6221 / 64 63 62 Email: aschoess@hdeng.de Establishment Registration No.: 8043762
#### OFFICIAL CONTACT PERSON
Lena Sattler Orasi Consulting, LLC. 226 1st Street Bonita Springs, FL 34134 Telephone: (440) 554-3706 Facsimile: (866) 904-4315 E-mail: lena@orasiconsulting.com
# COMMON/USUAL NAME
Optical Coherence Tomography
#### PROPRIETARY OR TRADE NAMES
ANTERION
#### CLASSIFICATION INFORMATION
| Regulation Number: | 21 CFR 886.1570 |
|-----------------------|-------------------------------------|
| Classification name: | Ophthalmoscope |
| Device Class: | II |
| Common name: | Optical Coherence Tomography |
| Product Codes, Name: | OBO (Tomography, Optical Coherence) |
| Medical Specialty: | Ophthalmic |
| Classification Panel: | Ophthalmic Device Panel |
Heidelberg Engineering GmbH
Traditional 510(k): ANTERION
{4}------------------------------------------------
#### PREDICATE DEVICE
Visante OCT (K051789), Carl Zeiss Meditec Inc.
#### REFERENCE DEVICE
CIRRUS HD-OCT 5000 (K150977), Carl Zeiss Meditec Inc.
### INDICATIONS FOR USE
The ANTERION is a non-contact ophthalmic imaging and analysis device for the eye. It is intended for visualization of the anterior segment.
### GENERAL DEVICE DESCRIPTION
The ANTERION is a diagnostic imaging device for the eye. The technology is based on swept-source optical coherence tomography (SS-OCT) technology. The device itself has two basic component groups:
- . ANTERION Hardware (Imager/Basis) with integrated forehead/ chin rest: The hardware includes imaging hardware (e.g., laser, LEDs, optics, detectors, hardware for spatial encoding) as well as a touch screen.
- ANTERION Software (PC): The ANTERION Software includes the main user . interface. The software allows for device control, such as selection of examination(s) and imaging parameter(s). The PC software additionally stores, receives, and displays the acquired data.
The ANTERION hardware is separated in three parts: the Basis (bottom part), the Imager (top part), and the Head Rest (forehead/chin rest).
For examinations, the patient places his/her head in the forehead/chin rest. The Head Rest is mechanically and electronically connected to the Basis and controlled via a ioystick. Placed within the stand are a stepper motor with additional mechanical parts and a controller board, allowing the operator to move the motorized chin rest up or down for optimally positioning the patients' eye. An external fixation light is mounted at the forehead rest.
The Basis mainly contains the power supply and PC connection of the device. In the Imager, the components for scanning, signal generation, and signal processing are contained.
Heidelberg Engineering GmbH
Traditional 510(k): ANTERION
{5}------------------------------------------------
The operator directly accesses two software modules, which are named AQM (acquisition module) and VWM (viewing module). The AQM allows selecting between examinations. The VWM shows acquired images, parameters, and reports.
The ANTERION device contains two imaging modalities, a scanning optical coherence tomography (OCT) modality and an infrared (IR) camera. The OCT modality allows for cross-sectional imaging, while the IR camera allows for en-face imaging of a patient's eye.
The Imaging App, the subject of this submission, is the foundation of the ANTERION platform, and focuses on the high-resolution visualization of the entire anterior segment, from the anterior surface of the cornea to the posterior surface of the lens.
The Imaging App delivers swept-source OCT images that allow visualization of anterior segment pathologies and evidence of surgical interventions, e.g. keratoplasty, LASIK, implanted IOLs, and phakic lenses.
### SUBSTANTIAL EQUIVALENCE
The Substantial Equivalence Summary tables below illustrate the comparisons of the ANTERION to the predicate device and the reference device.
The imaging function of the ANTERION is similar to the anterior imaging function of the predicate device Visante OCT (K051789) and the reference device CIRRUS HD-OCT 5000 (K150977) in intended use and technological characteristics.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a red square to the left of the first line and to the right of the second line.
| SUBJECT DEVICE | PREDICATE DEVICE | REFERENCE DEVICE | Discussion |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| ANTERION | Visante OCT (K051789) | CIRRUS HD-OCT 5000 (K150977) | |
| The ANTERION is a non-contact<br>ophthalmic imaging and analysis<br>device for the eye. It is intended<br>for visualization of the anterior<br>segment. | The Visante™ OCT is a non-contact, high<br>resolution tomographic and biomicroscopic<br>device indicated for the in vivo imaging and<br>measurement of ocular structures in the<br>anterior segment, such as corneal and<br>LASIK flap thickness. | The CIRRUS™ HD-OCT is a non-contact,<br>high resolution tomographic and<br>biomicroscopic imaging device intended for<br>in-vivo viewing, axial cross-sectional, and<br>three-dimensional imaging of anterior and<br>posterior ocular structures. The device is<br>indicated for visualizing and measuring<br>anterior and posterior ocular structures,<br>including cornea, retina, retinal nerve fiber<br>layer, ganglion cell plus inner plexiform<br>layer, macula, and optic nerve head. The<br>CIRRUS normative databases are<br>quantitative tools indicated for the<br>comparison of retinal nerve fiber layer<br>thickness, macular thickness, ganglion cell<br>plus inner plexiform layer thickness, and<br>optic nerve head measurements to a<br>database of normal subjects. The CIRRUS<br>OCT Angiography is indicated as an aid in<br>the visualization of vascular structures of<br>the retina and choroid. The CIRRUS HD-<br>OCT is indicated as a diagnostic device to<br>aid in the detection and management of<br>ocular diseases including, but not limited to,<br>macular holes, cystoid macular edema,<br>diabetic retinopathy, age-related macular<br>degeneration, and glaucoma. | Similar. All devices<br>support imaging of the<br>anterior segment of the<br>eye. |
## INTENDED USE/INDICATIONS FOR USE STATEMENT CHART
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the words "HEIDELBERG" on the top line and "ENGINEERING" on the bottom line. There is a dark red square on the left side of the word "HEIDELBERG" and on the right side of the word "ENGINEERING". The text is in a bold, sans-serif font.
### TECHNOLOGICAL CHARACTERISTICS COMPARISON CHART
(*specific data for the Anterior Segment module of CIRRUS HD-OCT 5000)
| | SUBJECT DEVICE<br>ANTERION | PREDICATE DEVICE<br>Visante OCT (K051789) | REFERENCE DEVICE<br>CIRRUS HD-OCT 5000<br>(K150977) | Discussion |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device classification<br>name | Optical Coherence Tomographer<br>(OCT) | Optical Coherence Tomographer<br>(OCT) | Optical Coherence Tomographer<br>(OCT) | Same |
| Main Technology | Swept Source OCT Technology | Time Domain OCT Technology | Spectral Domain OCT<br>Technology | The OCT technologies<br>are different, but the<br>resulting images show<br>comparable structures |
| Supporting<br>Technologies | Infrared Camera | Video camera | Infrared Camera | Similar. Their main<br>usage is to support a<br>proper OCT<br>acquisition. |
| OCT center wave<br>length | 1310 nm | 1310 nm | 750 nm | Similar. The OCT<br>wavelength of the<br>ANTERION and<br>Visante OCT are<br>intended for anterior<br>segment imaging. The<br>CIRRUS HD-OCT<br>5000 is optimized for<br>posterior segment<br>imaging but can also<br>be used to create<br>anterior segment |
| | SUBJECT DEVICE<br>ANTERION | PREDICATE DEVICE<br>Visante OCT (K051789) | REFERENCE DEVICE<br>CIRRUS HD-OCT 5000<br>(K150977) | Discussion |
| | | | | images using a special<br>lens module. |
| OCT axial resolution | <10 um (tissue) | 18 um (in tissue) | 5 um (in tissue) | Similar |
| OCT lateral resolution | 30 um | 60 um | 20 um - 45 um* | Similar |
| Scan length | 5 - 16.5 mm | 10 mm – 16 mm | 9 mm - 15.5 mm* | Similar |
| Scan depth (in tissue) | 8 mm | 3 mm; 6 mm | 2 mm; 2.9 mm; 5.8 mm* | The ANTERION<br>provides images<br>showing the full<br>anterior segment<br>including the posterior<br>lens. |
| Number of A-Scans<br>per B-Scan | 256; 512; 768; 1024 | 128; 256; 512 | 1024; 2048* | Similar |
| A-scan rate | 50,000 Hz | 2,000 Hz | 27,000/68,000 Hz | The OCT technology<br>used in the Visante<br>cannot provide high<br>A-Scan rates |
| Number of B-Scans | 1-65 | 1; 2; 3; 4; 8; 16 | 1-24* | Adjustable in all<br>devices. |
| Scan Pattern | Line, Volume, Arc, Radial | Line, Line Raster Scan | Line, Line Raster Scan, Radial | Similar |
| Scan Center | Adjustable | Not adjustable | Not adjustable | The ANTERION<br>provides more<br>flexibility in defining<br>the scanned area. |
| | SUBJECT DEVICE<br>ANTERION | PREDICATE DEVICE<br>Visante OCT (K051789) | REFERENCE DEVICE<br>CIRRUS HD-OCT 5000<br>(K150977) | Discussion |
| Tracking | Available | Available | Not available for anterior<br>segment imaging | Tracking improves the<br>ability to correlate<br>acquired B-Scans to<br>each other and enables<br>scanning of larger<br>areas. |
| Number of averaged<br>Scans per B-Scan/ | 1; 2; 4; 8 | 1; 4 | 20* | All devices provide<br>some sort of scan<br>averaging. |
| IR camera image size<br>(px) | 768 x 576 | No information found | 1280 x 1024 | Differences in the<br>camera specifications<br>are not relevant |
| Patient/ User Interface | | | | |
| Patient groups | No restrictions regarding patient<br>groups | No restrictions regarding patient<br>groups | No restrictions regarding patient<br>groups | Same |
| Dilation of pupil<br>required? | No | No | No | Same |
| Eye contact required? | No | No | No | Same |
| Fixation light | Internal, external | Internal, external | Internal, external | Same |
| Working position | Upright sitting position of the<br>patient, using chin rest and<br>forehead rest | Upright sitting position of the<br>patient, using chin rest and<br>forehead rest | Upright sitting position of the<br>patient, using chin rest and<br>forehead rest | Same |
| User Interface | Joystick for user to move and<br>align device, device has GUI<br>(Graphical user interface) for<br>display and analysis of data | Graphical user interface for user<br>to move and align device, device<br>has GUI for display and analysis<br>of data | Graphical user interface for user<br>to move and align device, device<br>has GUI for display and analysis<br>of data | All systems provide a<br>display attached to the<br>device, showing a GUI<br>to supervise and |
| SUBJECT DEVICE<br>ANTERION | PREDICATE DEVICE<br>Visante OCT (K051789) | REFERENCE DEVICE<br>CIRRUS HD-OCT 5000<br>(K150977) | Discussion | |
| | | | control the acquisition<br>and analyze the data. | |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
## NON-CLINICAL PERFORMANCE SUMMARY
Software documentation was provided, and software verification and validation was conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
Documentation regarding cybersecurity was submitted as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" and an overall risk assessment regarding security and safety of the device was conducted due to ISO 14971:2007.
Laser safety testing for the light sources used in ANTERION was provided and showed that the requirements according to FDA recognized standards IEC-60825-1:2007 and ANSI Z80.36:2016 were fulfilled.
Biocompatibility of the device was demonstrated by cytotoxicity testing and chemical analysis according to ISO 10993-5:2009 and ISO 10993-18:2005. respectively, and supported by biocompatibility assessment according to 10993-1:2018.
Tests for electrical safety (ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012) and electromagnetic radiation (EC 60601-1-2 Edition 4.0 2014-02) were performed with the Anterion device and passed the relevant requirements of the applied standards.
## CLINICAL EVALUATION
A prospective, clinical study was conducted at a single clinical site in the United States to compare the clinical performance of the ANTERION Imaging App between the ANTERION and the CIRRUS HD-OCT 5000 devices. 87 participants age 22 or older meeting eligibility criteria in at least one eye were enrolled into two study cohorts, the "Normal Anterior Segment group" (without abnormality present in the anterior segment of the eye) and the "Abnormal Anterior Segment group" (those anterior segment abnormalities identified with the slit lamp exam such as glaucoma and corneal surgeries). One eye of each subject was selected to be the study eye, (randomly determined when both eyes were eligible).
The following endpoints were evaluated:
- The image quality grade of the acquired images. ●
- Agreement in identification of abnormality between the B-Scan images using the ● slit lamp examination as the reference.
{12}------------------------------------------------
An eye examination was performed to determine study eligibility, then inclusion and exclusion criteria were assessed. The anterior segment abnormalities and a primary Abnormality of Interest (AOI) were identified during the eye examination by the investigator (when applicable). The device order and the acquisition type order was randomized.
Imaging was performed by the site operators based on randomization
The first acceptable scans were sent to the Reading Center for grading. Each grader graded each image for image quality and for identification and assessment of abnormality (if applicable). The images were presented to three independent graders, who were blinded to the subject identification, population allocation, eye examination results and to each other's grading results. Majority rule was applied to get consensus. For each device and acquisition type (if applicable), three qualified graders graded the OCT images for quality into categorical groups (1 = poor, 2 = fair, 3 = good).
In each image the visibility of following structures was assessed:
- Cornea/Conjunctiva ●
- Iris (temporal/nasal) ●
- Anterior Lens Surface
- . Scleral Spur (temporal/nasal)
- . Angles (temporal/nasal)
The graders also graded for presence or absence of pre-defined key abnormalities, specified on the grading form.
Eighty-seven subjects were enrolled in this study. Thirty-four in the Normal Anterior Segment population and fifty-three in the Abnormal Anterior Segment population.
The mean age was 45.7 ± 16.1 years overall, with the Abnormal Anterior Segment population subjects being older. The study population gender distribution was 67.9% male and 32.1% female. Sixty-seven percent (67%) were not Hispanic or Latino, 34% were Caucasian and 42% African American. The study eyes were fairly equally distributed between right and left eyes (51% right, 49% left).
To compare the image quality of the ANTERION and the CIRRUS devices, key anatomical structures were evaluated for image quality (Good/Fair/Poor) by the independent Reading Center by three masked graders.
{13}------------------------------------------------
Figure 1 below summarizes the structure image quality for all study eyes of the effectiveness analysis population. Overall, the ANTERION shows superiority in image quality compared to each of the three CIRRUS scans.
Image /page/13/Figure/3 description: The image is a set of bar charts that show the image quality of different structures in the eye, including the cornea/conjunctiva, anterior lens surface, iris, angle, and scleral spur. The image quality is measured using four categories: good, fair, poor, and no acceptable scans. The bar charts show the percentage of scans that fall into each category for each structure. For example, for the cornea/conjunctiva, 83.3% of scans are rated as good using the ANT scan type, while 52.6% of scans are rated as good using the CCN scan type.
FIGURE 1: IMAGE QUALITY OF EACH STRUCTURE BY DEVICE AND SCAN TYPE
Heidelberg Engineering GmbH
Traditional 510(k): ANTERION
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a red square on the left side of the first line and on the right side of the second line.
The slit lamp examination, as well as the OCT imaging as assessed by the Reading Center, have been assessed for the presence or absence of the predefined abnormalities. Figure 2 below illustrated the comparison among the different OCT scan types, using the slit lamp results as the reference. Overall, the ANTERION shows superiority compared to each of the three CIRRUS scans in identifying the abnormality in reference with the slit lamp examination.
{15}------------------------------------------------
Image /page/15/Figure/2 description: This image contains multiple bar charts that show the percent of frequency of different eye conditions. The eye conditions include cornea pterygium, cornea corneal inlay, cornea corneal transplant, cornea corneal scar, conjunctiva trabeculectomy bleb, iris nasal peripheral iridotomy, iris temporal peripheral iridotomy, angle nasal glaucoma drainage device, and angle temporal glaucoma drainage device. Each bar chart shows the percent of frequency for match, no match, UNA, and NAS. The x-axis of each bar chart shows ANT, CAC, and CAG/CCN.
FIGURE 2: OCT ABNORMALITY IDENTIFICATION USING SLIT LAMP IDENTIFICATION AS REFERENCE BY DEVICE AND SCAN TYPE
Heidelberg Engineering GmbH
{16}------------------------------------------------
Image /page/16/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a dark red square to the left of the first line and to the right of the second line.
There were no adverse event occurring during the course of the study.
The results of this study illustrated that when using OCT to visualize the anterior segment, ANTERION's overall OCT image quality and ability to visualize key anatomical structures on OCT images was similar to or better than CIRRUS for all scan patterns and all subject populations. Furthermore, using the slit lamp examination as a reference, the results indicate that ANTERION offers an equal or superior ability in identifying each pre-specified abnormality as compared to CIRRUS. Therefore, the study results support substantial equivalence of ANTERION and the CIRRUS 5000 with Anterior Segment Premier Module in regard to image quality, the ability to discern key anatomical structures, and the ability to identify various anterior segment abnormalities.
### CONCLUSION
Comparison of intended use, technological characteristics, and clinical performance data as well as evaluation of non-clinical testing support substantial equivalence of the ANTERION subject device to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.