OptoMedical Technologies OCT-Camera is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SD-OCT). The OCT-Camera is indicated for the use as an aid in the diagnosis of physiologic and pathologic conditions of the eye through non-contact optical imaging. Imaging of the various tissues of the eye is supported through the use of interchangeable lenses. It is indicated for use on patient populations from premature and neonatal infants to adult, and is suitable for patients ambulatory or confined. The system is indicated for use in supine imaging, mounted to a surgical microscope HS Hi-R NEO 900A NIR (Haag-Streit), and is suited for imaging patients under anesthesia.
Device Story
OCT-Camera is an intraoperative SD-OCT system attaching to surgical microscope camera ports. It captures, processes, and displays real-time depth-resolved images of ocular tissue microstructure; enables visualization of surgical steps (e.g., membrane transplantation, micro-implants) without disrupting microscopic view. Operated by surgeons in OR settings. Input: SLED light source; output: live OCT images on display. Benefits: real-time intraoperative guidance for ocular microsurgery. System uses galvanometric mirror scanners and spectrometer-based detection.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including electrical safety, electromagnetic compatibility, risk management, and light hazard protection testing.
Indicated for premature/neonatal infants to adults, ambulatory or confined, for aid in diagnosis of physiologic/pathologic eye conditions via non-contact SD-OCT imaging. Suitable for supine imaging under anesthesia when mounted to specified surgical microscope.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2015
OptoMedical Technologies GmbH c/o Mr. Richard A. Vincins, CQA, CBA RAC (US, EU) Vice President, Quality Assurance Emergo Global Consulting Group 816 Congress Avenue, Suite 1400 Austin, TX 78701
Re: K142953
Trade/Device Name: OCT-Camera, Model 21101A1 Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope, AC-Powered Regulatory Class: Class II Product Code: HLI, OBO Dated: January 27, 2015 Received: January 29, 2015
Dear Mr. Vincins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
Page 2 - Mr. Richard A. Vincins
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142953
Device Name OCT-Camera
#### Indications for Use (Describe)
OptoMedical Technologies OCT-Camera is intended to acquire, process, display and save depth-resolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SD-OCT). The OCT-Camera is indicated for the use as an aid in the diagnosis of physiologic conditions of the eye through non-contact optical imaging of the various tissues of the eye is supported through the use of interchangeable lenses. It is indicated for use on patient populations from premature and neonatal infants to adult, and is suitable for patients ambulatory or confined. The system is indicated for use in supine imaging, mounted to a surgical microscope HS Hi-R NEO 900A NIR (Haag-Streit), and is suited for imaging patients under anesthesia.
### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
for
# OCT-Camera
# K142953
# 1. Submission Sponsor
OptoMedical Technologies GmbH Maria-Goeppert-Strasse 1 Lübeck 23562 GERMANY Phone: 0049 451 2903 348 Fax: 0049 451 2903 333 Contact: Alexander HOLZHEY, Quality and Regulatory Affairs Manager
### 2. Submission Correspondent
Emergo Europe Consulting Prinsessegracht 20 2514AP The Hague The Netherlands Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Dirk Meier, Senior Consultant, QA Email: project.management@emergogroup.com
### 3. Date Prepared
2 March 2015
# 4. Device Identification
| Trade/Proprietary Name: | OCT-Camera |
|----------------------------|--------------------------------------------------------------|
| Common/Usual Name: | intraoperative Optical Coherence Tomography (iOCT) |
| Classification Name: | Ophthalmoscope, AC-powered;<br>Tomography, Optical Coherence |
| Classification Regulation: | 886.1570 |
| Product Code: | HLI, OBO |
| Device Class: | Class II |
| Classification Panel: | Ophthalmic |
{4}------------------------------------------------
# 5. Predicate Device
| Device trade name: | Envisu™ SDOIS |
|--------------------|----------------|
| Manufacturer name: | Bioptigen Inc. |
| 510(k) number: | K120057 |
## 6. Device Description
The OCT-Camera can be attached to the camera port of surgical microscopes. The OCT-Camera is completely integrated into the surgical procedure by enabling the OCT imaging before, during, and after microsurgery without disrupting the microscopic view. The individual steps and the outcome of the surgical procedures, such as transplantation of the thin membranes or micro implants, are visualized in real time.
The OCT-Camera by OptoMedical Technologies GmbH facilitates the intraoperative use of OCT (iOCT). It is called OCT-Camera because it can be attached to the camera port of an operating microscope like any common camera that are used for the purpose of providing live view images of the surgical field.
# 7. Indication for Use Statement
OptoMedical Technologies OCT-Camera is intended to acquire, process, display and save depthresolved images of ocular tissue microstructure using Spectral Domain Optical Coherence Tomography (SD-OCT). The OCT-Camera is indicated for the use as an aid in the diagnosis of physiologic and pathologic conditions of the eye through non-contact optical imaging. Imaging of the various tissues of the eye is supported through the use of interchangeable lenses. It is indicated for use on patient populations from premature and neonatal infants to adult, and is suitable for patients ambulatory or confined. The system is indicated for use in supine imaging, mounted to a surgical microscope HS Hi-R NEO 900A NIR (Haag-Streit), and is suited for imaging patients under anesthesia.
# 8. Substantial Equivalence Discussion
The following table compares the OCT-Camera to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Table 3A - Comparison of Characteristics | | | |
|------------------------------------------|----------------------------------|-----------------------------|----------------------------|
| Manufacturer | OptoMedical Technologies<br>GmbH | Bioptigen | Significant<br>Differences |
| Trade Name | OCT-Camera<br>(subject) | SDOIS Envisu<br>(predicate) | |
| 510(k)<br>Number | N/A | K120057 | Same |
| Product Code | HLI, OBO | HLI, OBO | Same |
| Regulation<br>Number | 21 CFR 886.1570 | 21 CFR 886.1570 | Same |
{5}------------------------------------------------
| Manufacturer | OptoMedical Technologies<br>GmbH | Bioptigen | Significant<br>Differences |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Trade Name | OCT-Camera<br>(subject) | SDOIS Envisu<br>(predicate) | |
| Regulation<br>Name | Ophthalmoscope | Ophthalmoscope | Same |
| Indications for<br>Use | OptoMedical Technologies OCT-<br>Camera is intended to acquire,<br>process, display and save depth-<br>resolved images of ocular tissue<br>microstructure using Spectral<br>Domain Optical Coherence<br>Tomography (SD-OCT). The OCT-<br>Camera is indicated for the use as<br>an aid in the diagnosis of<br>physiologic and pathologic<br>conditions of the eye through non-<br>contact optical imaging. Imaging of<br>the various tissues of the eye is<br>supported through the use of<br>interchangeable lenses. It is<br>indicated for use on patient<br>populations from premature and<br>neonatal infants to adult, and is<br>suitable for patients ambulatory or<br>confined. The system is indicated<br>for use in supine imaging, mounted<br>to a surgical microscope HS Hi-R<br>NEO 900A NIR (Haag-Streit), and is<br>suited for imaging patients under<br>anesthesia. | Bioptigen Envisu™ SDOIS is intended to<br>acquire, process, display and save<br>depth-resolved images of ocular tissue<br>microstructure using Spectral Domain<br>Optical Coherence Tomography (SD-<br>OCT). The Envisu™ SDOIS is indicated<br>for the use as an aid in the diagnosis of<br>physiologic and pathologic conditions<br>of the eye through non-contact optical<br>imaging. Imaging of the various tissues<br>of the eye is supported through the<br>use of interchangeable lenses. It is<br>indicated for use on patient<br>populations from premature and<br>neonatal infants to adult, and is<br>suitable for patients ambulatory or<br>confined. The system is indicated for<br>use in upright or supine imaging,<br>handheld or mounted, and is suited for<br>imaging patients under anesthesia. | Similar |
| Method of<br>Operation | SD-OCT | SD-OCT | Same |
| Light Source | SLED | SLED | Same |
| Light Source<br>Classification | Class 1 LED | Class 1 LED | Same |
| Optical Power<br>(OCT Light) | ≤2350 µW at cornea during OCT<br>scan<br>≤47 µW at cornea without scanning<br>Complies with the requirements of<br>IEC 60825-1:2007, Part 1 and 2 | ≤750 µW at cornea | Similar |
| Optical Power<br>(Pilot Light) | <50 µW at cornea | -option not available- | Similar |
| Resolution,<br>Lateral | Retina: 10.6 to 74 µm in tissue,<br>dependent on magnification of<br>microscope and its retina lens<br>Anterior Segment: 10.6 to 37 µm,<br>dependent on magnification of<br>microscope | Retina: 20 µm in tissue<br>Anterior Segment: 9, 12 and 25 µm | Similar |
{6}------------------------------------------------
| Manufacturer | OptoMedical Technologies<br>GmbH | Bioptigen | Significant<br>Differences |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------|
| Trade Name | OCT-Camera<br>(subject) | SDOIS Envisu<br>(predicate) | |
| Resolution,<br>Axial | $≤$ 7.5 µm in tissue | $≤$ 6 µm in tissue | Similar |
| Depth Range<br>(in tissue/in<br>air) | 3.1 / 4.2 mm | 1.7 / 2.3 mm | Similar |
| Scanner Type | Galvanometric Mirror Pair | Galvanometric Mirror Pair | same |
| Scan Patterns | Line, rectangular volume | Line, rectangular volume, circle,<br>concentric rings, radial lines | Similar |
| Scan Pixels | Axial (depth): 1024<br>Lateral: fixed to 1,000 A-Scans/B-<br>Scan<br>Max. 30,000 total A-Scans/Volume-<br>Scan | Axial (depth): 512 or 1024<br>Lateral: User Selectable<br>Max. 5,000 A-Scans/B-Scan<br>Max. 150,000 total A-Scans | Similar |
| Scan Rate | 15,000 A-Scans/s | 32,000 A-Scans/s | Similar |
| Detection | Transmission Grating,<br>Spectrometer / Line Scan Camera | Transmission Grating, Spectrometer /<br>Line Scan Camera | Same |
| Scanner<br>Ergonomics | Mounted to camera port of a<br>surgical microscope | Mounted (tabletop) or Handheld | Similar |
| Patient<br>Interface | Not required | Optional Chin Rest Assembly | Similar |
| Operating<br>System | Win XP | Win XP | Similar |
| Processor | 2.66 GHz Dual Core | Dual 2.0 GHz Quad Core | Similar |
| Memory | 4 GB | 4 GB | same |
| AC Powered | Yes | Yes | Same |
| Battery<br>Operated | No | No | Similar |
| Electrical<br>Safety Testing<br>Passed | IEC 60601-1, EN 60601-1-2 | IEC 60601-1, EN 60601-1-2 | Same |
# 9. Non-Clinical Performance Data
The OCT-Camara mets all the requirements for overall design, performance and electrical safety confirms that the output meets the design inputs and specifications. The OCT-Camara passed all testing and supports the claims of substantial equivalence and safe operation.
The OCT-Camera complies with the applicable voluntary standards for light hazard protection and safe laser products. The device passed all the testing in accordance with national and international standards.
In the following please find a listing of testing performed at OptoMedical Technologies GmbH all of which have been passed positively to demonstrate safety and effectiveness.
The following testing has been performed to support substantial equivalence:
{7}------------------------------------------------
- . IEC 60601-1: 2005 Medical electrical equipment – Part 1-2, General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 60601-1-2: 2007 Medical electrical equipment – Part 1-2, General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- . ISO 14971: 2007 Medical devices -- Application of risk management to medical devices
- IEC 60825-1: 2007 Safety of laser products - Part 1: Equipment classification and requirements
- ISO 15004-2: 2007 Ophthalmic instruments -- Fundamental requirements and test methods -- Part 2: Light hazard protection
# 10. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
# 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the difference between the OCT-Camera and the predicate devices do not raise any questions regarding its safety and effectiveness. The OCT-Camera, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.