K200966 · Icare Finland OY · HKY · May 7, 2020 · Ophthalmic
Device Facts
Record ID
K200966
Device Name
Icare HOME Tonometer
Applicant
Icare Finland OY
Product Code
HKY · Ophthalmic
Decision Date
May 7, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The Icare HOME tonometer is a prescription device intended as an adjunct to the routine clinical monitoring of intraocular pressure (IOP) of adult patients.
Device Story
Handheld, battery-operated rebound tonometer for home use; measures intraocular pressure (IOP) without topical anesthesia. Device uses sterile, single-use probe making brief corneal contact; measures probe deceleration and rebound time to calculate IOP. Sequence consists of six measurements; device stores final IOP, timestamp, eye identification, and measurement quality (standard deviation). Data uploaded to Icare CLINIC via Icare EXPORT (desktop) or Icare PATIENT (mobile) applications. Used by patients at home under healthcare professional supervision. Healthcare providers access data via browser-based Icare CLINIC for clinical monitoring and decision-making. Benefits include convenient, non-invasive IOP tracking outside clinical settings.
Clinical Evidence
No clinical studies were performed to test this device modification. Evidence consists of bench testing and compliance with recognized consensus standards including ANSI AAMI ES60601-1, ISO 15004-1, IEC 60601-1-6, IEC 62304, IEC 62366-1, IEC 60601-1-11, and IEC 62471.
Technological Characteristics
Handheld, battery-powered (2x CR123) rebound tonometer. Dimensions ~110mm x 80mm x 30mm; weight ~150g. Uses lightweight, disposable, gamma-sterilized plastic probes. Connectivity via USB (including Android support). Software lifecycle per IEC 62304 (Class B). No maintenance calibration required. Measurement range 5-50 mmHg.
Indications for Use
Indicated for adult patients requiring adjunct routine clinical monitoring of intraocular pressure (IOP).
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 7, 2020
Icare Finland Oy Hannes Hyvönen Regulatory Affairs Manager Äyritie 22 Vantaa, 01510 Finland
Re: K200966
Trade/Device Name: Icare HOME Tonometer Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and Accessories Regulatory Class: Class II Product Code: HKY Dated: April 7, 2020 Received: April 10, 2020
Dear Hannes Hyvönen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Ng Acting Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K200966
Device Name Icare HOME Tonometer
Indications for Use (Describe)
The Icare HOME tonometer is a prescription device intended as an adjunct to the routine clinical monitoring of intraocular pressure (IOP) of adult patients.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# ical
# 510(k) Summary
### 1.1 Submitter Name and Address
Icare Finland Oy Äyritie 22 Vantaa, Finland FI-01510
Contact: Hannes Hyvönen Phone: 358 9 8875 1150 Fax: 358 9 728 6670 Email: regulatory(@icarefinland.com Date prepared: May 4, 2020
### 1.2 Device Name
| Trade Name: | Icare HOME tonometer |
|--------------------------|---------------------------|
| Common/Usual Name: | Tonometer |
| Classification Name: | Tonometer and Accessories |
| Regulation No: | 21 CFR 886.1930 |
| Device Regulatory Class: | II |
| Review Panel: | Ophthalmic |
| Product Code: | HKY |
Premarket Notification 510(k) Number: K200966 This premarket notification type is Special 510(k).
# 1.3 Predicate Device
Icare HOME tonometer (type TA022), K163343 This predicate has not been subject to a design-related recall.
{5}------------------------------------------------
## 1.4 Indications for Use / Intended Use
"The Icare HOME tonometer is a prescription device intended as an adjunct to the routine clinical monitoring of intraocular pressure (IOP) of adult patients."
The intended use of the Icare HOME tonometer (type TA022) as well as the method used to obtain IOP measurement remain unchanged by this design modification.
# 1.5 Device Description
The Icare HOME tonometer is a home-use handheld, battery operated device that measures the intraocular pressure (IOP) without the need for topical anaesthesia. It is for prescription use only and to be used under the supervision of a healthcare professional. The device is intended to be used by patients at home.
The tonometer uses the rebound method. A small and light, sterile, single-use probe makes brief contact with the eye. The device measures the deceleration of the probe and the rebound time and calculates the IOP from these parameters. Deceleration of the probe is slower at low IOP compared to high IOP.
The measurement sequence includes six measurements. The probe moves to the cornea and back during every measurement. The tonometer stores information on every complete measurement sequence of six measurements. The stored information includes the calculated final IOP, time and date of the measurement, identification of the eye (right or left) and the quality level of the measurement (i.e. the standard deviation of the six individual measurements).
The measurement data can be uploaded to Icare CLINIC for further analysis using either Icare EXPORT (desktop application) or Icare PATIENT (mobile application). Icare CLINIC is a browser-based software designed for managing patient information and the IOP measurement data.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "icare" in a bold, sans-serif font. The word is black, and there is a thin blue line underneath it. The letters are closely spaced, and the overall appearance is clean and modern.
### 1.6 Comparison of Technological Characteristics with the Predicate
### Device
Comparison of the modified device to the cleared device is presented in the table below:
| # | Characteristic | Modified device (subject of this 510k) | Cleared device (K163343) |
|-----|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Product/Device Identification | Same | Icare HOME tonometer (type TA022) |
| 2. | Intended Use / Indications for Use Statement | Same | IOP Measurement<br>"The Icare HOME tonometer is a prescription device intended as an adjunct to the routine clinical monitoring of intraocular pressure (IOP) of adult patients." |
| 3. | Intended Users | Same | Patients at home |
| 4. | Design | Same | Handheld microprocessor based |
| 5. | Calibration | Same | No maintenance calibration required |
| 6. | Measurement range | Same | 5-50 mmHg |
| 7. | Measurement method | Same | Rebound tonometry |
| 8. | Contact tip (probe) | Same | Lightweight, disposable, single use, plastic probe |
| 9. | Contact tip sterilization | Same | Gamma-sterilized |
| 10. | Anaesthesia required | Same | No |
| 11. | Power supply | Same | 2 x CR123 batteries |
| 12. | Device dimensions<br>and weight | Same | Dimensions:<br>~110mm x 80mm x 30mm<br>Weight: ~150g |
| 13. | Device Software<br>Accessories | Difference, Icare LINK replaced<br>with Icare CLINIC, Icare<br>EXPORT and Icare PATIENT<br>software | Icare LINK software |
| 14. | Connectivity<br>interface | Equivalent, added support for<br>Android smartphone or tablet<br>with USB OTG micro B male –<br>micro B male cable and USB<br>micro B to C adapter | USB |
| 15. | User interface<br>sounds | Equivalent, improved indication<br>sounds (errors in 3kHz, all other<br>indications remain the same<br>4kHz) | All indication sounds 4 kHz. |
| 16. | Product<br>appearance | Equivalent, narrow collar<br>changes device appearance<br>slightly<br>Image: icare device TO with red circle around collar | Image: icare device FROM with red circle around collar |
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "icare" in a bold, sans-serif font. The letters are black, and there is a thin blue line underneath the word. The background is white. The word appears to be a logo or brand name.
Icare HOME 510(k)
{8}------------------------------------------------
### 1.7 Performance and Safety Data
The device has been tested according to relevant FDA recognized consensus standards. The following performance and safety data of the modified device are provided in support of the substantial equivalence:
- . ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . ISO 15004-1:2006 Ophthalmic instruments – Fundamental requirements and test methods - Part 1: General requirements
- IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment Part 1-6: General ● requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62304:2006 + A1:2015 Medical device software Software life-cycle . processes (Device firmware and software level of concern: Class B)
- IEC 62366-1:2015 Medical devices Application of usability engineering to ● medical devices
- . IEC 60601-1-11:2015 Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment
- . IEC 62471:2006 Photobiological safety of lamps and lamp systems
There were no significant changes related to electromagnetic compatibility (EMC), sterilization or biocompatibility. Device and software risk analysis has been performed in accordance with ISO 14971 risk management standard. No clinical studies were performed to test this device modification.
{9}------------------------------------------------
### 1.8 Substantial Equivalence
The modified Icare HOME is substantially equivalent to the predicate device. The technological differences between the Icare HOME and its predicate device raise no new issues of safety and effectiveness. Performance and safety data demonstrate that the Icare HOME is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.