K063873 · Tiolat OY · HKY · May 23, 2007 · Ophthalmic
Device Facts
Record ID
K063873
Device Name
ICARE TONOMETER TA01I
Applicant
Tiolat OY
Product Code
HKY · Ophthalmic
Decision Date
May 23, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The iCare Tonometer TA01i is intended to be used for the measurement of intraocular pressure of the human eye.
Device Story
The iCare Tonometer TA01i is a portable, handheld, battery-operated device used to measure intraocular pressure (IOP). It utilizes a rebound measurement principle: a lightweight (26.5 mg) disposable plastic probe is propelled at a slow speed (0.25-0.30 m/s) to make momentary contact with the cornea. The device measures the probe's speed during contact; this data is processed by an internal microprocessor to calculate IOP, which is then displayed on an LCD. The device is designed for use without topical anesthesia. It is operated by healthcare professionals in clinical settings. The low weight and speed of the probe ensure minimal force (8-16 mN) is applied to the eye, potentially increasing patient comfort and safety compared to traditional applanation tonometry.
Clinical Evidence
Clinical performance was evaluated by comparing the iCare TA01i against predicate devices, including the Goldmann-type applanation tonometer. Bench testing confirmed the device met acceptance criteria for operating range, accuracy, and repeatability. Biocompatibility of the probe was verified per ISO 10993-1:2003. Electrical safety and electromagnetic compatibility were confirmed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Handheld, battery-operated (4x AA). Rebound measurement principle using a 26.5 mg disposable plastic probe (0.9 mm tip radius). Microprocessor-based signal processing. LCD display. No maintenance calibration required. Biocompatible materials per ISO 10993-1. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for the measurement of intraocular pressure (IOP) in the human eye. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{0}------------------------------------------------
063873
## 5. 510(k) Summary
# MAY 2 3 2007
#### Summary prepared:
December 21, 2006
#### Submitted by
Company name: Company address:
Contact person: Phone number: Fax number:
Tiolat Oy Nuijamiestentie 5 B 00400 Helsinki Finland Kari Serjamaa, CEO +358 9 8775 1150 +358 9 728 6670
#### US Agent to respond to FDA requests:
Contact person: Company name: Company address:
Phone number:
Fax number:
Nelson Tobin Diagnostic Instrument Group 8404 Sunstate Street Tampa, FL 33634 (813) 926-3447 (813) 926-7954
#### Proposed Device
Tiolat Oy Manufacturer: Device name: iCare Tonometer Model name: TA01i Classification name: 86HKX Regulation number: 886.1930 Device class: Class II
{1}------------------------------------------------
#### 5.1 Intended Use
The iCare Tonometer TA01i is intended to be used for the measurement of intraocular pressure of the human eye.
### 5.2 Device Description
The iCare Tonometer TA01i is a portable, handheld, battery operated tonometer, which measures intraocular pressure (IOP). To measure IOP, a very light (26.5 mg) and a slow moving (0.25-0.30 m/s) probe makes momentary contact with the cornea of the eye. The device measures the speed of the probe during the time probe is in contact with cornea. Resulting IOP is displayed on a liquid crystal display (LCD). The force used for the measurement which is directed onto the eye is extremely small, because of the low weight and slow speed of the probe.
#### 5.3 Predicate Devices
The iCare Tonometer TA01 is judged to be substantially equivalent in safety, effectiveness and intended use to the following legally marketed devices:
| Manufacturer: | HAAG-STREIT |
|----------------------|----------------------------------|
| Device name: | APPLANATION TONOMETER AT 900 |
| Classification name: | 86HK Y |
| Regulation number: | 886.1930 |
| Device class: | Class II |
| 510(k) number: | K981432 |
| | |
| Manufacturer: | MEDTRONIC |
| Device name: | MEDTRONIC TONO-PEN XL |
| Classification name: | 86HK Y |
| Regulation number: | 886.1930 |
| Device class: | Class II |
| 510(k) number: | K053430 |
| | |
| Manufacturer: | SMT SWISS MICROTECHNOLOGY AG |
| Model name: | PASCAL DYNAMIC CONTOUR TONOMETER |
| Classification name: | 86HKY |
| Regulation number: | 886.1930 |
| Device class: | Class II |
| 510(k) number: | K032967 |
| | |
| Manufacturer: | RYAZAN STATE INSTRUMENT-MAKING |
| | ENTERPRISE |
| Model name: | Tonometer TGDc-01 "PRA" |
| Classification name: | 86HKX |
| Regulation number: | 886.1930 |
| Device class: | Class II |
| 510(k) number: | K021937 |
{2}------------------------------------------------
## 5.4 Substantial Equivalence Comparison
In Tables 5.1, 5.2, 5.3 and 5.4 the iCare Tonometer TA01i is compared to predicate devices to show the equivalence. The comparison of intended use and technological features of iCare Tonometer TA01 indicate that the device is substantially equivalent to legally marketed predicate devices.
Table 5.1. iCare Tonometer TA01i vs. Haag-Streit Applanation Tonometer AT-900 (Goldmann type) . .
| | iCare Tonometer TA01i | Haag-Streit AT-900 |
|---------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Indication | Intraocular pressure (IOP)<br>measurement | Same |
| Design | Hand-held microprocessor<br>based | Slit lamp mounted manual dial |
| Measurement technique | Rebound | Applanation |
| Calibration | No maintenance calibration<br>required | |
| Contact tip | Lightweight, disposable, single<br>use, plastic probe (26.5mg) | Prism as applanation surface |
| Contact tip sterilization | Sterilized, disposable, single<br>use | Prisms need to be disinfected<br>after each patient |
| Patient interface | Probe briefly touches the eye | Held on patient's eye while<br>dial is adjusted |
| Force caused to eye by<br>measurement | 8-16 mN | 10-50 mN (1g = 1/100N<br>=10mN) |
| Display | 4 digit LCD (2 digit reading) | Scribed dial |
| Range of measurement | 7-50 mmHg (display range 1-<br>99 mmHg) | 0-80 |
| Versatility | Patient can be measured in any<br>position, but the tonometer<br>must be oriented horizontally | Patient can be measured in<br>sitting position |
| Anesthesia required | No | Yes |
| Weight | 5.47 oz. without batteries<br>(155 g) | 1.10 lb<br>(0.5 kg) |
| Dimensions | 1.26" × 3.15" × 9.06"<br>(32 mm × 80 mm × 230 mm) | (250 mm × 115 mm × 89 mm) |
| Power source | 4 × 1.5 Volt AA batteries | |
11
{3}------------------------------------------------
| | iCare Tonometer TA0li | Tono-Pen XL |
|------------------------------------|------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Indication | Intraocular pressure (IOP) measurement | Same |
| Design | Hand-held microprocessor based | Same |
| Measurement technique | Rebound | Applanation |
| Calibration | No maintenance calibration required | Same |
| Contact tip | Light weight, disposable, single use, plastic probe with tip radius of 0.9 mm | 1.0 mm transducer tip, applanation surface approximately 3 mm |
| Contact tip sterilization | Sterilized, disposable, single use | Sanitized natural latex tip covers |
| Patient interface | Probe briefly touches eye | Briefly touched against eye |
| Force caused to eye by measurement | 8-16 mN | Depending on user, uncontrolled |
| Display | 4 digit LCD (2 digit reading) | Same |
| Range of measurement | 7-50 mmHg (display range 1-99 mmHg) | 5-80 mmHg |
| Versatility | Patient can be measured in any position, but the tonometer must be oriented horizontally | Patient can be measured in any position |
| Anesthesia required | No | Same |
| Weight | 5.47 oz. without batteries (155 g) | 2.25 oz. (64 g) |
| Dimensions | 1.26" × 3.15" × 9.06" (32 mm × 80 mm × 230 mm) | 7 1/4" × 1" × 7/8" (184.2 mm × 25.4 mm × 22.2 mm) |
| Power source | 4 × 1.5 Volt AA batteries | 2 × OCU-CELTM XL batteries |
Table 5.2. iCare Tonometer TA01i vs. Tono-Pen XL
. .
.
:
{4}------------------------------------------------
| | iCare Tonometer TA0li | Pascal DCT |
|------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------|
| Indication | Intraocular pressure (IOP) measurement | Same |
| Design | Hand-held microprocessor based | Slit lamp mounted microprocessor based |
| Measurement technique | Rebound | Dynamic contour |
| Calibration | No maintenance calibration required | Self calibrating |
| Contact tip | Light weight, disposable, single use, plastic probe with tip radius of 0.9 mm | Sensor tip 7 mm, pressure sensor 1.2 mm |
| Contact tip sterilization | Sterilized, disposable, single use | Sterile, single use disposable tip cover |
| Patient interface | Probe briefly touches the eye | |
| Force caused to eye by measurement | 8-16 mN | 10 mN (constant 1 gram) |
| Display | 4 digit LCD (2 digit reading) | LCD display |
| Range of measurement | 7-50 mmHg (display range 1–99 mmHg) | 5-200 mmHg |
| Versatility | Patient can be measured in any position, but the tonometer must be oriented horizontally | Same |
| Anesthesia required | No | Yes |
| Weight | 5.47 oz. without batteries (155 g) | 7.41 oz. (210 g) |
| Dimensions | 1.26" × 3.15" × 9.06" (32 mm × 80 mm × 230 mm) | 6.69" × 3.46" × 1.57" (170 mm × 88 mm × 40 mm) |
:
:
:
,
.
Table 5.3. iCare Tonometer TA01i vs. Pascal dynamic contour tonometer (DCT)
:
:
:
.
:
{5}------------------------------------------------
| | iCare Tonometer TA01i | Tonometer TGDc-01 "PRA" |
|---------------------------|------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Indication | Intraocular pressure (IOP) measurement | Same |
| Design | Hand-held microprocessor based | Hand-held microprocessor based |
| Measurement technique | Rebound | Same, but measurement to upper eyelid |
| Calibration | No maintenance calibration required | No maintenance calibration required |
| Contact tip | Light weight, disposable, single use, plastic probe with tip radius of 0.9 mm | – |
| Contact tip sterilization | Sterilized, disposable, single use | – |
| Patient interface | Probe briefly touches the eye | Touches the upper eyelid |
| Display | 4 digit LCD (2 digit reading) | 4 digit LCD (2 digit reading) |
| Range of measurement | 7-50 mmHg (display range 1–99 mmHg) | 0-60 |
| Versatility | Patient can be measured in any position, but the tonometer must be oriented horizontally | Patient can be measured in any position, but the tonometer must be oriented downwards |
| Anesthesia required | No | No |
| Weight | 5.47 oz. without batteries (155 g) | 2.79 oz. (79g) |
| Dimensions | 1.26" × 3.15" × 9.06" (32 mm × 80 mm × 230 mm) | 6.83" × 1.00" × 0.77" (173.5mm × 25.5mm × 19.5mm) |
| Power source | 4 × 1.5 Volt AA batteries | 4 × 1.5 Volt batteries (CR2032) |
Table 5.4. iCare Tonometer TA01ivs Tonometer TGDc-01 "PRA"
#### 5.5 Summary of Non-Clinical and Clinical Testing
The performance of iCare Tonometer TA01i has been evaluated in bench testing. The units under test met the acceptance criteria of the operating range, accuracy and repeatability.
In clinical evaluation, the performance of iCare Tonometer TA01i was compared to predicate devices, including Goldmann type tonometer.
Based on the results of non-clinical and clinical testing, the performance of iCare Tonometer TA01i is judged to be substantially equivalent to predicate devices. The patient contacting probe is considered biocompatible. The appropriate tests for biocompatibility were performed according to "ISO 10993-1:2003, Biological evaluation of medical devices - Part 1 Evaluation and Testing".
The electromagnetic compatibility and electrical safety of the iCare Tonometer TA01i was tested and found to conform to the requirements of IEC 60601-1 and IEC 60601-1-2.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Tiolat Oy c/o Mr. Nelson Tobin Diagnostic Instrument Group 8404 Sunstate Street Tampa, Fl 33634
MAY 2 3 2007
Re: K063873
Trade/Device Name: iCare Tonometer TA01i Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer Regulatory Class: Class II Product Code: HKX Dated: May 9, 2007 Received: May 10, 2007
Dear Mr. Tobin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
#### Page 2 – Mr. Nelson Tobin
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Egelman, M.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## 4. Indications for Use Statement
Indications for Use
510(k) Number (if known): not assigned yet
Device Name: iCare Tonometer TA01i
Indications for Use:
The iCare Tonometer TA01i is intended to be used for the measurement of intraocular pressure of the human eye.
Prescription Use __ X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Dan L. Kautz
(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises
510(k) Number K063873
8
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.