HAAG-STREIT FUNDUS MODULE 300

K163021 · Optomed OY · HKI · Apr 11, 2017 · Ophthalmic

Device Facts

Record IDK163021
Device NameHAAG-STREIT FUNDUS MODULE 300
ApplicantOptomed OY
Product CodeHKI · Ophthalmic
Decision DateApr 11, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1120
Device ClassClass 2

Indications for Use

FUNDUS MODULE 300 is an ophthalmic camera that is used as an accessory with dedicated slit lamps to capture digital images and video of the fundus of the human eye.

Device Story

FUNDUS MODULE 300 is an ophthalmic camera accessory for slit lamps; captures digital images/video of human fundus. Uses infrared light for non-mydriatic targeting; visible light flash for image capture. Features 9 internal fixation targets for patient alignment. Operated by qualified personnel in clinical settings. Integrates with Haag-Streit slit lamps, Release Module RM02, and EyeSuite software. Captured data transfers automatically to PC via USB or stored on micro SDHC card. Provides clinicians with retinal documentation for consultation; aids in diagnosis and patient management by enabling high-resolution fundus visualization without requiring pupil dilation.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), optical radiation safety (ISO 15004-2), mechanical/environmental testing (IEC 60601, IEC 60068), and usability engineering (IEC 62366-1). Software verification and validation conducted per IEC 62304.

Technological Characteristics

Ophthalmic camera; color CMOS sensor (5 Mp, 1.77 Mp effective). Illumination: White/NIR LEDs (OSRAM Oslon). Diopter compensation: -20D to +20D. Field of view: 40 degrees. Materials: plastic, aluminum, silicon. Power: Rechargeable Li-Ion battery (3.6V, 2350 mAh). Connectivity: USB 1.1. Standards: IEC 60601-1, IEC 60601-1-2, ISO 15004-2, ISO 10940, IEC 62366-1.

Indications for Use

Indicated for use by qualified and trained personnel to capture digital images and video of the fundus of the human eye for documentation and consultation. Intended for use with dedicated slit lamps in a medical environment. Prescription use only.

Regulatory Classification

Identification

An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.

Special Controls

*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with a design that resembles a bird or a wing above them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 11, 2017 Optomed Oy Mr. Jyri Leskelä Quality Manager Yrttipellontie 1 Oulu. 90230 FI Re: K163021 Trade/Device Name: HAAG-STREIT FUNDUS MODULE 300 Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic camera Regulatory Class: Class II Product Code: HKI Dated: February 23, 2017 Received: March 1, 2017 Dear Mr. Leskelä: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {2}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Denise L. Hampton -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163021 Device Name HAAG-STREIT FUNDUS MODULE 300 Indications for Use (Describe) FUNDUS MODULE 300 is an ophthalmic camera that is used as an accessory with dedicated slit lamps to capture digital images and video of the fundus of the human eye. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary | 510(k) submitter: | Optomed Oy | |----------------------|--------------------------------------------------------------------------------------| | Address: | Yrttipellontie 1, FI-90230 Oulu, Finland | | Company phone: | +358 20 741 3380 | | Contact person: | Mr. Jyri Leskelä, Quality Manager | | Date: | April 6, 2017 | | Subject device: | | | Trade name: | HAAG-STREIT FUNDUS MODULE 300 | | 510(k) number: | K163021 | | Common/usual name: | Ophthalmic camera | | Classification name: | camera, ophthalmic, ac-powered<br>(21 CFR 886.1120) | | Regulatory Class: | II | | Product Code: | HKI | | Predicate device: | | | Trade name: | Optomed Smartscope M5 digital camera<br>with Optomed Smartscope EY4 optics<br>module | | 510(k) number: | K132186 | | Regulatory Class: | II | | Product code: | HKI | www.optomed.com {5}------------------------------------------------ # Description of the device FUNDUS MODULE 300 device set for retinal imaging consists of FUNDUS MODULE 300 Camera with base plate, lens cover, battery and camera USB cable (0.6 m), and FUNDUS MODULE 300 Cradle with power supply and cradle USB cable (1.5 m). FUNDUS MODULE 300 is used with dedicated HAAG-STREIT slit lamps. FUNDUS MODULE 300 is designed to be used in medical environment and it is intended for nonmydriatic imaging of the fundus of the human eve for documentation and consultation. In nonmydriatic imaging no mydriatic is needed because infrared light is used for the targeting image to the fundus and visible light is flashed when an image is taken. The pupil does not respond to the infrared light so examination is convenient for the patient. Constant white light illumination for targeting can be selected from device menu if mydriatic drops are used. Images can also be taken using infrared lightning for both aiming and capturing. The FUNDUS MODULE 300 has 9 internal fixation targets for the patient to fixate at during imaging. The middle fixation target provides a macula centred image. FUNDUS MODULE 300 can be used together with HAAG-STREIT Release Module RM02 and HAAG-STREIT EyeSuite software. Controlling the FUNDUS MODULE 300 can be performed by Camera keys or by Release Module RM02. Captured images and recorded videos are automatically transferred to the PC (EyeSuite software) after imaging. Images and videos are stored on the micro SDHC memory card and can be also transferred to the PC by using Cradle USB connection. FUNDUS MODULE 300 Camera has rechargeable Li-lon battery and the device is charged by using the Cradle. Materials of outer shells of the FUNDUS MODULE 300 are plastic, aluminum and silicon. There is no applied parts (no parts which are designed to come into physical contact with the patient) in FUNDUS MODULE 300. # Indications for use FUNDUS MODULE 300 is an ophthalmic camera that is used as an accessory with dedicated slit lamps to capture digital images and video of the fundus of the human eye. FUNDUS MODULE 300 is classified as prescription device. Federal law restricts this device to sale by or on the order of a physician or licensed practitioner. This equipment must only be operated by qualified and trained personnel. The owner is responsible for the training. This device may only be used in accordance with intended use. The operating principle and the intended use of the FUNDUS MODULE 300 is the same than Smartscope M5 with Smartscope EY4. The main difference between subject device and predicate device is that FUNDUS MODULE 300 is intended to be used as slit lamp accessory instead of being used as hand-held. There is incremental changes in mechanics, software and hardware. There is no changes in optics. These changes do not affect safety and effectiveness of the product. {6}------------------------------------------------ # Comparison of technological characteristics with the predicate device Table 1 below includes a summary of the technical information used in the substantial equivalence discussion. | Point of<br>comparison | FUNDUS MODULE 300 | Optomed Smartscope M5 with<br>Smartscope EY4 | |-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) number | K163021 | K132186 | | Indications of use | FUNDUS MODULE 300 is an<br>ophthalmic camera that is used as an<br>accessory with dedicated slit lamps<br>to capture digital images and video<br>of the fundus of the human eye. | Optomed Smartscope M5 digital<br>camera with optics module<br>Smartscope EY4 is intended to<br>capture digital images and video of<br>the fundus of the human eye. | | Usage | Prescription use | Prescription use | | Use condition | Intended to be used without<br>mydriatic but can be used also with<br>mydriatic. | Intended to be used without<br>mydriatic but can be used also with<br>mydriatic. | | Observation light<br>source | Visible and infrared LEDs:<br>White: OSRAM Oslon LUW-H9GP<br>NIR: OSRAM Oslon SFH-4715<br>Target LEDs: OSRAM LR QH9F | Visible and infrared LEDs:<br>White: OSRAM Oslon LUW-H9GP<br>NIR: OSRAM Oslon SFH-4715<br>Target LEDs: OSRAM LR QH9F | | Observation and<br>display system | 3.97", TFT-LCD, 800x480 pixels, 16.7<br>M colors, anti-glare coating | 2.4", TFT-LCD, active matrix color,<br>antireflective coating | | Photographing<br>light source | Visible and infrared LED | Visible and infrared LED | | Camera<br>specification | Color CMOS camera maximum<br>resolution 5 Mp. Optics uses 1.77<br>Mp. | Color CMOS camera maximum<br>resolution 5 Mp. EY4 uses 1.77 Mp. | | Diopter<br>compensation | at least from –20 D to +20 D | at least from -20 D to +20 D | | Table 1. Summary of technical information used in the substantial equivalence discussion. | | |-------------------------------------------------------------------------------------------|--| |-------------------------------------------------------------------------------------------|--| {7}------------------------------------------------ | Picture angle | 40 degrees | 40 degrees | |-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Storage media | MicroSDHC memory card | MicroSDHC memory card | | Image data format | JPEG, MPEG-4 | JPEG, MPEG-1, MPEG-4 | | Weight | 1200 g | M5: 400g,<br>EY4: 300g | | Power<br>consumption | Re-chargeable Li-Ion battery,<br>NCA103450, 3.6 V, 2350 mAh | Re-chargeable Ni-MH Battery,<br>HR4U700AAA, 4.8V, 1000 mAh;<br>Charging unit 44 VA | | Output terminals<br>and data collection | USB (1.1) terminal (B-connector).<br>Compatible with Windows®<br>VISTA/7/8.1/10 | USB (1.1) terminal (B-connector).<br>Compatible with Windows®<br>VISTA/7/8.1/10. | | Exposure<br>parameters | "Exempt Group" (no risk) LED<br>product according to IEC 62471:2006<br>Group 2 instrument according to ISO<br>15004-2:2007 | "Exempt Group" (no risk) LED<br>product according to EN 62471:2008<br>Group 2 instrument according to ISO<br>15004-2:2007 | | Standards | IEC 60601-1:2005 (edition 3.1)<br>IEC 60601-1-2:2014 (edition 4.0)<br>IEC 62471:2006<br>ISO 15004-1:2006<br>ISO 15004-2:2007<br>ISO 10940:2009<br>IEC 62366-1:2015 | IEC 60601-1:2005 (edition 3.0)<br>IEC 60601-1-2:2007 (edition 3.0)<br>IEC 62471:2006<br>ISO 15004-1:2006<br>ISO 15004-2:2007<br>ISO 10940:2009<br>ISO 10993-5:2009 | www.optomed.com {8}------------------------------------------------ The modifications in FUNDUS MODULE 300 compared to Smartscope M5 with Smartscope EY4 optics are: #### ● Mechanics - FUNDUS MODULE 300 has new structure where camera and optics are integrated. o - FUNDUS MODULE 300 has shaft: electronics and mechanics interface part to base O plate. - o FUNDUS MODULE 300 has base plate: mechanical interface part to slit lamp, electrical interface to medical PC. - Optics ● - No changes. o - . SW - FUNDUS MODULE 300 only supported optics is EY4. о - FUNDUS MODULE 300 GUI is different: Display is larger with higher resolution o (800x480). Colors and fonts differ. - FUNDUS MODULE 300 has no patient list support. O - FUNDUS MODULE 300 can be controlled with Haag-Streit Release Module RM02. O - FUNDUS MODULE 300 has slit lamp USB mode support: Images are transferred right O after capture to the PC. - FUNDUS MODULE 300 is used together with Haag-Streit EyeSuite software. o - HW ● - FUNDUS MODULE 300 has camera board (main board, sensor board and AF board O combined) and also shaft board and base plate board. - FUNDUS MODULE 300 battery is different (Li-ion) O - FUNDUS MODULE 300 display is different O - FUNDUS MODULE 300 cradle board is different o - FUNDUS MODULE 300 AF module different O {9}------------------------------------------------ # Performance data The following performance data were provided in support of the substantial equivalence determination. ## Electrical safety and electromagnetic compatibility (EMC) FUNDUS MODULE 300 was tested according to all suitable clauses of IEC 60601-1:2005+A1:2012 (safety) and IEC 60601-1-2:2014 (EMC). FUNDUS MODULE 300 fulfills the requirements of the standards. #### Optical radiation safety Smartscope EY4 was tested according to the standard ISO 15004-2:2007 (optical radiation safety) and FUNDUS MODULE 300 (identical to Smartscope EY4 optics module) is classified as Group 2 instrument. #### Software Verification and Validation testing The level of concern of software included in the FUNDUS MODULE 300 is moderate. Software verification and validation testing was conducted to ensure that design requirements and specified indications for use were fulfilled. FUNDUS MODULE 300 complies with standards IEC 60601-1-4:1996+A1:1999 and IEC 62304:2006+A1:2015. ## Mechanic and environmental testing FUNDUS MODULE 300 was tested according to all suitable clauses of IEC 60601 and IEC 60068. FUNDUS MODULE 300 fulfills the requirements of the standards. #### Usability (Human factors) FUNDUS MODULE 300 was designed according to the usability engineering process. FUNDUS MODULE 300 complies with standards IEC 60601-1-6:2010 and IEC 62366-1:2015. #### Stability FUNDUS MODULE 300 was tested for in use stability to correspond the expected life time of five years. Shipping stability was tested according to IEC 60068-2-27:2008. # Conclusions Based on the comparison and supportive information, the FUNDUS MODULE 300 is substantially equivalent to the predicate device. www.optomed.com
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...