Geri Embryo Incubator and Geri Dish

K180304 · Genea Biomedx Pty, Ltd. · MQG · Apr 20, 2018 · Obstetrics/Gynecology

Device Facts

Record IDK180304
Device NameGeri Embryo Incubator and Geri Dish
ApplicantGenea Biomedx Pty, Ltd.
Product CodeMQG · Obstetrics/Gynecology
Decision DateApr 20, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 884.6120
Device ClassClass 2

Intended Use

The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours. The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.

Device Story

Benchtop incubator for human embryo development; features six independent incubation chambers. Each chamber includes heating elements, gas inlet, and integrated CMOS camera/optics (591 nm orange LED) for time-lapse imaging. Input: embryos placed in Geri Dish. Operation: device maintains controlled temperature, gas mixture (CO2/O2/N2), and humidity (via disposable water bottle). Firmware controls incubator settings; software manages patient information and imaging parameters. Output: time-lapse images of embryos viewed by embryologists/clinicians via software interface. Benefit: allows continuous embryo monitoring without removing them from the controlled environment, potentially improving selection and development outcomes. Used in clinical IVF laboratories.

Clinical Evidence

Bench testing only. Includes electrical safety (IEC 61010-1), EMC (IEC 61326-1), gas/temperature control validation, and time-lapse imaging performance. Embryo safety validated via Mouse Embryo Assay (MEA) showing ≥80% development to blastocyst. Light toxicity testing confirmed exposure levels are within limits of standard IVF procedures. Humidity control system verified via software V&V and bench performance testing. Sterilization and package integrity validated per ISO 11137 and ASTM standards.

Technological Characteristics

Benchtop incubator; 6 chambers; heated aluminum stage/lid. CMOS camera (2560x1928, 2 pixels/µm); 591 nm LED illumination. Humidity control via disposable water bottle. Gas supply: premixed medical gas. Connectivity: firmware/software for control. Sterilization: Radiation (SAL 10^-6) for Geri Dish and Geri Water Bottle. Materials: Polystyrene (dish). Standards: IEC 61010-1, IEC 61326-1, ISO 11137-2, ASTM F1929, ASTM F88.

Indications for Use

Indicated for the incubation, preparation, storage, and time-lapse imaging of human embryos in an assisted reproduction laboratory setting.

Regulatory Classification

Identification

Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.

Special Controls

*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. April 20, 2018 Genea Biomedx Pty Ltd % Roger Grav VP, Quality Assurance and Regulatory Affairs Donawa Lifescience Consulting Srl Piazza Albania, 10 00153 Rome ITALY Re: K180304 > Trade/Device Name: Geri Embryo Incubator and Geri Dish Regulation Number: 21 CFR 884.6120 Regulation Name: Assisted reproduction accessories Regulatory Class: Class II Product Code: MQG, MQK, MTX Dated: March 19, 2018 Received: March 22, 2018 Dear Roger Gray: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Charles Viviano -S For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K180304 Device Name Geri Embryo Incubator and Geri Dish #### Indications for Use (Describe) The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours. The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------| | <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The logo is simple and modern, with a focus on the company's name. ## 510(k) Summary (K180304) #### I. General Information on Submitter | Submitter/Address: | Genea Biomedx Pty Ltd<br>Level 2, 321 Kent Street<br>Sydney<br>NSW 2000<br>Australia | |--------------------|-------------------------------------------------------------------------------------------------------------------------------| | Contact person: | Ms. Kea Dent | | Phone: | +61 2 8484 7677 | | Fax: | +61 2 9229 6478 | | Correspondent: | Mr. Roger Gray<br>VP, Quality and Regulatory<br>Donawa Lifescience Consulting Srl<br>Piazza Albania 10<br>00153 Rome<br>Italy | | Phone: | +39 06 578 2665 | | Fax: | +39 06 574 3786 | | Email: | rgray@donawa.com | #### II. Date Prepared: April 19, 2018 #### III. General Information on Devices | Device Name: | Geri Embryo Incubator and Geri Dish | |--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Embryo Incubator | | Regulation Number: | 21 CFR 884.6120 | | Regulation Name: | Assisted Reproduction Accessories | | Product Code: | MQG (Accessory, Assisted Reproduction), MQK (Labware, Assisted<br>Reproduction), MTX (Microscope and Microscope Accessories, Reproduction,<br>Assisted) | | Regulatory Class: | II | #### IV. Predicate Device Geri Embryo Incubator (K171736) manufactured by Genea Biomedx, Australia. The predicate device has not been subject to any design related recalls. #### V. Device Description The Geri Embryo Incubator is a benchtop incubator with six modular incubation chambers each with its own temperature control capability and separate gas inlet feed. Each chamber contains heating elements in its lid and base, together with an orange light source (591 nm) and camera with integrated optics that take time-lapse images of embryos and allows operators to view embryos without removing them from the incubation chamber. Inside each chamber is a filter the gas mixture entering the incubation chamber from the gas supply. The gas is supplied from standard premixed medical gas cylinders. Each chamber can also contain an optional water bottle (Geri Water Bottle) to generate humidity. The Geri Water Bottle is supplied sterile with a sterility assurance level (SAL) of 10°, and has a shelf-life of three years. The Geri Water Bottle is designed to be used with the Geri Embryo Incubator only. The embryos are maintained in Geri Dishes that are supplied separately. The Geri Dish is an optically-clear polystyrene dish designed to be compatible with the Geri Embryo Incubator. Up to 16 embryos can be stored in {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for Genea BIOMEDX. The logo consists of a green square with rounded corners and a white curved line inside. To the right of the square, the word "Genea" is written in gray, and below it, the word "BIOMEDX" is written in green. one Geri Dish. Geri Dishes are supplied sterile with a sterility assurance level (SAL) of 10°, and have a shelflife of 12 months. The Geri Embryo Incubator and Geri Dish are intended to be used together for embryo imaging purposes. The Geri Dish is not compatible with other embryo time-lapse incubators. However, other assisted reproduction culture dishes may be used in the Geri Embryo Incubator in separate non-time-lapse positions located within each chamber. The Geri Embryo Incubator includes firmware to control the incubator settings, and software to control patient information and settings. #### VI. Indications for Use: The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours. The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator. #### VII. Comparison of Intended Use and Technological Characteristics of the Subject Device and Predicate Device | Parameter | Subject device (K180304) | Predicate device (K171736) | |--------------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | Same as the predicate device | The Geri Embryo Incubator is intended to provide an<br>environment with controlled temperature and mixed gas<br>(CO2 and other gases) for the development of embryos.<br>The Geri Embryo Incubator has an integrated camera and<br>optics for imaging and viewing embryos during incubation,<br>for a maximum time of 120 hours.<br><br>The Geri Dish is intended to be used for preparation,<br>storage and imaging of human embryos. The Geri Dish is<br>intended to be used only with the Geri Embryo Incubator. | | Number of incubation chamber | Same as the predicate device | Six | | Incubation chamber heating | Same as the predicate device | Heated aluminium stage and lid | | Incubation chamber gas supply | Same as the predicate device | Flow controlled positive pressure system with fast purge<br>flow rate for rapid recovery | | Incubation chamber humidity | Disposable water bottle to<br>generate humidity (~80% RH) in<br>each incubation chamber | N/A | | Gas supply type | Same as the predicate device | Premixed blend of 6% CO2, 5% O2, 89% N2 or high purity<br>6% CO2 | | Gas supply pressure | Same as the predicate device | $150 \pm 15$ kPa ( $218 \pm 2.2$ psi) | | Gas flow rate capability | Same as the predicate device | 10 mL/min per chamber. Purge at 160 mL/min per<br>chamber for 3 min | | Gas flow rate accuracy | Same as the predicate device | $\pm$ 15% of flow per chamber (normal flow) and $\pm$ 10 mL/min<br>per chamber (purge) | | HEPA Filter | Same as the predicate device | Retains 99.97% of particles >0.3 µm | | Chamber temperature capability | Same as the predicate device | 35.0-40.0°C in 0.1°C increments at 20-28°C. At set point of<br>37°C, the ambient temperature range is 18-30°C | | Chamber temperature accuracy | Same as the predicate device | $\pm$ 0.2°C at calibration point | | Microscope/camera | Same as the predicate device | CMOS microscope camera (2560x1928 pixels<br>monochrome ) | | Image resolution | Same as the predicate device | 2 pixels/µm | | Image acquisition frequency | Same as the predicate device | 5 min | | Illumination | Same as the predicate device | Single orange LED (591nm) | | Number of image focal planes | Same as the predicate device | 11 | | Light exposure | Same as the predicate device | ~45s/day | | Alarm (for out of range) | Same as the predicate device,<br>with additional alarm for humidity | Chamber temperature, CO2 concentration, CO2 pressure<br>O2 concentration, N2 pressure, Load door open >30s | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside on the left, and the text "Genea" in gray above the text "BIOMEDX" in green on the right. The logo is clean and modern, with a focus on the company name and a simple graphic element. | Parameter | Subject device (K180304) | Predicate device (K171736) | |--------------------------|------------------------------|-------------------------------------------------------------------------------------------------------------------------| | Culture dish (Geri Dish) | Same as the predicate device | 16-well culture dish (40-80 $\mu$ l fill volume/well) made of<br>polystyrene. Each incubation chamber can hold one dish | The subject and predicate devices have the same Intended Use for preparing and storing human embryos undergoing time-lapse imaging procedures. They also have the same technological characteristics, except that the subject device has an additional humidity control system, including a water container, humidity sensor, and software component to manage this new feature. The addition of humidification system to the device does not raise different questions of safety and effectiveness as compared to the predicate device. In addition, similar humidification systems are present on other cleared devices of this type. #### VIII. Summary of Non-Clinical Performance Testing 1. The following studies that were provided in the predicate submission (K171736) were leveraged in the current submission as the additional humidity control system did not impact the ability of these studies to support substantial equivalence: #### Testing for Geri Embryo Incubator - Electrical safety testing per IEC 61010-1:2010 - Electromagnetic Compatibility testing per IEC 61326-1:2012 ● - . Cleaning and disinfection validation testing per FDA Guidance Document, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling: Guidance for Industry and Food and Drug Administration" issued on March 17, 2015 - Bench performance tests that met design specifications described in Section VII, as follows: ● - o Gas maintenance testing to ensure that gas flow rates, gas recovery rates and CO2 levels in chambers meet design specifications - o Temperature control testing to ensure that culture chamber temperature conditions were within defined ranges and were stable over time - Time-lapse testing to demonstrate that all design specifications of time-lapse function are met O (dish presence detection, dish position accuracy, image quality and resolution, imaging of embryos in all wells on Geri Dishes within time and focal plane settings) - Embryo development test demonstrating that ≥70% mouse 1-cell embryos developed to blastocyst on ● day 5 following exposure to the maximum time lapse exposure conditions - . Embryo light energy exposure testing comparing the potential for light toxicity associated with timelapse imaging under worst-case exposure conditions (maximum light exposure over the full duration of use [120 hours]) to that of total light exposure during in assisted reproduction procedures (IVF and ICSI) using typical laboratory microscopes. Results showed total light energy exposure following timelapse imaging should not exceed levels associated with typical assisted reproduction imaging procedures. #### Testing for Geri Dish - . Radiation sterilization validation testing per ISO 11137-2:2009 - Package integrity testing: - Dye penetration testing per ASTM F1929-12 O - Peel strength testing ASTM F88/F88M-09 o - Endotoxin testing per USP <85> and ANSI/AAMI ST72: 2011. The testing demonstrated that the ● device met the specification of <20 EU/device. - Mouse embryo assay (MEA): One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for Genea BIOMEDX. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The green square is on the left side of the logo, and the text is on the right side. hours were assessed in comparison with the control group. The testing demonstrated that the device met acceptance criterion of "1-cell MEA ≥80% embryos developed to blastocyst in 96 hours." - Shelf-life testing (accelerated) demonstrating that visual requirements, media fill volume, time-lapse . alignment and MEA met design specifications before and after aging 2. The following studies on the subject device have been conducted to support the addition of the humidity control system to the Geri Embryo Incubator: Testing for Geri Embryo Incubator: - Software verification and validation testing that met the requirements of the FDA Guidance Document, ● "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005 - Humidity control testing to ensure that the humidity levels in chambers meet design specifications Testing for Geri Water Bottle: - . Radiation sterilization validation testing per ISO 11137-2:2013 - . Package integrity testing: - Dye penetration testing per ASTM F1929-15 O - Seal strength testing per ASTM F88/F88M-15 o #### IX. Conclusion The subject and predicate devices have the same intended use and fundamental technological characteristics. The differences in technological characteristics between subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...