The Embryo Real-time Incubator (TLS301) is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Embryo Real-time Incubator (TLS301) has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours. The Embryo Real-time Culture Dish (MC 2004) is intended for preparing, storing and transferring human embryos. It is intended to be used only with the Embryo Real-time Incubator (TLS301). Embryo Realtime 2S Software and Embryo Realtime 3E Software accessories for the Embryo Realtime Incubator (TLS301). The Embryo Realtime 2S Software is intended to store, archive and transfer data. In addition, the Embryo Realtime 2S Software includes functions for managing models and performing calculations based on image data and embryo development parameters. Embryo Realtime 3E Software is intended for viewing and recording embryo development events from images captured using the Embryo Real-time Incubator (TLS301). Embryo Realtime 3E Software includes a user annotation for capturing information on embryo development parameters and a user-defined modeling function that allows the user to combine annotated information on embryo development parameters to aid in embryo selection. Embryo Realtime 2S Software and Embryo Realtime 3E Software do not control any hardware components in the Embryo Real-time Incubator (TLS301). Embryo Realtime 2S Software and Embryo Realtime 3E Software are provided in different software package and must be used together.
Device Story
Benchtop tri-gas incubator (CO2, N2, air) with integrated inverted microscope and camera for time-lapse imaging of embryos. System includes Embryo Real-time Incubator (TLS301), Embryo Real-time Culture Dish (MC 2004), and two software packages (2S and 3E). Incubator maintains temperature and gas concentrations; camera captures images at multiple focal planes using 635 nm red LED illumination. Images are transmitted to a workstation for display, storage, and analysis. Embryologists use software to annotate embryo development events and apply user-defined models to assist in embryo selection. Software does not control incubator hardware. Used in clinical embryology laboratories. Benefits include continuous monitoring of embryo development without removing dishes from the controlled environment, potentially aiding in selection of viable embryos for transfer.
Clinical Evidence
Bench testing only. Includes electrical safety (IEC 61010-1, 61010-2-010), EMC (IEC 60601-1-2), software validation, cybersecurity, and performance testing (temperature/gas stability, imaging, auto-focus). Culture dish testing includes radiation sterilization (ISO 11137-2), transportation (ASTM D4169), shelf-life (ASTM F1980), endotoxin (USP <85>), and Mouse Embryo Assay (MEA) showing ≥80% blastocyst development at 96 hours.
Technological Characteristics
Benchtop tri-gas incubator; heating via aluminum container/lid; gas control via HEPA/VOC filters. Imaging: inverted microscope, monochrome CMOS camera, 20x magnification, 635 nm red LED. Connectivity: workstation-based. Culture dish: polystyrene, radiation sterilized. Software: non-hardware controlling, used for data management, annotation, and modeling.
Indications for Use
Indicated for the culture and monitoring of human embryos in Assisted Reproductive Technology (ART) procedures for a maximum duration of 120 hours.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of an eagle, while the FDA name on the right is written in blue and consists of the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
May 7, 2024
Wuhan Huchuang Union Technology Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 CHINA
Re: K232493 Trade/Device Name: Embryo Real-time Incubator (TLS301), Embryo Real-time Culture Dish (MC 2004) Regulation Number: 21 CFR 884.6120 Regulation Name: Assisted Reproduction Accessories Regulatory Class: II Product Code: MQG, MQK, MTX Dated: April 7, 2024 Received: April 8, 2024
Dear Ivy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510/k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael T. Bailey -S
For Monica D. Garcia, PhD Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K232493
#### Device Name
Embryo Real-time Incubator (TLS301), Embryo Real-time Culture Dish (MC 2004)
#### Indications for Use (Describe)
The Embryo Real-time Incubator (TLS301) is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Embryo Real-time Incubator (TLS301) has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.
The Embryo Real-time Culture Dish (MC 2004) is intended for preparing, storing and transferring human embryos. It is intended to be used only with the Embryo Real-time Incubator (TLS301).
Embryo Realtime 2S Software and Embryo Realtime 3E Software accessories for the Embryo Realtime Incubator (TLS301).
The Embryo Realtime 2S Software is intended to store, archive and transfer data. In addition, the Embryo Realtime 2S Software includes functions for managing models and performing calculations based on image data and embryo development parameters.
Embryo Realtime 3E Software is intended for viewing and recording embryo development events from images captured using the Embryo Real-time Incubator (TLS301). Embryo Realtime 3E Software includes a user annotation for capturing information on embryo development parameters and a user-defined modeling function that allows the user to combine annotated information on embryo development parameters to aid in embryo selection.
Embryo Realtime 2S Software and Embryo Realtime 3E Software do not control any hardware components in the Embryo Real-time Incubator (TLS301). Embryo Realtime 2S Software and Embryo Realtime 3E Software are provided in different software package and must be used together.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------|
| <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# K232493
#### I. Submitter Information
Name: Wuhan Huchuang Union Technology Co., Ltd. Address: Building10, No.818 Gaoxin Avenue Wuhan Hubei 430206 China Name of contact person: Zhan Yuanyuan Tel: +86-13554695434 E-Mail: hc@huchuang.com
## II. Date Prepared
May 7, 2024
## III.Subject Device Information
Device trade name: Embryo Real-time Incubator (TLS301), Embryo Real-time Culture Dish (MC 2004) Common Name: Embryo Incubator Regulation number: 21 CFR 884.6120 Regulation Name: Assisted Reproduction Accessories Regulation class: II Product code: MQG (Accessory, Assisted Reproduction), MQK (Labware, Assisted Reproduction), MTX (Microscope and Microscope Accessories, Reproduction, Assisted)
# IV. Predicate Device
EmbryoScope+ (K173264) manufactured by Vitrolife A/S
The predicate device has not been subject to any design related recalls.
#### V. Device Description
The Embryo Real-time Incubator (TLS301) consists of the following components:
- Embryo Real-time Incubator, integrated with temperature control system, gas ● supply control system and time-lapse imaging system.
- Image capture software
- Workstation, composed of server and workstation software. ●
- Embryo Real-time Culture Dish (MC 2004) ●
The Embryo Real-time Incubator (TLS301) is a benchtop tri-gas (CO2, N2, and air) embryo incubator with a built-in microscope for time-lapse imaging intended to be used for the culture and monitoring of embryos used in Assisted Reproductive Technology (ART) procedures. It provides temperature control, gas control, and time-lapse imaging at multiple focal planes. The incubator chamber has the capacity to hold up to ten
{5}------------------------------------------------
Embryo Real-time Culture Dishes (MC 2004). Each chamber contains a heating plate to maintain chamber temperatures. Gas (CO2, N2, and air) is mixed in a mixing chamber and passed through a HEPA/VOC filter prior to delivery to the incubation chamber. The built-in microscope consists of an illumination unit (red LED, 635 nm) and an inverted microscope/camera unit. The imaging system is mobile and is controlled through moving guide rail to automatically position the camera to the designated culture plate/well. The camera can acquire images in multiple focal planes. The time-lapse imaging system in the incubator along with the image capture software capture timelapse images of the embryos and transmit captured images to the computer for display and storage. The image capture software also reads the patient labels on the Embryo Real-time Culture Dish (MC 2004) and incorporates patient information into the imaging record.
The workstation includes a server and workstation software (Embryo Realtime 3E Software). It includes graphical user interface and receives and stores images. It also supports query, retrieval and display of the embryo images. The workstation software allows for manual annotation of the series of images obtained through time-lapse imaging into a user-defined model for the assessment of embryo's development. In addition, the server software (Embryo Realtime 2S Software), that does not include a graphical user interface, is designed to archive, transfer and store images of embryos from the time lapse incubator and performs user model management and calculations based on image data and user inputted embryo development parameters. The two software are used together.
The Embryo Real-time Culture Dish (MC 2004) is a single use, single-patient, polystyrene, radiation sterilized culture dish intended for preparing, storing, and transferring human embryos. It is intended to be used only with the Embryo Real-time Incubator (TLS301). It contains two types of wells for rinsing and handling the embryos before or after incubation and 16 wells for culturing the embryos during incubation. Each culture well is used to culture one embryo and a total of 16 embryos from a single patient can be cultured on one dish. The culture and rinsing wells have a volume of 30 µL and 50 µL, respectively. There is a central depression in the center of each culture well, where the embryo resides. The Embryo Real-time Culture Dishes (MC 2004) has a label area for unique identification of the culture dish.
#### VI. Indications for Use
The Embryo Real-time Incubator (TLS301) is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Embryo Real-time Incubator (TLS301) has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.
The Embryo Real-time Culture Dish (MC 2004) is intended for preparing, storing and transferring human embryos. It is intended to be used only with the Embryo Real-time
{6}------------------------------------------------
Incubator (TLS301).
Embryo Realtime 2S Software and Embryo Realtime 3E Software are optional software accessories for the Embryo Real-time Incubator (TLS301).
The Embryo Realtime 2S Software is intended to store, archive and transfer data. In addition, the Embryo Realtime 2S Software includes functions for managing models and performing calculations based on image data and embryo development parameters.
Embryo Realtime 3E Software is intended for viewing and recording embryo development events from images captured using the Embryo Real-time Incubator (TLS301). Embryo Realtime 3E Software includes a user annotation for capturing information on embryo development parameters and a user-defined modeling function that allows the user to combine annotated information on embryo development parameters to aid in embryo selection.
Embryo Realtime 2S Software and Embryo Realtime 3E Software do not control any hardware components in the Embryo Real-time Incubator (TLS301). Embryo Realtime 2S Software and Embryo Realtime 3E Software are provided in different software package and must be used together.
| Attribute | Subject device<br>(K232493) | Predicate device<br>(K173264) | Comparison |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Embryo Real-time<br>Incubator (TLS301) is<br>intended to provide an<br>environment with<br>controlled temperature<br>and mixed gas (CO2 and<br>other gases) for the<br>development of embryos.<br>The Embryo Real-time<br>Incubator (TLS301) has<br>an integrated camera and<br>optics for imaging and<br>viewing embryos during<br>incubation, for a<br>maximum time of 120<br>hours.<br>The Embryo Real-time | The EmbryoScope+<br>incubator provides an<br>environment with<br>controlled temperature and<br>gas concentrations (CO2<br>and O2) for the<br>development of embryos at<br>or near body temperature.<br>Use of the EmbryoScope+<br>incubator is limited to five<br>days (120 hr) covering the<br>time from post insemination<br>to day five of development.<br><br>The EmbryoSlide+ culture<br>dish is intended for<br>preparing, storing, and<br>transferring human | The indications for<br>use statements for<br>the subject and<br>predicate devices<br>are not identical;<br>however, their<br>intended uses are<br>the same (i.e.,<br>time-lapse imaging<br>of embryos<br>maintained in a<br>device-specific<br>culture dish). |
| Culture Dish (MC 2004)<br>is intended for preparing,<br>storing and transferring<br>human embryos. It is<br>intended to be used only<br>with the Embryo Real-<br>time Incubator (TLS301). | embryos. The<br>EmbryoSlide+ culture dish<br>must be used together with<br>the EmbryoScope+<br>incubator. | | |
| Embryo Realtime 2S<br>Software and Embryo<br>Realtime 3E Software are<br>optional software<br>accessories for the<br>Embryo Real-time<br>Incubator (TLS301). | The EmbryoViewer<br>software is intended for<br>displaying, comparing,<br>storing, and transferring<br>images generated by the<br>EmbryoScope+ incubator.<br>This software includes a<br>user annotation function for<br>capturing information on<br>embryo development<br>parameters as well as a<br>user-defined modeling<br>function, which allows the<br>user to combine annotated<br>information on embryo<br>development parameters to<br>aid in embryo selection.<br>The EmbryoViewer<br>software does not control<br>any hardware components<br>in the EmbryoScope+<br>incubator. | | |
| The Embryo Realtime 2S<br>Software is intended to<br>store, archive and transfer<br>data. In addition, the<br>Embryo Realtime 2S<br>Software includes<br>functions for managing<br>models and performing<br>calculations based on<br>image data and embryo<br>development parameters. | The ES Server software is<br>intended to store, archive<br>and transfer data. In<br>addition, this software<br>includes functions for<br>managing models and<br>performing calculations<br>based on image data and<br>embryo development<br>parameters. | | |
| Embryo Realtime 3E<br>Software is intended for<br>viewing and recording<br>embryo development<br>events from images<br>captured using the<br>Embryo Real-time<br>Incubator (TLS301).<br>Embryo Realtime 3E<br>Software includes a user<br>annotation function for<br>capturing information on<br>embryo development<br>parameters and a user-<br>defined modeling<br>function that allows the | The EmbryoScope+<br>incubator, EmbryoViewer<br>software, and ES Server<br>software must be used | | |
| | user to combine<br>annotated information on<br>embryo development<br>parameters to aid in<br>embryo selection.<br><br>Embryo Realtime 2S<br>Software and Embryo<br>Realtime 3E Software do<br>not control any hardware<br>components in the<br>Embryo Real-time<br>Incubator (TLS301).<br>Embryo Realtime 2S<br>Software and Embryo<br>Realtime 3E Software are<br>provided in different<br>software package and<br>must be used together. | together to export embryo<br>images from the<br>EmbryoScope+ incubator.<br>The EmbryoViewer<br>software and ES Server<br>software must be used<br>together to analyze the<br>embryo images. | |
| | Incubator, integrated computer and microscopy system | | |
| Culture dish<br>capacity | 10 dishes | 15 dishes | Different |
| Heating<br>mechanism | Heated aluminum<br>container and lid | Direct heat transfer | Different |
| Temperature range | 35-40 °C | 36-39 °C | Similar |
| Temperature<br>accuracy | $\pm$ 0.2 °C | $\pm$ 0.2 °C | Similar |
| CO2 accuracy | $\pm$ 0.1% for CO2<br>concentration at 6.0%,<br>$\pm$ 0.2% for other CO2<br>concentration | $\pm$ 0.3 % | Different |
| O2 accuracy | $\pm$ 0.1% for O2<br>concentration at 5.0%,<br>$\pm$ 0.2% for other O2<br>concentration | $\pm$ 0.5 % | Different |
| Recirculation rate | 150ml $\pm$ 7.5ml/min (full<br>purification of gas<br>volume every 6.3 min) | >100 L/h (full<br>purification of gas<br>volume every 6 min) | Different |
| Recovery time | CO2 (6% $\pm$ 0.3%) $\leq$ 90<br>seconds<br>O2(5% $\pm$ 0.3%) $\leq$ 90 | CO2 (5% $\pm$ 0.3 %) < 5 min<br>O2 (5% $\pm$ 0.5 %) <3 min | Different |
| | | load door opening | |
| | closed after a 15-second<br>load door opening (single<br>chamber) | | |
| Control of<br>temperature and<br>gas | Firmware | Firmware | Same |
| Computer | Integrated | Integrated | Same |
| Microscope | Inverted microscope | Inverted microscope | Same |
| Type of camera | Monochrome CMOS | Monochrome CMOS | Same |
| Magnification | 20× | 16× | Different |
| Focusing | Fully automated dish<br>detection and embryo<br>focusing | Fully automated dish<br>detection and embryo<br>focusing | Same |
| Numerical aperture | 0.45 | 0.5 | Similar |
| Number of pixels | 2048 × 1088 pixels | 2048 × 1088 pixels | same |
| Resolution | ≥ 2.7 pixels per µm | 3 pixels per µm | Different |
| Light source (for<br>imaging) | Low-power red LED 635<br>nm | Low-power red LED 627<br>nm | Similar |
| Time-lapse system | Time-lapse imaging<br>(Hoffman Modulation<br>Contrast Objective). 10<br>Min cycle time for 11<br>focal planes | Time-lapse imaging<br>(Hoffman Modulation<br>Contrast Objective). 10<br>Min cycle time for 11<br>focal planes for up to 5<br>days | Same |
| Embryo Realtime 2S Software and Embryo Realtime 3E Software | | | |
| Image display | High-resolution time-<br>lapse images of single<br>embryos | High-resolution time-<br>lapse images of single<br>embryos | Same |
| Data handling | Export, storage and<br>transfer of embryo data | Export, storage and<br>transfer of embryo data | Same |
| Embryo annotation<br>and data analysis | • Embryo annotation<br>tools which assist the user<br>in selecting embryos<br>• Model designer | • Embryo annotation<br>tools which assist the user<br>in selecting embryos<br>• Model designer<br>• Model management<br>• Data analysis using<br>user-defined modeling<br>function | Different |
| Incubation<br>condition<br>monitoring | Inspection of incubation<br>details, such as<br>temperature and gas<br>conditions | Inspection of incubation<br>details, such as<br>temperature and gas<br>conditions | Same |
| | Embryo Real-time Culture Dish (MC 2004) | | |
| General design | Optically clear culture dish with a lid | Optically clear culture dish with a lid | Same |
| Material | Polystyrene | Polystyrene | Same |
| Culture wells on the dish | 16 wells for individual embryo incubation | 16 wells for individual embryo incubation…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.