Guardia™ Access Nano and Soft-Trans Embryo Transfer Catheter Sets, Soft-Trans Embryo Transfer Catheter, Soft-Trans Guide Catheter

K172051 · Cook Incorporated · MQF · Jan 10, 2018 · Obstetrics/Gynecology

Device Facts

Record IDK172051
Device NameGuardia™ Access Nano and Soft-Trans Embryo Transfer Catheter Sets, Soft-Trans Embryo Transfer Catheter, Soft-Trans Guide Catheter
ApplicantCook Incorporated
Product CodeMQF · Obstetrics/Gynecology
Decision DateJan 10, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6110
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Guardia™ Access Nano and Soft-Trans Embryo Transfer Catheter Sets: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. Soft-Trans Embryo Transfer Catheter: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. This device is to be used in combination with a cleared, dimensionally compatible Embryo Transfer Guide Catheter for gaining access to the uterine cavity. Soft-Trans Guide Catheter: Used to supplement and assist uterine access of a cleared, dimensionally compatible Embryo Transfer Catheter for placement of in vitro fertilized (IVF) embryos into the uterine cavity.

Device Story

Catheter sets facilitate placement of IVF embryos into uterine cavity; used by clinicians during assisted reproduction procedures. System comprises transfer catheter, guide catheter, and optional trial catheter. Transfer catheters (polyurethane/stainless steel) deliver embryos; guide catheters (polyethylene/polymethylpentene) provide access to uterine cavity. Trial catheter (polycarbonate/polyurethane/stainless steel) assesses placement path prior to embryo transfer. Devices feature graduation marks, locking hubs, and depth indicators. Single-use, sterile, ethylene oxide-processed. Clinician inserts guide catheter to access uterus, then advances transfer catheter through guide to deposit embryos. Benefits include controlled, atraumatic embryo delivery to uterine cavity.

Clinical Evidence

Bench testing only. No clinical data. Testing included: Sterilization validation (ISO 11135-1:2007), biocompatibility (ISO 10993-5:2009, ISO 10993-10:2010), endotoxin testing (USP <85>), transportation simulation (ASTM D4169-16), package integrity (ASTM F2096-04, ASTM F88-09), and Mouse Embryo Assay (MEA). MEA results met acceptance criteria of ≥80% embryo development to blastocyst stage within 96 hours. Bench performance testing confirmed dimensional verification, aspiration integrity, leak resistance, and tensile strength met predetermined specifications.

Technological Characteristics

Materials: Polyurethane, stainless steel, polyethylene, polymethylpentene, polycarbonate. Dimensions: Transfer catheters 2.8-4.7 Fr (19.1-24 cm); Guide catheters 5.5-8.1 Fr (11.4-17.3 cm); Trial catheter 4.0 Fr (18 cm). Sensing/Actuation: Mechanical. Energy: None. Connectivity: None. Sterilization: Ethylene oxide. Standards: ISO 11135-1, ISO 10993, USP <85>, AAMI/ANSI ST72, ASTM D4169, ASTM F2096, ASTM F88.

Indications for Use

Indicated for patients undergoing in vitro fertilization (IVF) requiring embryo placement into the uterine cavity.

Regulatory Classification

Identification

Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 10, 2018 Cook Incorporated Naomi Funkhouser Regulatory Affairs Specialist 750 Daniels Way, P.O. Box 489 Bloomington, IN 47402 Re: K172051 > Trade/Device Name: Guardia™ Access Nano and Soft-Trans Embryo Transfer Catheter Sets, Soft-Trans Embryo Transfer Catheter, Soft-Trans Guide Catheter Regulation Number: 21 CFR§ 884.6110 Regulation Name: Assisted Reproduction Catheters Regulatory Class: II Product Code: MOF Dated: December 8, 2017 Received: December 11, 2017 Dear Naomi Funkhouser: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/6 description: The image shows the name "Joyce M. Whang -S" in a large, bold font. The name is written in black and is centered on a white background. The letters are clear and easy to read. for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number (if known) K172051 #### Device Name Guardia™ Access Nano and Soft-Trans Embryo Transfer Catheter Sets #### Indications for Use (Describe) Guardia™ Access Nano and Soft-Trans Embryo Transfer Catheter Sets: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K172051 #### Device Name Soft-Trans Embryo Transfer Catheter Indications for Use (Describe) Soft-Trans Embryo Transfer Catheter: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. This device is to be used in combination with a cleared, dimensionally compatible Embryo Transfer Guide Catheter for gaining access to the uterine cavity. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------| | <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # Indications for Use 510(k) Number (if known) K172051 Device Name Soft-Trans Guide Catheter Indications for Use (Describe) Soft-Trans Guide Catheter: Used to supplement and assist uterine access of a cleared, dimensionally compatible Embryo Transfer Catheter for placement of in vitro fertilized (IVF) embryos into the uterine cavity. | Type of Use (Select one or both, as applicable) | |--------------------------------------------------------------------------------------| | <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif font on the top portion of the logo. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font. COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA HONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM # 510(k) Summary – K172051 ### Submitted By: | Applicant: | Cook Incorporated | |----------------------------------------|----------------------------------------------| | Contacts: | Naomi Funkhouser | | | Rohini Patel | | Applicant Address: | Cook Incorporated | | | 750 Daniels Way | | | Bloomington, IN 47404 | | Contact Phone: | (812) 339-2235 x104371 | | Contact Fax: | (812) 332-0281 | | Date Prepared: | January 8, 2018 | | Device Information: | | | Trade Names: | Guardia™ Access Nano and Soft-Trans Embryo | | | Transfer Catheter Sets, Soft-Trans Embryo | | | Transfer Catheter, Soft-Trans Guide Catheter | | Common Name: | Embryo Transfer Catheter | | Classification Name: | Assisted Reproduction Catheters (21 CFR | | | 884.6110) | | Product Code: | MQF (Catheter, Assisted Reproduction) | | Regulatory Class/Classification Panel: | II | ## Predicate Device: Embryo Transfer Catheter/Sets (K983594) manufactured by Cook Ob/Gyn. This predicate device has not been subject to any design related recalls. ## Device Description: This 510(k) covers the following subject devices: | Device | RPN | Component | |--------------------------------------------------|-----------------|-----------------------------------------------------------------| | Guardia™ Access Nano<br>Embryo Transfer Catheter | K-JETS-551910-S | Embryo transfer catheter and guide catheter | | Soft-Trans Embryo Transfer<br>Catheter | K-SOFT-5000 | Embryo transfer catheter and guide catheter | | | K-SOFT-5000-ST | Embryo transfer catheter and guide catheter | | | K-SOFT-5010 | Guide catheter | | | K-SOFT-5100 | Embryo transfer catheter | | | K-SOFT-5000-TC | Embryo transfer catheter, guide catheter, and trial<br>catheter | {6}------------------------------------------------ The transfer catheters are manufactured from polyurethane. The Guardia™ Access Nano Embryo Transfer Catheter has a diameter of 2.8 Fr and is 24 cm in length. The Soft-Trans Embryo Transfer Catheters have a diameter of 4.7 Fr and range in length from 19 to 23 cm. Both transfer catheters also include a stainless steel cannula to provide additional support and depth indicators. The Soft-Trans Embryo Transfer Catheter Set (K-SOFT-5000-TC) also includes a closed-ended polycarbonate/polyurethane/stainless steel trial catheter that has a diameter of 4.0 Fr and is 18 cm in length. The guide catheter for the Guardia™ Access Nano Embryo Transfer Catheter is manufactured from polymethylpentene, and has a diameter of 5.5 Fr and is 17.3 cm in length. This guide catheter also features a bulb tip, silicone position marker, and is provided pre-curved. The guide catheter for the Soft-Trans Embryo Transfer Catheter Sets are manufactured from polyethylene, have a diameter of 8.1 Fr, and are 11.4 or 15.4 cm in length. This guide catheter is provided in a straight configuration. All subject devices are single-use devices and sterilized by ethylene oxide exposure. The transfer catheters, transfer catheter/sets undergo lot release Mouse Embryo Assay (MEA) for embryo toxicity and USP endotoxin (LAL) for pyrogenicity. The subject devices are packaged in peelopen sterile barrier pouches with a three-year shelf life. ### Indications for Use: Guardia™ Access Nano and Soft-Trans Embryo Transfer Catheter Sets: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. Soft-Trans Embryo Transfer Catheter: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. This device is to be used in combination with a cleared, dimensionally compatible Embryo Transfer Guide Catheter for gaining access to the uterine cavity. ### Soft-Trans Guide Catheter: Used to supplement and assist uterine access of a cleared, dimensionally compatible Embryo Transfer Catheter for placement of in vitro fertilized (IVF) embryos into the uterine cavity. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif, block letters on a red background. The red background is shaped like a banner. ## Comparison of Intended Use and Technological Characteristics with the Predicate Device: | Parameter | K172051 (subject device) | K983594 (predicate device) | |-----------------------------------|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Same as predicate | The embryo transfer catheters/sets are used<br>to place embryos into the uterine cavity. | | Design of transfer<br>catheter | Same as predicate | • Graduation marks<br>• Open end with side port<br>• Locking hubs | | Design of guide<br>catheter | Same as predicate | • Graduation marks<br>• Open end with side port<br>• Locking hubs<br>• Precurved or straight<br>• With or without positioner | | Design of trial catheter | • Positioning marks<br>• Closed end | N/A | | Dimension of transfer<br>catheter | OD 2.8-4.7 Fr<br>Length 19.1-24 cm | OD 2-8 Fr<br>Length 12-30 cm | | Dimensions of guide<br>catheter | OD 5.5-8.1 Fr<br>Length 11.4-17.3 cm | OD 5-8.5 Fr<br>Length 12-30 cm | | Dimensions of trial<br>catheter | OD 4.0 Fr<br>Length 18 cm | N/A | | Materials – Transfer<br>catheter | Polyurethane, Stainless Steel | Polyethylene, Teflon, Stainless Steel | | Materials - Guide<br>catheter | Polyethylene,<br>Polymethylpentene | Polyethylene, Echosight Polyethylene,<br>Teflon, Stainless Steel | | Material - Trial<br>catheter | Polycarbonate, Stainless Steel,<br>Polyurethane | N/A | The subject and predicate devices have intended use. They also have the same design. There are differences in dimensions and materials between the subject and predicate devices, but these differences do not raise any different questions of safety and effectiveness. The difference in dimensions can be addressed by bench performance testing. The differences in materials can be evaluated by biocompatibility testing and MEA. In addition, one version of the subject Soft-Trans Embryo Transfer Catheter Set (K-SOFT-5000-TC) includes a trial catheter that can be used to assess the placement of the device prior to conducting the actual embryo transfer procedure. The addition of the trial catheter does not raise different questions of safety and effectiveness as compared to the predicate device (ability to deliver the device to the uterine cavity, perforation risks, etc.). {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in smaller, white letters on a red background. The red background is shaped like a rectangle with a diagonal cut on the bottom right corner. ## Summary of Non-Clinical Performance Testing: The following studies have been performed to support substantial equivalence to the predicate device: - . Sterilization Validation testing per ISO 11135-1:2007 - Biocompatibility studies, as follows: . - Cytotoxicity testing per 10993-5:2009 о - Guinea Pig Maximization Sensitization testing per ISO 10993-10:2010 O - Intracutaneous Irritation testing per ISO 10993-10:2010 O - Endotoxin testing per USP <85> and AAMI/ANSI ST72:2011/(R)2016 (<20 . EU/device) - Transportation Simulation study per ASTM D4169-16 ● - Package Integrity testing after real-time aging: ● - Bubble Leak test per ASTM F2096-04 O - o Seal Strength testing per ASTM F88-09 - Visual Inspection: No package displayed damage (tears, folds, puncture holes, etc.) o - Mouse Embryo Assay (MEA) before and after aging: One-cell mouse embryos were exposed to subject devices and cultured at 37℃ in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed in comparison with the control group. The testing demonstrated that the devices met acceptance criterion of "1-cell MEA ≥80% embryos developed to blastocyst in 96 hours." - Bench Performance studies before and after real-time aging demondtrated that all . predetermined acceptance criteria were met in the following tests: - * Dimensional Verification of Guide Catheter - The guide catheters are measured and verified against device input requirements. - * Transfer Catheter Aspiration Test - Testing ensures there is no air leaking into the syringe through the transfer catheter. - * Transfer Catheter Leak Test - Testing ensures that fluid path catheter assembly does not leak under a predetermined injection pressure. - * Tensile Testing (Catheter Hub-shaft) - Testing demonstrated that the tensile strength value was greater than the predetermined acceptance criterion. ### Conclusion: The subject and predicate devices have the same intended use. Although there are differences in technological characteristics between the subject and predicate devices, these differences do not raise different questions of safety or effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...