Daylily Single Use Sterile Embryo Transfer Catheter

K220010 · Shanghai Horizon Medical Technology Co., Ltd. · MQF · Dec 16, 2022 · Obstetrics/Gynecology

Device Facts

Record IDK220010
Device NameDaylily Single Use Sterile Embryo Transfer Catheter
ApplicantShanghai Horizon Medical Technology Co., Ltd.
Product CodeMQF · Obstetrics/Gynecology
Decision DateDec 16, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6110
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Daylily Single Use Sterile Embryo Transfer Catheters are used to place in vitro fertilized (IVF) embryos into the uterine cavity.

Device Story

Daylily Single Use Sterile Embryo Transfer Catheters are sterile, single-use devices designed for IVF embryo delivery. The system comprises a transfer catheter and a guide catheter; specific models (Types II and IV) include an optional trial catheter. The guide catheter, featuring a rounded/blunt tip and marker bands, is inserted through the cervix to facilitate the placement of the transfer catheter. The transfer catheter, also featuring a rounded/blunt tip and marker bands, is pre-loaded with embryos and inserted through the guide catheter into the uterine cavity. A syringe (not provided) is attached to the transfer catheter connector to deliver the embryos. The trial catheter provides support and assesses placement prior to the actual transfer. Used in clinical settings by healthcare professionals. Benefits include precise, guided delivery of embryos to the uterine cavity to support assisted reproduction procedures.

Clinical Evidence

Bench testing only. Performance validated via Mouse Embryo Assay (MEA) showing ≥ 80% of embryos developed to expanded blastocyst at 96 hours. Biocompatibility testing (cytotoxicity, sensitization, irritation) per ISO 10993, sterilization validation per ISO 11135, and package integrity/transportation simulation per ASTM standards confirmed device safety and performance.

Technological Characteristics

Materials: Polyurethane, polycarbonate, stainless steel, polyether block polyamide, silicone. Components: Transfer catheter, guide catheter, optional trial catheter. Features: Rounded/blunt tips, marker bands, pre-curved guide catheters (select models), positioning rings. Sterilization: Ethylene oxide. Shelf life: 3 years. Standards: ISO 10993, ISO 11135, ASTM F2096, ASTM F88, ASTM F1929, ASTM D4169, ASTM F1980, USP <85>.

Indications for Use

Indicated for the placement of in vitro fertilized (IVF) embryos into the uterine cavity.

Regulatory Classification

Identification

Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 16, 2022 Shanghai Horizon Medical Technology Co., Ltd. % Zhixuan Zhang Senior Supervisor of Regulatory Affairs MicroPort Group Co., Ltd. 1601 ZhangDong Rd., ZJ Hi-Tech Park Pudong New District Shanghai, 201203 China - K220010 Re: Trade/Device Name: Daylily Single Use Sterile Embryo Transfer Catheters (Type I (ETC-1018 S. ETC-1018 L. ETC-1024 S. ETC-1024 Ll. Type II IETC-2018 S. ETC-2018 L. ETC-2024 S, ETC-2024 L], Type III [ETC-3018 S, ETC-3018 L, ETC-3024 S, ETC-3024 L, ETC-3024 S-X, ETC-3024 L-X], and Type IV [ETC-4018 S, ETC-4018 L, ETC-4024 S, ETC-4024 L, ETC-4024 S-X, ETC-4024 L-X]) Regulation Number: 21 CFR§ 884.6110 Regulation Name: Assisted Reproduction Catheters Regulatory Class: II Product Code: MQF Dated: November 14, 2022 Received: November 16, 2022 ## Dear Zhixuan Zhang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {2}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). # Sincerely, Michael T. Bailey -S For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220010 ### Device Name Daylily Single Use Sterile Embryo Transfer Catheters (Type I [ETC-1018 L, ETC-1024 S, ETC-1024 L, Type II [ETC-2018 S, ETC-2018 L, ETC-2024 L], Type III [ETC-3018 S, ETC-3018 L, ETC-3024 S, ETC-3024 L, ETC-3024 S-X, ETC-3024 L-X], and Type IV [ETC-4018 S, ETC-4018 L, ETC-4024 L, ETC-4024 S-X, ETC-4024 L-X]) ### Indications for Use (Describe) Daylily Single Use Sterile Embryo Transfer Catheters are used to place in vitro fertilized (IVF) embryos into the uterine cavity. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary - K220010 #### SUBMITTER I. | Applicant: | Shanghai Horizon Medical Technology Co., Ltd. | |--------------------|-----------------------------------------------------------------------------------------------------------------| | Applicant Address: | Room 401, building 24, No. 1-28, 588 Tianxiong Rd<br>Pudong New Area, Shanghai, P.R.China.<br>Postcode: 201318. | | Phone: | +86-21-38954600 | |-----------------|---------------------------| | Fax: | +86-21-50801305 | | Email: | BeiBei.Xing@microport.com | | Contact Person: | Beibei Xing | Date Prepared: December 14, 2022 #### II. DEVICE | Trade Name: | Daylily Single Use Sterile Embryo Transfer Catheters (Type I [ETC-1018 S, ETC-1018 L, ETC-1024 S, ETC-1024 L], Type II [ETC-2018 S, ETC-2018 L, ETC-2024 S, ETC-2024 L], Type III [ETC-3018 S, ETC-3018 L, ETC-3024 S, ETC-3024 L, ETC-3024 S-X, ETC-3024 L-X], and Type IV [ETC-4018 S, ETC-4018 L, ETC-4024 S, ETC-4024 L, ETC-4024 S-X, ETC-4024 L-X]) | |--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Embryo Transfer Catheters | | Regulation Name: | Assisted Reproduction Catheters | | Regulation Number: | 21 CFR 884.6110 | | Regulatory Class: | II | | Product Code: | MQF (Catheter, Assisted Reproduction) | #### III. PREDICATE DEVICE Guardia™ Access Nano and Soft-Trans Embryo Transfer Catheter Sets (K172051) manufactured by Cook Incorporated. The predicate device has not been subject to a design-related recall. #### IV. DEVICE DESCRIPTION Daylily Single Use Sterile Embryo Transfer Catheters are sterile single use catheters used to deliver in vitro fertilized embryos to the uterine cavity. All models of the Daylily Single Use Sterile Embryo Transfer Catheters consist of a transfer catheter and a guide {5}------------------------------------------------ catheter. Some device models (Type II and Type IV) also include a trial catheter. The guide catheter is composed of a catheter shaft, connector, and a positioning ring (for certain Type III and IV variants). Guide catheters for some Type III and Type IV variants are also provided pre-curved. Guide catheters have a rounded/blunt tip and marker bands at the distal tip to aid in catheter placement. The guide catheter is delivered through the cervix first and is used to guide the insertion of the transfer catheter holding the embryos into the uterine cavity. The transfer catheter is composed of a catheter shaft, connector, and a stainless-steel sleeve (for Type III and IV models). Transfer catheters have a rounded/blunt tip and marker bands at the proximal end of the catheter to aid in catheter placement. The transfer catheter is loaded with embryos prior to delivery through the guide catheter and into the uterine cavity. A syringe (not provided with catheters) connected to the connector of the transfer catheter is used to deliver the embryos into the uterine cavity. The trial catheter is an optional accessory for specific models (Types II and IV), composed of a catheter shaft, connector, and a polymer-coated stainless-steel core. Trial catheters have a rounded/blunt closed tip. The trial catheter is used to provide additional support during guide catheter placement and to assess the placement of the device prior to conducting an actual embryo transfer procedure. Device configurations and specifications for the Daylily Single Use Sterile Embryo Transfer Catheters (Types I-IV) are shown below: | | Components | | | | | | |----------|-----------------------------------------------------|----------------|-----------------------------------------------------|----------------|---------------------------|--------------| | Model | Transfer catheter | | Guide catheter | | Trial catheter | | | | Outer<br>Diameter<br>(mm)/Inner<br>Diameter<br>(mm) | Length<br>(cm) | Outer<br>Diameter<br>(mm)/Inner<br>Diameter(m<br>m) | Length<br>(cm) | Outer<br>Diameter<br>(mm) | Length (cm) | | Type I | | | 2.35 mm<br>OD/1.55-<br>1.75 mm ID | | N/A | N/A | | Type II | 0.95-1.50<br>mm<br>OD/0.55-<br>0.75 mm ID | 18-24 cm | | 12-17.3 cm | 1.40 mm | 15.3-20.6 cm | | Type III | | | | | N/A | N/A | | Type IV | | | 2.35-2.50/<br>1.55-1.90 ID | | 1.40 mm | 15.3-20.6 cm | Table 1. Device configurations and models {6}------------------------------------------------ The Daylily Single Use Sterile Embryo Transfer Catheters are intended for single use only and sterilized by ethylene oxide. They have a three-year shelf life. #### INDICATIONS FOR USE V. Daylily Single Use Sterile Embryo Transfer Catheters are used to place in vitro fertilized (IVF) embryos into the uterine cavity. ### COMPARISON OF THE INTENDED USE AND TECHNOLOGICAL VI. CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICES The following table compares the Daylily Single Use Sterile Embryo Transfer Catheters to the predicate device. | Comparison Item | Subject Device –<br>K220010 | Predicate Device –<br>K172051 | Comparison | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------| | Manufacturer | Shanghai Horizon<br>Medical Technology<br>Co., Ltd. | Cook Incorporated | Not applicable | | Trade name | Daylily Single Use<br>Sterile Embryo Transfer<br>Catheters<br>(Type I [ETC-1018 S,<br>ETC-1018 L, ETC-1024<br>S, ETC-1024 L], Type II<br>[ETC-2018 S, ETC-<br>2018 L, ETC-2024 S,<br>ETC-2024 L], Type III<br>[ETC-3018 S, ETC-<br>3018 L, ETC-3024 S,<br>ETC-3024 L, ETC-3024<br>S-X, ETC-3024 L-X],<br>and Type IV [ETC-4018<br>S, ETC-4018 L, ETC-<br>4024 S, ETC-4024 L,<br>ETC-4024 S-X, ETC-<br>4024 L-X]) | Guardia™ Access<br>Nano and Soft-Trans<br>Embryo Transfer<br>Catheter Sets | Not applicable | | Indications for Use | Daylily Single Use<br>Sterile Embryo Transfer<br>Catheters are used to<br>place in vitro fertilized | Used to place in<br>vitro fertilized (IVF)<br>embryos into the<br>uterine cavity. | Same | | | | Table 2. Intended use and technological characteristics comparison. | | | |--------------------------------------------|-------------------------------------------|----------------------------------------------------------------------------------|-----------------------------------------------------------------------|-----------| | | | | | | | Comparison Item | Subject Device –<br>K220010 | Predicate Device –<br>K172051 | Comparison | | | | (IVF) embryos into the<br>uterine cavity. | | | | | Dimension | Transfer<br>catheter | OD 0.95-1.5 mm (2.85-<br>4.5 Fr)<br>Length 18-24 cm | OD 2.8-4.7 Fr<br>Length 19.1-24 cm | Different | | | Guide<br>catheter | OD 2.35-2.5 mm (7.05-<br>7.5 Fr)<br>Length 12-17.3 cm | OD 5.5-8.1 Fr<br>Length 11.4-17.3 cm | Different | | | Trial<br>catheter | OD 1.4mm (4.2 Fr)<br>Length 15.3-20.6 cm | OD 4.0 Fr<br>Length 18 cm | Different | | Material | Transfer<br>catheter | Polyurethane,<br>polycarbonate, stainless<br>steel (with applicable<br>variants) | polyurethane,<br>stainless steel | Different | | | Guide<br>catheter | Polyether block<br>polyamide, silicone<br>(with applicable<br>variants) | Polyethylene,<br>polymethylpentene,<br>silicone | Different | | | Trial<br>catheter | Polyether block<br>polyamide, stainless<br>steel | Polycarbonate,<br>stainless steel,<br>polyurethane | Different | | Variants with pre-curved<br>guide catheter | | Yes | Yes | Same | | Variants with positioning<br>rings/markers | | Yes | Yes | Same | | Depth markers on<br>catheters | | Yes | Yes | Same | | Sterility | | Sterilized by ethylene<br>oxide | Sterilized by<br>ethylene oxide | Same | | Single Use | | Yes | Yes | Same | | Shelf Life | | 3 years | 3 years | Same | | Mouse Embryo Assay | | 1-Cell MEA: ≥ 80%<br>embryos developed to<br>expanded blastocyst at<br>96 hours. | 1-cell MEA ≥80%<br>embryos developed<br>to blastocyst in 96<br>hours. | Similar | | Endotoxin | | < 20 EU/device | < 20 EU/device | Same | {7}------------------------------------------------ The subject and predicate devices have identical indications for use statements and the same intended use. As shown in the table above, the subject and predicate device have many technological characteristics that are the same or similar. However, there are differences in dimensions and device materials. These differences do not raise {8}------------------------------------------------ different questions of safety and effectiveness. #### VII. SUMMARY OF NON-CLINICAL PERFORMANCE DATA The following studies have been performed to support of the substantial equivalence to the predicate device: - Sterilization Validation testing per: . - ISO 11135-1:2014 A - A AAMI TIR 28:2016 - ISO 10993-7: 2008 A - Package Integrity testing: - > Visual inspection - A Bubble Leak test per ASTM F2096-11 - > Seal Strength testing per ASTM F88/ F88M-15 - A Dye Penetration test per ASTM F1929-15 - Transportation Simulation testing per ASTM D4169-14 . - Biocompatibility studies conducted in accordance with the 2020 FDA guidance . document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process. " Testing included the following assessments: - > Cytotoxicity per ISO 10993-5: 2009 - A Sensitization ISO 10993-10: 2010 - A Irritation per ISO 10993-10: 2010 Testing showed the device material to be non-cytotoxic, non-sensitizing, and nonirritating. - Endotoxin testing per USP <85> and AAMI/ANSI ST72:2019 Specification: <20 ● EU/device - Bench performance studies before and after accelerated aging to the equivalent of three-years of real-time aging in accordance with ASTM F1980-16 demonstrated that all predetermined acceptance criteria were met in the following tests: - A Appearance - A Taper/Syringe compatibility - A Dimensional analysis - > Distance indication marker location and durability - > Dislodgement of positioning ring - > Tip drop when held horizontally - > Bonding strength of device connections/bonds - > Aspiration and leakage testing - A Mouse Embryo Assay (MEA) per the 2021 FDA guidance Mouse Embryo Assay for Assisted Reproduction Technology Devices: Specification - 1-Cell {9}------------------------------------------------ MEA: ≥ 80% embryos developed to expanded blastocyst at 96 hours. ## VIII. CONCLUSIONS The results of the testing described above demonstrate that the Daylily Single Use Sterile Embryo Transfer Catheters are as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...