The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours. The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.
Device Story
Geri Embryo Incubator is a benchtop incubator with six independent modular chambers; each chamber features individual temperature control and gas inlet feeds. Device utilizes an integrated CMOS camera (2560x1928 pixels) and orange LED (591 nm) illumination to capture time-lapse images of embryos across 11 focal planes every 5 minutes. Geri Dish, an optically-clear polystyrene dish, holds up to 16 embryos per unit. System is operated by embryologists in a laboratory setting. Firmware controls incubator hardware settings; software manages patient data and imaging parameters. Operators view time-lapse images on a connected interface to monitor embryo development without removing samples from the controlled environment. This continuous monitoring facilitates embryo selection and assessment, potentially improving clinical outcomes in assisted reproduction.
Clinical Evidence
Bench testing only. Performance testing included electrical safety (IEC 61010-1), EMC (IEC 61326-1), and cleaning/disinfection validation. Bench tests confirmed gas flow/recovery, temperature stability, and time-lapse imaging accuracy. Embryo development testing showed ≥70% mouse 1-cell embryos reached blastocyst stage after 120 hours of imaging. Light toxicity testing confirmed total energy exposure does not exceed levels associated with standard IVF/ICSI microscopy. Geri Dish testing included radiation sterilization (ISO 11137-2), package integrity (ASTM F1929-12, ASTM F88/F88M-09), endotoxin testing (USP <85>), and Mouse Embryo Assay (MEA) showing ≥80% blastocyst development.
Technological Characteristics
Benchtop incubator with six modular chambers; heated aluminum stage and lid. CMOS camera (2560x1928 pixels); 591 nm orange LED illumination. Gas supply: premixed CO2/O2/N2. Geri Dish: optically-clear polystyrene. Sterilization: radiation (SAL 10^-9). Connectivity: firmware for hardware control, software for patient/setting management. Temperature range: 35.0-40.0°C. Dimensions/specs: 11 focal planes, 2 pixels/µm resolution.
Indications for Use
Indicated for use in assisted reproduction procedures for the incubation, preparation, storage, and time-lapse imaging of human embryos for a maximum duration of 120 hours.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
EmbryoScope (Version D) and EmbryoViewer Software (K113075)
K173264 — EmbryoScope+ · Vitrolife A/S · May 11, 2018
K242684 — Harioculture TL-16 Time-lapse Incubator · Hua Yue Medical Technology Co., Ltd. · May 22, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2017
Genea Biomedx Pty Ltd % Roger Gray VP, Quality Assurance and Regulatory Affairs Donawa Lifescience Consulting Srl Piazza Albania 10 00153 Rome Italy
Re: K171736
Trade/Device Name: Geri Embryo Incubator and Geri Dish Regulation Number: 21 CFR& 884.6120 Regulation Name: Assisted Reproduction Accessories Regulatory Class: II Product Code: MOG, MOK, MTX Dated: October 12, 2017 Received: October 16, 2017
Dear Roger Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171736
Device Name
Geri Embryo Incubator and Geri Dish
#### Indications for Use (Describe)
The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.
The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The logo is simple and modern, with a focus on the company's name.
## 510(k) Summary (K171736)
#### I. General Information on Submitter
| Submitter/Address:<br>Contact person:<br>Phone:<br>Fax: | Genea Biomedx Pty Ltd<br>Level 2, 321 Kent Street<br>Sydney<br>NSW 2000<br>Australia<br>Ms. Kea Dent<br>+61 2 8484 7677<br>+61 2 9229 6478 |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Correspondent:<br>Phone:<br>Fax:<br>Email: | Mr. Roger Gray<br>VP, Quality and Regulatory<br>Donawa Lifescience Consulting Srl<br>Piazza Albania 10<br>00153 Rome<br>Italy<br>+39 06 578 2665<br>+39 06 574 3786<br>rgray@donawa.com |
#### II. Date Prepared: November 14, 2017
#### III. General Information on Devices
| Device Name: | Geri Embryo Incubator and Geri Dish | |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Common Name: | Embryo Incubator | |
| Classification Name: | Assisted Reproduction Accessories (21 CFR 884.6120) | |
| Product code: | MQG (Accessory, Assisted Reproduction), MQK (Labware, Assisted<br>Reproduction, MTX (Microscope and Microscope Accessories,<br>Reproduction, Assisted) | |
| Regulatory Class: | II | |
#### IV. Predicate Device
EmbryoScope (Version D) and EmbryoViewer Software (K113075) and EmbryoScope and EmbryoSlide (K092183), which are manufactured by Unisense FertillTech A/S. The predicate devices have not been subject to any design related recalls.
#### V. Device Description
The Geri Embryo Incubator is a benchtop incubator with six modular incubation chambers each with its own temperature control capability and separate gas inlet feed. Each chamber contains heating elements in its lid and base, together with an orange light source (591 nm) and camera with integrated optics that take time-lapse images of embryos and allows operators to view embryos without removing them from the incubation chamber. Inside each chamber is a filter the gas mixture entering the incubation chamber from the gas supply. The gas is supplied from standard premixed medical gas cylinders.
The Geri Embryo Incubator includes firmware to control the incubator settings, and software to control patient information and settings.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The logo is simple and modern, with a focus on the company's name.
The embryos are maintained in Geri Dishes that are supplied separately. The Geri Dish is an optically-clear polystyrene dish designed to be compatible with the Geri Embryo Incubator. Up to 16 embryos can be stored in one Geri Dish. Geri Dishes are supplied sterile with a sterility assurance level (SAL) of 10-9, and have a shelf-life of 12 months.
The Geri Embryo Incubator and Geri Dish are intended to be used together for embryo imaging purposes. The Geri Dish is not compatible with other embryo time-lapse incubators. However, other assisted reproduction culture dishes may be used in the Geri Embryo Incubator in separate non-time-lapse positions located within each chamber.
#### VI. Indications for Use:
The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.
The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.
#### VII. Comparison of Intended Use and Technological Characteristics of the Subject and Predicate Devices
| Device/Predicate<br>Devices | Subject device - Geri Embryo<br>Incubator | Predicate device - EmbryoScope (Version D)<br>and EmbryoViewer Software (K113075) |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Geri Embryo Incubator is intended<br>to provide an environment with<br>controlled temperature and mixed gas<br>(CO2 and other gases) for the<br>development of embryos. The Geri<br>Embryo Incubator has an integrated<br>camera and optics for imaging and<br>viewing embryos during incubation, for<br>a maximum time of 120 hours. | The Embryoscope (Version D) provides an<br>environment with controlled temperature, CO2<br>(and other gases) for the development of<br>embryos. This model has an integrated inverted<br>microscope and imaging system for embryo<br>viewing. Device use is limited to five days (120<br>hr) covering the time from post-fertilization to day<br>5 of development.<br>EmbryoViewer is an optional accessory software<br>package for use in displaying, comparing, storing,<br>and transferring EmbryoScope (Version D)<br>generated images. The software includes a user<br>annotation function for capturing information on<br>embryo development parameters, treatment data,<br>and outcome data. The EmbryoViewer software<br>does not control any hardware components in the<br>EmbryoScope (Version D) device. |
| Incubation chamber<br>heating | Heated aluminium stage and lid | Heated chamber with recirculation |
| Incubation humidity | Same as predicate | No chamber humidification |
| Gas supply type | Premixed blend of 6% CO2, 5% O2,<br>89% N2 or high purity 6% CO2 | Internally regulated mixture of CO2, O2 and N2 |
| Gas supply pressure | 150 ± 15 kPa (218 ± 2.2 psi) | Unclear |
| Gas flow rate capability | 10 mL/min per chamber. Purge at 160<br>mL/min per chamber for 3 min | Unclear |
| Gas flow rate accuracy | ±15% of flow per chamber (normal<br>flow) and ±10 mL/min per chamber<br>(purge) | Unclear |
| HEPA Filter | Same as predicate | Retains 99.97% of particles >0.3 µm |
| Chamber temperature<br>capability | 35.0-40.0°C in 0.1°C increments at 20-<br>28°C. At set point of 37°C, the ambient<br>temperature range is 18-30°C | 30-45 ± 0.2°C |
| Chamber temperature<br>accuracy | ±0.2°C at calibration point | During incubation ±0.1°C |
#### Comparison of Geri Embryo Incubator with predicate device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Genea Biomedx. The logo consists of a green square with a white curve on the left side, followed by the text "Genea" in gray and "BIOMEDX" in green. The logo is simple and modern, with a focus on the company's name.
| Device/Predicate<br>Devices | Subject device - Geri Embryo<br>Incubator | Predicate device - EmbryoScope (Version D)<br>and EmbryoViewer Software (K113075) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|-----------------------------------------------------------------------------------|
| Microscope/camera | CMOS microscope camera (2560×1928<br>pixels monochrome | CCD camera (1280×1024 pixels monochrome) |
| Image resolution | 2 pixels/um | 3 pixels/um |
| Image acquisition<br>frequency | 5 min | 10 min |
| Illumination | Single orange LED (591nm) | Single red LED (635nm) |
| Number of image focal<br>planes | 11 | 7 |
| Light exposure | ~45s/day | Unclear |
| The subject and predicate devices have intended use - culturing embryos and capturing, storing and<br>viewing embryo images. They also have the same fundamental technology - incubator with temperature and gas | | |
control and integrated microscope/camera for time-lapse imaging. There are differences in design and performance specifications between the subject and predicate devices. However, these different questions of safety and effectiveness.
#### Comparison of Geri Dish with predicate device
| Device/Predicate<br>Devices | Subject device – Geri Dish | Predicate device – EmbryoSlide (K092183) |
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Indications for Use | The Geri Dish is intended to be used<br>for preparation, storage and imaging of<br>human embryos. The Geri Dish is<br>intended to be used only with the Geri<br>Embryo Incubator. | Preparing, storing, and transferring human embryos.<br>To be used only with the EmbryoScope device. |
| Fundamental design | Same as predicate | Dish with multiple microwells intended to house<br>individual embryos |
| Design feature | 16 culture wells<br>Well base diameter 0.43mm<br>Well depth 0.4mm<br>3 outer rings for wash or rinse media<br>Fill volume 40-80µl<br>Optically clear with a lid | Unclear |
| Material | Same as predicate | Polystyrene |
| The subject and predicate devices have the same intended use - preparing and storing human embryos undergoing | | |
preparing and storing human devices have the same intended use time-lapse imaging procedures. They also have the same fundamental technology – culture dishes with multiple small microwells intended to house individual embryos. The subject device is different from the predicate device regarding microwell number, dimensions, and fill volume. However, these different questions of safety and effectiveness.
#### VIII. Summary of Non-clinical Performance Testing
The following studies have been performed to support substantial equivalence to the predicate devices:
#### Testing for Geri Embryo Incubator
- Electrical safety testing per IEC 61010-1:2010 ●
- Electromagnetic Compatibility testing per IEC 61326-1:2012 .
- Software verification and validation testing that met the requirements of the FDA Guidance . Document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005
- Cleaning and disinfection validation testing per FDA Guidance Document, "Reprocessing Medical . Devices in Health Care Settings: Validation Methods and Labeling: Guidance for Industry and Food and Drug Administration" issued on March 17, 2015
- Bench performance tests that met design specifications described in Section VII, as follows:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Genea Biomedx. The logo consists of a green square with a white curved line inside, followed by the text "Genea" in gray and "BIOMEDX" in green. The logo is clean and modern, with a focus on the company's name and a simple graphic element.
- Gas maintenance testing to ensure that gas flow rates, gas recovery rates and CO2 levels in chambers meet design specifications
- * Temperature control testing to ensure that culture chamber temperature conditions were within defined ranges and were stable over time.
- * Time-lapse testing to demonstrate that all design specifications of time-lapse function are met (dish presence detection, dish position accuracy, image quality and resolution, imaging of embryos in all wells on Geri Dishes within time and focal plane settings)
- Embryo development test demonstrating that ≥70% mouse 1-cell embryos developed to blastocyst ● on day 5 following exposure to the maximum time lapse exposure conditions.
- Embryo light energy exposure testing comparing the potential for light toxicity associated with time-. lapse imaging under worst-case exposure conditions (maximum light exposure over the full duration of use [120 hours]) to that of total light exposure during in assisted reproduction procedures (IVF and ICSI) using typical laboratory microscopes. Results showed total light energy exposure following time-lapse imaging should not exceed levels associated with typical assisted reproduction imaging procedures.
#### Testing for Geri Dish
- Radiation sterilization validation testing per ISO 11137-2:2009 ●
- . Package integrity testing:
- * Dye Penetration Test per ASTM F1929-12
- * Peel strength testing ASTM F88/F88M-09
- Endotoxin testing per USP <85> and ANSI/AAMI ST72: 2011. The testing demonstrated that the ● device met the specification of <20 EU/device.
- Mouse embryo assay (MEA): .
One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed in comparison with the control group. The testing demonstrated that the device met acceptance criterion of "1-cell MEA ≥80% embryos developed to blastocyst in 96 hours."
- Shelf-life testing (accelerated) demonstrating that visual requirements, media fill volume, timelapse alignment and MEA met design specifications before and after aging
#### IX. Conclusion
The subject and predicate devices have the same intended use and fundamental technological characteristics. The differences in technological characteristics between subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.