Momcozy Wearable Breast Pump (BP311); Momcozy Wearable Breast Pump (BP223); Momcozy Wearable Breast Pump (BP324); Momcozy Wearable Breast Pump (BP334); Momcozy Wearable Breast Pump (BP430)
Momcozy Wearable Breast Pump (BP311); Momcozy Wearable Breast Pump (BP223); Momcozy Wearable Breast Pump (BP324); Momcozy Wearable Breast Pump (BP334); Momcozy Wearable Breast Pump (BP430)
Applicant
Shenzhen Root Innovation Technology Co., , Ltd.
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Dec 20, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.5160
Device Class
Class 2
Indications for Use
The Momcozy Wearable Breast Pump (Model: BP311, BP223, BP334, BP334, BP430) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.
Device Story
Momcozy Wearable Breast Pump (models BP311, BP223, BP324, BP334, BP430) is a powered breast pump designed for single-user, over-the-counter use by lactating women. Device operates by applying suction to the breast to express and collect milk. It is a wearable, portable system intended to facilitate milk expression for the user. The device is operated by the user herself. Output is collected milk. Benefits include convenience and portability for breastfeeding mothers.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Powered breast pump; wearable form factor; intended for single-user use. Class II device (21 CFR 884.5160).
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for single-user, over-the-counter use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Related Devices
K241680 — Momcozy Wearable Breast Pump (BP311) · Shenzhen Root Innovation Technology Co., Ltd. · Dec 20, 2024
K230776 — Momcozy Wearable Electric Breast Pump ( Model: S9 Pro, S12 Pro) · Shenzhen Lutejiacheng Technology Co., Ltd. · Jun 9, 2023
K252192 — Wearable Breast Pump (P9, P10, P16, P20, P21, P22, P23, P25) · Huizhou Lvb Maternal and Infant Supplies Co., Ltd. · Aug 13, 2025
K221598 — Wearable breast pump · Mayborn (Uk) Limited · Dec 9, 2022
K250368 — Wearable Breast Pump (Model S21) · Shenzhen TPH Technology Co., Ltd. · May 30, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION". The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
Shenzhen Root Innovation Technology Co., Ltd. Amy Dong Regulatory Compliance Director #2-201, Floor 2, Hasee Computer Building, No.2 Beier Rd. Bantian Street, Longgang Shenzhen, Guangdong 518129 CHINA
Re: K243629
> Trade/Device Name: Momcozy Wearable Breast Pump (BP311); Momcozy Wearable Breast Pump (BP223); Momcozy Wearable Breast Pump (BP324); Momcozy Wearable Breast Pump (BP334); Momcozy Wearable Breast Pump (BP430) Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: Class II Product Code: HGX Dated: November 24, 2024 Received: November 25, 2024
## Dear Amy Dong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
announcements concerning your device in the Federal Register.
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/uniquedevice-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
{2}------------------------------------------------
3
(https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, PhD Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243629
#### Device Name
Momcozy Wearable Breast Pump (BP311); Momcozy Wearable Breast Pump (BP223); Momcozy Wearable Breast Pump (BP324); Momcozy Wearable Breast Pump (BP334); Momcozy Wearable Breast Pump (BP430)
#### Indications for Use (Describe)
The Momcozy Wearable Breast Pump (Model: BP311, BP223, BP334, BP334, BP430) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.