Aloe Glide Lubricant

K190323 · Desert Harvest, Inc. · NUC · Sep 25, 2019 · Obstetrics/Gynecology

Device Facts

Record IDK190323
Device NameAloe Glide Lubricant
ApplicantDesert Harvest, Inc.
Product CodeNUC · Obstetrics/Gynecology
Decision DateSep 25, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

Aloe Glide Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Device Story

Aloe Glide Lubricant is a non-sterile, water-based personal lubricant for over-the-counter use. It is applied topically to the penis or vagina to provide lubrication and moisture during intimate sexual activity. The device is packaged in polyethylene terephthalate (PET) bottles. It functions by providing a physical barrier and lubrication to reduce friction. The user applies the lubricant manually as needed. The device is compatible with natural rubber latex and polyisoprene condoms, but not polyurethane condoms. It benefits the patient by supplementing natural lubrication, thereby enhancing comfort during sexual activity.

Clinical Evidence

No clinical data. Evidence consists of bench testing: biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity), condom compatibility per ASTM D7661-10, and accelerated aging stability studies to support a 1-year shelf life. All testing met predefined specifications.

Technological Characteristics

Water-based lubricant; ingredients include water, xanthan gum, licorice/marshmallow extract, aloe barbadensis leaf extract, potassium sorbate, citric acid, and Natapres. Viscosity 10,000-15,000 cps; osmolality 300-400 mOsm/kg; pH 4.5-5.8. Packaging: PET bottles. Biocompatibility per ISO 10993-5:2009 and ISO 10993-11:2017. Condom compatibility per ASTM D7661-10. Microbial testing per USP <61>, <62>, and <51>.

Indications for Use

Indicated for individuals requiring personal lubrication for penile and/or vaginal application to moisturize, lubricate, and enhance comfort during intimate sexual activity. Compatible with natural rubber latex and polyisoprene condoms; contraindicated for use with polyurethane condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. September 25, 2019 Desert Harvest, Inc % Abhishek Gurnani Partner Amin Talati Wasserman, LLP 100 S. Wacker Drive, Suite 2000 Chicago, IL 60606 Re: K190323 Trade/Device Name: Aloe Glide Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: August 23, 2019 Received: August 26, 2019 Dear Abhishek Gurnani: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Sharon M. Andrews Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190323 Device Name Aloe Glide Lubricant Indications for Use (Describe) Aloe Glide Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. Type of Use (Select one or both, as applicable) | Registration Use (Part 21 CFR 201.2 Select B) | On The Counter Use (21 CFR 201.2 Select C) | |-----------------------------------------------|--------------------------------------------| |-----------------------------------------------|--------------------------------------------| _ | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary Aloe Glide Lubricant - K190323 This summary uses the format provided in 21 CFR 807.92: | (a)(1) Submitter/Owner: | Desert Harvest, Inc.<br>P.O. Box 1412 Hillsborough,<br>NC 27278 Phone: 919-245-<br>1853 | |-------------------------|--------------------------------------------------------------------------------------------------------------| | Preparer/Contact: | Abhishek K. Gurnani<br>Amin Talati Wasserman, LLP<br>100 South Wacker Drive, Suite 2000<br>Chicago, IL 60606 | | | Phone: 312-327-3325 | | | Fax: 312-884-7352 | | | Email: Abhishek@AminTalati.com | | Summary Prepared: | September 24, 2019 | | (a)(2) Trade Name: | Aloe Glide Lubricant | | Common Name: | Personal Lubricant | | Regulation Number: | 21 CFR 884.5300 | | Regulatory Name: | Condom | | Regulatory Class: | Class II | | Product Code: | NUC (lubricant, personal) | - (a)(3) Identification of Predicate Device: Aloe Cadabra Personal Lubricant (K124044). The predicate device has not been subject to a design related recall. - (a)(4) Device Description: Aloe Glide Lubricant is a personal lubricant that is non-sterile, waterbased, and provides lubrication during intimate sexual activity. This device is compatible with natural rubber latex and polyisoprene condoms, and is not compatible with polyurethane condoms. Its formulation consists of water, xanthan gum, licorice and marshmallow extract, aloe barbadensis leaf extract, potassium sorbate, citric acid, and natapres. Aloe Glide is packaged in 2, 4, and 8 fl. oz bottles composed of polyethylene terephthalate (PET). Aloe Glide lubricant is a personal lubricant for over-the-counter (OTC) use. {4}------------------------------------------------ Device specifications are listed in Table 1 below. | Property | Specification | |---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | Appearance | Clear and transparent | | Odor | Characteristic | | Viscosity | 10,000 - 15,000 cps | | Osmolality | 300 – 400 mOsm/kg | | pH at 25 °C | 4.5 – 5.8 | | Total Aerobic Microbial Count (USP<br><61>) | <100 cfu/g | | | | | Total Yeast & Mold Count (USP <61>) | <10 cfu/g | | | <100 cfu/g | | Absence of Pathogenic Organisms (USP<br><62>) | | | | | | Pseudomonas aeruginosa | Absent | | Staphylococcus aureus | Absent | | Candida albicans | Absent | | Escherichia coli, Salmonella, Clostridium<br>Species | Absent | | Antimicrobial Effectiveness (USP<51>) | | | Escherichia coli, Pseudomonas<br>aeruginosa, Staphylococcus aureus, | NLT a 2.0 log reduction from initial<br>count at 14 days and no increase from<br>the 14 day count at 28 days | | Candida albicans, Aspergillus niger | No increase from the initial calculated<br>count at 14 and 28 days | ## Table 1: Subject Device Specifications - (a)(5) Indications for Use Statement: Aloe Glide Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. The subject and predicate device have the same intended use. {5}------------------------------------------------ ## (a)(6) Comparison of Technological Characteristics | | K190323 | K124044 | |---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Subject Device | Predicate | | Sponsor | Desert Harvest, Inc. | Seven Oaks Ranch, Inc. | | Regulation Number<br>Product Code<br>Device Class | 844.5300<br>NUC<br>II | 844.5300<br>NUC<br>II | | Indications for Use | Aloe Glide Lubricant is a<br>personal lubricant, for penile<br>and/or vaginal application,<br>intended to moisturize and<br>lubricate, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication. This<br>product is compatible with<br>natural rubber latex and<br>polyisoprene condoms. This<br>product is not compatible with<br>polyurethane condoms. | Aloe Cadabra® Lubricant and<br>Aloe Cadabra®<br>Flavored/Scented Lubricants<br>are personal<br>lubricants, for penile and/or<br>vaginal application, intended to<br>moisturize and lubricate, to<br>enhance<br>the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication.<br>This product is compatible<br>with natural rubber latex and<br>polyisoprene condoms. This<br>product is<br>not compatible with<br>polyurethane condoms. | | Physical Features | Clear liquid | Not known | | Water-based | Yes | No (Aloe-based) | | Primary<br>Ingredients | Water<br>Xanthan Gum<br>Natapres* | Aloe Extract | | Sterile | No | No | | Condom<br>Compatibility | Natural Rubber Latex<br>Polyisoprene | Natural Rubber Latex<br>Polyisoprene | | Biocompatibility<br>Tested | Yes | Yes | | Antimicrobial<br>Tested | Yes | Yes | ## Table 2. Technological Characteristics of Aloe Glide and Compared to the Predicate *Natapres contains glycerin, leuconostoc/radish, root ferment filtrate, lonicera japonica (honeysuckle) flower extract, lonicera caprifolium (honeysuckle) extract, populus tremuloides bark extract, and gluconolactone {6}------------------------------------------------ The subject and predicate devices have the same indications for use statements. As noted in the table above, the subject and predicate device have different formulations. The differences in technological characteristics between the subject and predicate device do not raise different questions of safety and effectiveness. #### (b) Summary of Performance Data Biocompatibility: Aloe Glide has undergone biocompatibility testing including cytotoxicity based on alternate testing comparable to ISO 10993-5: 2009, human repeat insult patch testing (sensitization and irritation), and acute systemic toxicity testing per ISO 10993-11:2017. The testing found that Aloe Glide is non-cytotoxic, non-sensitizing, non-irritating, and non-systemically toxic. Condom Compatibility: The compatibility of the subject devices with natural rubber latex. polyisoprene and polyurethane condoms was evaluated in accordance with ASTM D7661-10 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms" and was determined to be compatible with natural rubber latex and polyisoprene condoms. The subject device was determined not to be compatible with polyurethane condoms. Shelf Life: Aloe Glide has a 1-year shelf life in accordance with the results of an accelerated aging stability study. Results from testing demonstrated that the device can maintain its specifications (as shown in Table 1) over the duration of its shelf life. Conclusion: The results of the performance testing described above demonstrate that the Aloe Glide is as safe and effective as the predicate device and supports a determination of substantial equivalence
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...