K141581 · Biofilm, Inc. · NUC · Aug 6, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K141581
Device Name
ASTROGLIDE NATURAL
Applicant
Biofilm, Inc.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Aug 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Astroglide Natural is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is ONLY compatible with natural rubber latex and polyisoprene condoms.
Device Story
Astroglide Natural is a non-sterile, water-based personal lubricant intended for OTC use to enhance comfort during sexual activity. The device is applied topically to the penis or vagina to supplement natural lubrication. It functions as a physical lubricant; it is not a spermicide or contraceptive. The product is packaged in a PET bottle with a polypropylene flip-top cap. Users apply the lubricant manually as needed. The device benefits patients by reducing friction and increasing comfort during intimate activity. It is compatible with natural rubber latex and polyisoprene condoms, but incompatible with polyurethane condoms.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993, including cytotoxicity, guinea pig maximization, vaginal and penile irritation in rabbits, and systemic toxicity. Condom compatibility testing performed per ASTM D7661-10. Stability testing confirmed a 3-year shelf life.
Technological Characteristics
Water-based personal lubricant. Ingredients: Water, Xylitol, Hydroxyethylcellulose, Phenoxyethanol, Aloe Barbadensis Leaf Juice Extract, Chamomilla Recutita Flower Extract, Pectin, Potassium Ascorbyl Tocopheryl Phosphate, Lactic Acid. pH 4.0-7.0; Osmolarity 769 mOsm/kg; Viscosity 2200-3400 cP. Packaging: PET bottle, polypropylene cap. Compatible with natural rubber latex and polyisoprene condoms per ASTM D7661-10.
Indications for Use
Indicated for penile and/or vaginal application to moisturize and lubricate during intimate sexual activity. Compatible with natural rubber latex and polyisoprene condoms; not compatible with polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized human figure with three faces in profile, representing health and well-being. The emblem is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 6, 2014
BioFilm. Inc. Sherry Castello Regulatory Affairs and Ouality Assurance Associate 3225 Executive Ridge Vista, CA 92081
Re: K141581 Trade/Device Name: Astroglide® Natural Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: June 10, 2014 Received: June 13, 2014
Dear Sherry Castello,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K141581
Device Name Astroglide Natural
Indications for Use (Describe)
Astroglide Natural is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is ONLY compatible with natural rubber latex and polyisoprene condoms.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
# Page 2
{4}------------------------------------------------
#### 510(k) Summary Astroglide® Natural
#### General Information on Submitter
Applicant: Address:
- Telephone: Fax: Contact Person: Email:
Date Prepared: Establishment Registration:
BioFilm, Inc. 3225 Executive Ridge Vista, CA 92081 USA 760-727-9030 760-727-8080 Sherry Castello Regulatory Affairs Associate sherry@biofilm.com Aug 5, 2014 2025771
#### General Information on Device
Proprietary Name: Common Name: Classification Name: Astroqlide® Natural
Personal Lubricant Lubricant, Personal, 21 CFR 884.5300 Condom, Product Code: NUC
#### Predicate Device
| Predicate Device | 510(k) Number | |
|--------------------------------------|---------------|--|
| Glycerin and Paraben Free Astroglide | K072647 | |
| Original Applicant: BioFilm, Inc. | | |
### Description of Device
Astroglide Natural personal lubricant is non-sterile, clear and water based. Astroglide® Natural is a proprietary blend consisting mainly of water soluble ingredients similar to other lubricants currently on the market. This product is not a spermicide or contraceptive. It is compatible with natural rubber latex and polyisoprene condoms. It is not compatible with polyurethane condoms. This product primary packaging is a clear PET bottle with a screw on flip top cap comprised of polypropylene. Each bottle is packaged into a cardboard carton, which constitutes the device outer packaging.
### Indications for Use
Astroglide® Natural is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is ONLY compatible with natural rubber latex and polyisoprene condoms.
{5}------------------------------------------------
### Technological Characteristics
There are no fundamental differences in Astroglide® Natural as compared to the predicate device Glycerin and Paraben Free Astroglide. Both lubricants are clear, water based lubricants formulated without glycerin and paraben. Both formulas use Xylitol as a humectant, thickener, and preservative aid.
| Attribute | Astroglide® Natural<br>Lubricant | Predicate: Glycerine and<br>Paraben Free Astroglide |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | To Be Determined | K072647 |
| Indications for use | Astroglide® Natural is a<br>personal lubricant for penile<br>and/or vaginal application,<br>intended to moisturize and<br>lubricate, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br>This product is ONLY<br>compatible with natural rubber<br>latex and polyisoprene<br>condoms. | Glycerin and Paraben<br>Free Astroglide® is a<br>personal lubricant for<br>penile, anal, or vaginal<br>application, intended to<br>moisturize and lubricate,<br>to enhance the ease<br>and comfort of intimate<br>sexual activity and<br>supplement the body's<br>natural lubrication. This<br>product is compatible<br>with latex condoms. |
| Ingredients | Water, Xylitol ,<br>Hydroxyethylcellulose,<br>Phenoxyethanol, Aloe<br>Barbadensis Leaf Juice<br>Extract, Chamomilla Recutita<br>(Matricaria) Flower Extract,<br>Pectin, Potassium Ascorbyl<br>Tocopheryl Phosphate, Lactic<br>Acid | Water, Butylene Glycol, Xylitol,<br>Propylene Glycol,<br>Polyquaternium-15 |
| Storage<br>Instruction | Room Temperature | Room Temperature |
| pH | 4.0 - 7.0 | 3.5 - 6.0 |
| Osmolarity | 769 mOsm/kg | 727 mOsm/kg |
| Viscosity | 2200-3400 cP | 1100 - 1500 cP |
| Condom<br>compatibility | Natural rubber latex and<br>polyisoprene | Latex |
Table 5-1. Technological Characteristics Comparison
#### Biocompatibility
Biocompatibility testing was performed in accordance with ISO 10993, Biological Evaluation of Medical Devices.
{6}------------------------------------------------
Table 5-2. Biocompatibility of Astroglide® Natural
| Biocompatibility Testing | Astroglide Natural Results |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity | Passed Test - Non-cytotoxic |
| Guinea Pig Maximization, ISO 10993-10 | Passed test - The test article did not<br>elicit any reactions |
| Vaginal Irritation, ISO 10993-10 | The test article was considered non-<br>irritating to the vaginal mucosa in New<br>Zealand White Rabbits as compared to<br>the control article. |
| Penile Irritation, ISO 10993-10 | The test article was considered non-<br>irritating to the penile tissue in New<br>Zealand White Rabbits as compared to<br>the control article. |
| Systemic Toxicity, ISO 10993-11 | All extracts of the test article met the<br>requirements of ISO 10993-11. No test<br>animals exhibited any biological<br>reactivity. |
## Specifications
Astroglide Natural has the following lot release specifications: color, clarity, odor, absence of particulate matter, pH, viscosity, total yeast/mold count, total aerobic microbial count, and absence of pathogens including Staphylococcus aureus, Pseudomonal aeruginosa, and Candida albicans.
### Shelf Life
Astroglide Natural shows a shelf-life of 3 years based on a 3-year real-time stability study and a 1 year accelerated aging study. Preservative effectiveness was demonstrated at critical time-points throughout the stability testing.
# Condom Compatibility
Astroglide® Natural was tested for compatibility with natural rubber latex, polyisoprene, and polyurethane condoms using ASTM D7661-10, Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. Results show that Astroglide® Natural is compatible with latex and polyisoprene condoms but not compatible with polyurethane condoms.
### Substantial Equivalence
Astroglide® Natural personal lubricant has the same intended use and basic technological characteristics as the predicate device. Astroglide® Natural performed well in biocompatibility and performance bench testing proving it is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.