The Lansinoh® Smartpump 3.0 Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for a single user.
Device Story
Double electric breast pump; expresses/collects breastmilk for infant feeding. System consists of diaphragm-type vacuum pump driven by microprocessor-controlled DC electric motor. User interface includes front panel keypad and LCD display. Powered by rechargeable 7.4V 1,500 mAh lithium-ion battery or AC adapter. User adjusts cycle mode and vacuum level via keypad. Device features Bluetooth BLE 4.2 for optional mobile app connectivity. Used in home healthcare environment by nursing women. Closed system design with diaphragm barrier prevents milk backflow into tubing/pump. Output is expressed breastmilk collected in containers. Provides user control over suction/cycle parameters to support comfort and milk expression.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: suction pressure stability, battery life, backflow protection, electrical safety (IEC 60601-1, 60601-1-2, 60601-1-11), and battery safety (IEC 62133-2).
Technological Characteristics
Powered breast pump; microprocessor-controlled DC motor; diaphragm-type vacuum pump. Power: 7.4V 1,500 mAh Li-ion battery or AC adapter. Connectivity: Bluetooth BLE 4.2. Materials: Patient-contacting components identical to K182749. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), IEC 62133-2 (battery). Software: Moderate level of concern, IEC 62304 compliant. Reusable, non-sterile.
Indications for Use
Indicated for lactating women for the expression and collection of breastmilk for infant feeding. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Lansinoh® Smartpump™ Double Electric Breast Pump (DEBP 2.2) (K182749)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 23, 2022
Lansinoh Laboratories Lindsay Ewers Director of Quality Assurance 99 Canal Center Plaza. Suite 550 Alexandria, VA 22314
Re: K222726
> Trade/Device Name: Smartpump 3.0 Double Electric Breast Pump Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: December 2, 2022 Received: December 2, 2022
Dear Lindsay Ewers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Reginald K. Avery -S
for
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K222726
Device Name
Smartpump 3.0 Double Electric Breast Pump
Indications for Use (Describe)
The Lansinoh® Smartpump 3.0 Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for a single user.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 1. SUBMITTER
| Applicant: | Lansinoh Laboratories, Inc.<br>99 Canal Center Plaza, Suite 550<br>Alexandria, VA 22314 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Contact: | Lindsay Ewers<br>Director of Quality Assurance<br>Lansinoh Laboratories, Inc.<br>(727)-542-3743<br>lewers@lansinoh.com |
| Submission Correspondent: | Lindsay Ewers<br>Director of Quality Assurance<br>Lansinoh Laboratories, Inc.<br>-(727)-542-3743<br>lewers@lansinoh.com |
| Date Prepared: | December 22, 2022 |
#### 2. DEVICE
| Device Trade Name: | Smartpump 3.0 Double Electric Breast Pump |
|---------------------|-------------------------------------------|
| Device Common Name: | Powered breast pump |
| Regulation Name: | Powered breast pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulatory Class: | Class II |
| Product Code: | HGX |
#### 3. PREDICATE DEVICE
Predicate Device: Lansinoh® Smartpump™ Double Electric Breast Pump (DEBP 2.2) [K182749]
The predicate device has not been subject to a design-related recall.
#### DEVICE DESCRIPTION 4.
The Lansinoh Smartpump 3.0 Double Electric Breast Pump is a double electric breast pump intended to express the breast milk of a nursing woman. The pumping can be performed on one breast or on both breasts at the same time. The pumps can be powered by a rechargeable 7.4 V, 1,500 mAh lithium-ion battery or by an AC adapter that is provided with each pump. The pumping system consists of a diaphragm-type vacuum pump which is driven by a microprocessor-controlled DC electric motor. The user interface consists of a front panel keypad
{5}------------------------------------------------
and LCD display. The user is able to adjust cycle mode and vacuum level based on personal comfort and preference.
#### INTENDED USE/INDICATIONS FOR USE ર.
The Lansinoh® Smartpump 3.0 Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for a single user.
#### SUBSTANTIAL EQUIVALENCE 6.
### Technological Comparisons
The table below compares the key technological feature of the subject devices to the predicate device (Lansinoh® Smartpump™ Double Electric Breast Pump (DEBP 2.2) [K182749]).
| | Subject Device<br>Lansinoh® Smartpump 3.0 Double<br>Electric Breast Pump | Predicate Device<br>Lansinoh® Smartpump™ Double<br>Electric Breast Pump (DEBP 2.2) |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K222726 | K182749 |
| Indications For Use | The Lansinoh® Smartpump 3.0<br>Double Electric Breast Pump is<br>intended to express and collect the<br>breastmilk of a nursing woman for the<br>purpose of feeding collected breastmilk<br>to a baby. The pump is intended for a<br>single user. | The Lansinoh® Smartpump™ Double<br>Electric Breast Pump is intended to<br>express and collect the breastmilk of a<br>nursing woman for the purpose of<br>feeding collected breastmilk to a baby.<br>The pump is intended for multiple<br>users, and single users. |
| Applicant | Lansinoh Laboratories | Lansinoh Laboratories |
| Classification<br>Regulation | 884.5160 | 884.5160 |
| Product Code | HGX | HGX |
| Regulatory Class | Class II | Class II |
| Patient Population | Lactating Women | Lactating Women |
| Suction Modes | Stimulation and Expression | Stimulation and Expression |
| Suction Levels<br>(stimulation) | 59-236 mmHg | 46-140 mmHg |
| Cycles per Second<br>(stimulation) | 1.36-2.69 | 1.61-2.33 |
| Suction Levels<br>(expression) | 89 mmHg - 280 mmHg | 95 mmHg - 280 mmHg |
| Cycles per Second<br>(expression) | 0.53-1.83 | 0.58 - 1.69 |
Table 1: Technological Comparison
{6}------------------------------------------------
| | Subject Device<br>Lansinoh® Smartpump 3.0 Double<br>Electric Breast Pump | Predicate Device<br>Lansinoh® Smartpump™ Double<br>Electric Breast Pump (DEBP 2.2) |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Suction Settings | 8 | 8 |
| Expression modes | 3 | 3 |
| Power Supply | AC Adapter<br>Replaceable, rechargeable 7.4 V, 1,500<br>mAh lithium-ion polymer battery | AC Adapter<br>6 AA alkaline batteries |
| Pumping Option | Single or Double | Single or Double |
| Back Flow Protection | Yes | Yes |
| Let Down Function | Yes | Yes |
| Cycling/Suction Control<br>Mechanism | Microprocessor with internal program<br>memory | Microprocessor with internal program<br>memory |
| Program Memory | 32K flash ROM, 2K RAM memory | 32K flash ROM, 2K RAM memory |
| Communication with<br>mobile app | Bluetooth BLE 4.2 | Bluetooth version 4.1 |
| Expected Use Life | 500 hours | 500 hours |
| Accessories | Tubing with Velcro Strap; Flange Cap;<br>Diaphragm; Breast Flange Body;<br>Comfort Fit Breast Flange (Cone);<br>Cone Soft Edge; White Valve; 5 oz<br>Container; Container Ring; Sealing<br>Disk; Storage Lid | Tubing with Velcro Strap; Flange Cap;<br>Diaphragm; Breast Flange Body;<br>Comfort Fit Breast Flange (Cone);<br>Cone Soft Edge; White Valve; 5 oz<br>Container; Container Ring; Sealing<br>Disk; Storage Lid |
| Wireless Connectivity | Bluetooth BLE 4.2 with an optional<br>mobile app | Bluetooth with an optional mobile app |
The subject device has the same intended use, but different technological characteristics compared to the predicate. The subject and predicate device operate at different cycle speeds, levels, and suction strengths for stimulation and expression modes, and power supply. The predicate device is also indicated for multiple and single users while the subject device is indicated for single users only. The differences in the indications and technological characteristics do not raise different questions of safety and effectiveness.
#### 7. PERFORMANCE DATA
## Biocompatibility Testing
The patient contacting components, in their final finished form, are identical to the Smartpump 1.0 Double Electric Breast Pump cleared in K182749 in formulation, processing, sterilization, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents). Therefore, no additional biocompatibility testing was conducted.
{7}------------------------------------------------
The milk contacting components are those components that are provided in the pump kit. All materials in contact with milk have been tested to meet FDA's Food Additive criteria (21 CFR 175-179)
## Electrical safety and electromagnetic compatibility (EMC)
The Smartpump 3.0 Double Electric Breast Pump was tested in accordance with IEC 60601-1:2005 (3rd ed) + CORR. 1:2006 + CORR.2:2007+A1:2012 Medical electrical equipment: Part 1: General requirements for basic safety and essential performance including US deviations, with the exception of Clause 11.7 regarding biocompatibility. The device passed all tests.
The Smartpump 3.0 Double Electric Breast Pump was tested in accordance with the FDArecognized standard IEC 60601-1-2:2014+A1:2021, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests. The device passed all tests.
The Smartpump 3.0 Double Electric Breast Pump was tested in accordance with IEC 60601-1-11:2015, Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. The device passed all tests
The Smartpump 3.0 Double Electric Breast Pump was tested in accordance with the FDArecognized standard, IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems, and found to comply with all relevant sections.
## Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." and in accordance with IEC 62304:2016, Medical device software - Software life cycle processes. The software for this device was considered as a Moderate level of concern.
Cybersecurity documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"
## Sterilization, Cleaning, and Shelf-Life
### Sterilization and Cleaning
The device is reusable, provided non-sterile, and is not sterile when used. Cleaning instructions are provided in the labeling.
### Shelf-Life
Shelf-life is not applicable due to the low likelihood of time-dependent product degradation. However, in accordance with IEC 60601-1:2005/(R)2012 the subject devices expected use-life is 500 hours. In testing, the devices were demonstrated to operate within specifications for up to 500 hours of continuous operation at the maximum setting (Expression mode 3, Level 8).
{8}------------------------------------------------
# Bench Testing
The following bench testing was performed to demonstrate substantial equivalence:
- Battery life Testing .
Battery life was measured with the pumps at the highest level ((Expression Mode 3, Level 8). The "fail time" was defined as the number of minutes the pump could run at this level before dropping below specifications. Testing supports that the battery life of the device, as noted in the labeling, is approximately 2 hours.
- Suction Pressure Stability Testing ●
Devices were evaluated for suction performance at each of the available settings. The suction curves for each cvcle mode and suction level demonstrated that the device meets its specifications and performs within the specified working ranges of pressure and cycle speed for each mode/level.
- Backflow Testing ●
The Lansinoh pumps are designed as a closed milk collection system. The diaphragm provides a physical barrier, preventing breastmilk from flowing into the tubing or pump body. There have been no changes to the Backflow Protection mechanism materials used in the subject device since its previous clearance in K182749. Therefore, Backflow Protection data from K182749 can be leveraged to support the biocompatibility of the subject devices. The purpose of the backflow test is to demonstrate that the design prevents backflow into the tubing and pump. Devices were tested at maximum pressure/cycle settings (Expression Mode 3, Level 8) at various orientations to simulate worst-case conditions. The testing demonstrated that no milk was present in test devices' tubing during and following the test; therefore, the diaphragm was demonstrated to prevent back flow of milk into the tubing and pump.
#### CONCLUSION 8.
The results of the testing described above demonstrate that the Smartpump 3.0 Double Electric Breast Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.