ASPIRATION/COAGULATION ELECTRODE ACE-311 & ACE-511
Device Facts
| Record ID | K955093 |
|---|---|
| Device Name | ASPIRATION/COAGULATION ELECTRODE ACE-311 & ACE-511 |
| Applicant | Utah Medical Products, Inc. |
| Product Code | HGI · Obstetrics/Gynecology |
| Decision Date | Apr 29, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.4120 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Aspiration/Coag Electrode is intended for use in electrosurgical procedures to provide an effective tool for allaying excessive bleeding by fulgurating and desiccating tissue damaged in electrosurgical procedures.
Device Story
Device consists of stainless steel spherical ball electrode with attached external PEBAX aspiration tube; connects to standard electrosurgical pen, electrosurgical generator, and vacuum suction system. Used by trained clinicians in clinical settings to manage bleeding during electrosurgical procedures. Device enables simultaneous or sequential tissue coagulation and fluid aspiration. Benefits include improved hemostasis and site visualization during surgery.
Clinical Evidence
Bench testing only; no clinical data provided. Biocompatibility and sterilization data maintained by manufacturer.
Technological Characteristics
Stainless steel spherical ball electrode; PEBAX aspiration tube; compatible with standard electrosurgical pens, generators, and vacuum systems. Gamma sterilization. Materials selected for biocompatibility and medical-grade standards.
Indications for Use
Indicated for use by trained clinicians during electrosurgical procedures to control bleeding via tissue fulguration and desiccation.
Regulatory Classification
Identification
A gynecologic electrocautery is a device designed to destroy tissue with high temperatures by tissue contact with an electrically heated probe. It is used to excise cervical lesions, perform biopsies, or treat chronic cervicitis under direct visual observation. This generic type of device may include the following accessories: an electrical generator, a probe, and electrical cables.
Predicate Devices
- Utah Medical Products, Inc. DBL-311 ball electrode
- Utah Medical Products, Inc. DBL-511 ball electrode
- Valleylab E2505-10FR disposable electrode
Related Devices
- K033003 — SUCTION COAGULATOR · Modern Medical Equipment Mfg., Ltd. · May 13, 2004
- K982746 — MONOPOLAR SUCTION COAGULATOR, MODEL # SB 110-1 · New Deantronics Taiwan , Ltd. · Sep 14, 1998
- K091223 — VALLEYLAB SUCTION COAGULATOR, MODEL E3305 · Covidien LP · Sep 11, 2009
- K052104 — ULTRACLEAN HANDCONTROLLED SUCTION COAGULATOR, MODEL 130182, 130183, 130184, 130185; ULTRACLEAN FOOTCONTROLLED SUCTION · Conmed Corporation · Oct 21, 2005
- K203745 — AURORA Evacuator +Coag · Rebound Therapeutics · Dec 2, 2021