PeriCALM Patterns 3.0

K241009 · Perigen, Inc. · HGM · Jan 10, 2025 · Obstetrics/Gynecology

Device Facts

Record IDK241009
Device NamePeriCALM Patterns 3.0
ApplicantPerigen, Inc.
Product CodeHGM · Obstetrics/Gynecology
Decision DateJan 10, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.2740
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAcceptanceObservedDev DSTest DS
Fetal Heart Rate Acceleration and Deceleration Detection (Term Gestation)Non-inferiority to Clinician Readers with an inferiority margin of 15% for Sensitivity (Se), Specificity (Sp), and Positive Predictive Value (PPV).Passed all acceptance criteria.Not specified.Retrospective multi-reader/multi-case study: 40 term subjects (≥36 weeks) with one tracing per subject, compared against a panel of 3 expert 'Truthers' and Clinician Readers.
Fetal Heart Rate Acceleration and Deceleration Detection (Preterm Gestation)Non-inferiority to Clinician Readers with an inferiority margin of 15% for Sensitivity (Se), Specificity (Sp), and Positive Predictive Value (PPV).Passed all acceptance criteria.Not specified.Retrospective multi-reader/multi-case study: 30 preterm subjects (32-35 weeks) with one tracing per subject, compared against a panel of 3 expert 'Truthers' and Clinician Readers.
Fetal Heart Rate Baseline MeasurementNon-inferiority to Clinician Readers with an inferiority margin of 10% for baseline level; 95% CI of Bias ≤ 5 bpm.Passed all acceptance criteria.Not specified.Retrospective multi-reader/multi-case study: 70 subjects (30 preterm, 40 term) with one tracing per subject, compared against a panel of 3 expert 'Truthers' and Clinician Readers.

Indications for Use

PeriCALM Patterns is intended for use to provide additional secondary information as an adjunct to qualified clinical decision-making during antepartum or intrapartum obstetrical monitoring at ≥32 weeks gestation for annotation and summary of the fetal heart rate recording for baseline, accelerations and decelerations and the uterine pressure recording for contractions. WARNING: Evaluation of FHR during labor and patient management decisions should not be based solely on PeriCALM Patterns annotations or summaries and should include inspection of the fetal monitor tracing and consideration of all pertinent clinical information.

Device Story

Software device for obstetrical care; processes fetal heart rate (FHR) and uterine activity data from external electronic fetal/maternal monitors. Uses Long and Short-Term Memory (LSTM) neural networks to detect, label, and measure FHR features (accelerations, decelerations, baseline) and uterine contractions. Operates in networked or standalone workstation environments. Provides clinicians with summarized measurements in 15 or 30-minute intervals; displays long-term compressed tracing views. Clinicians use output as secondary information to support decision-making; can modify or delete annotations. Benefits include automated pattern recognition and summary to assist in monitoring fetal status during labor or antepartum periods.

Clinical Evidence

Retrospective multi-reader/multi-case study; 70 subjects (30 preterm, 40 term). Compared PeriCALM Patterns 3.0 performance against a panel of 3 expert 'Truthers' and clinician readers. Endpoints included sensitivity, specificity, and PPV for accelerations/decelerations, and Bland-Altman analysis for baseline measurements (bias, LoA). Non-inferiority margins set at 15% for detection and 10% for baseline. Device met all 16 co-primary endpoints, demonstrating non-inferiority to clinician readers across both term and preterm subgroups.

Technological Characteristics

Software-based pattern recognition system. Utilizes signal processing and LSTM neural networks. Inputs: FHR (min 4 samples/sec) and uterine activity (min 1 sample/sec) from external monitors. Connectivity: Networked or standalone workstation. Software evaluated per 2023 FDA guidance for software functions and cybersecurity.

Indications for Use

Indicated for use in pregnant patients at ≥32 weeks gestation during antepartum or intrapartum obstetrical monitoring. Used as an adjunct to clinical decision-making for annotation and summary of fetal heart rate and uterine pressure recordings.

Regulatory Classification

Identification

A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 10, 2025 PeriGen, Inc. Candace Alva Director of Regulatory Affairs and Quality Assurance Sderot Nim 2, P.O. Box 110 Rishon LeTziyon, 7510002 Israel Re: K241009 > Trade/Device Name: PeriCALM Patterns 3.0 Regulation Number: 21 CFR§ 884.2740 Regulation Name: Perinatal Monitoring System and Accessories Regulatory Class: II Product Code: HGM Dated: December 11, 2024 Received: December 11, 2024 Dear Candace Alva: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. {2}------------------------------------------------ 3 For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Monica D. Garcia -S Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K241009 Device Name PeriCALM Patterns 3.0 #### Indications for Use (Describe) PeriCALM Patterns is intended for use to provide additional secondary information as an adjunct to qualified clinical decision-making during antepartum or intrapartum obstetrical monitoring at _32 weeks gestation for annotation and summary of the fetal heart rate recording for baseline, accelerations and the uterine pressure recording for contractions. WARNING: Evaluation of FHR during labor and patient management decisions should not be based solely on PeriCALM Patterns annotations or summaries and should include inspection of the fetal monitor tracing and consideration of all pertinent clinical information. | | Type of Use (Select one or both, as applicable) | | |--|-------------------------------------------------|--| | | | | X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for PeriGen, a company specializing in advanced perinatal systems. The logo features the company name in a teal sans-serif font, with the "P" extending into a partial red circle on the left side. Below the name, in a smaller font, are the words "Advanced Perinatal Systems." ## 510(k) Summary PeriCALM Patterns 3.0 - K241009 | Date Summary Prepared: | January 10, 2025 | |------------------------------|--------------------------------------------------------------------------------------------| | Submitter Information | | | Submitted by: | PeriGen Solutions, Ltd<br>Sderot Nim 2, PO Box 110,<br>Rishon LeTziyon,<br>7510002, Israel | | Telephone: | +972-3-905-9900 | | Contact Person: | Natan Blanks<br>GM, Israel Operations | | Subject Device Information | | | Device Trade Name: | PeriCALM Patterns 3.0 | | Common Name: | Computer-based information management system for<br>obstetrical care | | Regulation Name: | Perinatal monitoring system and accessories | | Regulation Number | 21 CFR 884.2740 | | Device Panel: | Obstetrics/Gynecology | | Product code: | HGM (system, monitoring, perinatal) | | Regulatory Class: | II | | Predicate Device Information | | | Predicate Device: | CALM Patterns PeriGen Solutions, Ltd. (Also listed as<br>PeriCALM Patterns) | | | The predicate device has not been subject to a design-<br>related recall | | 510(k) Number | K040788 | | Device Description | | PeriCALM Patterns 3.0 is a software device to be used with fetal/maternal monitoring systems. The subject device is a software algorithm to detect, label and measure features (accelerations, decelerations, baseline, and contractions) in electronic fetal monitoring (EFM) records. PeriCALM Patterns 3.0 uses fetal monitor data imported through an interface with an external source or with a third-party clinical information system. PeriCALM Patterns can function in a networked environment or as a standalone workstation. The subject device includes present-day Long and Short-Term Memory (LSTM) neural networks to identify segments of a fetal heart rate tracing corresponding to accelerations, decelerations, baseline as well as uninterpretable segments where is missing tracing. Contraction detection is achieved using the same processes as the predicate device. # PeriGen Solutions Ltd. Azrieli Rishonim | 2 Nim Blvd. POB 110 Rishon LeZion 7510002 Israel {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for PeriGen, a company that specializes in advanced perinatal systems. The word "PeriGen" is written in a combination of blue and red colors, with the "i" in "PeriGen" being replaced by a red arc. Below the company name, the words "Advanced Perinatal Systems" are written in a smaller, gray font. #### Indications for Use PeriCALM Patterns is intended for use to provide additional secondary information as an adjunct to qualified clinical decision-making during antepartum or intrapartum obstetrical monitoring at ≥32 weeks gestation for annotation and summary of the fetal heart rate recording for baseline, accelerations and decelerations and the uterine pressure recording for contractions. WARNING: Evaluation of FHR during labor and patient management decisions should not be based solely on PeriCALM Patterns annotations or summaries and should include inspection of the fetal monitor tracing and consideration of all pertinent clinical information. | Comparison of Intended Use and Technological Characteristics of the Subject and Predicate Device | Subject Device<br>PeriGen – PeriCALM Patterns<br>3.0<br>K241009 | Predicate Device<br>PeriGen - PeriCALM<br>Patterns<br>K040788 | |--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | PeriCALM Patterns is intended for<br>use to provide additional<br>secondary information as an<br>adjunct to qualified clinical<br>decision-making during<br>antepartum or intrapartum<br>obstetrical monitoring at ≥32<br>weeks gestation for analysis of the<br>fetal heart rate recording for<br>baseline, accelerations and<br>decelerations and the uterine<br>pressure recording for<br>contractions.<br><br>WARNING: Evaluation of FHR<br>during labor and patient<br>management decisions should not<br>be based solely on PeriCALM<br>Patterns annotations or summaries<br>and should include inspection of<br>the fetal monitor tracing and<br>consideration of all pertinent<br>clinical information. | CALM Patterns is intended for<br>use as an adjunct to qualified<br>clinical decision-making during<br>antepartum or intrapartum<br>obstetrical monitoring at ≥36<br>weeks gestation to obtain<br>annotation of the FHR for<br>baseline, accelerations, and<br>decelerations.<br><br>WARNING: Evaluation of FHR<br>during labor and patient<br>management decisions should not<br>be based solely on CALM<br>Patterns annotations. | | Applicable gestational age | ≥32 weeks | ≥36 weeks | | Data Collection | Collects FHR and uterine activity<br>data from electronic<br>maternal/fetal monitors. | Collects FHR and uterine<br>activity data from electronic<br>maternal/fetal monitors. | | Sample Rate | Samples FHR at a minimum of<br>four samples per second. Samples<br>uterine activity at a minimum or 1<br>sample per second | Samples FHR at a minimum of<br>four samples per second. Samples<br>uterine activity at a minimum or 1<br>sample per second | ## PeriGen Solutions Ltd. Azrieli Rishonim | 2 Nim Blvd. POB 110 Rishon LeZion 7510002 Israel {6}------------------------------------------------ | Pattern recognition module core<br>technology | Signal processing and neural<br>networks | Signal processing and neural<br>networks | |-----------------------------------------------------------------------------------------------------|------------------------------------------|---------------------------------------------------------| | Neural network interference:<br>combined accel-decel-baseline<br>+ non-interpretable detection | Yes | No | | Measurement and labeling<br>accelerations on the tracing | Yes | Yes | | Baseline measurements | Yes | Yes | | Contraction measurements | Yes | Yes | | Analysis provided and<br>summarized in time intervals | 15 or 30 minutes | 20 minutes | | Provides users with the option to<br>select, modify and export a set of<br>summarized measurements. | Yes | Yes | | Displays a long-term<br>compressed view of tracing | Yes | Yes | | Summarizes measurements with<br>the long-term view. | Yes | Yes | | Handling of disagreements<br>between algorithm detected<br>patterns and clinician opinion | User can delete a feature<br>annotation | User can add, modify or delete a<br>feature annotation. | | Non-interpretable segments<br>where there is missing tracing | Yes | No | The indications for use of PeriCALM Patterns 3.0 are similar to the predicate device and the subject and predicate device have the same intended use - as an adjunct to clinical decision making during antepartum, or intrapartum monitoring. The subject device has an extended gestational age from > 32 weeks, compared to the predicate device which has a gestational age from ≥ 36 weeks. This difference in gestational age does not raise different questions of safety and effectiveness. The technological differences between the subject and predicate device include the addition of neural network interference (combined accel-decel-baseline + non-interpretable detection), different time intervals of analysis, and differences in the managements of non-interpretable segments for a missing trace. These differences do not raise different questions of safety and effectiveness as compared to the predicate device. ## Summary of Non-clinical Performance Testing #### Software and Cybersecurity Perinatal Systems Software was evaluated as recommended in the 2023 FDA guidance document "Content of Premarket Submissions for Device Software Functions." The following activities were successfully completed: - Risk Analysis / Management - Requirements Reviews ● - Design Reviews ● - Software Verification and Validation ## PeriGen Solutions Ltd. Azrieli Rishonim | 2 Nim Blvd. POB 110 Rishon LeZion 7510002 Israel {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for PeriGen. The logo consists of the word "PeriGen" in blue text, with the "P" partially enclosed by a red semi-circle. Below the word "PeriGen" are the words "Advanced Perinatal Systems" in a smaller font size. The successful V&V tests demonstrate that the device performs in accordance with specifications. Cybersecurity was evaluated as recommended in the 2023 FDA guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions." # Summary of Clinical Performance Testing A retrospective study with multi-reader/multi-case technique was used to determine if the performance of PeriCALM Patterns 3.0 is not inferior to that of a set of qualified and experienced Obstetrician Gynecologists called the Clinician Readers. The predicate device was cleared for use in gestational ages of ≥36 weeks comparing to Ground Truth and Clinician Readers. The same approach was taken for this study, comparing PeriCALM Patterns 3.0 to Ground Truth and Clinician Readers for two gestational age subgroups: ≥36 weeks (term) and 32-35 weeks (pre-term). The ground truth for this study is established by a panel of 3 experts, called the "Truthers". To resolve the inter-observer variation, a majority opinion approach was used. The results of the comparison of the subject device to ground truth is to be non-inferior to the results of the comparison of Readers to ground truth. A total of 70 subjects (30 preterm and 40 term), with one tracing per subject, were used in the study. Tracings were obtained from hospitals using a variety of fetal monitor models and manufacturers including Corometrics 170, 250 series by GE HealthCare, Avalon FM50 and FM40 by Philips Medical, and S1 from Neoventa and included both antepartum and intrapartum recordings. The tracings were selected from pregnancies with gestational ages from 32 weeks to term, covering a wide range of operating and clinical conditions with respect to maternal BMI, delivery modes, fetal birth weights, hospital types, Apgar scores and antepartum or intrapartum usage. The study included sixteen (16) co-primary endpoints with different acceptance criteria for noninferiority. Twelve (12) co-primary endpoints pertained to accelerations and decelerations for both gestational groups (i.e., Term and Preterm) and included performance metrics (Sensitivity (Se), Specificity (Sp), Positive Predictive Value (PPV)) of the subject device compared to those of the Readers. Remaining four co-primary endpoints focused on the FHR baseline measurements and were based on Bland-Altman analyses to determine bias, the mean difference between two methods and the Limits of Agreement (LoA) which include 95% of the differences between the two measurement methods. Inferiority margin of 15% for acceleration and deceleration detection and 10% for baseline level was used. The acceptance criteria for 95% CI of Bias for the subject device was set to threshold of ≤ 5 bpm. PeriCALM Patterns 3.0 passed all acceptance criteria on performance testing regarding pattern detection. # Conclusion The performance data described above demonstrate that the PeriCALM Patterns 3.0 is as safe and effective as the PeriCALM Patterns (K040788) and supports a determination of substantial equivalence.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...