K032610 · Life Measurement, Inc. · OAC · Mar 1, 2004 · Cardiovascular
Device Facts
Record ID
K032610
Device Name
MODIFIED SONAMET BODY COMPOSITION ANALYZER
Applicant
Life Measurement, Inc.
Product Code
OAC · Cardiovascular
Decision Date
Mar 1, 2004
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The modified Sonamet Body Composition Analyzer, also known as the PEA POD® Infant Body Composition System (the PEA POD), is indicated for measuring body mass and estimating the body composition (i.e., the body fat and lean body mass) of infants between 1 and 8 kilograms. It is not intended for use with infants requiring life support.
Device Story
PEA POD Infant Body Composition System measures infant body mass and composition; utilizes densiometric approach. Inputs: infant body mass via integrated weighing apparatus; body volume via air displacement plethysmography. Device transforms inputs into density calculations; applies scientific research-derived algorithms to estimate body fat and lean body mass. Housed in movable cart containing reference chamber, calibration volume, air circulation, temperature control, CPU, and printer. Used in clinical settings; operated by healthcare professionals. Output provided via printer/display; assists clinicians in assessing infant nutritional status and body composition. Benefits: non-invasive, accurate body composition assessment for infants.
Clinical Evidence
Bench testing only. Verification testing demonstrated that the device satisfies functional performance requirements safely and accurately.
Technological Characteristics
Densiometric body composition analyzer; utilizes air displacement plethysmography and integrated weighing scale. Components: reference chamber, calibration volume, air circulation system, Air Temperature Control System, CPU, printer. Form factor: movable cart. Software-based algorithm for density and body composition calculation.
Indications for Use
Indicated for measuring body mass and estimating body fat and lean body mass in infants weighing 1 to 8 kg. Contraindicated for infants requiring life support.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
EM-SCAN HP-2 Pediatric Body Composition Analyzer (K902042)
Related Devices
K060848 — BOD POD · Life Measurement, Inc. · Jun 27, 2006
K040978 — TANITA SCALE PLUS BODY FAT MONITOR WITH BODY WATER PERCENTAGE · Tanita Corp. of America · Jul 22, 2004
K994242 — BODYSTAT 1500MDD BODY COMPOSITION MONITORING UNIT · Bodystat , Ltd. · Mar 15, 2000
K040778 — TANITA INNERSCAN BODY COMPOSITION MONITOR, MODEL BC-53X (WHERE X=0 TO 9) · Tanita Corp. of America · Dec 23, 2004
K030349 — HEALTHOMETER PROFESSIONAL BODY FAT MONITOR AND SCALE, MODELS BFM 940, BFM 945 & BFM 960 · Bonso Electronics International, Inc. · May 2, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
MAR - 1. 2004
K032610
page 1 of 2
# Section 6 – 510(k) Summary
# SUBMITTER INFORMATION
Company Name:
Establishment Registration Number:
Company Address:
Company Phone:
Company Facsimile:
Contact Person:
Life Measurement, Inc.
3003873943
1850 Bates Avenue Concord, CA 94520
(925) 676-6002
(925) 676-6005
Michael Sullivan Vice President of Operations
### DEVICE IDENTIFICATION
| Device Trade Name: | - Modified Sonamet Body Composition Analyzer<br>Note : The trade name for the original Sonamet Body<br>Composition analyzer is currently the “BOD POD”. The<br>modified version will be referred to as the “PEA POD -<br>Infant Body Composition System” when it is introduced into<br>the market. |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Generic Name: | Body Composition Analyzer |
| Device Classification: | Classification code: 21 CFR 870.2770<br>Code MNW |
| Classification Panel | Cardiovascular |
{1}------------------------------------------------
K032610
page 2 of 2
#### IDENTIFICATION OF PREDICATE DEVICES
The modified Sonamet Body Composition System is substantially equivalent to the following devices, which have received FDA clearance:
| Device Name | Manufacturer | K Number |
|--------------------------------------------------------|---------------------------|----------|
| Sonamet Body<br>Composition Analyzer | Life Measurement,<br>Inc. | K924972 |
| EM-SCAN HP-2<br>Pediatric Body<br>Composition Analyzer | EM-SCAN, Inc. | K902042 |
The intended use and technological characteristics of the modified Sonamet Body Composition System are the same as the above predicate devices. Modifications for the pediatric use indication do not raise additional questions regarding safety and efficacy,
#### DEVICE DESCRIPTION
The modified Sonamet Body Composition System ("The PEA POD") is designed to measure the mass and estimate the body composition of infants with body weights ranging between 1 and 8 kilograms, who do not require life support. The PEA POD estimates body composition using a densiometric approach (i.e. by determining the density of the entire body). A weighing apparatus is used to-measure the subject's mass. Air displacement plethysmography is used to measure the subject's volume. Using this data, the subject's density is calculated. The - subject's body composition is then estimated using several algorithms derived from scientific research. The device components are housed in a movable cart, which contains the reference chamber, calibration volume, air circulation system, Air Temperature Control System, electronic components, printer and CPU.
#### INTENDED USE
The modified Sonamet Body Composition Analyzer, also known as the PEA POD® Infant Body Composition System (the PEA POD), is indicated for measuring body mass and estimating the body composition (i.e., the body fat and lean body mass) of infants between 1 and 8 kilograms. It is not intended for use with infants requiring life support.
#### CONCLUSIONS DRAWN FROM STUDIES
The results of verification testing demonstrate that the PEA POD Body Composition Analyzer is substantially equivalent to the predicate devices. Test results indicate that the device satisfies functional performance requirements safely and accurately when used as indicated.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle.
MAR - 1. 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Life Measurement, Inc. c/o Dr. Sheila Pickering 2081 Longden Circle LOS ALTOS CA 94024
Re: K032610
Trade/Device Name: Modified Sonamet Body Composition Analyzer (PEA POD) Regulation Number: 21 CFR \$870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: 78 MNW Dated: January 5, 2004 Received: January 9, 2004
Dear Dr. Pickering:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I ins leter will anow you to begin maining of substantial equivalence of your device to a legally premarket notheadon: "The PPP intenty since of the more and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice to: your of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 10. questions of (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general by releveloc to promance nonities under the Act may be obtained from the Division of Strail mornation on your responsibility on the Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdoninal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
## ATTACHMENT 14
### FDA Submission Cover Sheet
510(k) Number: K032610
Device Name: Modified Sonamet Body Composition Analyzer
Indication For Use:
The modified Sonamet Body Composition Analyzer, also known as the PEA POD® Infant Body Composition System (the PEA POD), is indicated for measuring body mass and estimating the body composition (i.e., the body fat and lean body mass) of infants between 1 and 8 kilograms. It is not intended for use with infants requiring life support.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence Of CDRH, Office Of Device Evaluation (ODE)
Prescription Use
OR Over-The-Counter Use
(Per 21CFR
801)
Nancy C. Broadon
(Division Sign-Off Division of Reproductive. Abde and Radiological Devices 510(k) Number.
143
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.