K990569 · Huntleigh Healthcare, Inc. · HGM · Jan 24, 2000 · Obstetrics/Gynecology
Device Facts
Record ID
K990569
Device Name
BABY DOPPLEX 4000, MODEL BD 4000
Applicant
Huntleigh Healthcare, Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Jan 24, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
The Baby Dopplex® 4000 is a fetal monitor, which produces CTG's and gives indication of uterine activity by processing received ultrasound/electrical signals. In basic format, the device performs antepartum fetal monitoring and uterine activity functions. Connecting separate modules to the Baby Dopplex® 4000 main unit allows it to perform Intrapartum monitoring operations. In instances of twin presentations, two basic Baby Dopplex® 4000 units can be linked together so that both fetuses can be monitored simultaneously.
Device Story
BD4000 is a fetal monitor processing ultrasound and electrical signals to produce cardiotocographs (CTG) and uterine activity data. Basic unit performs antepartum monitoring; optional modules enable intrapartum monitoring. Two units can be linked for simultaneous twin monitoring. Used in clinical settings by healthcare professionals. Output displayed for clinician review to support fetal status assessment and clinical decision-making during pregnancy and labor.
Clinical Evidence
Bench testing only. Compliance with IEC601-1, BS EN60601-1, UL2601-1, BS EN 60601-1-2, BS EN55011, and DIN VDE 0750-1. Biocompatibility assessed per BS EN30993-1 and ISO10993-1.
Technological Characteristics
Fetal monitor utilizing ultrasound and electrical signal processing. Modular design for intrapartum and twin monitoring. Compliant with IEC601-1, BS EN60601-1, UL2601-1, BS EN 60601-1-2, BS EN55011, DIN VDE 0750-1. Biocompatibility per BS EN30993-1/ISO10993-1.
Indications for Use
Indicated for antepartum and intrapartum fetal monitoring and uterine activity assessment in pregnant patients, including twin pregnancies.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
K052190 — FETAL MONITOR, MODELS BT-300 AND BT-200 · Bistos Co., Ltd. · Oct 4, 2005
K043598 — FC 1400 FETAL MONITOR · Bionet Co., Ltd. · Dec 16, 2005
K082602 — FETAL & MATERNAL MONITOR, MODEL F6, F9 · Edan Instruments, Inc. · Nov 26, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
JAN 2 4 2000
# 510(k) Summary
1 of 2
# Baby Dopplex® 4000
| Submitter's Name: | Ms Audrey A. Witco,<br>Director of Corporate Affairs,<br>Huntleigh Healthcare,<br>227 Route 33 East,<br>Manalapan, NJ 07726<br>USA |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Telephone Nº: | (800) 223 1218 ext 127 |
Fax Nº: (732) 446 1938
Name of Device: Baby Dopplex® 4000 (BD4000)
Huntleigh Diagnostics Ltd Manufactured by: 35, Portmanmoor Road, Cardiff South Glamorgan CF24 5HN Wales, U.K.
#### Contact Person at Manufacturing Facility:
B.J.Colleypriest +44 (0) 1222 485885 Telephone Nº: +44 (0) 1222 490170 Fax Nº: bryn.colleypriest@huntleigh-diagnostics.co.uk e-mail:
Date summary prepared: 01 December 1999
## Classification Name
Fetal Ultrasonic Monitor and Accessories (21 CFR § 884.2660)
#### Predicate Devices
Baby Dopplex 3000 (ABD3000) K963711. Hewlett Packard GmbH's Series 50 range of Fetal Monitors Model Nºs M1351A & M1353A (K921957 & K921956 respectively).
{1}------------------------------------------------
## Device Description
The BD4000 is a fetal monitor, which produces cardiotocographs (CTG) and gives indication of uterine activity by processing received ultrasound/electrical signals. In basic format, the device performs antepartum fetal monitoring and uterine activity functions. Connecting separate modules to the BD4000 main unit allows it to perform Intrapartum monitoring operations.
In instances of twin presentations, two Baby Dopplex® 4000 units can be linked together so that both fetuses can be monitored simultaneously.
The complete BD4000 system has been tested to ensure compliance with internationally recognised medical electrical equipment standards. Tests have been performed in accordance with the requirements of IEC601-1. BS EN60601-1. UL2601-1. BS EN 60601-1-2. BS EN55011 & DIN VDE 0750-1. A checklist of Essential Requirements tests has been compiled in accordance with the direction of the Medical Devices Directive 93/42/EEC. Biocompatability has been assessed in line with the principles contained within BS EN30993-1 & ISO10993-1.
### Intended use
The BD4000 device performs the same functions as the Hewlett Packard predicate devices. The BD4000 can also monitor twins by linking two unmodified devices together.
The BD4000 is also an upgrade of the predicate ABD3000 (K963711), and includes intrapartum and twins monitoring functions that were not available on the previous model.
The fetal parameters are displayed in the same way as those of the predicate devices.
The following table briefly compares the monitoring characteristics of the applicant device to its predicate devices.
| Monitoring Mode | Baby Dopplex®<br>4000 | Baby Dopplex®<br>3000<br>(K963711) | Hewlett Packard<br>M1351A<br>(K921957) | Hewlett Packard<br>M1353A<br>(K921956) |
|---------------------------------------------------------|-----------------------|------------------------------------|----------------------------------------|----------------------------------------|
| Monitoring fetal<br>heart rate using<br>ultrasound | Yes | Yes | Yes | Yes |
| Monitoring twin<br>fetal heart rate<br>using ultrasound | Yes(1) | No | Yes(1) | No |
| Monitor fetal heart<br>rate using FECG | Yes(1) | No | Yes | Yes |
NB: (1) optional feature.
{2}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 4 2000
Ms. Audrey A. Witco Director of Corporate Affairs Huntleigh Healthcare 227 Route 33 East Manalapan, NJ 07726
Re: K990569 Baby Dopplex® 4000 Dated: December 1. 1999 Received: December 9, 1999 Requiatory Class: II 21 CFR §884.2740/Procode: 85 HGM
Dear Ms. Witco:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Kaaooeea 510(k) Number (if known):
Device Name:
Baby Dopplex® 4000
# Indications for Use
The Baby Dopplex® 4000 is a fetal monitor, which produces CTG's and gives indication of uterine activity by processing received ultrasound/electrical signals.
In basic format, the device performs antepartum fetal monitoring and uterine activity functions.
Connecting separate modules to the Baby Dopplex® 4000 main unit allows it to perform Intrapartum monitoring operations.
In instances of twin presentations, two basic Baby Dopplex® 4000 units can be linked together so that both fetuses can be monitored simultaneously.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sig
Division of Abdominal, ENT,
and Radiolo
510(k) Number K9905695002
Prescription Use
(Per 21 CFR 801.109)
OR Over the counter use
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches