K143695 · Edan Instruments, Inc. · HGM · Mar 30, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K143695
Device Name
CENTRAL MONITORING SYSTEM
Applicant
Edan Instruments, Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Mar 30, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Maternal Fetal Monitoring – Central Nurse System (hereinafter called "MFM-CNS") is a clinical data managing software application and is indicated for antepartum and intrapartum monitoring of pregnant women in a healthcare setting. The MFM-CNS is intended to manage perinatal monitoring data acquired from bedside monitors or manual input for viewing at the central nursing station. The system also produces an electronic medical record. The MFM-CNS has display fields for the following obstetric data: - patient demographics - provider notes - fetal heart rate (FHR) - uterine activity (via tocodynamometry or IUP) - fetal movement - maternal heart rate - SpO2 - non-invasive blood pressure (NIBP) - respiratory rate - temperature - pulse
Device Story
MFM-CNS is a clinical data management software application for perinatal monitoring. It receives fetal heart rate, uterine activity, maternal vital signs (SpO2, NIBP, respiratory rate, temperature, pulse), and fetal movement data from bedside monitors or manual input via Ethernet. The system processes this data to display real-time obstetric information, generate electronic medical records, and provide visual alarms for out-of-limit heart rates or poor signal quality. Used in hospital clinical areas (e.g., delivery rooms) by healthcare professionals. The system provides a central nursing station interface for viewing multiple patients simultaneously. It does not replace bedside monitoring; users must verify results at the bedside. The system aids clinical decision-making by centralizing patient data and trends, potentially improving monitoring efficiency and record-keeping.
Clinical Evidence
No clinical data. Bench testing only. Software validation performed per FDA guidance (Jan 11, 2002), risk analysis per ISO 14971:2007, usability per IEC 62366:2007, and software life cycle per IEC 62304:2006. All testing passed.
Technological Characteristics
Software-only clinical data management application. Operates on off-the-shelf PC hardware. Connectivity via Ethernet to bedside monitors. Complies with IEC 62304 (software life cycle) and IEC 62366 (usability).
Indications for Use
Indicated for antepartum and intrapartum monitoring of pregnant women in a healthcare setting.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Philips OB TraceVue Obstetrical Information Management System (K081203)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 30, 2015
Edan Instruments Inc. Doug Worth Sr. Director US RA/OA 1200 Crossman Ave, Suite 200 Sunnyvale, California 94089
> K143695 Trade/Device Name: Central Monitoring System Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal Monitoring System and Accessories Regulatory Class: Class II Product Code: HGM Dated: December 19, 2014 Received: January 5, 2015
Dear Doug Worth,
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Joyce M. Whang -S
for
Benjamin Fisher Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143695
Device Name Central Monitoring System
The Maternal Fetal Monitoring – Central Nurse System (hereinafter called "MFM-CNS") is a clinical data managing software application and is indicated for antepartum and intrapartum monitoring of pregnant women in a healthcare setting.
The MFM-CNS is intended to manage perinatal monitoring data acquired from bedside monitors or manual inout for viewing at the central nursing station. The system also produces an electronic medical record.
The MFM-CNS has display fields for the following obstetric data:
- patient demographics
- provider notes
- fetal heart rate (FHR)
- uterine activity (via tocodynamometry or IUP)
- fetal movement
- maternal heart rate
- SpO2
- non-invasive blood pressure (NIBP)
- respiratory rate
- temperature
- pulse
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (K) Summary
# Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc.<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>Shekou, Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: +86(0755) 26858736<br>Fax: +1 (408) 418-4059 |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Queena Chen |
| Preparing date: | March 24, 2015 |
| 2. Device name and<br>classification: | Device Name: Central Monitoring System<br>Model: MFM-CNS<br>Classification Name:<br>21 CFR 884.2740 Perinatal monitoring system and<br>accessories<br>Product code: HGM<br>Regulatory Class: Class II<br>Review Panel: Obstetrics/Gynecology |
| 3.Premarket<br>Notification Class III<br>Certification and<br>Summary | Not applicable, the subject device is Class II. |
| 4. Predicate Device(s): | CIV-ob Obstetrical Monitoring Software Application<br>/K103172/ CIVNET Communication Ltd.<br>Philips OB TraceVue Obstetrical Information Management<br>System/K081203/ Philips Medizin System |
| 5. Device Description: | The Maternal Fetal Monitoring – Central Nurse System<br>(hereinafter called "MFM-CNS") is a clinical data managing<br>software application. Its function is to manage clinical data<br>of fetal heart and maternal vital signs (CTG -<br>Cardiotocography), which is automatically acquired from<br>bedside monitors, for the purpose of collecting, processing<br>and saving the patient and/or clinical data that is normally<br>provided on record papers and/or separate bedside monitors.<br>It provides electronic medical records and operates with |
{4}------------------------------------------------
off-the-shelf software and hardware.
The MFM-CNS is intended to be used in hospital clinical areas such as monitor units, delivery room, etc. It is intended to be operated by or under guidance of qualified healthcare professionals, not intended for home healthcare environment. During monitoring, the user should check the results on the bedside monitor in person, even though they could observe the results on the MFM-CNS system interface. The user cannot only depend on the MFM-CNS system to obtain monitoring data, because whether the data provided by the system is accurate depends on the stability of the operating system, the performance of PC station and the network. Although the software has its independent alarm system, the alarm information provided by the system is just for reference.
### 6. Indications for Use: The Maternal Fetal Monitoring – Central Nurse System (hereinafter called "MFM-CNS") is a clinical data managing software application and is indicated for antepartum and intrapartum monitoring of pregnant women in a healthcare setting.
The MFM-CNS is intended to manage perinatal monitoring data acquired from bedside monitors or manual input for viewing at the central nursing station. The system also produces an electronic medical record.
The MFM-CNS has display fields for the following obstetric data:
- ●patient demographics
- ●provider notes
- ●fetal heart rate (FHR)
- ●uterine activity (via tocodynamometry or IUP)
- ●fetal movement
- ●maternal heart rate
- ●SpO2
- ●non-invasive blood pressure (NIBP)
- ●respiratory rate
- ●temperature
- ●pulse
## 7. Predicate Device Comparison
{5}------------------------------------------------
| Item | CIV-obTM (plus) | MFM-CNS 3.82 | SE |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Manufacturer/K# | CIVNET Communication<br>Ltd./ K103172 | EDAN Instruments/N/A | |
| Classification | | | |
| Classified as per<br>FDA regulation | Class II | Class II | Same |
| Network and Hardware | | | |
| Hardware | Off-the-shelf computers<br>and accessories | Off-the-shelf computers<br>and accessories | Same |
| Network<br>connecting to<br>bedside monitor | Ethernet | Ethernet | Same |
| Software | | | |
| Display | Fetal heart rate, TOCO,<br>maternal vital signs,<br>patient demography data,<br>and notes.<br>Providing the means to<br>display multiple beds<br>simultaneously. | Fetal heart rate, TOCO,<br>maternal vital signs,<br>patient demography data,<br>and notes.<br>Providing the means to<br>display multiple beds<br>simultaneously. | Same |
| Print | Print (locally or<br>remotely) CTG, patient<br>records, and CIV-obTM<br>(plus) data base definition<br>(e.g. item names). | Print (locally or<br>remotely) CTG and<br>patient records. | Different |
| Archive | CTG and maternal vital<br>signs.<br>Providing the ability to<br>archive files to a<br>secondary or tertiary<br>storage medium (i.e.<br>optical disk).<br>Providing automatic<br>archiving of the data. | CTG and maternal vital<br>signs.<br>Providing the ability to<br>archive files to a<br>secondary or tertiary<br>storage medium (i.e.<br>optical disk).<br>Saving data<br>automatically. | Same |
| Alarm | Visual alerts of<br>fetal/maternal monitor<br>such as out-of-limit heart<br>rate or poor signal<br>quality. | Visual alerts of<br>fetal/maternal monitor<br>such as out-of-limit heart<br>rate or poor signal<br>quality. | Same |
| Electronic<br>patient record. | Easy interfacing with any<br>IT patient record system<br>for data acquisition,<br>viewing and storage of | Easy interfacing with<br>any IT patient record<br>system for data<br>acquisition, viewing and | Same |
{6}------------------------------------------------
| | electronic patient record. | storage of electronic patient record. | |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Notes | Providing the user the ability to enter comments and specific data. | Providing the user the ability to enter comments and specific data. | Same |
| | | | |
| Remote Access | Review fetal/maternal monitor data remotely over the TCP/IP. | Review fetal/maternal monitor data remotely over the TCP/IP. | Same |
| | | | |
| Standards compliance | | | |
| Detail | IEC 62304<br>IEC 62366 | IEC 62304<br>IEC 62366 | Same |
| Intended Use | | | |
| Intended use | The CIVNET CIV- ob TM<br>(plus) is a clinical data managing software application and is indicated for antepartumn and intrapartumn monitoring of pregnant women in a healthcare setting | The Maternal Fetal Monitoring - Central Nurse System (hereinafter called "MFM-CNS") is a clinical data managing software application and is indicated for antepartum and intrapartum monitoring of pregnant women in a healthcare setting. | Same |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | The CIVNET CIV- ob TM<br>(plus) is indented to manage perinatal monitoring data acquired from bedside monitors or manual inputs for viewing at the central nursing station. The system also produces an electronic medical record.<br>The CIYNET CIV ob TM<br>(plus) has display fields for the following obstetric data: patient demographics, provider notes, fetal heart rate | The MFM-CNS is intended to manage perinatal monitoring data acquired from bedside monitors or manual input for viewing at the central nursing station. The system also produces an electronic medical record.<br>The MFM-CNS has display fields for the following obstetric data: | |
{7}------------------------------------------------
| (FHR), uterine activity<br>(via tocodynamometry or<br>IUP), etc. | patient demographics,<br>provider notes, fetal<br>heart rate (FHR), uterine<br>activity (via<br>tocodynamometry or<br>IUP), etc. |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Item | The Philips OB TraceVue<br>Obstetrical Information<br>Management System | MFM-CNS | SE |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Manufacturer/K# | Philips Medizin<br>System/K081203 | EDAN Instruments/None | |
| | Classification | | |
| Classified as per<br>FDA regulation | Class II | Class II | Same |
| | Software | | |
| CTG | The Philips OB TraceVue<br>Obstetrical Information<br>Management System can<br>analyze signal loss,<br>contractions, basal heart<br>rate, accelerations,<br>decelerations, short term<br>variation, long term<br>variation and other<br>parameters. | The MFM-CNS can<br>analyze signal loss,<br>contractions, basal heart<br>rate, accelerations,<br>decelerations, short term<br>variation, long term<br>variation and other<br>parameters. | Same |
| NICHD | The Philips OB TraceVue<br>Obstetrical Information<br>Management System can<br>analyze FHR baseline and<br>its scope, FHR baseline<br>variation and its scope,<br>accelerated number, early<br>deceleration, late<br>deceleration, variable<br>deceleration, prolonged<br>deceleration, sine curve and<br>other parameters. | The MFM-CNS can<br>analyze FHR baseline and<br>its scope, FHR baseline<br>variation and its scope,<br>accelerated number, early<br>deceleration, late<br>deceleration, variable<br>deceleration, prolonged<br>deceleration, sine curve<br>and other parameters. | Same |
The subject device shares the same characteristics in most items with the predicate device except in the following one aspect: The predicate device prints the CIV ob ™ (plus) data base definition (e.g. item names) but the subject device does not. MFM-CNS only prints fetal and/or maternal reports such as fetal monitoring graphs or maternal trend lists.
The comparison above shows that the differences do not affect the safety and
{8}------------------------------------------------
effectiveness of the MFM-CNS and there are no safety and effectiveness issues relating to the MFM-CNS.
# 8. Effectiveness and Safety Considerations:
## Clinical test:
Clinical testing is not required.
# Non-clinical test:
Since the subject is a software only product, EMC and Electrical Safety Evaluation are not required. But the following quality assurance measures were applied to the development of the MFM-CNS to ensure its safety and effectiveness:
- Software testing according to FDA Guidance General Principles of Software Validation dated on Jan. 11, 2002.
- Risk analysis according to ISO 14971: 2007
- Usability analysis according to IEC 62366: 2007
- Software life cycle management according to IEC 62304: 2006
The subject device passed all testing. The tests and analysis were all conducted to ensure the safety and effectiveness, and results show substantial equivalence between the subject device and the predicates.
# 9. Substantially Equivalent Determination
Verification and validation testing was done on the MFM-CNS and all testing passed pre-specified criteria. Since the testing passed, this premarket notification submission demonstrates that the subject device MFM-CNS is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.