PHILIPS AVALON FETAL MONITORS FM20, FM30, FM40, AND FM50
Applicant
Philips Medizin Systeme Boeblingen GmbH
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Feb 17, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
The Philips Avalon FM20, FM30, FM40 and FM50 Fetal/Maternal Monitors are intended for non-invasive monitoring of the physiological parameters of pregnant women during antepartum testing and labor and delivery. All monitors are intended for monitoring fetal and maternal heart rates, fetal movement profiles, maternal uterine activity, maternal noninvasive blood pressure, and additionally for the FM30, FM40 and FM50 maternal oxygen saturation (SpO2). The FM30 and FM50 are additionally intended for maternal ECG and for invasive monitoring of fetal Direct ECG and intrauterine pressure. All monitors are intended for generating alarms from fetal and maternal parameters, for displaying, storing and recording patient waves, data and related patient information and transmitting patient data to a surveillance system on a network, and for postpartum monitoring of the mother. All monitors are intended for use by trained health care professionals. They are intended for use in labor and delivery rooms, antepartum testing areas and during postpartum recovery in the hospital environment. They are not intended for use in intensive care units or operating rooms. The FM20 and FM30 are additionally intended for use in healthcare facilities and in private households.
Device Story
Avalon Fetal Monitors (FM20, FM30, FM40, FM50) provide multi-parameter monitoring for pregnant women. Inputs include ultrasound (fetal heart rate), fetal scalp electrodes (Direct ECG), Toco/IUP transducers (uterine activity), SpO2 sensors, and ECG electrodes. Device processes signals to display real-time waveforms, numeric data, and trends; generates alarms for abnormal parameters. Used by clinicians in labor/delivery, antepartum, and postpartum settings; FM20/FM30 also used in private homes. Output supports clinical decision-making regarding fetal/maternal well-being. Modification allows use of specific SpO2 sensors previously cleared for other Philips modules.
Clinical Evidence
Bench testing only. Verification and validation activities, including system-level regression testing and hazard analysis, were performed based on cleared specifications for predicate devices. Results demonstrated that the modified monitors meet all reliability requirements and performance claims.
Technological Characteristics
Multi-parameter fetal/maternal monitors. Sensing: ultrasound, ECG, pressure, pulse oximetry. Connectivity: network transmission to patient information/surveillance systems. Software-based signal processing for fetal/maternal heart rate, uterine activity, and SpO2. Form factor: bedside monitor units. Sterilization: N/A (non-invasive/external accessories).
Indications for Use
Indicated for trained healthcare professionals to monitor physiological parameters (uterine activity, heart rate, NIBP, pulse rate, SpO2, ECG) and fetal heart rates (single, twins, triplets) in pregnant women during antepartum testing, labor, delivery, and postpartum recovery. FM20/FM30 used in hospitals and private homes; FM40/FM50 used in hospitals. Contraindicated for use during defibrillation, electro-surgery, MRI, or on patients with cardiac pacemakers/external electrical stimulators.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 (K052795)
Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 (K062137)
Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 (K071800)
Reference Devices
Philips picoSATIIplus SpO2 pulse oximetry module
Philips Intellivue X2 (M3002A) multi measurement module
Philips M1020B SpO2 plug-in module
Related Devices
K071800 — PHILIPS AVALON FETAL MONITORS FM20 AND FM30 · Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · Sep 27, 2007
K111083 — PHILIPS AVALON FETAL/MATERNAL MONITORS FM20, FM30, FM40 AND FM50 · Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · Aug 17, 2011
K052795 — PHILIPS AVALON FETAL MONITOR, MODELS FM20 AND FM30 · Philips Medizin Systeme Boeblingen GmbH · Dec 30, 2005
K062137 — MODIFICATION TO PHILIPS AVALON FETAL MONITORS FM20 AND FM30 · Philips Medizin Systeme Boeblingen GmbH · Aug 24, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
KD92028.-p.1/4
8.0 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
FEB 1 7 2010
The submitter of this premarket notification is: Michael Asmalsky Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany
Fax: +49 7031 463-2442 Tel: +49 7031 463-1277 Email: michael.asmalsky@philips.com
This revised summary was prepared on August 14th, 2009.
8.1 Device Names
The names of the devices are the Philips Avalon Fetal Monitors FM20, FM30,FM40 and FM50. Classification names are as follows.
| Device Panel | Classification | ProCode | Description |
|-----------------------------------------------------------|----------------|---------|----------------------------------------------------------------|
| Obstetrical and<br>Gynecological<br>Monitoring<br>Devices | §884.2660, II | MAA | Fetal ultrasonic monitor and<br>accessories |
| | §884.2660, II | HEL | |
| | §884.2660, II | HEK | accessories |
| | §884.2660, II | KNG | |
| | §884.2675, II | HGP | Fetal scalp circular (spiral)<br>electrode and applicator |
| | §884.2700, II | HGS | |
| | §884.2700, II | KXO | Intrauterine pressure monitor<br>and accessories |
| | §884.2700, II | HFO | |
| | §884.2720, II | HFN | |
| | §884.2720, II | HFM | External uterine contraction<br>monitor and accessories |
| | §884.2740, II | HGM | Perinatal monitoring system<br>and accessories |
| | §884.2960, II | HGL | Obstetric ultrasonic<br>transducer and accessories |
| Circulatory<br>System Devices | §870.1100, II | DSJ | Alarm, Blood Pressure |
| | §870.1110, II | DSK | Computer, Blood Pressure |
| | §870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive |
| | §870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer<br>Signal |
| | §870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate<br>Alarm) |
| | §870.2340, II | DPS | Electrocardiograph |
| | §870.2600, I | DRJ | System, Signal Isolation |
| | §870.2700, II | DQA | Oximeter |
| | §870.2810, I | DSF | Recorder, Paper Chart |
| | §870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector |
•
{1}------------------------------------------------
8.2 Subject devices
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The subject devices Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 are substantially equivalent to the previously cleared Philips Avalon Fetal Monitors FM20 . FM30, FM40 and FM50 marketed pursuant to K052795, K062137 and K071800.
K0920Z8
The subject devices Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 are a modification of the legally marketed Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 and offer monitoring of fetal heart rates, fetal movement profiles, fetal Direct ECG, maternal uterine activity, maternal heart rates, maternal pulse rates, maternal ECG, maternal non-invasive blood pressure (NIBP) and maternal oxygen saturation (Sp02) during antepartum testing and labor and delivery.
8.3 Modifications
The modification of the Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 only permits specific Sp02 sensors, which are already cleared for the Philips picoSATIIplus SpO2 pulse oximetry module, the Philips Intellivue X2 (M3002A) multi measurement module and the Philips M1020B Sp02 plug-in module, to be also used with the Philips Avalon Fetal monitors FM30, FM40 and FM50.
8.4 Intended uge
modified devices have the same intended use as The previously cleared for the legally marketed predicate devices Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 (K052795, K062137 and K071-800). .
Intended Use: The Philips Avalon FM20, FM30, FM40 and Fetal/Maternal Monitors are intended for non-FM50 invasive monitoring of the physiological parameters of pregnant women during antepartum testing and labor and delivery.
monitors are intended for monitoring fetal and All Arr monitors rates, fetal movement profiles, maternal maternal issues, uterine activity, maternal noninvasive blood pressure, and additionally for the FM30, FM40 and FM50 maternal oxygen saturation (Sp02).
The FM30 and FM50 are additionally intended for maternal ECG and for invasive monitoring of fetal Direct ECG and intrauterine pressure.
All monitors are intended for generating alarms from fetal and maternal parameters, for displaying, storing and related waves, data recording patient and transmitting patient data to a patient information and surveillance system on a network, and for postpartum monitoring of the mother.
All monitors are intended for use by trained health care professionals.
They are intended for use in labor and delivery rooms, antepartum testing areas and during postpartum recovery in the hospital environment. They are not intended for use in intensive care units or operating rooms. The FM20 and FM30 are additionally intended for use in healthcare and 1.50 are are are and for example in doctors' offices, and in private households.
{2}------------------------------------------------
K092028 D. 3/4
Contraindications: All monitors are NOT intended for: conviduring defibrillation, electro-surgery, or magnetic resonance imaging (MRI).
ECG measurements on patients connected to external । electrical stimulators or with cardiac pacemakers.
use with the IUP/ECG patient module (M2738A) in । domestic establishments and those connected directly to public low-voltage supply network that supplies the buildings used for domestic purposes.
The before is the same intended use as previously cleared for the Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 (K052795, K062137 and K071800). The intended use and 1430 (Results) is a subset of the intended use of the FM20 and FM30.
As a summary, the following table provides an overview about the monitored parameters per fetal monitor.
| Monitored Parameters | FM20 | FM30 | FM40 | FM50 |
|-------------------------------------------------------------------|------|------|------|------|
| Up to three<br>Fetal Heart<br>Rates (FHR) via<br>ultrasound (US)* | Yes | Yes | Yes | Yes |
| Single Fetal<br>Heart Rate via<br>direct ECG<br>(DFHR) * | -- | Yes | -- | Yes |
| Fetal Direct ECG<br>(DECG) | -- | Yes | -- | Yes |
| Fetal Movement<br>Profile | Yes | Yes | Yes | Yes |
| Uterine activity<br>via external<br>Toco | Yes | Yes | Yes | Yes |
| Uterine activity<br>via intrauterine<br>pressure (IUP) | -- | Yes | -- | Yes |
| Pulse oximetry<br>(maternal SpO2) | -- | Yes | Yes | Yes |
| Maternal Pulse<br>Rate | Yes | Yes | Yes | Yes |
| Maternal Heart<br>Rate (MHR) via<br>maternal ECG | Yes | Yes | Yes | Yes |
| Maternal ECG<br>(MECG) | -- | Yes | -- | Yes |
| Non-invasive<br>blood pressure<br>(NIBP) | Yes | Yes | Yes | Yes |
*: A maximum of three fetal heart rates can be monitored (including max. one DECG)
{3}------------------------------------------------
8.5 Technological Characteristics The subject devices Philips Avalon Fetal Monitors FM30, FM40 and FM50 have the same technological characteristics as the legally marketed predicate devices Avalon Fetal Monitors FM20, FM30, FM40 and FM50.
109 20 28
8 . ୧ Verification and Validation
testing activities validation, and Verification, establish the performance, functionality, and reliability characteristics of the modified devices with respect to predicate. Testing involved level and system the regression tests as well as testing from the hazard criteria were based on the Pass/Fail analysis. specifications cleared for the predicate devices and test results showed substantial equivalence. The results demonstrate that the Philips Avalon Fetal Monitors meet all reliability requirements and performance claims.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing right, with flowing lines beneath them, possibly representing water or movement.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Michael Asmalsky Senior Regulatory Affairs Engineer Philips Medizin Systeme Böblingen GmbH Hewlett-Packard-Str. 2 71034 Böblingen GERMANY
FEB 1 7 2010
Re: K092028 · ·
Trade Name: Philips Avalon Fetal Monitors, model FM20, FM30, FM40 and FM50 Regulation Number: 21 CFR §884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: October 19, 2009 Received: October 22, 2009
Dear Mr. Asmalsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devise-Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{5}------------------------------------------------
Page 2 -
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
anine M. Morri Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K092028
Device Name: Philips Avalon Fetal Monitors FM20(M2702A), FM30(M2703A), FM40(M2704A) and FM50(M2705A) .
## Indications for Use:
### Avalon Fetal Monitor FM20:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.
# Avalon Fetal Monitor FM30:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.
#### Avalon Fetal Monitor FM40:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
#### Avalon Fetal Monitor FM50:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use No (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
| 11 A1 11 | |
| 100 | |
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number ***_***
Page 15
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.