Fluent Fluid Management System

K180825 · Hologic, Inc. · HIG · Aug 3, 2018 · Obstetrics/Gynecology

Device Facts

Record IDK180825
Device NameFluent Fluid Management System
ApplicantHologic, Inc.
Product CodeHIG · Obstetrics/Gynecology
Decision DateAug 3, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1700
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Fluent Fluid Management System is intended to provide liquid distension of the uterus during diagnostic and operative hysteroscopy, and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus while providing drive, control and suction for hysteroscopic morcellators.

Device Story

System consists of console, single-use procedure kit (In-FloPak, Out-FloPak), and waste bag; console includes motors for fluid inflow/outflow, pressure sensors, and touchscreen interface. Device provides liquid uterine distension; monitors fluid volume deficit; provides drive, control, and suction for hysteroscopic morcellators. Used in clinical settings by physicians. Console connects to MyoSure tissue removal devices and pneumatic foot pedal. Fluid management uses peristaltic pumps for irrigation and vacuum. Output includes real-time pressure and fluid deficit data on GUI. Overpressure protection and deficit alerts assist clinical decision-making to prevent complications during hysteroscopic procedures.

Clinical Evidence

Bench testing only. Included design verification of tubing, fluid management validation (inflow/outflow rates, suction), intrauterine pressure control, system performance, and cutting performance. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, irritation, acute systemic toxicity). Electrical safety and EMC testing per IEC 60601 series.

Technological Characteristics

Console-based system with peristaltic pumps for fluid management and morcellator drive. Pressure control range 40-120 mmHg; max flow 650 mL/min. Single-use sterile tubing sets (EtO sterilized per ISO 11135:2014). Includes pressure sensors, touchscreen GUI, and pneumatic foot pedal. Software classified as Major Level of Concern. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6.

Indications for Use

Indicated for patients undergoing diagnostic and operative hysteroscopy requiring uterine distension and/or tissue removal via hysteroscopic morcellators.

Regulatory Classification

Identification

A hysteroscopic insufflator is a device designed to distend the uterus by filling the uterine cavity with a liquid or gas to facilitate viewing with a hysteroscope.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue. August 3, 2018 Hologic, Inc. Christine M. Cameron Manager, Regulatory Affairs 250 Campus Drive Marlborough, MA 01752 Re: K180825 > Trade/Device Name: Fluent Fluid Management System Regulation Number: 21 CFR& 884.1700 Regulation Name: Hysteroscopic Insufflator Regulatory Class: II Product Code: HIG, HIH Dated: July 3, 2018 Received: July 5, 2018 Dear Christine M. Cameron: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Michael T. Bailey -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K180825 Device Name Fluent Fluid Management System Indications for Use (Describe) The Fluent Fluid Management System is intended to provide liquid distension of the uterus during diagnostic and operative hysteroscopy, and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus while providing drive, control and suction for hysteroscopic morcellators. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------|--| | <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | | __ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY Date: July 31. 2018 #### 5.1. 510(k) Submitter: Hologic, Inc. 250 Campus Drive Marlborough, MA 01752 Telephone: 508.263.8654 Establishment Registration Number: 1222780 Contact: Christine Cameron, Regulatory Affairs Manager Email: Christine.Cameron@hologic.com Phone: 508-263-8654 #### 5.2. Device: Trade Name: Fluent Fluid Management System Common/Usual Name: Hysteroscopic Insufflator Regulation Number: 21.CFR.Reg 884.1700 Regulation Name: Hysteroscopic insufflator Product Code: HIG (insufflator, hysteroscopic), HIH (hysteroscope (and accessories)) Device Class: Class II #### 5.3. Predicate Devices: Predicate 1 Tradename: Aquilex Fluid Control System H112 Submitter / 510(k) Holder: W.O.M. World of Medicine AG 510(k) #'s: K112642 #### Predicate 2 Tradename: MyoSure Tissue Removal Device System and MyoSure Tissue Removal Devices Submitter / 510(k) Holder: Hologic, Inc. 510(k) #'s: K172566 The predicate devices have not been subject to a design related recall. {4}------------------------------------------------ #### 5.4. Device Description: The Fluent Fluid Management System is a hysteroscopic fluid management system and drive console for hysteroscopic morcellators. The Fluent Fluid Management System consists of a console and single-use procedure kit. The single use procedure kit consists of a sterile Inflow Tube Set (In-FloPak) and Outflow Tube Set (Out-FloPak) with a nonsterile Waste Bag. The two FloPaks and Waste Bag connect to the Console for performing hysteroscopic procedures. The console includes motors that control inflow and outflow of fluid for hysteroscopic insufflation, as well as a pressure sensor monitoring system and graphical user interface touchscreen. In addition, the console includes a connection for MyoSure Tissue Removal Devices and a pneumatic foot pedal to perform tissue removal procedures. The Fluent Fluid Management System is compatible with currently marketed MyoSure Tissue Removal Devices. The primary functions of the subject device are fluid distension, fluid deficit tracking, and morcellator drive control. The subject device includes a deficit alert feature as well as an overpressure protection mechanism. ## 5.5. Indications for Use: The Fluent Fluid Management System is intended to provide liquid distension of the uterus during diagnostic and operative hysteroscopy, and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus while providing drive, control and suction for hysteroscopic morcellators. The subject device has the same general intended use as the primary and secondary predicate devices, respectively: (1) to provide liquid distension of the uterus while monitoring irrigation fluid and (2) to provide drive control for morcellators during hysteroscopic procedures. The subject device indications include reference to both the fluid management function of the primary predicate and the "drive, control, and suction for hysteroscopic morcellators" function of the secondary predicate to reflect the combination of these two device functions into the subject device. These features do not interfere with each other and do not change the overall intended use relative to the predicate devices. ## 5.6. Comparison of Characteristics: The method of use, mechanism of action, mode of operation and functional performance of the Fluent Fluid Management System are substantially equivalent to the Aquilex Fluid Control System H112 (K112642) for fluid management utilized for uterine distention as well as the MyoSure Tissue Removal Device System, specifically the console for providing drive, control and suction for the MyoSure Tissue Removal Devices for tissue removal (K172566). {5}------------------------------------------------ | Table 5-1: Device Comparison of Technological Characteristics | | | | | |---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | Device and<br>Predicate<br>Devices | Aquilex Fluid<br>Control System<br>K112642<br>Predicate 1 | MyoSure<br>Tissue<br>Removal<br>System<br>K172566<br>Predicate 2 | Fluent Fluid<br>Management<br>System<br>Subject Device | Comparison to<br>Predicates | | General Device Characteristics | | | | | | Method of<br>Use | Uterine<br>distention | Drive, control, and<br>suction for<br>hysteroscopic<br>morcellators and<br>hysteroscopic<br>resection and<br>removal of tissue via<br>tissue removal<br>devices | Uterine distension<br>(fluid management)<br>and drive, control,<br>and suction for<br>hysteroscopic<br>morcellators | Similar | | Mechanism of<br>Action | Fluid flow with a<br>single roller<br>peristaltic pump<br>for irrigation<br>and a<br>diaphragm<br>pump for<br>vacuum. | The console<br>provides<br>mechanical<br>control, drive, and<br>suction for the<br>hysteroscopic<br>morcellator. | Fluid flow with a<br>single peristaltic<br>pump for irrigation<br>and single peristaltic<br>pump for vacuum. | Similar | | Mode of<br>Operation | Provide liquid<br>distension of<br>the uterus<br>during<br>diagnostic and<br>operative<br>hysteroscopy,<br>and to monitor<br>the volume<br>differential<br>between the<br>irrigation fluid<br>flowing into<br>and out of the<br>uterus | Vacuum is used to<br>pull tissue into<br>the side- facing<br>cutting window<br>where it is cut and<br>aspirated through<br>the cutter and<br>deposited into a<br>collection<br>canister. | Provide liquid<br>distension of the<br>uterus during<br>diagnostic and<br>operative<br>hysteroscopy, and<br>to monitor the<br>volume differential<br>between the<br>irrigation fluid<br>flowing into and out<br>of the uterus.<br>Vacuum is created<br>to pull tissue into<br>the side-facing<br>cutting window<br>where it is cut and<br>aspirated through<br>the cutter and<br>deposited into a<br>collection canister. | Similar | | Fluid Management System | | | | | | Pressure<br>Control<br>Range | 40-150mmHg | N/A | 40-120mmHg | Different | | Maximum<br>Intrauterine<br>Pressure | 150mmHg | N/A | 120mmHg | Different | | Flow rate<br>(maximum) | 800mL/min | N/A | 650mL/min | Different | | Overpressure<br>Protection | Yes | N/A | Yes | Same | | Fluid Deficit<br>Tracking | Yes | N/A | Yes | Same | | Drive Control Console for Hysteroscopic Morcellators | | | | | | Morcellator<br>drive<br>Rotation<br>Speed | N/A | 8075±1000 RPM | 8075±1000 RPM | Same | {6}------------------------------------------------ {7}------------------------------------------------ The descriptive characteristics of the Fluent Fluid Management System and the predicate devices are shown in the table above. K112642 is the primary predicate used to compare the technological characteristics of the fluid management function (hysteroscopic insufflator) of the subject device, while K172566 is used only to compare the morcellator drive-control console function of the subject device. The combination of these features does not change the intended use or risk profile relative to the independent predicate devices. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety or effectiveness. ## 5.7. Performance Testing: ## Sterility and Shelf Life For the sterile Inflow/Outflow tubing, EtO sterilization per ISO 11135:2014 and EtO and ECH residual testing in accordance with ISO 10993-7:2008 were conducted. In addition, package integrity testing per ASTM F1980-16 and transport validation and accelerated aging testing with functional testing were performed for the sterile components in order to confirm the three-year shelf life. ## Biocompatibility: - Cytotoxicity - ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity - Sensitization - ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization - Irritation ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests ● for irritation and skin sensitization - Acute System Toxicity - ISO 10993-11:2006 - Part 11: Tests for systemic toxicity ## Software The Fluent Fluid Management System successfully performed system design control verification and validation testing, which included software documentation for a Major Level of Concern in accordance with FDA's Guidance for the Content of premarket Submissions for Software Contained in Medical Devices (issued on May 11, 2005). ## Electrical Safety and EMC: - IEC 60601-1:2005 + A1:2012, C1:2009 + A2:2010 Medical Electrical Equipment - Part 1 General requirements for safety and essential performance. - IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements for basic . safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests - IEC 60601-1-6:2010 (Third Edition) + A1:2013 Medical electrical equipment - Part 1-6: Particular requirements for the basic safety and essential performance: Usability ## Performance Testing Non-clinical performance testing was conducted to verify the functional performance of the subject device. The performance testing included the following: {8}------------------------------------------------ - Design verification (mechanical testing) of the inflow/outflow tubing sets - . Fluid management validation testing to assess inflow rate, outflow rate, and TRD suction - Intrauterine pressure control testing ● - System performance testing ● - System cutting performance for the drive control system #### 5.8. Conclusion: The performance data demonstrate that the Fluent Fluid Management System is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...