Myosure Hysteroscopic Tissue Removal System and Myosure Tissue Removal Devices

K172566 · Hologic, Inc. · HIH · Sep 20, 2017 · Obstetrics/Gynecology

Device Facts

Record IDK172566
Device NameMyosure Hysteroscopic Tissue Removal System and Myosure Tissue Removal Devices
ApplicantHologic, Inc.
Product CodeHIH · Obstetrics/Gynecology
Decision DateSep 20, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 884.1690
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Myosure Hysteroscopic Tissue Removal System and Myosure Tissue Removal Devices are intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: Submucous myomas Endometrial Polyps Retained products of conception

Device Story

System uses mechanical resection for intrauterine tissue removal; includes tissue removal drive system, tissue removal device, and foot pedal. Surgeon operates device hysteroscopically; targeted tissue drawn into cutting window via suction; inner blade performs mechanical resection. Provides precise control over location and extent of tissue removal. Used by gynecologists in clinical settings. Benefits include minimally invasive removal of submucous myomas, endometrial polyps, and retained products of conception.

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing (cytotoxicity, systemic toxicity, intracutaneous reactivity, sensitization) performed per ISO 10993-5, 10993-11, and 10993-10 to validate modified cleaning processes.

Technological Characteristics

Mechanical tissue resection system. Components: drive system, removal device, foot pedal. Materials: stainless steel. Manufacturing: laser cutting of window; alkaline cleaning, nitric acid electropolish, and passivation per ASTM-A967/A967M-13. Sterilization: same as predicate. Connectivity: N/A.

Indications for Use

Indicated for trained gynecologists to perform hysteroscopic resection and removal of submucous myomas, endometrial polyps, and retained products of conception in patients requiring intrauterine tissue removal.

Regulatory Classification

Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a design that suggests movement or flow. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 20, 2017 Hologic, Inc. Catherine Sanford Regulatory Affairs Specialist 250 Campus Drive Marlborough, MA 01752 Re: K172566 > Trade/Device Name: Myosure Hysteroscopic Tissue Removal System and Myosure Tissue Removal Devices Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: August 28, 2017 Received: August 30, 2017 Dear Catherine Sanford: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172566 #### Device Name Myosure Hysteroscopic Tissue Removal System and Myosure Tissue Removal Devices Indications for Use (Describe) The Myosure Hysteroscopic Tissue Removal System and Myosure Tissue Removal Devices are intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: Submucous myomas Endometrial Polyps Retained products of conception | Type of Use (Select one or both, as applicable) | <div> <input type="checkbox"/> </div> | <div> <input type="checkbox"/> </div> | |-------------------------------------------------|---------------------------------------|---------------------------------------| |-------------------------------------------------|---------------------------------------|---------------------------------------| X | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) SUMMARY Date: September 18th, 2017 #### 510(k) Submitter: Hologic, Inc. 250 Campus Drive Marlborough, MA 01752 Attn: Catherine Sanford P: 508.263.8604 F: 508.263.2403 #### Establishment Registration Number: 1222780 #### Subject Device: Trade Name: Myosure Hysteroscopic Tissue Removal System and Myosure Tissue Removal Devices Common/Usual Name: Hysteroscope and Accessories Product Code: HIH Classification Name: Hysteroscope and Accessories, 21.CFR.Reg 884.1690 Classification: Class II Panel: Obstetrics/Gynecology #### Predicate Device: Trade Name: Myosure Tissue Removal Device System and Myosure Tissue Removal Devices Submitter / 510(k) Holder: Hologic, Inc. 510(k) #: K152723 Classification code: HIH, Hysteroscope and Accessories, and Regulation: 21.CFR.884.1690 This predicate has not been subject to a design-related recall. #### DEVICE DESCRIPTION The Myosure Hysteroscopic Tissue Removal System consists of the following procedural components which are identical to those found in the sponsor's previously cleared device: - Tissue Removal Drive System O - Tissue Removal Device o - o Foot Pedal {4}------------------------------------------------ Hologic. Inc. Special 510(k) Submission September 18th, 2017 The Myosure Hysteroscopic Tissue Removal System uses mechanical resection to remove endometrial polyps, submucous myomas, and retained products of conception hysteroscopically from the uterus. Mechanical resection allows the surgeon to have precise control of the locations and extent of tissue resected by drawing the targeted tissue into the cutting window under suction while the inner blade cuts the tissue. There have been no major changes in design or materials in the subject Myosure System or its associated tissue removal devices since their market clearance. ## Indications for Use: The Myosure Hysteroscopic Tissue Removal System and Myosure Tissue Removal Devices are intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: > Submucous myomas Endometrial Polyps Retained products of conception ## Comparison of Characteristics: The principles of operation and primary functional specifications of the modified Myosure Hysteroscopic Tissue Removal System are identical to those of the predicate Myosure Hysteroscopic Tissue Removal System. The modified Myosure Hysteroscopic Tissue Removal System is different from the predicate Myosure Hysteroscopic Tissue Removal System as follows: - . The process of manufacturing the outer sheath of the device handpiece has been brought in-house to Hologic Inc. - . The cutting window of the outer tube is being cut into the tube using a laser rather than the previously used electronic discharge machining (EDM) - . The process used to clean the outer tube following cutting of the window has been modified to include alkaline cleaning, nitric acid electropolish, and passivation in accordance with ASTM-A967/A967M -13 Standard Specification for Chemical Passivation Treatments for Stainless Steel Parts. The fundamental scientific technology of the proposed device has not changed relative to the predicate device (K152723): - . Has the same Indications for Use, - Has the same mechanism of action and mode of operation, ● - Is manufactured using the same raw materials, {5}------------------------------------------------ Hologic. Inc. Special 510(k) Submission September 18th, 2017 Page 3 of 3 - Is packaged and sterilized using the same process, ● - Has the same shelf life. ## Biocompatibility Testing: Biocompatibility testing of the modified MyoSure Hysteroscopic Tissue Removal Device outer tubes was conducted to demonstrate that the modified cleaning process does not negatively impact the biocompatibility of the device. Testing was completed using the same methodology as was used in support of the predicate MyoSure Hysteroscopic Tissue Removal System 510(k) submission and in accordance with both AAMI/ISO 10993-1:2009/(R) 2013 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process and the FDA Guidance for Industry and Food and Drug Administration Staff titled Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and dated June 16, 2016. Cytotoxicity, systemic toxicity, intracutaneous reactivity, and maximization sensitization studies were completed in accordance with the standards ISO 10993-5:2009. ISO 10993-11:2006. and ISO 10993-10:2010 respectively. Results from this testing demonstrated that the modified cleaning process does not negatively impact the biocompatibility of the device for the specified use conditions and is therefore substantially equivalent to that of the predicate device. ## Performance Testing: Performance verification testing of the modified Myosure Hysteroscopic Tissue Removal System was not required because there is no modification to the design of the outer tube of the device. An engineering assessment of the cutting method determined there was no impact to the cutting performance of the device as a result of the change in manufacturing process. ## Conclusion: The proposed Myosure Tissue Removal System met all acceptance criteria for design verification and validation, as specified by applicable standards, guidance, test protocols and/or customer inputs. The proposed Myosure Tissue Removal System is substantially equivalent to the legally marketed predicate device (Hologic's Myosure Tissue Removal Device System and Tissue Removal Devices).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...