IMAGYN MICROLAP AND MICRO LAP-GOLD LAPAROSCOPES
Device Facts
| Record ID | K965055 |
|---|---|
| Device Name | IMAGYN MICROLAP AND MICRO LAP-GOLD LAPAROSCOPES |
| Applicant | Imagyn Medical, Inc. |
| Product Code | HET · Obstetrics/Gynecology |
| Decision Date | Mar 6, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.1720 |
| Device Class | Class 2 |
Indications for Use
The MicroLap™ and MicroLap Gold Laparoscopes are used to permit viewing of the abdominal cavity for the purpose of performing diagnostic and surgical procedures.
Device Story
Small diameter, rigid, fiberoptic laparoscopes; no through lumens. Stainless steel shaft; inner illumination fibers; imaging fiber bundle; distal lens; rotatable eyepiece. Connects to external endoscopic light source via light cables; optional connection to camera head via focusing optical coupler for video display/recording. Used in clinical settings by physicians for abdominal cavity visualization; facilitates diagnostic and surgical procedures; provides direct optical or video-assisted visualization.
Clinical Evidence
Bench testing only. Evaluated optical and mechanical characteristics, effects of simulated reuse, thermal characteristics, and biocompatibility.
Technological Characteristics
Rigid, fiberoptic laparoscope; stainless steel shaft; fiber bundle imaging; fiber illumination. External light source (e.g., xenon) required. No through lumens. Optical transmission via distal/proximal lens systems.
Indications for Use
Indicated for patients requiring visualization of the abdominal cavity during diagnostic and surgical procedures.
Regulatory Classification
Identification
A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.
Predicate Devices
- Imagyn MicroLap Laparoscope
- Medical Dynamics Optical Catheter
- Optimized Technologies Model F18-200 Laparoscope
- Origin Medsystems, Inc. Small Diameter Endoscope
- Galileo Endoscope
Related Devices
- K962246 — RIGID FIBER OPTIC LAPAROSCOPE · Health Care International Assoc. · Feb 7, 1997
- K030174 — GENICON LAPAROSCOPES · Genicon, LC · Jun 19, 2003
- K955845 — LL 100, 103, 104 DIAGNOSTIC LAPAROSCOPE RIGID ROD LENS · Surgical Image Laboratories, Inc. · Jul 5, 1996
- K201151 — O-Mec 690 Series Laparoscopes · Omec Medical, Inc. · Jun 22, 2020
- K971150 — MIST PANORAMIC LAPAROSCOPE A RIGID LAPAROSCOPE · Mist, Inc. · Sep 23, 1997