INSUFFLATION NEEDLE
Device Facts
| Record ID | K963359 |
|---|---|
| Device Name | INSUFFLATION NEEDLE |
| Applicant | Applied Medical Resources |
| Product Code | HET · Obstetrics/Gynecology |
| Decision Date | Sep 24, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.1720 |
| Device Class | Class 2 |
Indications for Use
The AMR Insufflation Needle is a disposable device indicated for use for insufflating the peritoneum prior to laparoscopic surgery.
Device Story
Disposable 14-gauge stainless steel needle; plastic handle; spring-loaded safety stylet; color-coded external indicator for stylet position; stopcock with luer port for insufflation pump connection. Used by surgeons in clinical/OR settings to establish pneumoperitoneum. Needle inserted into peritoneum; stylet retracts upon resistance; indicator confirms position; gas delivered via pump to facilitate laparoscopic access.
Clinical Evidence
Bench testing only; no clinical data.
Technological Characteristics
14-gauge stainless steel needle; plastic handle; spring-loaded safety stylet; color-coded indicator; stopcock with luer port; 120mm and 150mm lengths; disposable.
Indications for Use
Indicated for patients undergoing laparoscopic surgery requiring peritoneal insufflation.
Regulatory Classification
Identification
A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.
Related Devices
- K172835 — GRI-Alleset Veress Needle · Gri Medical and Electronic Technology Co., Ltd. · Jun 11, 2018
- K970788 — PNEUMO-MATIC INSUFFLATION NEEDLE · Apple Medical Corp. · Mar 27, 1997
- K150068 — Unimicro Veress Needle · Unimicro Medical Systems (Shenzhen) Co., Ltd. · Jul 31, 2015
- K983925 — ENDOPATH ULTRA VERESS NEEDLE · Ethicon Endo-Surgery, Inc. · Feb 3, 1999
- K172120 — Veress Needle · Wickimed (Huizhou) Medical Equipment Manufacturing Co.,Ltd · Sep 14, 2017