The Utepreva Endometrial Sampler (UP01) is used to obtain endometrial cytological and histological samples.
Device Story
Single-use, sterile endometrial sampler; used by clinicians to obtain cytological and histological samples. Device inserted through cervix until guard reaches cervix; brush extended to uterine fundus; sponge tip minimizes perforation risk and absorbs fluid/cells. Device rotated 4-5 times (360°); brush/tip withdrawn into sheath. Integrated plunger creates suction to aspirate cells and prevent sample loss. Sample immersed in preservative solution for pathology lab analysis. Benefits include minimally invasive tissue collection for diagnostic evaluation.
Clinical Evidence
Bench testing only. Performance verified via biocompatibility (ISO 10993-5, ISO 10993-10), visual/dimensional inspection, tensile strength, torque, bristle retention, fluid retention, and fluid transfer testing. All tests passed.
Indicated for obtaining endometrial cytological and histological samples in patients requiring endometrial sampling.
Regulatory Classification
Identification
An endometrial brush is a device designed to remove samples of the endometrium (the mucosal lining of the uterus) by brushing its surface. This device is used to study endometrial cytology (cells).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” and
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(2) Labeling:
(i) Indication: Only to evaluate the endometrium, and
(ii) Contraindications: Pregnancy, history of uterine perforation, or a recent cesarean section, and
(3) Design and testing:
(i) The sampling component is covered within the vagina, and
(ii) For adherence of the bristles and brush head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 14, 2024
Utepreva LLC % Melissa Paffenroth Senior Regulatory Consultant MEDIcept, LLC. 200 Homer Avenue Ashland, Massachusetts 01721
Re: K240595
Trade/Device Name: Utepreva Endometrial Sampler (UP01) Regulation Number: 21 CFR 884.1100 Regulation Name: Endometrial Brush Regulatory Class: II Product Code: HFE Dated: March 4, 2024 Received: July 1, 2024
Dear Melissa Paffenroth:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdth/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Ouality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/trainingand-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory- assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240595
## Device Name
Utepreva Endometrial Sampler (UP01)
Indications for Use (Describe)
The Utepreva Endometrial Sampler (UP01) is used to obtain endometrial cytological and histological samples.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary Utepreva, LLC Utepreva Endometrial Sampler (UP01)
#### I. SUBMITTER
| 510(k) Holder: | Utepreva, LLC<br>100 Jericho Quadrangle<br>Suite 333<br>Jericho, NY 11753 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Melissa Paffenroth, PhD, RAC-Devices<br>Senior Consultant, Regulatory<br>MEDIcept, Inc.<br>Tel: (617) 512-9943<br>Email: mpaffenroth@medicept.com |
Date Prepared: March 4, 2024
#### II. DEVICE
| Name of Device: | Utepreva Endometrial Sampler (UP01) |
|----------------------|-------------------------------------|
| Common Name: | Endometrial brush |
| Classification Name: | Brush, Endometrial |
| Regulation Number: | 21 CFR 884.1100 |
| Regulatory Class: | Class II |
| Product Code: | HFE |
#### III. PREDICATE DEVICE
K170603, Tao BrushTM I.U.M.C Endometrial
The predicate device identified above has not been subject to design-related recalls.
#### IV. DEVICE DESCRIPTION
The Utepreva Endometrial Sampler (UP01) is a single-use, sterile device designed to obtain endometrial cytological and histological samples. The device is an Rx device that functions as follows: The Utepreva Endometrial Sampler is gently inserted though the cervix into the uterus. The skirt (guard) will reach the cervix indicating to the clinician to stop inserting the device. The brush is then extended by pushing the handle until resistance is felt when the tip meets the fundus of the uterus. The sponge on the tip is intended to minimize the risk of piercing of the fundus by the tip of the brush and enables additional tissue sampling by absorption of the fluid and the cells from the uterus. The Utepreva Endometrial Sampler is rotated in a clockwise or counterclockwise manner for four or five 360° rotations. Upon completion of the rotations, the handle is pulled back so that the brush and tip are withdrawn into the outer sheath. The suction
{4}------------------------------------------------
created by the plunger located inside of the sheath provides aspiration of the cells and prevents loss of the sample from the brush, sponge and any fluid that was aspirated can then be immersed into the preservative solution to be used by the pathology lab.
The device materials consist of polydimethylsiloxane, high density polyethylene, cyanoacrylate adhesive, Loctite, Permabond polyolefin primer, stainless steel, Nylon, polycarbonate, and polypropylene. The device is packaged in a Tyvek® pouch. One pouch is then packaged in a unit box. Fifteen unit boxes are packaged within a single master box to create the final shipped product.
#### V. INDICATIONS FOR USE
The Utepreva Endometrial Sampler (UP01) is used to obtain endometrial cytological and histological samples.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
In accordance with the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] issued July 28, 2014, the comparison between the predicate and the subject devices is shown to be substantially equivalent by comparing the indications for use, principles of operation, technological characteristics, and performance testing similarities and differences (Table 1). The technological characteristic differences rationale demonstrates that the subject device is safe and effective as a legally marketed device and does not raise different questions of safety and effectiveness than the predicate device.
| Characteristic | Subject Device<br>Utepreva Endometrial<br>Sampler (UP01)<br>(K240595) | Predicate Device<br>Tao Brush™ I.U.M.C.<br>Endometrial Sampler<br>(K170603) | Comparison |
|------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------------|------------|
| Regulatory Class | Class II | Class II | Same |
| Regulatory Number | 884.1100 | 884.1100 | Same |
| Regulation Name | Endometrial Brush | Endometrial Brush | Same |
| Product Code | HFE | HFE | Same |
| Manufacturer | Utepreva LLC | Cook Incorporated | - |
| Rx/OTC | Rx | Rx | Same |
| Sterile | Yes | Yes | Same |
| Sterilization | EtO | EtO | Same |
| Shelf-Life | 3 months | 3 years | Different |
| Insertion Depth<br>Indicator | Skirt (Guard) on Sheath | Black Ink Mark on Sheath | Different |
| Dimensions | | | |
| Working Length | 26.8 cm | 26.9 cm | Similar |
| Ball Tip | 10.2 Fr | 9.5 Fr | Similar |
Table 1. Comparison Table for Subject Device Utepreva Endometrial Sampler and Predicate Device Tao Brush™ I.U.M.C. Endometrial Sampler
{5}------------------------------------------------
| Characteristic | Subject Device<br>Utepreva Endometrial<br>Sampler (UP01)<br>(K240595) | Predicate Device<br>Tao Brush™ I.U.M.C.<br>Endometrial Sampler<br>(K170603) | Comparison |
|----------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------------|------------|
| Sheath OD | 11.43 Fr | 9.3 Fr | Similar |
| Outer Sheath Length | 22.9 cm | 21.5 cm | Similar |
| Brush Length | 3.2 cm | 3.5 cm | Similar |
| Brush Diameter | 3 mm | 6 mm | Different |
| Insertion Depth<br>Indicator Distance from<br>Distal Tip | 6.58 cm | 7 cm | Similar |
| <b>Materials</b> | | | |
| Insertion Depth<br>Indicator | Skirt: Polycarbonate,<br>acrylic adhesive | Black ink mark on sheath | Different |
| Wire Core | Stainless steel | Stainless steel | Same |
| Plunger | Polydimethylsiloxane | NA | - |
| Silicone Fluid | Polydimethylsiloxane | NA | - |
| Brush Bristles | Nylon | Nylon | Same |
| Backstop | Nylon 11 | NA | - |
| Ball Tip | Polyurethane,<br>Polypropylene | Acrylonitrile butadiene<br>styrene (ABS) | Different |
#### PERFORMANCE DATA VII.
Non-clinical performance testing was conducted to verify the Utepreva Endometrial Sampler performance. The following performance data was provided in support of the substantial equivalence determination (Table 2).
Table 2. Non-Clinical Performance Testing
| Test | Conclusion |
|------------------------------------------------------------------------------------------------------------------|------------|
| Biocompatibility | Pass |
| Cytotoxicity Testing (ISO 10993-5:2009) Sensitization (ISO 10993-10:2010) Vaginal Irritation (ISO 10993-10:2010) | |
| Visual & Dimensional Inspection | Pass |
| Tensile Strength | Pass |
| Torque | Pass |
| Bristle Retention | Pass |
| Fluid Retention | Pass |
| Fluid Transfer | Pass |
## VIII. CONCLUSIONS
The information provided and summarized in the substantial equivalence table and performance testing results demonstrate that the Utepreva Endometrial Sampler is as safe and effective as the predicate device and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.