Last synced on 13 December 2024 at 11:05 pm

Brush, Endometrial

Page Type
Product Code
Regulation Medical Specialty
Obstetrical and Gynecological
Review Panel
Obstetrical and Gynecological
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
884.1100
GMP Exempt?
No
Summary Malfunction Reporting
Eligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Third Party Eligible

CFR § 884.1100 Endometrial brush

§ 884.1100 Endometrial brush.

(a) Identification. An endometrial brush is a device designed to remove samples of the endometrium (the mucosal lining of the uterus) by brushing its surface. This device is used to study endometrial cytology (cells).

(b) Classification. Class II. The special controls for this device are:

(1) FDA's:

(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” and

(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”

(2) Labeling:

(i) Indication: Only to evaluate the endometrium, and

(ii) Contraindications: Pregnancy, history of uterine perforation, or a recent cesarean section, and

(3) Design and testing:

(i) The sampling component is covered within the vagina, and

(ii) For adherence of the bristles and brush head.

[45 FR 12684, Feb. 26, 1980, as amended at 52 FR 17741, May 11, 1987; 65 FR 17146, Mar. 31, 2000]

Brush, Endometrial

Page Type
Product Code
Regulation Medical Specialty
Obstetrical and Gynecological
Review Panel
Obstetrical and Gynecological
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
884.1100
GMP Exempt?
No
Summary Malfunction Reporting
Eligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Third Party Eligible

CFR § 884.1100 Endometrial brush

§ 884.1100 Endometrial brush.

(a) Identification. An endometrial brush is a device designed to remove samples of the endometrium (the mucosal lining of the uterus) by brushing its surface. This device is used to study endometrial cytology (cells).

(b) Classification. Class II. The special controls for this device are:

(1) FDA's:

(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” and

(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”

(2) Labeling:

(i) Indication: Only to evaluate the endometrium, and

(ii) Contraindications: Pregnancy, history of uterine perforation, or a recent cesarean section, and

(3) Design and testing:

(i) The sampling component is covered within the vagina, and

(ii) For adherence of the bristles and brush head.

[45 FR 12684, Feb. 26, 1980, as amended at 52 FR 17741, May 11, 1987; 65 FR 17146, Mar. 31, 2000]