Hysteroscope: Provides illumination and visualization of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. Gynecologic Laparoscope: Used to permit direct viewing of the organs within the peritoneal cavity for the purpose of performing diagnostic and surgical procedures.
Device Story
Rigid endoscope system comprising hysteroscope and gynecologic laparoscope; features compact objectives and rod-lens system. Device body constructed of surgical steel with internal light-carrying fibers. Used by physicians in clinical/surgical settings for direct visualization of cervical canal, uterine cavity, or peritoneal cavity during diagnostic and surgical procedures. Device provides optical illumination and image transmission to the clinician's eye via an eyepiece. Facilitates minimally invasive surgical access and diagnostic assessment. Sold non-sterile; requires user-performed sterilization prior to each procedure.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence established through design comparison, material biocompatibility, and validation of sterilization protocols.
Technological Characteristics
Rigid endoscope; surgical steel construction (DIN 17442); rod-lens optical system; fiber-optic illumination. Complies with IEC 60601-2-18, DIN 58105, DIN 17442, and DIN 980. Non-sterile; user-sterilizable.
Indications for Use
Indicated for patients requiring diagnostic or surgical procedures involving the cervical canal, uterine cavity, or peritoneal cavity. Used by clinicians for direct visualization and illumination.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{1}------------------------------------------------
## 510/k) Summary of safety and effectiveness
#### Applicant:
SEP 2 9 2003
asap endoscopic products GmbH Tullastr. 87 a 79108 Freiburg / Germany
#### Contact:
Dr. Martina Günderoth C.R.C. Partnerschaftsgesellschaft Katharinenstr. 5 23554 Lübeck, Germany
Phone: +49 (451) 388 2864 Fax: +49 (451) 388 2867 Email: crc@crc-online.de
#### Device name
| Hysteroscope, Types: | 10-0018-00, 10-0019-00, 10-0020-00, 10-0021-00,<br>10-0022-00, 10-0023-00, 10-0024-00, |
|-------------------------|----------------------------------------------------------------------------------------|
| Gyn. Laparoscope, Types | 10-0015-00, 10-0016-00, 10-0089-00, |
#### Common name
Hysteroscope, Gyn. Laparoscope
#### Predicate device name
- Baho Hysteroscope (K # not available) and Galileo hysteroscope (K962256)
- Baho Gyn. Laparoscope (K982276)
#### Code of Federal Regulations (CFR) number
888.1690 and 884.1720
#### General device description
The asap hysteroscope and gynecologic laparoscope (herinafter: asap endoscope) is a rigid type endoscope with a new generation of compact objectives and a newly developed rod-lens system.
The basic design of the asap endoscope is similar to those legally available for sale in the U.S.A.. It consists of an eyepiece and the body with light guide and rod-lens system. The body is designed of an outer and an inner tube of surgical steel. The light carrying fibers are sandwiched between these tubes. The inner tube of the body contains the rod-lens system.
{2}------------------------------------------------
510(k) Sum
#### Indications for use
Like the predicate device, the asap hysteroscope is used to illuminate and visualize the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
The asap gynecologic laparoscope is - like the predicate device - used to permit direct viewing of the organs wihin the peritoneal cavity for the purpose of performing diagnostic and surgical procedures.
### Voluntary standard compliance
The asap endoscopse comply with
- applicable portions of voluntary standards IEC 60601-2-18
- DIN 58105, part 1 and 2
- DIN 17442 (medical steel), as well as applicable portions of
- DIN 980
#### Substantially equivalence - Safety and effectiveness
The specifications and intended use of the asap endoscopes are the same to those of the claimed predicate devices. There are no significant differences between the asap endoscopes and the claimed predicates in design or conditions of intended use.
The asap endoscope is constructed of materials of the same specifications as the predicate devices to ensure biocompatibility. The asap endoscope conforms to applicable ISO standards.
The device will be sold non-sterile, to be sterilized prior to each procedure by the user. The ability to repeatedly adequately sterilize the devices has been confirmed by validation protocol.
#### Conclusion
In all respects, the asap endoscope is substantially equivalent to one or more rigid endoscopes currently marketed in the USA. It is constructed of materials of the same specifications as the predicate devices to ensure biocompatibility and it conforms to applicable ISO standards.
The ability to repeatedly adequately sterilize the asap endoscopes has been confirmed by validation protocol.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal. The seal is circular and contains text around the perimeter. The text reads "Department of Health & Human Services USA". In the center of the seal is an abstract image of an eagle.
# SEP 2 9 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Horst Baholzer General Manager asap endoscopic products GmbH Tullastr. 87a 79108 freiburg GERMANY
Re: K031974
Trade/Device Name: asap Hysteroscope, Models 10-0018-00, 10-0019-00, 10-0020-00, 10-0021-00, 10-0022-00, 10-0023-00, 10-0024-00; and, asap Gynecologic Laparoscope (Surgery), Models 10-0015-00, 10-0016-00, and 10-0089-00 Regulation Number: 21 CFR §884.1690 Regulation Name: Hysteroscope and accessories
Product Code: 85 HIH Regulation Number: 21 CFR §884.1720 Regulation Name: Gynecologic laparoscope and accessories
Product Code: 85 HET
Regulatory Class: II
Dated: June 10, 2003
Received: July 1, 2003
Dear Mr. Baholzer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to de vices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 510(k) Number (if known):_______K031974
# Device Name:
Indications For Use:
Hysteroscope: Provides illumination and visualization of the cervical canal and the r frorerosevity for the purpose of performing diagnostic and surgical procedures.
Gynecologic Laparoscope: Used to permit direct viewing of the organs within the Cyrioosie Laparess a performing diagnostic and surgical procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel h. Inson
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Dev
510(k) Number K031974
**Prescription Use**
**(Per 21 CFR 801.109)**
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.