GALILEO HYSTEROSCOPES AND LAPAROSCOPES
K962256 · Galileo Electro-Optics Corp. · HIH · Dec 6, 1996 · Obstetrics/Gynecology
Device Facts
| Record ID | K962256 |
| Device Name | GALILEO HYSTEROSCOPES AND LAPAROSCOPES |
| Applicant | Galileo Electro-Optics Corp. |
| Product Code | HIH · Obstetrics/Gynecology |
| Decision Date | Dec 6, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.1690 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Galileo Hysteroscopes are rigid endoscopes intended for direct visualization of cervical canal and uterine cavity for diagnostic and surgical procedures during gynecological procedures. The Galileo Laparoscopes are intended for direct visualization of the organs within the peritoneum for diagnostic and surgical procedures on the female genital organs. The Galileo Laparoscopes are designed to be introduced through natural body cavities or through introducers, needles, trocars, catheters, sheaths or other devices with thru-lumens having inside diameters larger than the outside diameter of the endoscope.
Device Story
Reusable rigid endoscopes; rod-lens or fiber-optic imaging bundles; external high-intensity light source transmits light via optical fibers to distal tip; objective lens captures target image; image transmitted to proximal eyepiece for direct viewing or video monitor connection. Used in clinical settings by physicians for gynecological diagnostic and surgical procedures. Enables visualization of body cavities, tissues, and organs; optional working channels allow passage of surgical instruments. Benefits include minimally invasive access to internal anatomy for diagnosis and treatment.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Rigid endoscope; stainless steel construction; rod-lens or fiber-optic imaging bundles; external high-intensity light source; optical fiber illumination; optional working channels for instruments; various lengths and diameters; reusable.
Indications for Use
Indicated for direct visualization of the cervical canal, uterine cavity, and organs within the peritoneum for diagnostic and surgical procedures on female genital organs.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
- Karl Storz Hysteroscopes and Laparoscopes (K88270)
- Karl Storz Hysteroscopes and Laparoscopes (K935277)
- Richard Wolf Hysteroscopes and Laparoscopes (K880314)
- Richard Wolf Hysteroscopes and Laparoscopes (K770378)
Related Devices
- K251652 — Strauss Surgical Cystoscopes & Hysteroscopes (STS-2040-302ASA HD-endoscope, STS-2040-302AA HD-endoscope, STS-2040-302BA HD-endoscope, STS-2040-302BSA HD-endoscope, STS-2040-302ESA HD-endoscope, STS-2040-302CA HD-endoscope, STS-2029-302AA HD-endoscope, STS-2029-302BA HD-endoscope, STS-2040-298DSA-RWF-endoscope, STS-2040-298ASA-RWF-endoscope, STS-2040-300GSA-RWF-endoscope, STS-2040-303CA-RWF-endoscope, STS-2040-280ASA-OLY-endoscope, STS-2040-280ESA-OLY-endoscope, STS-2040-283BSA-OLY-endo · American Medical Endoscopy, Inc. · Aug 1, 2025
- K962116 — GALILEO ENDOSCOPES · Galileo Electro-Optics Corp. · Nov 27, 1996
- K980972 — RIGID CULDOSCOPE AND ACCESSORIES · Circon Corp. · Jun 12, 1998
- K252012 — SPY Cystoscope/Hysteroscope · Stryker Endoscopy · Jul 18, 2025
- K031141 — ASAP CYSTOSCOPE; ASAP LAPAROSCOPE; ASAP LARYGOSCOPE; ASAP OESOPHAGOSCOPE · Asap Endoscopic Products GmbH · Jun 30, 2003
Submission Summary (Full Text)
{0}
DEC - 6 1996
# 510(k) Summary
## Galileo Electro-Optics Corporation
## Galileo Hysteroscopes and Laparoscopes
1. Sponsor/Applicant Name and Address
K962256 p193
Galileo Electro-Optics Corporation
Galileo Park
P.O. Box 550
Sturbridge, MA 01566
Telephone (508) 347-9191
Contact Person
Kin M. Wong, Director of Quality Assurance
Date of Summary Preparation
June 11, 1996
2. Device Name
Proprietary Name: Galileo Hysteroscopes and Laparoscopes
Common/Usual Name: Hysteroscopes and Laparoscopes
Classification Name: Hysteroscopes and accessories
Laparoscopes and accessories
3. Identification of Predicate or Legally Marketed Device(s)
The Galileo Hysteroscopes and Laparoscopes are substantially equivalent to several legally marketed endoscopes including the Karl Storz Hysteroscopes and Laparoscopes, (K88270 and K935277) and the Richard Wolf Hysteroscopes and Laparoscopes (K880314 and K770378).
Galileo Hysteroscopes and Laparoscopes 510(k) 6/11/96 Page D-1
{1}
K962256 p293
## 4. Device Description
The Galileo Hysteroscopes and Laparoscopes are a line of reusable rigid endoscopes based on existing endoscope technology. The Galileo Hysteroscopes and Laparoscopes will be available in various lengths, diameters and configurations for the convenience of the user. Two designs of endoscopes are available:
### Galileo Rod and Lens Design Endoscope:
The rod and lens design Galileo Endoscope functions by light being transmitted from a standard external high intensity light source through optical fibers to the distal tip of the endoscope. The image of the target is then transmitted from the distal end via an objective lens and an alternating set of rod and lenses to a proximal eyepiece. The image can be viewed directly or it may be transmitted through a video camera to a video monitor.
### Galileo Fiber Optic Design Endoscope:
The fiber optic design Galileo Endoscope functions by light being transmitted from a standard external high intensity light source through optical fibers to the distal tip of the endoscope. The image of the target is then transmitted from the distal end via an objective lens and a fiberoptic imaging bundle (instead of a rod and lens) to a proximal eyepiece. The image can be viewed directly or it may be transmitted through a video camera to a video monitor.
## 5. Intended Use
The Galileo Hysteroscopes are rigid endoscopes intended for direct visualization of cervical canal and uterine cavity for diagnostic and surgical procedures during gynecological procedures. The Galileo Laparoscopes are intended for direct visualization of the organs within the peritoneum for diagnostic and surgical procedures on the female genital organs. The Galileo Laparoscopes are designed to be introduced through natural body cavities or through introducers, needles, trocars, catheters, sheaths or other devices with thru-lumens having inside diameters larger than the outside diameter of the endoscope.
Galileo Hysteroscopes and Laparoscopes 510(k) 6/11/96 Page D-2
{2}
K962257 p 3 of 3
# 6. Comparison of Technological Characteristics
The Galileo Hysteroscopes and Laparoscopes and the substantially equivalent devices are identical in intended use in that they are all rigid endoscopes intended to be passed through a lumen of an introducer or into natural body cavities for visualization of body cavities, tissues, organs or canals.
The Galileo Hysteroscopes and Laparoscopes and the substantially equivalent devices are similar in designs in that they all offer various configurations including rod/lens or fiberoptic design, optional working channels, several OD’s and lengths, and use external light sources, and similar stainless steel materials.
The Galileo Hysteroscopes and Laparoscopes and the substantially equivalent devices are similar in technological characteristics in that they offer a channel for either viewing body cavities, tissues, organs or canals and an optional channel for passing instruments into the desired anatomical sites.
Galileo Hysteroscopes and Laparoscopes 510(k) 6/11/96 Page D-3