NeuroStar TMS Therapy System, NeuroStar Advanced Therapy System, NeuroStar, NeuroStar Advanced Therapy for Mental Health
Applicant
Neuronetics, Inc.
Product Code
OBP · Neurology
Decision Date
Jul 15, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NeuroStar Advanced Therapy System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.
Device Story
NeuroStar Advanced Therapy System is a computerized, electromechanical transcranial magnetic stimulation (TMS) device. It delivers non-invasive magnetic pulses via an electromagnetic coil placed against the patient's scalp to induce electrical currents in the cerebral cortex, specifically targeting the left dorsolateral prefrontal cortex (DLPFC). This process causes localized axonal depolarization to activate neurons. Used in clinical settings (hospitals, clinics, physician offices) by trained operators. The system includes a mobile console, treatment chair, head support, and TrakStar PC/software for data management. The provider uses the system to deliver repetitive pulse trains; clinical output is measured via standardized depression (PHQ-9) and anxiety (GAD-7) scales. The device aims to improve depressive and comorbid anxiety symptoms in patients resistant to prior antidepressant medication, potentially offering a non-pharmacological treatment benefit.
Clinical Evidence
Evidence includes 2 RCTs (N=491) and supportive data from 874 patients (open-label and retrospective RWD). RCTs (O'Reardon 2007, George 2010) showed statistically significant (p<0.05) improvement in HAMD-A/S F scores for active vs. sham. Retrospective RWD (TrakStar 2022, N=664) showed 65.5% of patients achieved 6+ point reduction in GAD-7 (p<0.0001).
Technological Characteristics
Biphasic figure-8 coil with ferromagnetic core; air-cooled. Operates via repetitive magnetic pulse trains (10 Hz, 120% intensity). Complies with ISO 13485:2016, IEC 60601-1, and IEC 60601-1-2. System includes mobile console, treatment chair, and TrakStar PC/software. Software-controlled stimulation parameters.
Indications for Use
Indicated for adult patients with Major Depressive Disorder (MDD) who failed to achieve satisfactory improvement from previous antidepressant medication in the current episode, including those with comorbid anxiety symptoms.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
Predicate Devices
Brainsway Deep Transcranial Magnetic Stimulation (K210201)
K231926 — NeuroStar Advanced Therapy System (All previously cleared models) · Neuronetics, Inc. · Mar 22, 2024
K213543 — NeuroStar TMS Therapy System, NeuroStar Advanced Therapy System, NeuroStar, NeuroStar Advanced Therapy for Mental Health · Neuronetics, Inc. · Dec 10, 2021
K161519 — NeuroStar TMS Therapy System · Neuronetics, Inc. · Sep 11, 2016
K233621 — NeuroStar Advanced Therapy System (Version 3.8) · Neuronetics · Dec 13, 2023
K220819 — BrainsWay Deep TMS System · Brainsway , Ltd. · Aug 26, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 14, 2022
Neuronetics, Inc Robin Fatzinger Director, Regulatory Affairs 3222 Phoenixville Pike Malvern, Pennsylvania 19355
Re: K220127
Trade/Device Name: NeuroStar TMS Therapy System, NeuroStar Advanced Therapy System, NeuroStar, NeuroStar Advanced Therapy for Mental Health Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: June 11, 2022 Received: June 14, 2022
Dear Robin Fatzinger :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220127
#### Device Name
NeuroStar TMS Therapy System, NeuroStar Advanced Therapy System, NeuroStar Advanced Therapy for Mental Health
Indications for Use (Describe)
The NeuroStar Advanced Therapy System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="text-decoration: overline;">☒</span> Prescription Use (Part 21 CFR 201 Subpart D)</span> |
| <span style="font-family: Arial, sans-serif;"> ☐ Over-The-Counter Use (21 CFR 201 Subpart G)</span> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters "NS" in a bold, purple font. Above the letters, there is a stylized star shape, also in purple, with curved lines that resemble human figures reaching upwards. The overall design is simple and modern, with a focus on the letters and the star-like symbol.
| 510(k) Number: | K220127 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | 14 July 2022 |
| Applicant: | Neuronetics, Inc.<br>3222 Phoenixville Pike<br>Malvern, PA 19355 |
| Primary Contact: | Robin Fatzinger, RAC<br>Sr. Director, Regulatory Affairs<br>Phone: 610-981-4027<br>Email: robin.fatzinger@neurostar.com |
| Secondary Contact: | Cory Anderson<br>VP, Clinical Affairs & Product Development<br>Phone: 610-981-4104<br>Email: cory.anderson@neurostar.com |
| Device Trade Names: | NeuroStar, NeuroStar TMS Therapy System, NeuroStar Advanced Therapy<br>System, NeuroStar Advanced Therapy for Mental Health |
| Device Common Name: | Transcranial Magnetic Stimulator |
| Classification: | 21 CFR 882.5805 |
| Product Code: | ОВР |
| Predicate Devices: | Primary Predicate: Brainsway Deep Transcranial Magnetic Stimulation<br>(K210201)<br>Predicate: NeuroStar Advanced Therapy System (K201158, K161519,<br>K160703, K133408, K130233, K083538) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters 'NS' in purple. Above the letters, there are three stylized figures, also in purple, arranged in a semi-circle. The figures appear to be reaching upwards, creating a star-like shape above the letters. The overall design is simple and modern, with a focus on the letters and the figures.
#### Device Description
The NeuroStar Advanced Therapy System is a transcranial magnetic stimulation device. Specifically, it is a computerized, electromechanical medical device that produces and delivers non-invasive magnetic fields to induce electrical currents targeting specific regions of the cerebral cortex. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold and is directed in an appropriate orientation relative the brain's neuronal pathway, localized axonal depolarizations are produced, thus activating neurons in the targeted brain region.
The NeuroStar System consists of a combination of hardware, disposable, and consumable supplies, which are required for the operation of the system. The basic configuration includes the following components:
- Mobile Console
- System Software
- Treatment Chair
- Head Support System
- TrakStar PC
- TrakStar Software
#### Indications for Use:
The NeuroStar Advanced Therapy System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.
The indications for use for the subject device (NeuroStar Advanced Therapy System) is identical to that of the Primary Predicate Device.
## Technological Characteristics and Substantial Equivalence:
NeuroStar TMS Therapy system has previously obtained FDA clearance for treatment of major depressive disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode (K061053, K083538, K130233, K133408, K160703, K161519, K201158).
The subject device is substantially equivalent to the primary predicate device that was cleared under K210201. Neuronetics has implemented minor labeling changes to update the indications for use and clinical summaries. None of these changes alter the technical specifications for the subject device. Consistent with the primary predicate and predicate devices, the NeuroStar Advanced Therapy System enables non-invasive activation of brain regions.
The components of, and mechanisms of operation for, the subject device are identical to the
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters "NS" in a stylized font. The letters are purple and are positioned to the left of a star-like symbol. The star is formed by three curved lines that converge at a central point, with small, circular shapes at the ends of each line, resembling stylized figures. The overall design is simple and modern.
previously cleared predicate device, NeuroStar Advanced Therapy System (K061053, K083538, K130233, K133408, K160703, K161519, K201158). The performance characteristics, including the Electrical and Magnetic Field Distribution testing are the same as the previously cleared NeuroStar Advanced Therapy System. The subject device has the following similarities to the primary predicate (Brainsway Deep TMS; K210201) and predicate NeuroStar Advanced Therapy System:
- Principles of operation
- Design for delivery of Transcranial Magnetic Stimulation (TMS)
- Materials
The stimulation parameters (frequency, train duration, interval, number of trains, number of pulses, and total duration of treatment) of the subject device are different from those of the primary predicate. Substantial equivalence between the subject device and the primary predicate device was established based on clinical performance data.
The proposed changes for the NeuroStar Advanced Therapy System are limited to labeling updates, specifically to include the treatment of adult patients with MDD who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode and may exhibit comorbid anxiety symptoms. The proposed changes are supported by information submitted in this premarket notification and with the following rationale:
- The subject device is substantially equivalent to the FDA-cleared Brainsway Deep TMS (K210201).
- The predicate device, NeuroStar Advanced Therapy System, was previously cleared by the FDA under DEN060153/K061053, K083538, K130233 and K133408, K160703, K161519, and K201158.
- . The subject device changes remain limited to labeling revisions, in support of the reference to treatment of adult patients with MDD who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode and may exhibit comorbid anxiety symptoms. No other changes are made to the device or product labeling.
Supportive clinical data has been provided to support the substantial equivalence of the subject NeuroStar Advanced Therapy System in terms of safety and effectiveness for the expanded indications for use. Therefore, the NeuroStar TMS Therapy System with the proposed changes to the product labeling is substantially equivalent to the primary predicate and predicate devices.
## Non-Clinical Testing:
The subject device has been tested in accordance with IEC60601 Edition 3.1 and verified to comply with the specified permissible sound pressure levels and permissible thresholds for exposure defined by the Occupational Safety and Health Administration (OSHA).
These tests provide evidence that the NeuroStar Advanced Therapy system does not pose any risk for potential hearing reduction or loss in either patients or operators.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters "NS" in a bold, sans-serif font, colored in purple. Above and around the letters, there is a stylized star shape, also in purple, formed by abstract figures that resemble people with outstretched arms. The overall design is simple and modern, suggesting a focus on community or people-oriented services.
## Performance Standards:
The NeuroStar Advanced Therapy System has been tested and conforms with the following standards:
- . ISO 13485:2016
- IEC60601-1
- IEC60601-1-2
Additionally, the contents of this 510(k) complies with the FDA Guidance Document: "Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems - Guidance for Industry and Food and Drug Administration Staff". The non-clinical performance testing of the components of NeuroStar Advanced Therapy System was conducted as required according to the standards listed above. All system components have been previously cleared by the FDA.
# Clinical Performance Data:
Clinical performance data was provided to support the safety and effectiveness of the NeuroStar Advanced Therapy System device for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD), as described in the Indications for Use statement. The clinical performance data was based on 2 Randomized Controlled Trials (RCTs) (George et al. (2010), O'Reardon et al. (2007)) in a total of 491patients and on supportive data from a total of 874 patients in one open-label study and four observational studies (retrospective medical chart review), including a large-scale analysis of real-world data (RWD).
The 2007 O'Reardon study was designed to evaluate the safety and efficacy of the NeuroStar Advanced Therapy System for the treatment of adult patients with major depressive disorder (MDD) (N=301) who failed to receive benefit from 1 to 4 prior antidepressant medications verified by the ATHF. The 2010 study by George et al. was a company-independent, NIMH-funded trial designed to evaluate the safety and efficacy of TMS in adult patients (N=190) with moderate to severe MDD who failed to receive benefit from 1 to 4 prior antidepressant medications verified by the ATHF (George, et al, 2010). Clinical data from these two randomized sham-controlled trials were originally used to establish the safety and efficacy effectiveness of NeuroStar Advanced Therapy for the treatment of adult patients with MDD (DEN060153/K061053) and as such, patient demographics and depression outcomes are presented in detail in earlier sections.
For the 2007 O'Reardon study and the 2010 George study, the effect sizes were identical, 0.36 (Cohen's d), representing a low effect size for decreasing anxiety symptoms in this patient population when compared to Sham control. These effect sizes compare favorably to the effect size reported in two of the three RCTs (Levkovitz et al. (2015) – 0.34 and Kaster et al. (2018) – 0.36) which formed the basis for clearance of Brainsway Deep TMS™ System (K210201), the primary predicate device, for the indications for use in section referenced above.
In both RCT studies, the effectiveness of NeuroStar Advanced Therapy for the treatment of comorbid anxiety symptoms in adult patients with MDD and comorbid anxiety symptoms was assessed using the Hamilton-D anxiety/somatization factor (HAMD – A/S F) Scale. The HAMD – A/S F Scale is a validated scale comprising six anxiety-specific items from the HAMD-17 or the HAMD-24. In O'Reardon et al. (2007), evaluation of change on the HAMD – A/S F was an a priori defined secondary endpoint. In both trials, the
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo with the letters "NS" in a bold, purple font. To the right of the letters, there is a stylized star shape, also in purple. The star is formed by three curved lines that resemble human figures with outstretched arms, giving the impression of people reaching towards the center of the star.
between-group difference in the mean change in HAMD – A/S F scores from baseline to the 6-week endpoint evaluation was statistically significant (p<0.05) in favor of the active treatment group.
Additional clinical evidence intended to support the safety and efficacy effectiveness of The NeuroStar Advanced Therapy System administered to the left dorsolateral prefrontal cortex (DFLPFC), as per the standard NeuroStar treatment protocol for MDD, in patients with MDD and comorbid anxiety symptoms included data from the following studies:
- 1) A retrospective analysis of medical records from 57 adult patients with major depressive disorder (MDD) and comorbid anxiety symptoms who received rTMS treatment with The NeuroStar Advanced Therapy System demonstrated a clinically significant reduction in anxiety symptoms over a 6-week period, as measured by the HAMD-21 – A/S F or Generalized Anxiety Disorder-7 (GAD-7) score. (Tunistra et al. (2022)).
- 2) A large-scale retrospective analysis of real-world data (RWD) derived from the TrakStar registry data of 664 patients who received the standard NeuroStar treatment protocol for MDD over a span of 13 years, beginning in 2008. This RWD was collected from patients across 75 TMS centers in the US. These 664 patient records were analyzed to determine the difference in measures of depression (PHQ-9 scores) and anxiety (GAD-7 scores) over a pre-post TMS treatment interval of 6 weeks. (TrakStar (2022) study).
Patients included in this RWD study had a primary diagnosis of MDD, had received at least 20 treatments with NeuroStar Advanced Therapy to the left dorsolateral prefrontal cortex (DLPFC); were required to exhibit baseline moderate or greater depression, defined as a score on the Physician Health Questionnaire-9 (PHQ-9) ≥ 10, and baseline moderate or greater anxiety symptoms, defined as a score on the Generalized Anxiety Disorder-7 (GAD-7) ≥ 10. The average patient age was 47.4 ± 13.9 years, and 67.9% were female.
A total of 65.5% of patients met the primary outcome measure for anxiety which was the proportion of subjects who achieved a 6+ point improvement (reduction) in the GAD-7 score at end of treatment relative to baseline. The mean change in GAD-7 scores from baseline to endpoint was -8.0 ± 5.7 (p<0.0001), and 34.0% of patients attained remission, with remission of anxiety symptoms defined as a score on the GAD-7 < 5 at study endpoint. These statistically significant and clinically meaningful improvements in anxiety symptoms were accompanied by comparable changes in depression scores on the PHQ-9.
The TrakStar registry did not account for medication or other treatment change for MDD or symptoms of anxiety that may have occurred during the treatment phase.
{8}------------------------------------------------
| Study | Device(s) | Study Design | Population | Sample size | Outcome<br>Measure(s) | Study<br>Results |
|---------------------------|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| O'Reardon<br>et al., 2008 | NeuroStar | • RCT<br>• Left-sided<br>dorsolateral<br>prefrontal cortex<br>(DLPFC) rTMS<br>Therapy per<br>standard NeuroStar<br>treatment protocol | • Mean age (years):<br>- Active: 47.9 ± 11.0<br>- Sham: 48.7 ± 10.6<br>• Females:<br>- Active: 55.5%<br>- Sham: 48.7%<br>• Primary diagnosis of<br>MDD. | 301:<br>- Active:155<br>- Sham: 146 | • HAMD-24<br>• HAMD-17<br>• HAMD AS/F (a<br>priori defined<br>secondary<br>endpoint) | Comparison between active and sham TMS groups in the<br>change from baseline to endpoint (6 weeks):<br><br>Response Rate: ≥ 50% decrease in end score relative to baseline<br>(6 weeks):<br>HAMD<br>17:<br>p<0.05<br>HAMD-<br>24:<br>p<0.05<br>Mean Change in Scores from Baseline to<br>Endpoint: HAMD-17: p<0.01<br>HAMD-24: p<0.05<br>HAMD-17-AS/F: p<0.05 (p=0.023)<br><br>Effect size (Cohen's d): 0.36 |
| | NeuroStar | • RCT<br>• Left-sided<br>dorsolateral<br>prefrontal cortex<br>(DLPFC) rTMS<br>Therapy per<br>standard NeuroStar<br>treatment protocol | • Mean age (years):<br>- Active: 47.7 ± 10.6<br>- Sham: 46.5 ± 12.3<br>• Females:<br>- Active: 63%<br>- Sham: 51%<br>• Primary diagnosis of<br>MDD. | 190:<br>- Active:92<br>- Sham: 98 | • HAMD-24<br>• HAMD AS/F | Comparison between active and sham TMS groups in the<br>change from baseline to endpoint (6 weeks):<br><br>Response Rate: ≥ 50% decrease in end score relative to baseline<br>(6 weeks):<br>HAMD-24: p<0.01<br>Remission:<br>HAMD-24 < 10: p<0.05<br>Mean Change in Scores from Baseline to<br>Endpoint (6 weeks): HAMD-17-AS/F:<br>p<0.05:<br>- Active: -1.3; = Sham: -0.4<br><br>Effect size (Cohen's d): 0.36 |
| TrakStar<br>(unpublished) | NeuroStar | • Retrospective<br>chart review &<br>analysis of RWD<br>since 2008.<br><br>• Left-sided DLPFC<br>rTMS Therapy per<br>standard<br>NeuroStar<br>treatment<br>protocol. | • Adults 22-70 years;<br>mean age of<br>$47.4\pm13.9$ years.<br>• 67.9% female<br>• Primary diagnosis of<br>MDD.<br>• Baseline score on the<br>GAD-7 ≥ 10 | 664 | • GAD-7<br>• PHQ-9 | Response Rate: ≥ 6 point decrease in end score relative<br>to baseline (6 weeks): PHQ-9: 58.9%<br>GAD-7: 65.5%<br>Remission:<br>PHQ-9: < 5: 30.4%<br>GAD-7 < 5: 34.0%<br>Mean Change in Scores from Baseline to<br>Endpoint (6 weeks): PHQ-9: $-10.5 \pm 6.7$ ,<br>p<0.0001<br>GAD-7: $-8.0 \pm 5.7$ : p<0.0001<br>Effect size (Hedges g): -1.4 |
| Tuinstra, et<br>al. 2022 | NeuroStar | • Retrospective<br>chart review and<br>analysis of RWD<br>over 3.5 years.<br><br>• Left-sided<br>dorsolateral<br>prefrontal cortex<br>(DLPFC) rTMS<br>Therapy per<br>standard NeuroStar<br>treatment protocol | • Adults 18-77 years;<br>mean age of<br>$45.3\pm16.7$ years.<br>• 61% female<br>• Primary diagnosis of<br>Treatment Resistant<br>Depressions (TRD).<br>• 52% had at least one<br>diagnosis of a<br>comorbid anxiety<br>disorder | 77<br>57 had<br>clinically<br>significant<br>anxiety<br>symptoms | • HAMD-17<br>Anxiety<br>Somatization<br>Factor (AS/F)<br>Subscale<br>• GAD-7<br>• PHQ-9 | Remission:<br>HAMD AS/F < 7: 87.5%<br>GAD-7 < 5: 22.6%<br>Response Rate: ≥ 50% decrease in end score relative to baseline<br>(6 weeks):<br>HAMD AS/F: 50%<br>GAD-7: 41.5%<br>Clinically Significant Change:<br>HAMD AS/F: final score ≤ 3.2: 37.5%<br>GAD-7: mean change from baseline to endpoint ≥ 6: 41.5%<br>Mean Change in Scores from Baseline to Endpoint<br>(6 weeks): HAMD AS/F: -3.53, p<0.001<br>GAD-7: -5.32: p<0.001<br>Effect size (Hedges g): -0.777 |
Table 1: Study Design and Outcomes from NeuroStar RCTs, TrakStar Data and Supportive Literature
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo with the letters "NS" in a stylized, purple font. Above the letters is a graphic of three figures with outstretched arms, also in purple, arranged in a star-like formation. The overall design is simple and modern, suggesting a focus on people and community.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a logo with the letters 'NS' in a stylized font. The letters are purple and are positioned to the left of a star-like symbol. The star is formed by four stylized figures with outstretched arms, also in purple. The overall design is simple and modern.
# Table 2: Substantial Equivalence Comparison
| | Subject Device<br>K220127 | Primary Predicate Device<br>K210201 | |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Specification | NeuroStar Advanced Therapy System | Brainsway Deep TMS System | Justification of Differences |
| Indications for use | The NeuroStar Advanced Therapy System is<br>indicated for the treatment of depressive episodes<br>and for decreasing anxiety symptoms for those<br>who may exhibit comorbid anxiety symptoms in<br>adult patients suffering from Major Depressive<br>Disorder (MDD) and who failed to achieve<br>satisfactory improvement from previous<br>antidepressant medication treatment in the<br>current episode. | The Brainsway Deep TMS™ System is indicated for<br>the treatment of depressive episodes and forexhibit comorbid anxiety symptoms in adult<br>patients suffering from Major Depressive Disorder<br>(MDD) and who failed to achieve satisfactory<br>improvement from previous antidepressant<br>medication treatment in the current episode. | No Difference |
| Intended Use | Major Depressive Disorder (MDD) and comorbid<br>anxiety symptoms | Major Depressive Disorder (MDD) and comorbid<br>anxiety symptoms | No Difference |
| Anatomical Sites | Left dorsolateral prefrontal cortex | Left dorsolateral prefrontal cortex | No Difference |
| Target Population | Adult Patients | Adult Patients | No Difference |
| Clinical Setting | Inpatient and outpatient settings including<br>physician's offices and clinics, hospitals, and<br>general medical/surgical hospitals | Inpatient and outpatient settings including<br>physician's offices and clinics, hospitals, and<br>general medical/surgical hospitals | No Difference |
| Treatment Schedule | 5 daily sessions per week for 6 weeks plus 6 taper<br>sessions over 3 weeks (36 total sessions) | 5 daily sessions for 4 weeks; bi-weekly sessions for<br>12 Weeks (optional maintenance treatments) | The difference in treatment<br>schedule between the subject and<br>predicate devices is dependent on<br>the device treatment protocol;<br>differences do not raise differences<br>in safety or effectiveness based on<br>clinical data. Clinical evidence for<br>the subject device has been included<br>and demonstrated safety and<br>effectiveness for the intended use. |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows a logo with the letters "NS" in a bold, purple font. Above the letters is a stylized star shape, also in purple, with three abstract figures incorporated into its design. The figures appear to be reaching upwards, adding a sense of dynamism and community to the logo. The overall design is simple yet eye-catching, using a consistent color scheme and a combination of text and abstract shapes.
| | Treatment parameters | | Treatment parameters | | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------------------------------------|----------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Magnetic Field Intensity | 120% | Magnetic Field Intensity | 120% | No Difference |
| Energy Delivered and<br>Performance | Repetition Rate | 10 Hz | Repetition Rate | 20 Hz | While the frequency is different<br>between the subject and predicate<br>devices, they are appropriate and<br>within safe limits based on the train<br>duration and intensity for each as<br>indicated in Table 2 in Section 4 of<br>the Special Controls Guidance.<br>Clinical evidence for the subject<br>device has been included and<br>demonstrated safety and<br>effectiveness for the intended use. |
| | Train Duration | 4 sec | Train Duration | 2 sec | While the train duration is different<br>between the subject and predicate<br>devices, they are appropriate and<br>within safe limits based on the<br>frequency and intensity for each as<br>indicated in Table 2 in Section 4 of<br>the Special Controls Guidance.<br>Clinical evidence for the subject<br>device has been included and<br>demonstrated safety and<br>effectiveness for the intended use. |
| | Inter-Train-Interval | 11-26 secs | Inter-Train-Interval | 20 secs | The difference between the inter-<br>train-intervals is appropriate for<br>each device design and treatment<br>protocol. These differences do not<br>raise different questions of safety<br>or effectiveness based on non-<br>clinical and clinical testing.<br>Clinical evidence for the subject<br>device has been included and<br>demonstrated safety and<br>effectiveness for the intended use. |
| | Number of Trains | 75 | Number of Trains | 55 | The difference between the<br>number of trains is appropriate for |
| | | | | | each device design and treatment.<br>protocol. This difference does not<br>raise different questions of safety or<br>effectiveness based on non-clinical<br>and clinical testing. Clinical<br>evidence for the subject device has<br>been included and demonstrated<br>safety and effectiveness for the<br>intended use. |
| | Number of Pulses | 3000 | Number of Pulses | 1980 | The difference between the number<br>of pulses is appropriate for each<br>device design and treatment protocol.<br>This difference does not raise<br>different questions of safety or<br>effectiveness based on non-clinical<br>and clinical testing. Clinical evidence<br>for the subject device has been<br>included and demonstrated safety and<br>effectiveness for the intended use. |
| | Treatment Duration | 18.75 min | Treatment Duration | 20.2 min | The difference between the<br>treatment duration is appropriate<br>for each device design and<br>treatment protocol. This difference<br>does not raise different questions of<br>safety or effectiveness based on<br>non- clinical and clinical testing.<br>Clinical evidence for the subject<br>device has been included and<br>demonstrated<br>safety and effectiveness for the<br>intended use. |
| | Treatment area of brain to be stimulated:<br>Left Dorsolateral Prefrontal Cortex | | Treatment area of brain to be stimulated:<br>Left Dorsolateral Prefrontal Cortex | | No Difference |
| | Output Stimulation Parameters<br>Available Stimulation Intensity in terms of Standard<br>Motor Threshold (SMT) units | | Output Stimulation Parameters:<br>Available Stimulation Intensity in terms of Standard<br>Motor Threshold (SMT) units | | The differences between the SMT<br>ranges are dependent on the device<br>design and treatment |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows a logo with the letters "NS" in purple. The letters are stylized and appear to be part of a larger design. Above the letters, there are three abstract figures that look like people with their arms raised. These figures are arranged in a curved line, suggesting movement or connection. The overall design has a modern and dynamic feel.
Page 10 of 13
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows a logo with the letters 'NS' in a bold, sans-serif font, colored in purple. Above the letters, there are three stylized figures that appear to be dancing or reaching upwards, also in purple. These figures are arranged in an arc shape, suggesting movement and energy. The overall design is simple yet dynamic, conveying a sense of community or collaboration.
| | Range: .22 - 2.08<br>Waveform: Biphasic | Range: 0.6-1.4<br>Waveform: Biphasic | protocols. Each range is within the<br>therapeutic stimulation range of the<br>device. The difference does not<br>raise new issues of safety or<br>effectiveness since the maximum<br>(100%) output level is unchanged<br>from the previously cleared<br>Neuronetics predicate device.<br>Clinical evidence for the subject<br>device has been included and<br>demonstrated safety and<br>effectiveness for the intended use. |
|--------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | The system consists of:<br>1. Mobile console<br>2. System software with GUI<br>3. Treatment chair<br>4. Head support system<br>5. Coil positioning system<br>6. Same Coil for both MT and treatment<br>7. Coil fixture<br>8. Data management system | The system consists of:<br>1. Mobile console<br>2. System software with GUI<br>3. Treatment chair<br>4. Head support system<br>5. Coil positioning system<br>6. Same Coil for both MT and treatment<br>7. Coil fixture<br>8. Data management system | No Difference |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows a logo with the letters "NS" in bold, purple font. Above the letters is a stylized star shape, also in purple, with three figures incorporated into the design. The figures appear to be reaching upwards, contributing to the star's overall shape. The logo has a clean and modern look.
| Coil | Biphasic Figure 8 Coil with Ferromagnetic Core | Biphasic H-Coil with Air Core | Both coils rely on transcranial<br>magnetic stimulation by means of<br>repetitive pulse trains at<br>predetermined frequency. Both<br>coils use the same mechanism of<br>action (electromechanical<br>instrument that produces and<br>delivers brief duration, pulsed<br>magnetic fields to induce electrical<br>currents in localized regions of the<br>prefrontal cortex). Differences in<br>design do not raise different<br>questions of safety or effectiveness<br>as demonstrated by the clinical<br>evidence presented within the<br>510(k). Clinical evidence for the<br>subject device has been included<br>and demonstrated safety and<br>effectiveness for the intended use. |
|---------------------------------------------------------|-------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cooling | Air cooled.<br>Used for both MT determination and treatment | Liquid cooled<br>Used for both MT determination and treatment | The H-Coil uses liquid cooling to<br>maintain safe operating<br>temperatures. The iron core figure-<br>8 coil uses air cooling to maintain<br>safe operating temperatures. Both<br>methods of cooling are sufficient for<br>the specific design. The differences<br>do not raise different questions of<br>safety or effectiveness as<br>demonstrated by the non-clinical<br>and clinical testing provided within<br>the 510(k). Clinical evidence for the<br>subject device has been included<br>and demonstrated safety and<br>effectiveness for the intended use. |
| Quality & Risk Standards | Company complies with ISO 13485:2016 and ISO<br>14971 | Company complies with ISO 13485:2016 and ISO<br>14971 | No Difference |
| Electrical Safety &<br>Electromagnetic<br>Compatibility | Complies with IEC60601-1 and IEC60601-1-2 | Complies with IEC60601-1 v. 3.1, and IEC60601-1-2 | No Difference |
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image shows a logo with the letters "NS" in a bold, purple font. The letters are stylized and slightly overlapping. Above the letters, there is a star-like shape formed by three abstract figures that appear to be people with their arms raised. The overall design is simple and modern, with a focus on the letters and the star-like symbol.
### Conclusion:
The NeuroStar Advanced Therapy System has the same intended use and indications, principles of operation, and similar technological characteristics as the primary predicate device and the previously cleared predicate device. The differences noted in the technical characteristics do not raise different questions of safety or effectiveness. Clinical performance data demonstrates that the subject device is safe and effective for its intended use. Thus, the NeuroStar Advanced Therapy System is substantially equivalent to the primary predicate and predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.